Steris Harmonyair E-Series Surgical Lighting System Operating instructions

Type
Operating instructions
OPERATOR MANUAL
HarmonyAIR® Surgical Lighting System
E Series
(Rev. A) 10099907
i
10099907
A WORD FROM STERIS CORPORATION
©2019, STERIS Corporation. All rights reserved. Printed in U.S.A.
This manual contains important information on proper use and care
of this surgical lighting system. All operators and department heads
are urged to carefully review and become familiar with the warnings,
cautions and instructions contained herein. Your new surgical
lighting fixture features an advanced, state-of-the-art design, with
cool, shadow-reduced light and ease of maneuverability. It produces
light of a quality necessary for the most demanding and complex of
surgical procedures.
A thorough preventive maintenance program is essential for safe
and proper operation of your surgical light. You are encouraged to
contact STERIS concerning annual maintenance agreements. Under
the terms of this agreement, preventive maintenance, adjustments
and replacement of worn parts are done on a scheduled basis to
verify lighting fixture performance according to its specifications and
to help avoid untimely or costly downtime. STERIS maintains a
nationwide staff of well-equipped, factory-trained technicians to
provide this service, as well as expert repair services. Contact
STERIS for details.
Indications for Use The HarmonyAIR® Surgical Lighting System E Series is a variable
pattern, variable intensity surgical lighting fixture designed to
provide visible illumination of the surgical field for the operating room
staff.
Advisory The following is an important message from STERIS about the
advantages and limitations associated with the use of high
intensity surgical lighting systems.
Because of the variety of surgical procedures performed and the
wide range of individual preferences of surgical staffs, it is desirable
that a surgical lighting system be capable of selective control across
a wide range of illumination intensities. The Illuminating Engineering
Society (IES) stresses that in addition to providing control of
intensity, surgical lighting systems should provide shadow control,
correct color rendition, and a suitable depth of field to provide sharp,
consistent lighting into deep body cavities. As illumination levels
increase, however, radiant heat also increases. Therefore, the IES
cautions that for most operations, radiant heat should be kept to a
minimum. The user of surgical lights should utilize the lowest
possible illumination level suitable for the procedure, especially in
certain neurological or intestinal procedures on delicate, thin, dry or
abnormal tissue. Furthermore, for the protection of surgically
exposed tissues and for the comfort and efficiency of the surgeon
and assistants, radiant energy can be effectively controlled by
limiting the time of exposure at higher illumination levels. Extra care
must be taken when the light fields from multiple lightheads are
overlapped on the surgical site, since this condition creates a risk of
too much heat and excessive UV irradiance.
ii
10099907 Operating Instructions Introduction
An international standard for the safety of surgical lights established
by the International Electrotechnical Commission (IEC) sets minimum
and maximum levels of illumination and maximum levels of radiant
heat that can be emitted from a single surgical luminaire. The
HarmonyAIR E Series has been designed to comply with this
international standard and to provide a wide range of illumination
levels while minimizing the potentially damaging infrared heat in the
surgical field.
The illumination level of the HarmonyAIR E Series Lights can be
adjusted through several intensity settings via conveniently located
controls on either the wall-mounted control center or the lighthead
handle. The illumination level also decreases as the pattern size
increases. Maximum illuminance can reach 160 klx for the smallest
pattern size of HarmonyAIR lighthead, and can be adjusted by
intensity control or pattern size control throughout the entire range
specified by the IEC.
Manufactured by:
STERIS Corporation
2720 Gunter Park East
Montgomery, AL 36109 • USA
334-277-6660/800-444-9009
www.steris.com
Class 2 Equipment
Ordinary Equipment (enclosed equipment without protection of
ingress of water)
Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or oxygen or nitrous oxide.
Suitable for continuous operation.
Equipment provided and incorporated by STERIS Corporation into
complete Optical Fiber Communication System (OFCS) satisfies
applicable laser safety requirements of the IEC 60825-1 (Class 1
Laser Product) and IEC 60825-2 (Hazard Level 1M for OFCS).
The base language of this document is
ENGLISH. Any translations must be made
from the base language document
iii
Table of Contents Operator Manual 10099907
TABLE OF CONTENTS
Section
Number Title Page
1 Safety Precautions................................................................................................................1-1
2 Symbol Definitions................................................................................................................2-1
3 Product Performance............................................................................................................3-1
3.1 Lighthead Optical Performance....................................................................................................3-1
3.2 Essential Performance.................................................................................................................3-2
3.3 Environmental
Conditions....................................................................................................................................3-2
3.4 Power Requirements....................................................................................................................3-2
4 Post Installation Verification................................................................................................4-1
4.1 Pre-operation Checklist................................................................................................................4-2
4.1.1 Check Suspension Movement............................................................................................4-2
4.1.2 Monitor Arms......................................................................................................................4-4
4.2 Check System Operation.............................................................................................................4-6
4.3 Check Optional Video Camera Module Operation.....................................................................4-10
4.4 Check Optional Monitor Support Arms.......................................................................................4-15
5 Operating Instructions..........................................................................................................5-1
5.1 Light Control Center – Control Screen.........................................................................................5-1
5.2 Intensity Controls .........................................................................................................................5-2
5.3 Lighthead Features......................................................................................................................5-6
5.4 Lighthead Fault Indications..........................................................................................................5-8
5.5 TPCU Fault Indication..................................................................................................................5-9
5.5.1 Fault Display.......................................................................................................................5-9
5.5.2 TPCU Fault Log Indications .............................................................................................5-10
5.6 Lighthead Positioning.................................................................................................................5-12
5.7 Lighthead Pattern Adjustment....................................................................................................5-13
5.8 Lighthandles...............................................................................................................................5-14
5.8.1 Plastic Lighthandle and Disposable Lighthandle Cover ...................................................5-14
5.8.2 Sterilizable Surgeon Control Handle ................................................................................5-14
5.9 Monitor Arms..............................................................................................................................5-17
5.10 Video Camera Module Installation or Removal..........................................................................5-20
5.11 Install Camera Disposable Sterile Cover ...................................................................................5-21
5.12 Video Camera............................................................................................................................5-22
5.12.1 Camera Operation...........................................................................................................5-24
5.12.2 HD Settings.....................................................................................................................5-29
5.12.3 Guidelines for Maximizing Video Image..........................................................................5-30
5.12.4 Engage Manual Focus....................................................................................................5-31
5.13 System Menu.............................................................................................................................5-32
iv
10099907 Operator Manual Table of Contents
Section
Number Title Page
TABLE OF CONTENTS (CONT’D)
5.14 System Standby.........................................................................................................................5-33
5.15 Ambient Light Fixture.................................................................................................................5-34
5.15.1 Ambient Light Operation Overview .................................................................................5-34
5.15.2 Configuring and Using the Ambient Light........................................................................5-34
5.16 ACT Enabled System.................................................................................................................5-35
6 Cleaning the Equipment.......................................................................................................6-1
6.1 Cleaning Equipment.....................................................................................................................6-1
6.2 General Cleaning/Disinfecting Procedure....................................................................................6-2
6.3 Areas to Be Cleaned Before Each Use........................................................................................6-4
7 Operator Troubleshooting....................................................................................................7-1
8 Maintenance ..........................................................................................................................8-1
9 Replacement Parts................................................................................................................9-1
10 Waste Management Guidelines.........................................................................................10-1
11 Appendix – EMC Compliance Technical Data for Wall Mounted TPCU with Integrated
Power Supplies ...................................................................................................................11-1
v
Table of Contents Operator Manual 10099907
LIST OF FIGURES
Figure Title Page
Figure 4-1. Check Suspension...............................................................................................................................4-3
Figure 4-2. Monitor Arms Positioning.....................................................................................................................4-5
Figure 4-3. Touch Panel Control Unit (TPCU) .......................................................................................................4-6
Figure 4-4. LED Module Failure Indicator..............................................................................................................4-6
Figure 4-5. Use TPCU to Check System Operation (Normal Operation Screen Shown with Ambient Light
Installed and Turned ON)....................................................................................................................4-7
Figure 4-6. Lighthead Exterior LED Indicator.........................................................................................................4-8
Figure 4-7. Check Lighthead Intensity Operation Using Surgeon’s Control Buttons .............................................4-9
Figure 4-8. Camera Control Menu 1 (Camera Menu Active with Camera Turned Off)........................................4-10
Figure 4-10. Camera Control Menu 1 (Camera On Screen)................................................................................4-11
Figure 4-9. Camera Control Menu 1 (Identifying Camera Screen) ......................................................................4-11
Figure 4-11. Camera Control Menu 1 ..................................................................................................................4-12
Figure 4-12. Camera Control Menu 2 ..................................................................................................................4-13
Figure 4-13. Camera Control Menu 3 ..................................................................................................................4-14
Figure 4-14. In-Light Video HD Camera ..............................................................................................................4-14
Figure 4-15. Monitor Support Arm (Typical Single Flat Panel Monitor Arm with Handle Kit Shown)...................4-15
Figure 5-1. Touch Panel Control Unit (TPCU) Light Control Screen......................................................................5-1
Figure 5-2. On/Off Toggle Button (Off Position Displayed)....................................................................................5-2
Figure 5-3. On/Off Toggle Button (On Position Displayed)....................................................................................5-2
Figure 5-4. Press On/Off Toggle Button to Select Lighthead.................................................................................5-3
Figure 5-5. Increase or Decrease Light Intensity...................................................................................................5-3
Figure 5-6. Increase or Decrease Light Intensity for All Lightheads Simultaneously.............................................5-4
Figure 5-7. Lighthead Surgeon’s Control Buttons..................................................................................................5-4
Figure 5-8. HarmonyAIR Surgical Lighting System E Series Lighthead................................................................5-6
Figure 5-9. Lighthead Exterior LED Indicator.........................................................................................................5-8
Figure 5-10. LED Module Failure Indicator............................................................................................................5-9
Figure 5-11. HarmonyAIR E Series Lighthead Positioning..................................................................................5-12
Figure 5-12. Light Pattern Size Adjustment .........................................................................................................5-13
Figure 5-13. Standard Handle..............................................................................................................................5-16
Figure 5-14. Sterilizable Surgeon Control Handle ...............................................................................................5-17
Figure 5-15. HarmonyAIR Surgical Lighting System E Series Monitor Arms, Part 1...........................................5-18
Figure 5-16. HarmonyAIR Surgical Lighting System E Series Monitor Arms, Part 2...........................................5-19
Figure 5-17. Install Optional Camera Module ......................................................................................................5-20
Figure 5-18. Install Camera Disposable Sterile Cover.........................................................................................5-21
Figure 5-19. Camera Control Menus 1, 2 and 3 ..................................................................................................5-24
Figure 5-20. Brightness Buttons ..........................................................................................................................5-25
Figure 5-21. Focus Buttons..................................................................................................................................5-26
Figure 5-22. Camera Control Screen 2................................................................................................................5-27
Figure 5-23. Camera Speed Adjustments............................................................................................................5-28
vi
10099907 Operator Manual Table of Contents
Figure Title Page
LIST OF FIGURES (CONT’D)
Figure 5-24. High Definition (HD) Settings...........................................................................................................5-29
Figure 5-25. System Menu Screen ......................................................................................................................5-32
Figure 5-26. System Standby Screen..................................................................................................................5-33
Figure 5-27. System Menu...................................................................................................................................5-34
Figure 6-1. Areas to Be Cleaned ...........................................................................................................................6-3
vii
Table of Contents Operator Manual 10099907
LIST OF TABLES
Table Title Page
Table 1-1. System Laser Components .............................................................................................................1-2
Table 2-1. Symbol Definitions...........................................................................................................................2-1
Table 3-1. HarmonyAIR Surgical Lighting System E Series Lighthead Optical Performance ..........................3-1
Table 5-1. Lighthead Features..........................................................................................................................5-7
Table 5-2. TPCU Fault Indications..................................................................................................................5-10
Table 5-3. Camera Functions .........................................................................................................................5-23
Table 5-4. Camera Control Screen 4 – Speed Adjustment Operation............................................................5-28
Table 5-5. Camera Control Screen 3 – HD Video Signal Adjustment Operation............................................5-30
Table 11-1. Guidance and Manufacturer’s Declaration – Electromagnetic Emissions – for all ME Equipment
and ME Systems (per IEC 60601-1-2 clause 5.2.2.1)...................................................................11-1
Table 11-2. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity for all ME Equipment
and ME Systems ...........................................................................................................................11-2
Table 11-3. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity for ME Equipment and
ME Systems that are not Life-Supporting......................................................................................11-3
Table 11-4. Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the HarmonyAIR Surgical Lighting System with or without Video Monitor and
Video-Monitor Related Components .............................................................................................11-4
viii
10099907 Operator Manual Table of Contents
Healthcare
Capital Equipment
1-1
Safety Precautions Operator Manual 10099907
1
The following Safety Precautions must be observed when operating or servicing this HarmonyAIR® Surgical
Lighting System E Series. WARNING indicates the potential for personal injury and CAUTION indicates the
potential for damage to equipment. For emphasis, certain Safety Precautions are repeated throughout the
manual. It is important to review all Safety Precautions before operating or servicing the unit.
Strictly following these Safety Precautions enhances your ability to safely and effectively use the unit and helps to
avoid improper maintenance methods which may damage the unit or render it unsafe. It is important to
understand that these Safety Precautions are not exhaustive; Customers are encouraged to develop their own
safety policies and procedures to enhance and complement these Safety Precautions.
NOTICE: This product contains a Class I laser product: No known biological hazard. The light is shielded from
any possible viewing by a person and the laser system is enclosed and requires a tool to access Fiber Optic
connections.
WARNING – POTENTIAL ENVIRONMENTAL HAZARD:
WARNING – EXPLOSION HAZARD:
Do not use lighting fixture in the presence of flammable anesthetics.
This product contains materials which may require disposal through appropriately licensed and permitted
hazardous waste management firms.
WARNING – ELECTRIC SHOCK HAZARD:
Do not remove covers or perform service other than as described in SECTION 8, MAINTENANCE of this
operator manual. Refer servicing to qualified service personnel. (MAINTENANCE MANUAL, P764337-464.)
Do not remove Touch Panel Control Unit (TPCU) bezel. Servicing must be performed by qualified service
personnel.
Avoid possibility of electric shock, and refer any servicing of the ambient light fixture to qualified service
personnel.
Avoid possibility of electric shock. This equipment must only be connected to a supply mains with
protective earth.
The main breaker panel is to be used as means of mains isolation prior to installing or working on the
ambient light system.
WARNING – PINCHING HAZARD:
Pinch points are created during extreme articulation of the suspension system. Do not place hands on or
near the suspension knuckle during lighthead articulations and in the area around the monitor yoke
mechanism cover.
!
!
!
!
!
!
!
!
SAFETY PRECAUTIONS 1
1-2
10099907 Operator Manual Safety Precautions
WARNING – IMPACT HAZARD:
Do not remove locking pin from the lighthead spring arm joint until the lighthead has been securely
installed onto the spring arm. Note: This warning applies at both installation and removal procedures.
Monitor spring arms must be locked (using accompanying locking pin) prior to installing and/or removing
the monitor and/or monitor yoke.
WARNING – POSSIBLE PATIENT INJURY HAZARD:
Failure to engage the lighthandle cover completely may result in cover falling from lighthead during the
procedure.
WARNING – BIOHAZARD:
Sterile disposable covers are intended for single use only.
Universal disposal precautions must be observed when disposing of any single-use disposable item.
WARNING – PERSONAL INJURY HAZARD:
When fiber optic video accessories are installed to the HarmonyAIR Surgical Lighting System, do not
operate system unless optical fiber cables are fully connected.
To maintain Hazard Levels (1M or less) and safety measures established by STERIS, end user must
assure that any laser device integrated into the system and supplied by sources other than STERIS does
not exceed the optical radiation power (per wavelength) of corresponding STERIS transmitters listed in
Table 1-1. It is the responsibility of the end user to ensure any device not supplied by STERIS
corresponds to the values listed in Table 1-1.
Do not attempt to clean lighthead unless power is turned off and the lighthead has cooled sufficiently.
Do not attempt to adjust suspension system. Refer servicing to qualified service personnel.
Avoid looking directly at illuminated lighthead. Looking directly into high-intensity light may result in eye
injuries.
!
!
!
!
!
!
!
Table 1-1. System Laser Components
Type of Laser
Source
Max. Output
/ Channel Channels Wavelength Notes
(dBm) (mW) (nm)
SDI Transmitter +7dBm 5.01 1 1310
Sources of radiation - self-
contained, Class 1 Laser
Products
S-Video Transmitter 0dBm 1.0 1 1310
VGA Transmitter +3dBm 2.0 3 1310
DVI Transmitter -3.6dBm 0.44 4 850
!
!
!
1-3
Safety Precautions Operator Manual 10099907
Do not attempt to replace LED modules on HarmonyAIR E Series lights. Refer servicing to qualified
service personnel.
Do not remove monitor arm locking pin until monitor is installed.
Do not modify this equipment without authorization of the manufacturer.
WARNING – STERILITY ASSURANCE HAZARD:
During surgical procedure, do not use the surgeon’s control buttons unless a disposable sterile cover is
installed. If the sterilizable lighthandle (metal, provided with lighthead) is used without a disposable cover,
the surgeon’s control buttons are not protected by a sterile covering.
Do not use disposable handle covers if the packaging has been damaged, torn or opened; as the sterility
of the cover may be compromised.
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
Cleaning and disinfecting agents used on this lighting system must be certified by their manufacturer to be
compatible with the following materials: polycarbonate (PC), thermoplastic elastomer (TPE), silicone,
syndiotactic polystyrene (SPS).
Use only recommended cleaning/disinfecting and/or anti-static agents on this light. Some degree of
staining, pitting and/or discoloration could occur if a phenolic-, iodophor- or glutaraldehyde-based
disinfectant is used on the surfaces of this lighthead. Also, alcohol-based or aerosol spray cleaner/
disinfectants containing a substantial amount of alcohol in the formula can damage the polycarbonate
lens.
Use of any disinfectant solution OTHER than those listed here may cause discoloration or deformation on
the lens surface and other system components:
• Coverage® Plus Germicidal Surface Wipes Disinfecting/Deodorizing/Cleaning Wipes
• Quaternary Ammonium Compound (Quats) with Ethanol solvent
• Quaternary Ammonium Compound (Quats) with Isopropyl Alcohol (IPA) solvent
• Quaternary Ammonium Compound (Quats) with IPA + 2-Butoxyethanol solvent
• Quaternary Ammonium Compound (Quats) + Biguanide
• H2O2 (Hydrogen Peroxide)
• Neutral Cleaners
The use of H2O2 + PAA (Hydrogen Peroxide + Peracetic Acid) is strongly discouraged for use on all
STERIS products.
Always follow manufacturer instructions for concentrations and use of cleaning products.
DO NOT SPRAY any cleaning product directly onto the lighthead, Touch Panel Control Unit (TPCU) or any
system components. Clean TPCU glass with a clean, lint-free cloth dampened with 90% isopropyl alcohol.
For other system components, dampen a clean, soft cloth with the cleaning solution and wring out the
excess moisture.
Do not scratch the lens when cleaning; always wear rubber gloves and use only a clean, white, lint-free
cloth when wiping external surfaces.
WARNING – PERSONAL INJURY HAZARD: (CONTINUED)
!
!
!
!
!
!
!
!
!
!
!
1-4
10099907 Operator Manual Safety Precautions
The camera module tends to roll if placed on a flat surface. Camera must be cradled in a carton, tray or
similar item when not in use or when transported.
Prevent leakage of fluids into interior of lighthead or Touch Panel Control Unit (TPCU). Ensure no excess
fluids remain on lighthead or panel during and after cleaning.
Avoid discoloration of bezel label. Do not clean Touch Panel Control Unit (TPCU) with povidone-iodine
solutions or allow such solutions to contact bezel label surfaces.
Appropriate components of this lighting system have been tested and found in compliance with IEC
60601-1-2:2014, Medical Electrical Equipment – Part 1: General Requirements for Safety;
Electromagnetic Compatibility (EMC). There is, however, a potential for electromagnetic or other
interference between this equipment and other devices. Should you experience interference, relocate this
device or minimize the use of the affected equipment while this device is in operation.
Do not bump lightheads into walls or other equipment. Always use handles or gripping surface when
positioning lighthead during surgical procedures, or when cleaning or servicing the lighting system.
Portable and mobile RF communications equipment may temporarily affect the operation of the
HarmonyAIR Surgical Lighting System equipment.
Use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of this device as replacement parts for internal
components, may result in increased EMISSIONS or decreased IMMUNITY of the HarmonyAIR Surgical
Lighting System.
The HarmonyAIR Surgical Lighting System E Series should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the equipment or system should be observed to verify
normal operation in the configuration in which it will be used.
Accessories or replacement parts not listed in the Operator Manual or Maintenance Manual should not be
used as it may affect EMC or result in equipment damage.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in this manual.
To avoid inadequate balancing, load added to single monitor yoke must not exceed 35 lb (16 kg).
To avoid inadequate balancing, load added to dual monitor yoke must not exceed 26 lb (12 kg) per
monitor.
Failure to replace stops on accessory arms after installation or adjustment results in cable malfunction.
Cables carrying voltages greater than 42 Vac must be double insulated construction and compliant to
relevant safety standards.
Portable RF communications equipment, including peripherals such as antenna cables and external
antennas, should be used no closer than 12" (305 mm) to any part of the equipment boom, including
cables specified by STERIS. Degradation of equipment performance could result.
CAUTION – POSSIBLE EQUIPMENT DAMAGE: (CONTINUED)
!
!
!
!
!
!
!
!
!
!
!
!
!
!
!
2-1
Symbol Definitions Operator Manual 10099907
The following symbols appear on the HarmonyAIR® Surgical Lighting System E Series.
Table 2-1. Symbol Definitions
Symbol Definition
Standby
Lighthead (Designation and Intensity) Symbol Definitions
Ambient Light Control – ON/OFF (Touch Panel Control Unit)
Camera Controls Button
Press button to access System Menu for system configuration or information
Indicates control fault. Refer situation to qualified, trained service technician.
Standby Symbol
LED Module Failure
SYMBOL DEFINITIONS 2
2-2
10099907 Operator Manual Symbol Definitions
Pinch Point Label
Protective Earth (Ground)
or
Attention, consult manual for further instructions
Consult instructions before use
Hot, Potential Burn Hazard
Potential Impact Hazard
Maximum Load Hazard (maximum load limit given on label)
Do not operate system unless optical fiber cables are fully connected
Serial Number of Unit
V~ Voltage Rating of Unit, Alternating Current
A Amperage Rating of Unit, Alternating Current
Hz Frequency Rating of Unit, Alternating Current
Table 2-1. Symbol Definitions
Symbol Definition
SN
2-3
Symbol Definitions Operator Manual 10099907
Increase Intensity (Surgeon's Control Buttons)a
Decrease Intensity (Surgeon's Control Buttons)a
a. Symbol is a fingertip "contour" molded into the lighthead surgeons's control bezel.
Table 2-1. Symbol Definitions
Symbol Definition
✙
2-4
10099907 Operator Manual Symbol Definitions
Healthcare
Capital Equipment
3-1
Product Specifications Operator Manual 10099907
3.1 Lighthead Optical
Performance
Important: Values are typical for the small pattern size at highest
intensity setting (unless otherwise noted) at 100 cm (39-3/8") from the
lighthead. Definitions and measurements are in accordance with
IEC 60601-2-41.
Table 3-1. HarmonyAIR Surgical Lighting System E Series Lighthead Optical Performance
Feature HarmonyAIR Surgical Lighting System E
Series Lighthead
Maximum Central Illuminance 160,000 Lux
Peak Total Irradiance <500 W/m2
Pattern Size Range 18 – 28 cm (7- 11")
D50 Diameter 63% of small beam pattern
(see Pattern Size Range)
Depth of Illumination (to 20%) 100 cm (39")
Depth of Illumination (to 60%) 52 cm (20")
Color Temperature (CCT) 4,400K ±300K
General Color Rendering Index (CRI) Up to 96
Deep Saturated Red Color Rendering Index (R9) Up to 98
Shadow Control
Single Mask 54%
Double Mask 46%
Cavity 100%
Single Mask w/ Cavity 54%
Double Mask w/ Cavity 46%
LED Life 60,000 hours for LEDs
PRODUCT PERFORMANCE 3
3-2
10099907 Operator Manual Product Specifications
3.2 Essential
Performance
Essential Performance (EP) for surgical luminaires:
1. Peak central illuminance (measured at 1m from the luminaire) within the
range from 40 klx to 160 klx.
2. Peak central irradiance (measured at any distance along the beam axis)
less than 1000 W/m2.
3. Provide white light to illuminate the operating field. (The apparent
whiteness of the light can be verified by visual observation.)
NOTE: In extreme cases of electromagnetic interference, the operator may
experience temporary loss of light function or camera operation. This can be
corrected by removing the source of interference and power cycling the light if
necessary.
3.3 Environmental
Conditions
3.4 Power
Requirements
Normal Operating Temperature
Full System 20 to 25°C (68 to 77°F), ambient
Recommended Transport/
Storage Temperature
(not exceeding 15 weeks)
2 to 38°C (36 to 100°F)
Recommended Relative
Humidity 30% to 75%
Atmospheric Pressure 500 to 1060 hPascals
Four-Lighthead System 120 - 240 Vac, 50/60 Hz
700 Watts
Typical Lighthead Wattage
(for both ac and dc backup
operation)
HarmonyAIR Surgical Lighting System E
Series Lighthead: 40 Watts
Optional LED Ambient
Light Kit Ambient light draw 24 Vdc, 15W
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90

Steris Harmonyair E-Series Surgical Lighting System Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI