Midmark VetPro 5000 Mounted Dental Delivery System Installation guide

Category
Sanitary ware
Type
Installation guide
1
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
003-1904-00 Rev J (11/11/15)
VetPro
®
5000 Wall / Cabinet Mount Installation
Applies to Models:
8001-001
8001-002
8001-005
8001-006
WARNING
Equipmentnotsuitableforuseinthepresenceofaammableanestheticmixture.
Note
Use clamps and fittings for tubing connections shown in
VetPro Solo and or VetPro 5000 Junction box installation
instructions. Clamp sleeves are only needed if you need to
shorten the water line and lose your existing clamp sleeves.
Water Bottles (2)
Grease (1)
P80 (PN SMSC-0031)
Dowel Pin (1)
Clamp Sleeves (5)
Barb Fittings (2)
Uni-Clamps (20)
Instruments and Accessories
May Vary Depending on Model
MaxTensionSprings
Loose Components
Bearing Plate and Cover
NextNext
BackBack
midmark.com
Style A
2
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
SuggestedMountingLocationFromProcedureTable
DeliveryUnitcanbeinstalledonleftorrightsideoftableorcabinet.
180° Arm Length
Standard - 28”
Short - 13 1/2”
250°
Arm Length
Standard-45.3”
Short - 37 5/16”
Mounting Requirements
MountingStructureNote:Themountingstructuremustbecapableofsupportingtheweightandtorquelistedbelowwithoutdeection.
Unit / Arm Assembly Maximum
Weight 50lbs (22.6 kgs)
Torque 170.8 ft. lbs. (231.5 N•M)
Note: To prevent unit from drifting, the mounting plate must deect no more than 1/16” (1.6 mm) over a 12” (30.5) distance.
Cabinet or Wall
Cabinet or Wall
Note
WHIP head folded against wall measures 30.6” from center of main axis to outermost point.
Space Requirements
NextNext
BackBack
midmark.com
3
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Range of Motion
NextNext
BackBack
midmark.com
4
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
VA1327
Cabinet Mount - WallMountStartsOnNextPage
Note
Delivery Unit can be installed on left or right side of cabinet.
The bearing plate can be mounted with the needle bearing at
the top or bottom. It is recommended that the needle bearing is
positioned at the Top.
If user desires, hang plastic sleeve for tri-fold quick reference
with training CD on side of cabinet next to bearing plate for
easy access.
Continue Cabinet Mount with Step 3
Step2:ForsteelcabinetsONLY.
A) Drill two # 21 (.221) holes in the side of the cabinet for the two
bottom backer plate screws. This is not required when steel cabinets
are supplied with a wood end panels.
Bearing Plate
Shim
Level
Needle Bearing
Equipment Alert
The cabinet must be fastened to a wall stud similar to the wall mount
installation. Failure to do that will cause cabinet to tip over.
Step1:Mounttocabinet.
A) Screw cabinet to wall studs on delivery mount side as close to corner as possible.
B) Mark and drill top two 5/16” mounting holes through bearing plate and cabinet.
C) Align mounting holes and install appropriate bolts (1” or 2 1/4”) in top two holes.
Repeat(B)and(C)forlowertwoholesonbearingplate.
D) Loosen all four bolts and level bearing mounting plate.
Use shims if necessary, then tighten bolts.
Mount Cabinet
to Wall
Backer
Plate
Drill Bottom Holes on
Steel Cabinets Only
Chase
Note
Refer to chase installation manual for mounting hole locations in chase cabinet.
To install VetPro 5000, use hardware included with chase cabinet. (1/4” screws, washers and nuts)
Chase
Cabinets
NextNext
BackBack
midmark.com
5
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Wall
Studs
Mounting
Cover
Mounting
Plate
Wall Mount
Note
The bearing plate can be mounted with the needle bearing at the top or bottom. It is recommended that the needle bearing is positioned at the Top.
If user desires, hang plastic sleeve for tri-fold quick reference with training CD on side of wall next to bearing plate for easy access.
Continue Wall Mount with Step 3
Step2:Installbearingplate.
A) Align mounting holes in bearing plate and frame cover.
B) Install four bolts.
C) Loosen bolts and level bearing mounting plate. Use shims,
if necessary, then tighten bolts.
Bearing
Plate
Shim
Needle
Bearing
Equipment Alert
Mounting holes MUST be aligned with wall studs. Failure to comply
could cause delivery unit to fall damaging unit and wall structure.
Step1:Mounttowall.
A) Locate wall studs and align mounting plate with studs.
B) Level then mark and drill 1/4” pilot holes (36” top to floor).
C) Secure frame to wall with (8) (5/16” x 2 1/2”) lag screws.
D) Install mounting cover using (4) (#10-32 x 3/8”) screws.
Level
NextNext
BackBack
midmark.com
6
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step3:Installunitarm.
A) Loosen set screw on bearing.
B) Apply grease (supplied) to fixed arm.
C) Route umbilical tubing thru bearing and slide fixed arm into bearing.
D) Size and trim umbilical sleeve length to Jbox or Solo unit. Slide sleeve
over tubing & wiring.
E) Install and secure umbilical retainer.
Umbilical Sleeve
Retainer
Apply Grease
Set Screw
NextNext
BackBack
midmark.com
7
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
VA1719
Adjustment
Screw
Step4:Adjustaretension,ifnecessary.
Adjust arms at pivot points using a 3/32 allen wrench (5 places).
• If arms have too much play, tighten them (Clockwise).
• If they do not move freely, loosen them (Counter-Clockwise).
Adjustment
Screws
Adjustment Screw
(ThisLocationOnlyUse1/8”AllenWrench)
Adjustment
Screw
NextNext
BackBack
midmark.com
8
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
6
Step5:Installbearingcoverandwaterbottle.
A) Install bearing cover.
B) Install water bottle.
Water Bottle
Bearing Cover
CABINETINSTALLSONLY!
Equipment Alert
Installing hard stop bracket on cabinet installations will prevent
flex arm from swinging into cabinet and causing damage to unit.
Step6:Installhardstopbracket.
A) Remove cap screw.
B) Install bracket.
C) Install and tighten screw.
Hard Stop Bracket
NextNext
BackBack
midmark.com
9
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Wall or Cabinet
Equipment Alert
Stop pins prevent arm from turning 360 degrees which twists and kink tubing. You may install one or two stop pins. Verify desired arm turning positions with customer.
Position may be changed by moving stop pins.
Step7:Installstoppins.
A) Position arm at 90 degree angle based on user preferred direction swing, right-hand swing or left-hand swing.
B) Lift trim ring located at bend on arm, closest to wall / cabinet.
C) Rotate arm to desired position and install stop pins in rear holes.
Trim
Ring
Stop Pin Locations
Rear Only
NextNext
BackBack
midmark.com
10
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step8:InstallJunctionBoxAssemblyORVetPro
®
Solo.
Refer to: Junction Box Installation Instructions (Document 003-1914-00 included with Junction Box) OR VetPro
®
Solo Installation Instructions (Document # 003-
1996-00 included with VetPro
®
Solo)
ContinuewithStep9afterinstallationofJunctionBoxorVetPro
®
Solo
Step9:Activateflexarmlock.
A) Remove end cap.
B) Slide cover back to access locking pawls.
C) Cut and remove wire tie.
D) Adjust jam nut if too much “drift” in the deliver head (if applicable).
E) Position arm going downward toward deliver head to slide cover in position and install end cap.
Locking Pawls
WireTie
End Cap
Jam Nut
Toadjustflexarmdrift...
A) Tighten jam nut clockwise if flex arm drifts DOWN.
B) Loosen jam nut counter clockwise if flex arm drifts UP.
NextNext
BackBack
midmark.com
11
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step10:Installinstrumenttubing.
A) Remove packaging from dental delivery head.
B) Move pivot assembly forward and remove set screw from pivot.
C) Slide arm into pivot on back of delivery head.
D) Tighten set-screw with 3/32” allen wrench.
E) Repeat with each arm.
Set-Screw
Arm
Arm Installation
NextNext
BackBack
midmark.com
12
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step11:Installprophyangle.
A) Rotate chuck release ring to the right to unlock.
B) Install Prophy angle, turn chuck release ring to
the left to tighten.
Step12:Installburinhighspeedhandpiece.
A) Push bur into handpiece. (The shank will only go in half-way)
B) Then push button on back of handpiece and push the bur in until shank is all the way in.
Note: Friction grip burs (FG Style) are designed so that the entire straight portion of the shaft should be inside the chuck
of the handpiece. You should only see the tapered portion of the bur if installed properly.
Push Bur Into Handpiece
Push Button
Finish Sliding Bur Into Handpiece
ThisPortion
of the Bur
Should be
Inside the
Handpiece
Equipment Alert
Handpieces will not work properly if not installed as shown.
Shaft Inside the Handpiece
Prophy
Angle
NextNext
BackBack
midmark.com
13
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step13:Installinstruments,adjusttubing.
A) Install instruments.
B) Adjust tubing length by sliding tube. (on scaler slide grommet).
C) Change arm spring if arm moves back and forth too freely. Maximum
tension springs are included with loose parts.
Handpiece
Connection
Tochangearmtensionspring
(ifmoretensiononarmiswanted)...
A) Remove tool pad cover and upper shroud.
B) Disconnect from pivot pin with pliers.
C) Remove allen screw with 7/64 allen wrench.
D) Replace spring and tighten allen screw.
E) Install upper shroud and tool pad.
Tension
Spring
Allen
Screw
Pivot Pin
ToolPad
Upper Shroud
Arm Adjustment
Swivel Coupler
Connection
Lightly spray the exposed o-rings
with one step conditioner.
Tighten coupler firmly
with wrench.
Line up components properly to prevent
damage resulting in hand piece failure.
NextNext
BackBack
midmark.com
14
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Note
The air adjustment valve that corresponds to the ultrasonic scaler
must be OPEN in order to operate the scaler with the foot pedal.
Step15:Checkhandpiecewaterflow.
A) Lift and pull handpiece forward.
B) Toggle foot control to Wet position. Depress foot control.
C) Regulate water pressure by adjusting the corresponding knobs.
D) Repeat for each handpiece that uses water.
Pressure Gauge
[Drive Air]
Coolant Water
Scaler
Adjustment
Syringe Water
Drive Air
Coolant Air
Front View
Bottom View
Syringe Air
Pressure
Coolant Water
Flush Button
Air
Water
Step14:Checkhandpieceairflow.
A) Lift and pull handpiece forward.
B) Toggle foot control to Dry. Depress foot control.
C) Regulate air pressure by adjusting the corresponding knobs.
D) Repeat for each handpiece.
NextNext
BackBack
midmark.com
15
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
InstallationTroubleshooting
Problem Probable Cause Check Correction
Nowatertothesystem.
City water and delivery head supply
tube installed incorrectly.
Verify City water is connected to the
ribbed portion of the tube.
Correct tubing.
Watertosyringebutnothingelse.
Water toggle is toggled to dry. Verify Toggle is towards the blue dot. Redirect toggle.
Water control valves are closed. Verify water control knob is turned
on.
Turn on water supply.
Waterisleakingfromthelowspeedinstrument.
Water supply is turned on to low
speed.
Verify water is turned off to low
speed.
Turn water off.
Fiberopticlightonhighspeedandscalerarenotworking.
GFI was tripped. Verify green light on GFI is on. Reset GFI.
Scalerhaswaterbutdoesnotvibrate.
Transformer or universal power supply
is not plugged in.
Verify power supply is plugged in. Connect scaler to power supply.
Inadequate PSI. Verify PSI is reaching mid 50’s to mid
60’s.
Adjust PSI
Tip is not tight. Verify tip is tight. Tighten tip with scaling tip wrench.
Frequency setting is not set correctly. Verify dial corresponds with tip color. Adjust dial.
Fiberopticlightisnotworkingonhighspeed.
High speed is installed on the wrong
tubing.
Verify high speed is installed on the
ber optic tubing.
Reinstall high speed.
Multiplehandpiecesrunatthesametime.
On / Off switch to the delivery head is
turned off.
Verify switch position. Turn switch on.
Actuator arm was fallen down on one
or more handpiece positions.
Verify actuator arm position. Correct positioning
Airleakingfromfootcontroler.
Foot controller tubes were routed
incorrectly.
Verify dashed line connects to Solo
or Junction Box. Verify brown tube is
not being used.
Correct tubing.
Deliveryheadwillnotlock.
Twist tie is still in place on the locking
claws inside ex arm.
Verify twist tie is removed. Remove tie.
JunctionBoxonly:Nocitywater.
City water regulator is turned down too
low.
Verify regulator is set to 30. Adjust regulator.
NextNext
BackBack
midmark.com
16
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
PostInstallation&TrainingChecklist
Check Verify the following:
Delivery Head
1 Delivery unit comes ON when master switch is turned ON. Verify user knows switch location.
2 Delivery head locks, unlocks and doesn’t drift. (Must have air supply to delivery head to function.)
3 Flex arm and delivery unit rotate easily and stay where placed.
4 Flex arm and delivery unit are level. Stop pins are in place.
5 User knows the coolant water ush button, potentiometer, and water control valves. Note:Keepwatershutofftolowspeedatalltimes.
6
Pressing foot control delivers air/water to handpieces as expected. Verify user knows the function of the wet/dry toggle on foot control.
(Effects water ow to low speed, high speed, and scaler only.)
7 Review quick reference guide with user, making sure they understand procedures for deliver head.
8
User knows prophy angle, burr, scaling tip and syringe installation and removal. Review using Midmark conditioner for instruments, Note: Burrs
shouldbechangedandinstrumentsconditionedaftereachpatient.Scalingtipsshouldbeoperatedwithinthe
correspondingcoloronthepotentiometerdial.
9 Wipe all components with a soft dry cloth to remove dirt and smudges.
Junction Box (if applicable)
10 Water and air supply regulator gauges in J-box read appropriate pressures. 32-40 for the high speed, 35-50 for the low speed, full ow for scaler.
Solo (if applicable)
11 Solo unit On/Off switch works and user is aware of switch location.
12 Drain valve is working properly. User can locate city water shut off valve.
13 User is aware of GFI (used with ber optics and scaler) plug on outside of Solo unit and plug on the inside of Solo unit.
14 Installers Signature_____________________________________________________ Date _________________________________
15 Users Signature________________________________________________________Date___________________________________
NextNext
BackBack
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16

Midmark VetPro 5000 Mounted Dental Delivery System Installation guide

Category
Sanitary ware
Type
Installation guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI