Midmark LED Dental Light Installation guide

Category
Wall & ceiling mounts accessories
Type
Installation guide

This manual is also suitable for

Midmark LED Dental Light offers precise illumination and versatility for various dental procedures. With its articulating arm and adjustable light intensity, it provides optimal visibility and flexibility during examinations and treatments. The LED technology ensures long-lasting, energy-efficient operation, reducing maintenance and replacement costs. Its ergonomic design allows for easy positioning and adjustment, enhancing comfort for both the dentist and the patient.

Midmark LED Dental Light offers precise illumination and versatility for various dental procedures. With its articulating arm and adjustable light intensity, it provides optimal visibility and flexibility during examinations and treatments. The LED technology ensures long-lasting, energy-efficient operation, reducing maintenance and replacement costs. Its ergonomic design allows for easy positioning and adjustment, enhancing comfort for both the dentist and the patient.

English-1
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Midmark
®
Cabinet / Wall Light Installation
Applies to Models:
153908 (-002)
153909 (-002)
153640 (-001)
153641 (-001)
Language of origin: English
WARNING
Equipment is not suitable for use in the
presence of a flammable anesthetic mixture.
003-1239-00 Rev. R (6/21/13)
Physical Requirements
Floor
[Full size template is available: 003-1289-00]
Note
Unit is shipped with a power cord and plug assembly. Where installation requires
a power outlet, the outlet must be within 12 feet of wall or cabinet mounting.
Equipment Alert
Before installing the Midmark Cabinet / Wall Light, ensure
the physical room requirements outlined are met.
93.5 in.
(237.5 cm)
Note
Recommended height may vary due
to 6 in. (15.2 cm) min. distance
required for installation or personal
preferences. All dimensions are based
on a min. ceiling height of 8 ft. (244 cm).
For every one inch below 8 ft. drop the
mounting one inch.
17.5 in.
(44.5 cm)
8.125 in.
(20.6 cm)
16 in. on center
(40.6 cm)
6.0 in.
(15.2 cm)
360° AXIS F
(constrained to
180° by wall)
56.0 in.
(142.2 cm)
5.5 in.
(14.0 cm)
18.0 in.
(45.7 cm)
Bottom of
Mounting Bracket
62 5/16 in.
(158.2 cm)
18.8 in.
(47.8 cm)
16.0 in.
(40.6 cm)
74.2 in.
(188.5 cm)
75.0 in.
(190.5 cm)
280°
AXIS E
AXIS D
(does not rotate)
14.7 in.
(37.3 cm)
6.0 in.
(15.2 cm)
96.0 in.
(243.8 cm)
Floor
Ceiling
Halogen LED
NOTE: This install guide applies to Halogen and LED-style lights.
Style A
midmark.com
English-2
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
____
Physical Requirements...continued
3.75 in.
(9.53 cm)
Equipment Alert
Before installing the Midmark Cabinet / Wall Light, ensure
the physical room requirements outlined are met.
Note
Recommended min. 8 ft. (244 cm.)
ceiling height due to the height of the
central free standing cabinet and the
6 in. (15.2 cm) min. distance required
between the light arm and the ceiling
for installation.
Center Column
Installation
4.625 in.
(11.75 cm)
8.125 in.
(20.6 cm)
Bottom of Mounting Bracket
6.0 in.
(15.2 cm)
360°
AXIS F
56.0 in.
(142.2 cm)
93.5 in.
(237.5 cm)
5.5 in.
(14.0 cm)
Floor Floor
Ceiling
62 5/16 in.
(158.2 cm)
7.6 in.
(19.3 cm)
Center
Column
70.1 in.
(178.1 cm)
75.0 in.
(190.5 cm)
280°
AXIS E
14.7 in.
(37.3 cm)
6.0 in.
(15.2 cm)
AXIS D
(does not rotate)
96.0 in.
(243.8 cm)
Style A
midmark.com
English-3
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
____
General Movement Characteristics
The support structure must be capable of supporting:
System (only)
Weight (Max.): .............................................66 lbs (29.9 kgs)
Torque (Max.): ................................... 170 ft-lbs. (230.5 N
·
m)
In addition to supporting the above listed weights, the
support structure must also be able to support the light
without deection. To prevent the suspension arm from
drifting, the assembly must deect no more than 1/16 in.
(1.6 mm) over a 12 in. (30.5 cm) distance.
General Information
AXIS E
180°
(no stops)
56.0 in.
(142.2 cm)
AXIS D
300°
AXIS C
50°
26.0 in.
(66.0 cm)
AXIS B
300°
AXIS A
180°
24.0 in.
(61.0 cm)
Style A
midmark.com
English-4
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
____
Wall Mount Installation
Step 1: Using template located on shipping box,
place in position, level, and mark wall
(outside) mount backer holes.
The wall mount assembly must be located so the operation
of the light and arm assembly does not interfere with
procedures or patient access. Power source must be within
12 feet of mounting location.
Location
Structure
The wall mount backer plate must be leveled and secured
with appropriate hardware to vertical wall studs or other
substantial support structure capable of supporting the
total weight of the light and arm assembly (refer to General
Information).
Power Source
Light may be plugged into an appropriate wall outlet or wired
to a suitable power source. The power source must be
grounded, and protected by a 15 amp circuit breaker for 115
VAC operation or a 7.5 amp breaker for 230 VAC operation.
All wiring must be applicable to state and local codes.
Step 2: Remove transformer cover, four screws.
Step 3: Slide wall mount cover
from light mount assembly.
Step 4: Remove light mount assembly from
wall mount backer, four screws.
* Recommended height may vary due to 6 in. (15.2 cm) min. distance required for installation or personal preferences.
All dimensions are based on a min. ceiling height of 8 ft. (244 cm). For every 1” (2.54 cm) below 8 ft. drop the mounting 1” .
16 in. (40.6 cm)
on Center
62 5/16 in.
(158.2 cm)
To Floor*
Transformer
Cover
Wall Mount
Cover
Light Mount
Assembly
Wall Mount
Backer
Wall Mount
Backer
Chase
Note
Refer to chase installation manual for mounting hole locations.
To install light, use hardware included with chase cabinet.
(3/8” bolts, washers and nuts)
Style A
midmark.com
English-5
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
____
Wall Mount Installation...continued
* Recommended height may vary due
to 6 in. (15.2 cm) min. distance
required for installation or personal
preferences. All dimensions are based
on a min. ceiling height of 8 ft. (244 cm).
For every one inch below 8 ft. drop
the mounting one inch.
16 in. (40.6 cm)
on Center
Note
If supply voltage is 230 VAC refer to ADJUSTING
TRANSFORMER FOR 230 VAC.
Equipment Alert
Failure to secure to a substantial structure
with appropriate hardware capable of
supporting the light assembly could result
in equipment malfunctioning and damage.
Transformer
Cover
Wall Mount
Cover
Light Mount
Assembly
Wall Mount
Backer
Step 5: Align mounting holes of wall mount backer
with location holes on wall and secure with
appropriate hardware. Check to assure
wall mount backer is level.
Step 6: Position light mount assembly on wall mount
backer and secure with four screws.
Step 7: Place wall mount cover in position
on light mount assembly.
Step 8: Refer to Light Head and Arm
Installation before installing
transformer cover.
62 5/16 in.
(158.2 cm)
To Floor*
Style A
midmark.com
English-6
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
____
Cabinet Mount Installation
Step 1: Use existing holes from previous
light or use template located on
shipping box. Place template in
position, level, and mark cabinet
mounting holes.
The cabinet mount assembly must be located so the
operation of the light and arm assembly does not interfere
with procedures or patient access. Power source must be
within 12 feet of mounting location.
Location
Structure
The cabinet mount must be leveled and secured to cabinet
structure with appropriate hardware (1/4”-20, grade 8, socket
head cap screws) capable of supporting the total weight of
the light and arm assembly (refer to General Information).
Power Source
Light may be plugged into an appropriate wall outlet or wired
to a suitable power source. The power source must be
grounded, and protected by a 15 amp circuit breaker for 115
VAC operation or a 7.5 amp breaker for 230 VAC operation.
All wiring must be applicable to state and local codes.
Note
There are two sets of holes on the template for cabinet mounting.
Inside holes are for drilling clearance holes when installing screws
from the front into threaded holes or when using washers and nuts.
The outer holes are when installing 1/4”-20 screws through cabinet
into mounting plate.
Equipment Alert
Failure to secure to a substantial structure
capable of supporting the light assembly
could result in equipment malfunctioning and
damage. Mounting hardware must be 1/4 in.-20,
grade 8, socket head cap screws and lockwashers.
Step 2: Align mounting holes of mounting bracket
with location holes on cabinet and secure
with appropriate hardware. Check to assure
mounting bracket is level.
Note
If supply voltage is 230 VAC refer to
Adjusting Transformer for 230 VAC.
Step 3: Refer to Light Head
and Arm Installation
before installing post
mount cover.
Template
Inside
Holes
Outer
Holes
Outer
Holes
1/4” Nylock nuts can be
used when installation
is on products other
than Midmark.
Mounting
Bracket
Style A
midmark.com
English-7
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
____
Adjusting Transformer for 230 VAC (Halogen ONLY)
Step 1: Remove transformer cover, one screw.
Post Mount
Cover
Note
If supply voltage is 230 VAC, the following procedure
must be performed before connecting power to light.
Note
Cabinet units may not have the post mount cover.
Transformer
Cover
Brown
Lead
Step 2: Disconnect brown lead from 115 V tap
on transformer and connect to 230 V.
Step 3: Install transformer cover and, where
applicable, post mount cover.
Style A
midmark.com
English-8
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
____
Installing Plug to Power Cord
Note
If facility voltage is 230 VAC, install a properly rated
plug according to local codes and regulations.
Adjust transformer for 230 VAC operation.
Step 2:
A) Insert BROWN wire from power cord
into hole (A) & tighten GOLD screw.
B) Insert BLUE wire from power cord
into hole (B) & tighten SILVER screw.
C) Insert GREEN wire from power cord
into hole (C) & tighten GREEN screw.
Hole
(B)
Hole
(A)
Hole
(C)
Brown
Wire
Green
Wire
Blue
Wire
Silver
Screw
Gold
Screw
Green
Screw
Step 3:
A) Rotate three pronged portion of plug to desired
position in housing & secure with screw.
B) Tighten two screws so that power cord is
secured. Do not overtighten screws.
Note
The three pronged portion of plug may
be installed in four different positions.
Be sure two tabs on pronged portion
align with notches in housing.
Step 4: Plug power cord into properly
rated receptacle & check for
proper operation.
Step 1:
A) Remove screw as shown & three
pronged portion of plug.
B) Loosen two screws as shown &
insert power cord into housing.
Style A
midmark.com
English-9
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
____
Arm and Light Head Installation
Step 1: Place fixed arm assembly in position
& connecting harness plug (A) from
fixed arm to plug (B) on cabinet / wall
mount assembly.
Step 2: Insert fixed arm into bearing housing.
Step 3: Insert collar onto fixed arm, snugging
up set screw to hold it in place.
Step 4: Connect harness plug (C) from fixed arm
to connector (D) in flex arm assembly.
Fixed Arm
Assembly
A
B
Bearing
Housing
Collar
C
D
Flex
Arm
Note
Assure the collar is installed with the
set screw toward the bottom. Do Not
locate the set screw over the key.
Key
Style A
midmark.com
English-10
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
____
Arm and Light Head Installation...continued
Step 5: Insert end of flex arm into end of fixed
arm. Install the key, then move collar
down over the key & tighten set screw.
Note: The set screw cannot be located over key.
Step 6: Check level of fixed arm and,
if necessary loosen or tighten
four leveling screws.
Step 7: Tighten mount screws
to secure position.
Step 8: On wall mount units,
install transformer cover.
Step 9: Connect unit to the
appropriate electrical supply.
Fixed Arm
Assembly
Collar
Flex
Arm
Key
Leveling
Screws
Mount
Screws
Transformer
Cover
Note
Units are equipped with a separate 115 VAC plug for
installation on the cord. The unit can also be direct
wired to an appropriate circuit. Should a 230 VAC
plug be desired it must be purchased locally.
Equipment Alert
To ensure quiet operation, apply a thin coat of
lubricant (included w/light) to inside of fixed arm
and bearing shaft of flex arm assembly.
Style A
midmark.com
English-11
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Tension Adjustments
To adjust Flex Arm tension...
A) Remove end cap and top cover as shown.
B) Loosen or tighten adjustment nut as desired.
To adjust tension at Arm / Yoke pivot joints...
A) Remove neck cover as shown (see Note).
B) Loosen or tighten appropriate screws to adjust tension.
Note: Some halogen models have a short neck on the lighthead. It is not
necessary to remove the neck cover on these models.
To adjust tension at Light pivot joint...
A) Remove screw cover as shown.
B) Loosen set screw.
C) Loosen or tighten light pivot adjustment screw.
D) Tighten set screw.
Short Neck
Yoke Pivot
Adjustment Screw
Arm Pivot
Adjustment Screw
Adjustment Nut
Yoke Pivot
Adjustment Screws
Arm Pivot
Adjustment Screws
Light Pivot
Adjustment Screw
Set Screw
Neck Cover
(Long Neck)
Style A
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11

Midmark LED Dental Light Installation guide

Category
Wall & ceiling mounts accessories
Type
Installation guide
This manual is also suitable for

Midmark LED Dental Light offers precise illumination and versatility for various dental procedures. With its articulating arm and adjustable light intensity, it provides optimal visibility and flexibility during examinations and treatments. The LED technology ensures long-lasting, energy-efficient operation, reducing maintenance and replacement costs. Its ergonomic design allows for easy positioning and adjustment, enhancing comfort for both the dentist and the patient.