Steris Harmony La Surgical Lighting And Visualization System Operating instructions

Type
Operating instructions
OPERATOR MANUAL
Harmony® Lux Advantage LA
Surgical Lighting and Visualization System
(08/10/11) P129382-106
i
129382-106
A WORD FROM STERIS CORPORATION
©2011, STERIS Corporation. All rights reserved. Printed in U.S.A.
This manual contains important information on proper use and care of this
surgical lighting system. All operators and department heads are urged to
carefully review and become familiar with the warnings, cautions, and
instructions contained herein. Your new surgical lighting fixture features an
advanced, state-of-the-art design, with cool, shadow-reduced light, and
ease of maneuverability. It produces light of a quality necessary for the most
demanding and complex of surgical procedures.
A thorough preventive maintenance program is essential for safe and proper
operation of your surgical light. You are encouraged to contact STERIS
concerning our Annual Maintenance Agreement. Under the terms of this
agreement, preventive maintenance, adjustments and replacement of worn
parts are done on a scheduled basis to verify lighting fixture performance
according to its specifications and to help avoid untimely or costly downtime.
STERIS maintains a nationwide staff of well-equipped, factory-trained
technicians to provide this service, as well as expert repair services. Contact
STERIS for details.
The Harmony® Lux Advantage LA Surgical Lighting and Visualization
System is a variable pattern, variable intensity surgical lighting fixture
designed to provide visible illumination of the surgical field or the patient for
the operating room staff.
This product may contain communication devices. A medical image
communications device provides electronic transfer of medical image data
between medical devices. It may include a physical communications
medium, modems, interfaces, and a communications protocol.
A summary of the safety precautions to be observed when operating and
servicing this equipment can be found in SECTION 1, SAFETY PRECAUTIONS.
Do not operate or service the equipment until you have become familiar with
this information.
Any alteration of the lighting system not authorized or performed by STERIS
Engineering Service which could affect its operation will void the warranty,
could violate national, state and local regulations and jeopardize your
insurance coverage.
The following is an important message from STERIS about the
advantages and limitations associated with the use of high intensity
surgical lighting systems.
Because of the variety of surgical procedures performed and the wide range
of individual preferences of surgical staffs, it is desirable that a surgical
lighting system be capable of selective control across a wide range of
illumination intensities. The Illuminating Engineering Society (IES) stresses
that in addition to providing control of intensity, surgical lighting systems
should provide shadow control, correct color rendition, and a suitable depth
of field to provide sharp, consistent lighting into deep body cavities. As
illumination levels increase, however, radiant heat also increases. Therefore,
the IES cautions that for most operations, radiant heat should be kept to a
minimum. The user of surgical lights should utilize the lowest possible
Indications for Use
Advisory
ii
129382-106
Manufactured by:
STERIS Corporation
2720 Gunter Park East
Montgomery, AL 36109 • USA
Class 1 Equipment
Type B Equipment
Ordinary Equipment (enclosed equipment without protection of
ingress of water)
Equipment not suitable for use in the presence of a flammable anesthetic
mixture with air or oxygen or nitrous oxide.
Suitable for continuous operation.
Equipment provided and incorporated by STERIS Corporation into
complete Optical Fiber Communication System (OFCS) satisfies applicable
laser safety requirements of the IEC 60825-1 (Class 1 Laser Product) and
IEC 60825-2 (Hazard Level 1M for OFCS).
STERIS Ltd.
Chancery House
190 Waterside Road
Hamilton Industrial Park
Leicester
LE5 1QZ
United Kingdom
illumination level suitable for the procedure, especially in certain neurological
or intestinal procedures on delicate, thin, dry, or abnormal tissue.
Furthermore, for the protection of surgically exposed tissues and for the
comfort and efficiency of the surgeon and assistants, radiant energy can be
effectively controlled by limiting the time of exposure at higher illumination
levels. Extra care must be taken when the light fields from multiple lightheads
are overlapped on the surgical site, since this condition creates a risk of too
much heat.
An international standard for the safety of surgical lights established by the
IEC sets minimum and maximum levels of illumination and maximum levels
of radiant heat that can be emitted from a single surgical luminaire. The LA
Surgical Lighting and Visualization System has been designed to comply
with this international standard and to provide a wide range of illumination
levels while minimizing the potentially damaging infrared heat in the surgical
field.
The illumination level of the Harmony LA Surgical Lights can be adjusted
through several intensity settings via conveniently located controls on either
the wall-mounted control center or the lighthead handle. The illumination
level also decreases as the pattern size increases. Maximum illuminance
can exceed 12,000 fc for the smallest pattern size of the Harmony LA 500
or exceed 14,500 fc for the smallest pattern size of the Harmony LA 700
lighthead, and can be adjusted by intensity control or pattern size control
throughout the entire range specified by the IEC.
STERIS Corporation,
Montgomery, Alabama is an
ISO 13485 certified facility.
The base language of this document is
ENGLISH. Any translations must be made
from the base language document.
iii
129382-106
TABLE OF CONTENTS
Section Title Page
1 Listing Of Safety Precautions ......................................................... 1-1
1.1 Definition of Symbols................................................................................................... 1-5
2 Installation Verification .................................................................. 2-1
2.1 Pre-operation Checklist .............................................................................................. 2-1
2.1.1 Check Suspension Movement .......................................................................... 2-2
2.1.2 Harmony LA Fiber Optic Video Enabled System ............................................... 2-2
2.1.3 Check System Operation .................................................................................. 2-4
2.1.4 Check Optional Video Camera Operation ........................................................ 2-5
2.1.5 Check Optional Monitor Support Arms ............................................................. 2-7
2.1.6 Check Optional DeepSite Lighthead ................................................................ 2-8
2.1.7 Check Optional DeepSite Ceiling-mounted Illuminator Housing ....................... 2-9
2.2 Harmony LA Optical Performance ............................................................................ 2-10
3 Operating Instructions .................................................................... 3-1
3.1 Intensity Controls ........................................................................................................ 3-1
3.2 Comparison of Lighthead Features ............................................................................ 3-3
3.3 Harmony LA 500 and 700 Lightheads Lamp Failure Indications................................ 3-4
3.4 Lighthead Positioning ................................................................................................. 3-4
3.5 Harmony LA 500 and 700 Lighthead Pattern Adjustment ......................................... 3-6
3.6 Lighthandle and Lighthandle Cover ........................................................................... 3-7
3.6.1 Installing and Using Lighthandle and Lighthandle Cover ................................. 3-7
3.6.2 Metal Sterilizable Handle .................................................................................. 3-8
3.7 Harmony LA Monitor Arms ......................................................................................... 3-8
3.8 Install Flat Panel Monitor Yoke Handle Disposable Sterile Covers ........................... 3-11
3.9 Video Camera Installation or Removal ...................................................................... 3-13
3.10 Install Camera Disposable Sterile Cover ................................................................ 3-13
3.11 Video Camera Operation ........................................................................................ 3-14
3.12 Guidelines for Maximizing Video Image ................................................................. 3-19
3.13 System Standby...................................................................................................... 3-19
3.14 DeepSite Control Center Status Indicators ............................................................. 3-20
3.15 DeepSite Lighthead Positioning ............................................................................. 3-20
3.15 DeepSite Pattern Adjustment .................................................................................. 3-21
3.17 Disposable Sterile Sleeve ....................................................................................... 3-22
3.18 Ambient Light Fixture .............................................................................................. 3-23
3.18.1 Ambient Light Operation Overview ............................................................... 3-23
3.18.2 Configuring and Using the Ambient Light ...................................................... 3-23
3.19 ACT Enabled System ............................................................................................. 3-25
. . . Continued
iv
129382-106
4 Cleaning the Equipment .................................................................. 4-1
4.1 Cleaning Equipment ................................................................................................... 4-1
4.2 General Cleaning/Disinfecting Procedure .................................................................. 4-2
4.3 Areas To Be Cleaned Before Each Use ...................................................................... 4-4
5 Operator Troubleshooting ............................................................... 5-1
6 Maintenance .................................................................................... 6-1
6.1 Preventive Maintenance Record ................................................................................. 6-1
6.2 Inspect Lampholder Module....................................................................................... 6-2
6.3 Inspect Suspension .................................................................................................... 6-2
6.4 Inspect Control Center ................................................................................................ 6-2
6.5 Reattach Lighthead Access Hood............................................................................. 6-3
6.6 Lamp Replacement: Harmony LA 500 and LA 700 Lightheads ................................ 6-4
6.6.1 Reset Harmony LA 700 Lampholder................................................................. 6-6
6.7 Lamp Replacement: Harmony LA 300 Lightheads..................................................... 6-7
6.8 Lamp Replacement: Optional DeepSite Lighthead ..................................................... 6-9
7 Replacement Parts .......................................................................... 7-1
8 Waste Disposal Guidelines ............................................................. 8-1
9 Appendix–EMC Compliance Technical Data................................... 9-1
Section Title Page
Harmony LA Surgical Lighting and Visualization System (Typical)
1-1
Operator Manual Listing of Warnings and Cautions 129382-106
The following
Safety Precautions
must be observed when operating or servicing this Harmony® LA Surgical
Lighting and Visualization System . WARNING indicates the potential for personal injury and CAUTION indicates
the potential for damage to equipment. For emphasis, certain
Safety Precautions
are repeated throughout the
manual. It is important to review ALL
Safety Precautions
before operating or servicing the unit.
Strictly following these
Safety Precautions
enhances your ability to safely and effectively utilize the unit and helps
the customer avoid improper maintenance methods which may damage the unit or render it unsafe. It is important
to understand that these
Safety Precautions
are not exhaustive; customers are encouraged to develop their own
safety policies and procedures to enhance and complement these
Safety Precautions.
NOTICE: This product contains a Class I laser product: No known biological hazard. The light is shielded from any
possible viewing by a person and the laser system is enclosed and requires a tool to access Fiber Optic connections.
WARNING – EXPLOSION HAZARD:
Do not use lighting fixture in the presence of flammable anesthetics.
This product contains materials which may require disposal through appropriately licensed and
permitted hazardous waste management firms.
WARNING – ELECTRIC SHOCK HAZARD:
Do not remove covers or perform service other than as described in this operator manual. Refer
servicing to qualified service personnel. (Maintenance Manual P764330-226.)
Do not remove control center covers. Refer servicing to qualified service personnel.
Ambient light kit operates from a 100, 120, or 240 VAC power source in the ceiling. Refer any
servicing to qualified service personnel.
WARNING – PINCHING HAZARD:
Pinch points are created during extreme articulation of the suspension system. Do not place hands
on or near the suspension knuckle during lighthead articulations.
WARNING POSSIBLE PATIENT INJURY HAZARD:
Failure to engage the lighthandle cover completely may result in cover falling from lighthead during
the procedure.
WARNING – BIOHAZARD:
Sterile disposables are intended for single use only.
Universal precautions must be observed when disposing of any single use disposable item.
WARNING – PERSONAL INJURY HAZARD:
Do not attempt to replace the lamp unless lighthead is turned off and has cooled sufficiently.
LISTING OF SAFETY PRECAUTIONS 1
Continued . . .
1-2
129382-106 Listing of Warnings and Cautions Operator Manual
WARNING – PERSONAL INJURY HAZARD (Continued):
Do not attempt to clean lighthead unless power is turned off and the lighthead has cooled
sufficiently.
Do not attempt to adjust suspension system. Refer servicing to qualified service personnel.
Avoid looking directly at high-intensity light, whether at the lamp or directly at the lighthead. Eye
injury may result.
Do not attempt to replace lamp unless the power is turned off at the control center and the lamp
module has cooled sufficiently. Lamp contains xenon gas at high pressure. Persons changing or
disposing of the lamp should wear eye protection.
Do not remove monitor-arm safety pin until monitor is installed.
Monitor extension arm must have monitor (or equivalent counterweight) installed to keep arm in
required positions while installing sliding guards.
Do not operate system unless optical fiber cables are fully connected.
To maintain Hazard Levels (1M or less) and safety measures established by STERIS end user must
Type of Max. Output
Laser Source / Channel Channels Wavelength Notes
(dBm) (mW) (nm)
SDI Transmitter +7dBm 5.01 1 1310 Sources of radiation - self-contained,
S-Video Transmitter 0dBm 1.0 1 1310 Class 1 Laser Products
VGA Transmitter +3dBm 2.0 3 1310
DVI Transmitter -3.6dBm 0.44 4 850
Table 1. System Laser Components
assure that any laser device integrated into the system and supplied by sources other than
STERIS does not exceed the optical radiation power (per wavelength) of corresponding STERIS
transmitters listed in Table 1.
It is the responsibility of the end user to ensure any device not supplied by STERIS corresponds to
the values listed in Table 1.
WARNING – IMPACT HAZARD:
Do not remove tension screw from the spring arm joint until the lighthead has been securely
installed onto the spring arm. Note: This warning applies at both installation and de-installation
procedures.
WARNING – STERILITY ASSURANCE HAZARD:
Do not use the surgeon’s control buttons unless a disposable sterile cover is installed. If the
sterilizable lighthandle (metal) is used without a disposable cover, the surgeon’s control buttons
are not protected by a sterile covering.
1-3
Operator Manual Listing of Warnings and Cautions 129382-106
WARNING – EXPLOSION AND LACERATION HAZARD:
This lamp contains xenon gas at high pressure. Proper disposal procedures must be followed. Use
eye protection when working with these lamps.
CAUTION: POSSIBLE EQUIPMENT DAMAGE
Use of any disinfectant solution OTHER than those listed below may cause discoloration or
deformation on the lens surface: Germicidal Surface Wipes Disinfecting/Deordorizing/Cleaning
Wipes. Cleaning solutions other than those listed have NOT been tested for compatibility or
effectiveness. Always follow manufacturer instructions for concentrations and use of cleaning
products.
Use only recommended cleaning/disinfecting and/or anti-static agents on this light. Some degree of
staining, pitting, and/or discoloration could occur if a phenolic-, iodophor-, or glutaraldehyde-
based disinfectant is used on the surfaces of this light. Also, use of alcohol or aerosol spray
cleaner/disinfectants containing a substantial amount of alcohol in the formula can damage the
polycarbonate lens.
Prevent leakage of fluids into interior of lighthead, or control center.
Avoid discoloration of control center keypad and display. Do not clean control center with
Betadine®1 solutions or allow such solutions to contact keypad and display surfaces.
Do not scratch optical coating on accessible portions of optic assembly when cleaning; always
wear rubber gloves and use only a clean, white, lint-free cloth when wiping external surfaces.
Appropriate components of this lighting system have been tested and found in compliance with
IEC 60601-1-2 (2nd Edition: 2001), Medical Electrical Equipment – Part 1: General Requirements
for Safety; Electromagnetic Compatibility (EMC). There is, however, a potential for
electromagnetic or other interference between this equipment and other devices. Should you
experience interference, relocate this device or minimize the use of the affected equipment while
this device is in operation.
Do not touch the glass portion of the lamp with bare fingers. Skin oils can cause deterioration of
material leading to possible failure of the lamp.
Do not bump lightheads into walls or other equipment.
To avoid damage to CRT shelf, do not exceed 75 lbs (34 kg).
Portable and mobile RF communications equipment used in close proximity to the wall control or
canopy controls units may temporarily affect the operation of the Harmony LA System equipment.
Use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement
parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the
Harmony LA System.
The Harmony LA System should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it will be used.
Accessories or replacement parts not listed in the Operator or Maintenance Manuals should not be
used as it may affect EMC or result in equipment damage.
1. Betadine
®
is a registered trademark of Purdue Pharma L.P. Continued . . .
1-4
129382-106 Listing of Warnings and Cautions Operator Manual
CAUTION: POSSIBLE EQUIPMENT DAMAGE (Continued)
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in this manual.
Cleaning and disinfecting agents used on this lighting system must be certified by their
manufacturer to be compatible with the following materials: polycarbonate, polyetherimide,
santoprene.
To avoid damage to modem chassis, load added to chassis must not exceed 4.4 lbs (2 kg).
To avoid damage to yoke assembly, load added to yoke must not exceed 22 lbs (10 kg).
1-5
Operator Manual Listing of Warnings and Cautions 129382-106
1.1 Definition of
Symbols
Symbol Definition
ON
OFF
STANDBY
Lamp Out
Protective Earth (Ground)
Do not operate system unless optical fiber
cables are fully connected.
Maximum load hazard (maximum load limit
given on label)
or Attention, consult manual for further instructions
Hot, Potential Burn Hazard
SER. NO. Serial Number of Unit
V~ Voltage Rating of Unit, Alternating Current
A Amperage Rating of Unit
Hz Frequency Rating of Unit
Increase Intensity (Surgeon's Control Buttons)1
-Decrease Intensity (Surgeon's Control Buttons)1
1. Symbol is a fingertip "contour" molded into the lighthead surgeons's control bezel.
1-6
129382-106 Listing of Warnings and Cautions Operator Manual
Symbol Definition
Zoom
Rotate
Manual Focus
Auto Focus
Brightness Auto Mode
Potential Impact Hazard
Potential Pinch-Point Hazard
2-1
Operator Manual Installation Verification 129382-106
!
WARNING – ELECTRIC
SHOCK HAZARD: Do not
remove covers or perform
service other than as de-
scribed in this operator
manual. Refer servicing
to qualified service per-
sonnel (Maintenance
Manual P764330-226).
!
WARNING – IMPACT
HAZARD: Do not remove
screw from the spring arm
joint until the lighthead
has been securely in-
stalled onto the spring
arm. Note: This warning
applies at both installa-
tion and de-installation
procedures.
Equipment Drawings showing all of the space and utility requirements were
sent to the purchaser after the order for this surgical light was received. The
clearance space shown on the drawing is necessary for proper installation,
operation and maintenance of this fixture.
Installation and Uncrating Instructions were furnished with the Lighting
Fixture.
If any of these documents are missing or misplaced,
contact STERIS, giving
the serial and model numbers of the equipment. Replacement copies will be
sent to you promptly.
Before operating the equipment, complete the pre-operation checklist. It is
essential to the safe operation and continuing maintenance of this equipment
to verify that the installation is complete and correct. (Refer to Figures 2-1
through 2-3 to locate parts.)
INSTALLATION VERIFICATION
2.1 Pre-operation
Checklist
2
Tension Screw
Figure 2-1. Tension Screw Location
for Harmony® Lux Advantage LA 500 and LA 700
Lightheads
2-2
129382-106 Installation Verification Operator Manual
Check all suspension joints for compromised integrity, such as loose fasten-
ers or components.
Verify that suspension system moves through all articulations smoothly
without binding. Lightheads and monitors should move smoothly and easily.
When positioned, the lighthead and monitor support arms should not drift. If
binding or drifting is present in suspension movements, call your STERIS
service representative to make adjustments.
Figure 2-2. Check Suspension
Check these and all similar areas for ease of
movement, as well as suspension integrity
2.1.1 Check Suspension
Movement
!
WARNING – PINCHING
HAZARD: Pinch points are
created during extreme ar-
ticulation of the suspen-
sion system. Do not place
hands on or near the sus-
pension knuckle during
lighthead articulations.
Refer to Figure 2-3. Fiber optic flat panel monitor support arms are capable of
the following articulations: (1) rotate 270° at the secondary spindle; (2) rotate
330° at the central spindle; (3) rotate 310° at the horizontal extension arm; (4)
rotate 320° at the yoke transition; (5) move up or down by pivoting at spring
arm knuckle up to 40° up, and 40° down. Flat panel yoke also allows (6) tilt
forward (down) 15° or backward (up) 75° in yoke. The flat panel yoke can be
set for either portrait or landscape display (7).
For additional information, refer to separate operating instructions supplied
with monitor(s). (Monitor instructions not by STERIS.)
• Power supply and input signals to monitors can be routed through the
suspension wiring from an external video source (external video sources not
provided by STERIS).
Range of upward motion for spring arm can be adjusted to 20°, 30° or 40°
above horizontal to expand storage/parking options when monitors are not
in use.
NOTE: If weight or range of motion for flat panel monitor mount system is
changed in any way, refer to INSTALLATION INSTRUCTIONS, P129382-756
for procedures to help ensure proper balancing and motion.
2.1.2 Harmony LA Fiber
Optic Video Enabled
System
2-3
Operator Manual Installation Verification 129382-106
5
3
2
1
3
4
5
Figure 2-3. Harmony LA Monitor Arms Positioning
4
76
4
6
NOTE: For purposes of clarity, single and dual monitor arms shown without lighthead
spring arms or lightheads
Single Flat Panel Monitor Dual Flat Panel Monitor
Modem Chassis
2-4
129382-106 Installation Verification Operator Manual
Check Lamp Failure LED:
If the LED on any of the lightheads is blinking, one
or both of the fixture’s lamps may have to be replaced. (Refer to
SECTION 6,
MAINTENANCE.
) Verify lamp-out indicator on control center display is visible.
Check Intensity Controls at Each Surgeon's Control: Verify that the
intensity level can be increased and decreased at each lighthead using the
surgeon's control buttons (see Figure 2-5).
Figure 2-4. Use Control Center to Check System
Operation
Verify that electrical power to the control center is on.
Harmony® Lux Advantage LA Control Center:
Turn control on by pressing
“ON” touch pad. Verify that control cycles through all displays. Return to main
menu and check intensity levels for each lighthead in the system (See Figure
2-4.)
NOTE: Turn power OFF to each lighthead using intensity controls when testing
is complete.
Check Lamp Failure Indicators:
If any lamp-out graphic on the control center
display is visible, one or both of the indicated lighthead's lamps may have to
be replaced. (Refer to
SECTION 6, LAMP REPLACEMENT
.) Verify lamp failure LED on
lighthead housing is blinking.
2.1.3 Check System
Operation
2-5
Operator Manual Installation Verification 129382-106
Turn the control center ON. Install video camera in lighthandle
(see
SECTION 3.9, VIDEO CAMERA INSTALLATION OR REMOVAL
). Press the Camera
Control button on the touch pad to select the Camera Control Menu, select
and turn ON the camera. (See Figure 2-6.)
Video: Verify that a clear signal is reaching the video display device
(monitor) from the camera. (Check cable connection between control
center and monitor, if necessary.)
Control Center: Verify zoom, rotation and focus functions with the control
center switches. Select Camera Menu 2; verify Gain controls (if neces-
sary), verify Freeze Frame function, verify Time and Date displays (if
necessary).
Optional Wireless Hand-held Control: Verify zoom, rotation and focus
functions with the optional wireless control. (Check batteries in control if
necessary.)
NOTE: Transmitter on optional wireless control must be within 15 feet and
directed within 50° of receiver on control center.
Optional Foot Control: Verify zoom and rotation functions with the
optional foot control. (Check cable connection between foot control and
control center, if necessary.)
2.1.4 Check Optional
Video Camera Operation
Figure 2-5. Use Surgeon's Control Buttons to Check Lighthead
Intensity Operation (Harmony LA 500 Lighthead Shown)
Surgeon's
Control
Buttons
2-6
129382-106 Installation Verification Operator Manual
Figure 2-6. Camera Control Menus
CAMERA CONTROL 1 2 3 4
SLOW FAST
ZOOM SPEED
SLOW FAST
FOCUS SPEED
2-7
Operator Manual Installation Verification 129382-106
Refer to Figure 2-7.
The Harmony LA Surgical Lighting and Visualization System can include up to
two monitor arms on a single central mount. Verify smooth and easy movement
(without binding or drifting) through the monitor arm(s) ranges of articulations.
Verify that monitor, or monitors, receive video signals.
Figure 2-7. Harmony LA Monitor Support Arm
(Typical Dual Flat-Panel Monitor Arm Shown)
2.1.5 Check Optional
Monitor Support Arms
2-8
129382-106 Installation Verification Operator Manual
Refer to Figure 2-8.
Verify that DeepSite™ lighthead and arm move freely and easily through range
of articulations.
Turn lighthead ON at control center, verify intensity levels, and verify pattern
adjustment function.
Inboard Hub of
Centra Mount
Lighthead
Suspension
Knuckle
Adjustable Positioning
Gooseneck
Figure 2-8. DeepSite Lighthead and Arm
Adjustable (lower)
Suspension Arm
Spring Arm
2.1.6 Check Optional
DeepSite Lighthead
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72

Steris Harmony La Surgical Lighting And Visualization System Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI