0171--976 5
The calibration is done as follows:
Note! Make sure that the Unisab II is not in alarm.
1. Apply supply voltage minimum 5 minutes before calibration.
2. Press the calibration button for 5 seconds to get the transmitter in calibration
mode. The red LED changes from normal flash to OFF.
3. Decrease the capacity to its minimum, and then press the calibration button
once. The red LED turns immedially ON. After 10 seconds the red LED turns
OFF again, indicating that it is ready for 100% calibration.
4. Increase the capacity to its maximum, and then push the calibration button
twice. The red LED starts flashing quickly. After 10 seconds the red LED
changes to normal flash rate, indicating thatthe calibration has been completed.
5. The calibration will be remembered, even if a power failure occurs.
To return to “Factory Set Point”, press the CAL. Switch for 20 seconds.
Indication of this is shown by a short flash of the green and red led (one after the other).
The 4--20 mA signal from the transmitter is permanently available, however, during calibration
the signal is based on the default calibration values, so it will not show 0% at minimum capacity
nor 100% at maximum capacity! Note that it is possible to reset to the default calibration by
pressing the button pressed for 20 seconds, however, it is the transmitter manufacturer’s de-
fault values, which might differ very much from the proper values for the actual compressor.
The reset is indicated by the red and the green LED flashing briefly in turn a couple of times.