Midmark Procenter Systems offer a comprehensive solution for dental professionals seeking advanced and efficient dental delivery. These systems are designed with meticulous attention to detail, providing a wide array of features to enhance patient care and streamline dental procedures. With the ability to accommodate various delivery options, including handpieces, syringes, and water bottles, these systems adapt seamlessly to different clinical needs.

Midmark Procenter Systems offer a comprehensive solution for dental professionals seeking advanced and efficient dental delivery. These systems are designed with meticulous attention to detail, providing a wide array of features to enhance patient care and streamline dental procedures. With the ability to accommodate various delivery options, including handpieces, syringes, and water bottles, these systems adapt seamlessly to different clinical needs.

English-1
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Applies to Models:
153839 (-003)
Special Tools:
none
Language of origin: English
003-2793-00 Rev. A (8/28/14)
Procenter Delivery System Cabinet Installation
Equipment Alert
Must provide Air, Water and Electrical supply for Delivery System.
Dental System must be installed per local plumbing and electrical codes.
If using a Illuminated Handpiece, 115V~, 0.6 amp is required.
Note: Electrical only needed if until has optional Illuminated Handpiece.
Cabinet / Utilities Requirements
Minimum cabinet interior dimensions required are 10” wide x 17” deep x 25” high.
• Required: Cabinet floor opening or junction box (for mobile applications) to access utility connections and supplied regulators.
A 1” diameter hole is required in cabinet toekick area to route the delivery system foot control into utilities area.
• Depending upon the delivery system options selected,
utilities area must include connections for air, water,
electrical supply and vacuum. All utilities must be installed
per local plumbing and electrical codes by a licensed
contractor. If using Illuminated Handpiece option,
a medical-grade receptacle and circuit rated at 115V~,
60 Hz, 0.6 amp is required.
DA2638
Style G
midmark.com
English-2
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
4227
1-1/8”
Handpiece Pilot
Tubing Hole.
Step 1: Mount Holder bar for instruments.
A) Remove two #10 screws provided in instrument mounting bar.
B) Using the two #10 screws mount instrument bar to door.
C) Route handpiece pilot tubing through hole in door near holders
as shown.
Note: See door preparation if holes are not provided..
Door Preparation...
A) Mark mounting locations on cabinet. Must be located in front of control tray panel.
Installer must add hole in cabinet door or provide opening at the bottom of the
door (2” minimum) for instrument tubing to go inside cabinet if not provided.
B) Drill two 7/32” holes at 1 1/8” spread if not provided.
1-1/8”
4”
Center of Door
2-5/16”
3-1/2”
003-2793-00
Style G
midmark.com
English-3
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 3: Mount Water Bottle Assembly. (if applicable)
A) Mark mounting locations in cabinet. Must be located across from control
tray or preferably inside door.
B) Drill 1/8” pilot holes (refer to dimensions in picture).
C) Mount water bottle assembly using four #8 wood screws provided.
Note: May use bracket to mark pilot hole spread.
Water Bottle
Mounting Bracket
Solids Collector
1-1/2”
3-1/2”
1-3/8”
Step 2: Mount Solids Collector.
A) Mark mounting locations in cabinet. Must be located across
from control tray.
B) Drill two 1/8” pilot holes (refer to dimensions in picture).
C) Mount solids collector using two #8 wood screws provided.
Note: May use bracket to mark pilot hole spread.
003-2793-00
Style G
midmark.com
English-4
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
DA2604
Cabinet Members
#8 x 3/4” Screws
Drawer Members
Assembled View
Hole Position of
Drawer Member
Step 5: Slide installation.
A) Drill six 3/32 pilot hole locations for cabinet members with template
provided with Install Kit (002-1359-00) if installing in custom cabinet.
B) Screw slide cabinet members to cabinet.
Step 4: Assembling control tray to slides.
A) Screw slide drawer members to control tray assembly.
#10 x 3/8” Screws
003-2793-00
Style G
midmark.com
English-5
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 6: Slide installation.
A) Position bearing race at end of each slide that is attached to cabinet.
B) Align top and bottom slides on control tray with slides on cabinet.
Move carefully into position.
Step 7: Slide installation.
A) Route handpiece and syringe tubing from control tray through cutout
in door to appropriate holder on holder bar.
B) Loosen screws holding bracket to door, slip bracket off, route
tubing through and replace bracket onto door. The tubing must be
prevented from dragging of the floor outside the cabinet.
Bearing Race
003-2793-00
Style G
midmark.com
English-6
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 8: Umbilical Tubing (for optional mobile application).
A) Take the end of the umbilical with the quick connects on the internal tubing
and thread one umbilical cuff onto the end about 12” (this can be adjusted
once umbilical is secured to the cabinet).
B) If not already provided in the cabinet, drill a 2” Dia hole in the sides of the
cabinet where the umbilical will enter the cabinet.
C) Feed any electrical Flexible conduit through large corrugated tubing.
D) Route umbilical assembly (with electrical whip included) through the
cutouts in cabinet.
E) Use the other umbilical cuff to secure the tubing to the cabinet.
Step 9: Umbilical Tubing to cabinet (for optional mobile application).
A) Route umbilical assembly (with electrical whip included) through the cutouts in the
cabinet into compartment with the control tray (if applicable).
B) Use the other umbilical cuff to secure the tubing to the cabinet.
12”
Flat side of
Threaded Cuff
2nd Cuff
Flat side
003-2793-00
Style G
midmark.com
English-7
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 10: Handpiece Tubing Connections.
A) Take numbered red and purple tubing from control tray and connect to
numbered pilot tubing from handpieces.
Step 11: Water Bottle Assembly Tubing
Connections (if applicable).
A) Take the yellow tubing from control tray and pair with
the yellow tubing from water bottle assembly.
B) Trim the yellow tubing 18” from control tray and
connect the tubing from the control tray to the yellow
tubing from water bottle assembly.
C) Then tee excess yellow tubing that was trimmed off
onto tee’d assembly to route from the tee connection
to the air regulator.
D) Take the orange tubing from control tray and pair with
the orange tubing from water bottle assembly.
E) Trim the orange tubing 18” from the control tray and
connect the tubing from control tray to the orange
tubing from the water bottle assembly.
F) Then tee excess orange tubing that was trimmed off
onto tee’d assembly to route from the tee connection
to the foot control switch.
G) Connect blue tubing from control tray to blue tubing
from water bottle assembly.
H) Install water bottle.
Note: If water bottle assembly is not configured with unit, yellow and
orange tubing will go from the control tray to the air regulator and the
foot control switch.
Step 12: Solids Trap Tubing Connections.
A) Route corrugated vacuum tubing from vacuum supply line to solids trap.
B) Route HVE and SE tubing into cabinet through cutout in door/drawer,
then through bracket on the back of the door then hook up to solids trap.
Note: This may require removing red caps from HVE fitting on solids trap.
003-2793-00
Style G
midmark.com
English-8
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 14: Install Air and Water Regulators.
A) Refer to diagrams in this manual for other air and water connections.
Step 13: Install foot control and make tubing connections.
A) Route tubing to cabinet. On mobile models, route thru the cutout in the bottom of the cabinet. On static models,
route thru the grommet in the front of the toekick.
B) Make air and water connections according to tubing diagrams on the following pages.
Water Regulator
Air Regulator
Step 15: Organize Tubing.
A) Organize tubing from water bottle assembly and route through
supplied cable connectors inside the cabinet.
003-2793-00
Style G
midmark.com
English-9
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
153839
(-001 Thru -004)
Control
Panel
Schematic
DA2613
Handpiece
Drive Air
Handpiece
Coolant
Air
Syringe
Water
Syringe
Air
Drive Air
Pressure Gauge
Syringe
Tubing
1/8"
Yellow
Air / Oil
Separator
Check
Valve
Check
Valve
Handpiece
Drive Air
Handpiece
Drive Air
Handpiece
Coolant
Water
Handpiece
Coolant
Water
Handpiece
Coolant
Water
Kink
Valve
Kink
Valve
1/8" Blue
1/8"
Green
CW- Coolant Water
DA - Drive Air
CA - Coolant Air
Pilot
Air
From
Pilot
Air
To
Doctor's
Instrument
Handpiece
Valves
Pilot Air from
Master Switch on
Water Bottle Assy.
1/8"
Blue
1/8" Yellow
1/8" Green
1/8" Orange
1/8" Red
1/4" Clear
1/8"
Clear
1/8" Red
1/8" Orange
1/8" Orange
1/8"
Purple
1/8"
Grn
1/8"
Orange
1/4"
Clear
1/8" Red
1/8"
Clear
1/4"
Clear
1/4" Clear
1/4" Clear
1/4" Clear
1/4" Clear
1/8"
Wht
Flush Valve Pilot
Air (Control Panel)
Handpiece Coolant
Water Pilot Air
(Cabinet Base)
Check
Valve
CA
CW
DA
1/8"
Blue
1/8"
Blue
1/8"
Wht.
1/8"
Blue
1/4" Clear
1/8"
Clear
Exhaust Air
Exhaust
Air
1/8"
Blue
1/8"
Blue
1/8"
Wht
CA
CW
DA
1/4"
Clear
1/4"
Grey
Kink
Valve
1/4" Clear
1/8"
Purple
1/8"
Wht.
1/8" Orange
CA
CW
DA
1/4"
Clear
1/4"
Grey
1/4" Clear
1/8"
Wht.
1/8"
Purple
1/8"
Blue
1/8" Orange
1/4"
Clear
1/8"
Grn
Exhaust Air
Exhaust Air
Exhaust
Air
1/8"
Clear
1/8"
Wht
1/8"
Blue
1/8" Blue
1/8" Green
Manifold
(Air)
1/8"
Blue
1/8"
Grn
Manifold
(Water)
1/4"
Clear
1/8" Blue
1/8"
Wht
1/4" Clear
1/4"
Clear
1/4"
Clear
1/4"
Clear
1/8"
Red
1/8"
Purple
1/8"
Clear
1/8" Clear
1/8" Clear
Syringe
Supply Air
Instrument
Water Supply
Handpiece
Drive Air
Doctor's Syringe
Master
On/Off
Valve
Handpiece
Coolant Air
Supply
1/8 Red
1/8 Purple
Bleed
Air to Master
Switch
Pilot Air From
Master Switch
OFF
ON
003-2793-00
Style G
midmark.com
Color Page
English-10
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
003-2793-00
Style G
midmark.com
Color Page
English-11
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
003-2793-00
Style G
midmark.com
Color Page
English-12
TP201 Rev. A
© 2014 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
DA2474-1
153839 (-001 thru -004)
Water Bottle Assembly
1/8 Blue
City
Bottle
Water
Selector
Valve
30 PSI
Bottle
Water
City
Bottle
1/8 Brown
30 PSI
Air
1/8 Blue
1/8 Blue
Instrument
Air Supply
Instruments
Water Supply
30 PSI City Water
To / From
Air - Water
Junction
(located in
cabinet
base or
control
panel)
80 PSI Air
80 PSI Syringe Air
1/8
Clear
1/8 Brown
1/8 Yellow
1/8
Brown
Ribbed
Side
30
PSI
Bottle
Water
Instrument Water Supply
Out
Push
Flush Valve
(N.C.) 2-Way
Manual
Momentary
1/8 Orange
1/8 Yellow
1/8 Orange
1/8 Yellow
1/8 Blue
Flush Valve Pilot Air
(Control Panel)
Syringe Supply Air
Double
Pinch
Valve
Rib on Air Side
(Yellow Tubing)
Water
Air
1/8
Blue
To Assistant's
Instrument
Syringe
1/8
Blue
1/8
Yellow
1/8
Blue
In-Line
Regulator
003-2793-00
Color Page
Style G
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12

Midmark Procenter Systems Installation guide

Type
Installation guide

Midmark Procenter Systems offer a comprehensive solution for dental professionals seeking advanced and efficient dental delivery. These systems are designed with meticulous attention to detail, providing a wide array of features to enhance patient care and streamline dental procedures. With the ability to accommodate various delivery options, including handpieces, syringes, and water bottles, these systems adapt seamlessly to different clinical needs.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI