Midmark Artizan® / Artizan® Expressions Operatory Cabinetry - Pedo Bench Installation guide

Type
Installation guide
1
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
003-2198-00 Rev. E (1/25/13)
Procenter Delivery System Cabinet Installation
Applies to Models:
153839-004
Special Tools:
none
Language of origin: English
Equipment Alert
Must provide Air, Water and Electrical supply for Delivery System.
Dental System must be installed per local plumbing and electrical codes.
If using a Illuminated Handpiece, 115V~, 0.6 amp is required.
Note: Electrical only needed if until has optional Illuminated Handpiece.
Cabinet / Utilities Requirements
Minimum cabinet interior dimension required are 10” wide x 17” deep x 25” high.
• Required: Cabinet floor opening to access utility connections and supplied regulators.
A 1” diameter hole is required in cabinet toekick area to route the delivery system foot control into utilities area.
• Depending upon the delivery system options selected,
utilties area must include connections for air, water,
electrical supply and vacuum. All utilities must be installed
per local plumbing and electrical codes by a licenced
contractor. If using Illuminated Handpiece option,
a medical-grade receptacle and circuit rated at 115V~,
60 Hz, 0.6 amp is required.
DA2638
midmark.com
Style G
2
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
DA2604
Cabinet Members
Step 1: Drill six 3/32 pilot hole locations for cabinet members
with template provided with delivery system.
Step 2: Screw slide cabinet
members to cabinet.
Step 3: Screw slide drawer members to control tray assembly.
#8 x 3/4” Screws
Drawer Members
#10 x 3/8” Screws
Assembled View
Hole Position of
Drawer Member
midmark.com
Style G
3
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
DA2605
Step 4: Position bearing race at end of each slide that is attached to cabinet.
Bearing Race
Step 5: Align top and bottom slides on
control tray with slides on cabinet.
Move carefully into position.
midmark.com
Style G
4
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 6: Mount Holder bar for instruments.
A) Mark mounting locations on cabinet. Must be located in front of control tray panel.
Installer must add hole in cabinet door or provide opening at the bottom of the door for instrument tubing to go inside cabinet.
B) Drill two 7/32” holes at 1 1/8” spread.
C) Mount bar using two #10 screws provided.
Instrument Bar
Opening for Tubing
1 1/8”
7/32” Holes
2” MIN
EXAMPLES
Door With Hole
Short Door
midmark.com
Style G
5
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 7: Mount Water Bottle Assembly. (If Supplied)
A) Mark mounting locations in cabinet. Must be located across from control tray panel or inside door.
B) Drill four 3/32” pilot holes. Refer to dimensions in picture.
C) Mount water bottle assembly using four #8 x 3/4” screws provided.
Note: May use bracket to mark pilot hole spread.
Step 8: Mount Solids Collector. (If Supplied)
A) Mark mounting locations in cabinet. Must be located across from control tray panel.
B) Drill two 3/32” pilot holes. Refer to dimensions in picture.
C) Mount solids collector using two #8 x 3/4” screws provided.
Note: May use bracket to mark pilot hole spread.
DA2607
3 1/2”
1 3/8”
1 1/2”
Solids Collector
Water Bottle Assembly
midmark.com
Style G
6
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 9: Install Foot Control and Make Tubing Connections.
A) Route tubing through 1” hole in cabinet toe kick.
B) Make air and water connections according to tubing diagrams on the following pages.
DA2612
Foot Control
Tubing
Step 10: Install Air and Water Regulators.
Refer to diagrams in this manual for other air and water connections.
Regulator
Shut-off Valve
Regulator
Compression Fitting
Shut-off Valve
Compression
Fitting
Air Regulator
Water Regulator
To Handpiece
Coolant Air
To Pilot Air for
Coolant Water
To Handpiece
Drive Air
midmark.com
Style G
7
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
DA2474-1
153839 (-001 thru -004)
Water Bottle Assembly
1/8 Blue
City
Bottle
Water
Selector
Valve
30 PSI
Bottle
Water
City
Bottle
1/8 Brown
30 PSI
Air
1/8 Blue
1/8 Blue
Instrument
Air Supply
Instruments
Water Supply
30 PSI City Water
To / From
Air - Water
Junction
(located in
cabinet
base or
control
panel)
80 PSI Air
80 PSI Syringe Air
1/8
Clear
1/8 Brown
1/8 Yellow
1/8
Brown
Ribbed
Side
30
PSI
Bottle
Water
Instrument Water Supply
Out
Push
Flush Valve
(N.C.) 2-Way
Manual
Momentary
1/8 Orange
1/8 Yellow
1/8 Orange
1/8 Yellow
1/8 Blue
Flush Valve Pilot Air
(Control Panel)
Syringe Supply Air
Double
Pinch
Valve
Rib on Air Side
(Yellow Tubing)
Water
Air
1/8
Blue
To Assistant's
Instrument
Syringe
1/8
Blue
1/8
Yellow
1/8
Blue
In-Line
Regulator
midmark.com
Style G
Color Page
8
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
midmark.com
Style G
Color Page
9
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
midmark.com
Style G
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9

Midmark Artizan® / Artizan® Expressions Operatory Cabinetry - Pedo Bench Installation guide

Type
Installation guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI