English-7
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 10: Handpiece Tubing Connections.
A) Take numbered red and purple tubing from control tray and connect to
numbered pilot tubing from handpieces.
Step 11: Water Bottle Assembly Tubing
Connections (if applicable).
A) Take the yellow tubing from control tray and pair with
the yellow tubing from water bottle assembly.
B) Trim the yellow tubing 18” from control tray and
connect the tubing from the control tray to the yellow
tubing from water bottle assembly.
C) Then tee excess yellow tubing that was trimmed off
onto tee’d assembly to route from the tee connection
to the air regulator.
D) Take the orange tubing from control tray and pair with
the orange tubing from water bottle assembly.
E) Trim the orange tubing 18” from the control tray and
connect the tubing from control tray to the orange
tubing from the water bottle assembly.
F) Then tee excess orange tubing that was trimmed off
onto tee’d assembly to route from the tee connection
to the foot control switch.
G) Connect blue tubing from control tray to blue tubing
from water bottle assembly.
H) Install water bottle.
Note: If water bottle assembly is not configured with unit, yellow and
orange tubing will go from the control tray to the air regulator and the
foot control switch.
Step 12: Solids Trap Tubing Connections.
A) Route corrugated vacuum tubing from vacuum supply line to solids trap.
B) Route HVE and SE tubing into cabinet through cutout in door/drawer,
then through bracket on the back of the door then hook up to solids trap.
Note: This may require removing red caps from HVE fitting on solids trap.
003-2793-00