Steris Deepsite Fiber Optic Light Operating instructions

Type
Operating instructions
i
Table of Contents Operator Manual P-129380-133
A WORD FROM STERIS CORPORATION
This manual contains important information on proper use and maintenance
of the STERIS® DeepSiteTM Fiber Optic Light. All personnel involved in the
use and maintenance of this equipment must carefully review and
comply with the warnings, cautions, and instructions contained in this
manual. These instructions are important to protect the health and safety of
personnel operating a DeepSite lighting system and should be retained in a
conveniently accessible area for quick reference.
Complete instructions for uncrating and connecting utilities, as well as
equipment drawings, have been furnished. If missing, contact STERIS for
replacement copies, giving the serial number and model numbers of the unit.
STERIS carries a complete line of accessories for use with this lighting system.
A STERIS representative will gladly review these with you.
The STERIS DeepSite Fiber Optic Light is a variable pattern, variable intensity
fiber optic surgical light designed to provide visible illumination of the surgical
field or the patient. The DeepSite consists of a center-mounted suspension;
a flexible "gooseneck" lighthead for ease of positioning; an electronic, wall-
mounted control; and a ceiling-mounted illuminator box. This lighting system
is designed to augment other major surgical luminaries
Indications for Use
A thorough preventive maintenance program is essential to safe and proper
unit operation. This manual contains maintenance schedules and procedures
which should be followed for satisfactory equipment performance.
You are encouraged to contact STERIS concerning our comprehensive
Annual Maintenance Agreements. Under the terms of these agreements,
preventive maintenance, adjustments, and replacement of worn parts are
done on a scheduled basis to assure equipment performance at peak
capability and to help avoid untimely or costly interruptions. STERIS maintains
a global staff of well equipped, factory-trained technicians to provide this
service, as well as expert repair services. Please contact your STERIS
representative for details.
Service Information
A listing of the safety precautions to be observed when operating and
servicing this equipment can be found in Section 1 of this manual. Do not
operate or service the equipment until you have become familiar with this
information.
Any alteration of this equipment not authorized or performed by STERIS
Engineering Service which could affect its operation will void the warranty,
could violate national, state, and local regulations, and could jeopardize your
insurance coverage.
Advisory
©2001, STERIS Corporation. All rights reserved. Printed in U.S.A.
ii
P-129380-133 Operator Manual Table of Contents
Manufactured
Exclusively by
STERIS Corporation
Montgomery, AL
ISO 9001
EN 46001
ISO 13485
Certified
The base language of this document is
ENGLISH. Any translations must be
made from the base language document.
STERIS offices worldwide:
Canada 800 661 3937
Germany 49 2233 6999 0
Hong Kong 852 2 563 3623
Italy 39 0141 590429
Japan 81 78 321 2271
Korea 82 2 554 1661
Latin America 305 442 8202
Singapore 65 841 7677
Spain 34 91 658 5920
United Kingdom 44 (0)1256 840400
EC Authorized Representative
STERIS Ltd.
STERIS House
Jays Close
Viables
Basingstoke
Hampshire
RG22 4AX
Tel: +44 (0)1256 840400
Fax: +44 (0) 1256 866502
Manufactured by:
STERIS Corporation
2720 Gunter Park East
Montgomery, AL 36109 • USA
Tel: 334 277 6660
Fax: 334 271 5450
Class 1 Equipment
Type B Equipment
Ordinary Equipment (enclosed equipment without protection of ingress
of water)
Equipment not suitable for use in the presence of a flammable anes-
thetic mixture with air or oxygen or nitrous oxide.
Suitable for continuous operation.
Electrical Specifications: 115/230 Volts AC, 50/60 Hz, 800 Watts
iii
Table of Contents Operator Manual P-129380-133
TABLE OF CONTENTS
Section Title Page
1 LISTING OF WARNINGS AND CAUTIONS ....................................... 1-1
2 COMPONENT IDENTIFICATION....................................................... 2-1
General .............................................................................................................................. 2-1
Wall-mounted Control Panel .............................................................................................. 2-2
Controls ......................................................................................................................... 2-2
Indicators....................................................................................................................... 2-2
Lighthead and Suspension ................................................................................................ 2-4
Ceiling-mounted Illuminator Housing ................................................................................ 2-5
3 OPERATING INSTRUCTIONS .......................................................... 3-1
Wall Control........................................................................................................................ 3-1
Lamp ............................................................................................................................. 3-1
Illumination Intensity ...................................................................................................... 3-1
Check Lamp-Life Indicator ................................................................................................ 3-2
Lighthead Positioning ........................................................................................................ 3-2
Light Pattern Adjustment ............................................................................................... 3-3
Disposable Sterile Sleeve .................................................................................................. 3-4
4 CLEANING THE EQUIPMENT .......................................................... 4-1
General Cleaning/Disinfecting Procedure ......................................................................... 4-1
Cleaning Equipment .......................................................................................................... 4-1
Areas to be Cleaned Before Each Use .............................................................................. 4-2
5 OPERATOR TROUBLESHOOTING ................................................... 5-1
6 ROUTINE MAINTENANCE ............................................................... 6-1
Lamp Change Procedure .................................................................................................. 6-3
7 RECOMMENDED SPARE PARTS ...................................................... 7-1
8 WASTE DISPOSAL GUIDELINES ..................................................... 8-1
iv
P-129380-133 Operator Manual Table of Contents
This Page Intentionally Left Blank
1-1
Listing of Warnings and Cautions Operator Manual P-129380-133
The following is a listing of the safety precautions which must be observed when operating and servicing this equipment.
WARNINGS indicate the potential for danger to personnel, and CAUTIONS indicate the potential for damage to
equipment. These precautions are repeated (in whole or in part), where applicable, throughout the manual.
WARNING EXPLOSION HAZARD:
Do not use lighting fixture in the presence of flammable anesthetics.
WARNING ELECTRIC SHOCK AND BURN HAZARD:
Do not remove covers or perform service other than as described in this operator manual. Refer servicing to
qualified service personnel.
WARNING RISK OF FIRE:
When replacing FUSE (F-1), use only replacement parts as specified in Section 7 of this manual (or as specified
by STERIS Corporation).
WARNING PERSONAL INJURY HAZARD:
Eye injury may result from exposure to this lamp. Avoid looking directly at high-intensity light, whether at the lamp,
directly at the lighthead, or fiber optic cable.
Do not attempt to adjust suspension system. Refer servicing to qualified service personnel.
Person changing or disposing of the lamp must wear eye protection. Lamp contains xenon gas at high pressure.
Do not attempt to replace lamp unless the power is turned off at the wall control and the lamp module has cooled
sufficiently. Lamp contains xenon gas at high pressure. Persons changing or disposing of the lamp should wear
eye protection.
STERIS does not recommend using the light closer than 15" (381 mm) from the patient.
Read and follow warnings and instructions found on detergent labels and Material Safety Data Sheets.
WARNING – BIOHAZARD:
Sterile sleeve is intended for single use only.
Universal precautions must be observed when disposing of the sterile sleeve.
LISTING OF WARNINGS AND CAUTIONS 1
Listing Continued on Next Page.
1-2
P-129380-133 Operator Manual Listing of Warnings and Cautions
WARNING PINCHING HAZARD:
Pinch points are created during extreme articulation of the suspension system. Do not place hands on or near
the suspension knuckle during lighthead articulations.
CAUTION POSSIBLE EQUIPMENT DAMAGE:
Do not bump lighthead into walls or other equipment.
Use only a soft, clean cloth dampened in a solution of mild detergent and water to clean the exterior of this light.
Always follow manufacturer instructions for concentrations and use of cleaning products.
Prevent fluid leakage into the lighting system, do not use excess amounts of fluid during cleaning/disinfecting
procedure.
Do not touch the glass window of the lamp with bare fingers. Skin oils can cause deterioration of material leading
to possible failure of the lamp.
Repairs and adjustments to this equipment should be made only by fully qualified service personnel.
Maintenance performed by inexperienced, unqualified personnel or installation of unauthorized parts could
cause personal injury, invalidate the warranty, or result in costly equipment damage. Contact STERIS
Engineering Service regarding service options.
This device has been tested and found in compliance with IEC 60601-1-2 (1st Edition 1993-04), Medical
Electrical Equipment Part 1: General Requirements for Safety; Electromagnetic Compatibility (EMC). There is,
however, a potential for electromagnetic or other interference between this equipment and other devices. Should
you experience interference, relocate this device or minimize the use of the affected equipment while this device
is in operation.
1-3
Listing of Warnings and Cautions Operator Manual P-129380-133
Definition of
Symbols Symbol Definition
Power ON
Lamp OFF (Standby)
Lamp ON
Thermal Cutoff Indicator
Intensity ON Indicator
Intensity Increase/Decrease
New Lamp
Lamp Half Used
Change Spent Lamp
Protective Earth (Ground)
Attention, consult manual for further instructions
Hot
SER. NO. Serial Number of Unit
V~ Voltage Rating of Unit, Alternating Current
Listing Continued on Next Page.
1-4
P-129380-133 Operator Manual Listing of Warnings and Cautions
Symbol Definition
A Amperage Rating of Unit
Hz Frequency of Unit
Pinch Point Hazard
Wear Eye Protection
Electric Shock Hazard
Definition of
Symbols
2-1
Component Identification Operator Manual P-129380-133
COMPONENT IDENTIFICATION
Refer to Figure 2-1.
The DeepSite lighting system is made up of three primary components:
1. the wall-mounted control panel,
2. the adjustable lighthead and suspension, and
3. the ceiling-mounted illuminator housing.
2
2
General
Figure 2-1. STERIS DeepSite Fiber Optic Surgical Light
LAMP
INTENSITY
LAMP LIFE
SYSTEM
POWER
REPLACE LAMP
Fiber Optic Light
THERMAL
SHUTDOWN
DeepSite
TM
Wall-mounted
Control Panel
NOTE: System circuit
breakers are behind service
access door of ceiling-
mounted illuminator.
Inboard Hub of
Centra Mount
Upper Horizontal
Suspension Arm
Lighthead
Adjustable (lower)
Suspension Arm
WARNING – EXPLOSION
HAZARD: Do not use this
lighting fixture in the pres-
ence of flammable anes-
Suspension
Knuckle
Outboard
Hub
of Centra
Mount
Ceiling-mounted
Illuminator
Adjustable Positioning
Gooseneck
Vertical Suspension
Tube
2-2
P-129380-133 Operator Manual Component Identification
Refer to Figure 2-2.
The wall-mounted control panel should be located at an easily accessible point
on the wall of the operating room.
The controls on the wall panel allow the user to adjust or monitor the
performance characteristics of the DeepSite lighting system. The following
controls are provided:
Lamp Power On/Off
Illumination Intensity
The following indicators are provided:
Lamp Power Status
System Power Status
Thermal Shutdown Status
Intensity Level
Remaining Lamp-life
Lamp Press these membrane touch pads to turn the DeepSite lamp on or off.
[symbol] If off, press to turn ON
[symbol] If on, press to turn OFF
Illumination Intensity This control is used to adjust the illumination level of
the light pattern delivered by the DeepSite lighthead.
Press this membrane touch pad to DECREASE the illumination level.
Press this membrane touch pad to INCREASE the illumination level.
NOTE: The DeepSite illumination level is continuously adjustable between
minimum value and maximum value. LEDs on control panel indicate relative
intensity level.
System Power When LED is lit, system power is ON.
» Indicators
» Controls
Wall-mounted
Control Panel
WARNING – PERSONAL
INJURY HAZARD: Eye in-
jury may result from expo-
sure to this lamp. Avoid
looking directly at the high-
intensity light, whether at
lamp, directly at the
lighthead, or fiber optic
cable.
Lamp Power When LED is lit, lamp power is ON.
Thermal Shutdown If the ceiling illuminator becomes overheated, the
system automatically shuts down. If the system has shut down, this indicator
is illuminated.
IMPORTANT:
If thermal shutdown occurs,
refer to Operator Troubleshooting
(Section 5 of this manual), before continuing to use the light.
Intensity This indicator shows the relative intensity level of the light pattern.
When the lamp is turned off, this indicator also turns off.
Lamp-life This indicator shows the approximate remaining lamp-life time.
2-3
Component Identification Operator Manual P-129380-133
Figure 2-2. DeepSite Wall Control
LAMP
INTENSITY
LAMP LIFE
SYSTEM
POWER
REPLACE LAMP
Fiber Optic Light
THERMAL
SHUTDOWN
DeepSiteTM
New Lamp. This symbol indicates that the lamp is new or near the beginning
of its expected use cycle.
This symbol indicates the lamp has reached the approximate
half-way point in its expected use cycle.
This symbol indicates the lamp has reached the end of its expected use cycle.
When this indicator is lit, the lamp should be removed and
replaced with a new one, even if the lamp is still functioning.
Refer to Figure 2-3.
T h e lighthead is supported from a centra mount that allows 360°
Lamp ON/OFF
Control
Illumination
Intensity Controls
System Power
ON/OFF Indicator
Thermal
Shutdown
Indicator
Illumination
Intensity
Indicator
Lamp-life
Indicator
Lamp ON
Indicator
WARNING PERSONAL
INJURY HAZARD: Eye in-
jury may result from expo-
sure to this lamp. Avoid
looking directly at the high-
intensity light, whether at
lamp, directly at the
lighthead, or fiber optic
cable.
2-4
P-129380-133 Operator Manual Component Identification
rotation around the hubs center. The upper (horizontal) suspension arm ends
in a second (outboard) hub that allows 340° rotation of the lower suspension
arm and lighthead. The adjustable suspension arm is capable of up and down
articulation. The arm can be positioned up to 15° above horizontal and 90°
below horizontal.
The adjustable suspension arm directly supports the lighthead and adjustable
positioning gooseneck. The gooseneck can be used for fine positioning
without unwanted movement or spring back.
Lighthead pattern size is adjusted by rotating the end of the lighthead.
The section of the lighthead used to adjust the pattern is shown in the detail
of Figure 2-3.
The lighthead must always be positioned at working distance between 15" (381
mm) to 22" (559 mm) from the patient. The lighthead pattern size can be
adjusted from a diameter of 2 (50 mm) to 6 (152 mm) while positioned within
the working distance.
This ceiling-mounted component contains the light source of the DeepSite
system, as well as much of the essential system control circuitry and mechani-
Lighthead and
Suspension
Figure 2-3. Lighthead and Adjustable Positioning Gooseneck
Positioning
Gooseneck Rotate this Section of
Lighthead to Adjust
Light Pattern Size
WARNING PINCHING
HAZARD: Pinch points are
created during extreme ar-
ticulation of the suspen-
sion system. Do not place
hands on or near the sus-
pension knuckle during
lighthead articulations.
CAUTION: Do not bump
lighthead into walls or
other equipment.
WARNING PERSONAL
INJURY HAZARD: Eye in-
jury may result from expo-
sure to this lamp. Avoid
looking directly at the high-
intensity light, whether at
lamp, directly at the
lighthead, or fiber optic
cable.
WARNING PERSONAL
INJURY HAZARD: STERIS
does not recommend us-
ing the light closer than
15" (381 mm) from the pa-
tient.
2-5
Component Identification Operator Manual P-129380-133
Ceiling-mounted
Illuminator Housing
Figure 2-4. Ceiling-mounted Illuminator
Lamp Housing
Lock Handle*
Service Access Latch
(to Circuit Breakers)
cal components. The user accesses the housing to change lamps. Typically
this part of the system is only accessed when the lamp reaches the end of its
use cycle.
Circuit breakers for emergency ON/OFF are located behind the service
access panel.
* NOTE: Lamp-life indicator (on wall control)
automatically resets to New Lamp if the
lamp housing lock lever is open for 25
seconds while the circuit breaker is ON.
2-6
P-129380-133 Operator Manual Component Identification
This Page Intentionally Left Blank
3-1
Operating Instructions Operator Manual P-129380-133
3
OPERATING INSTRUCTIONS
Refer to Figure 3-1.
The wall control is used to turn the lamp in the illuminator housing on or off, and
to vary the intensity level of the light pattern.
[symbol] Press this membrane touch pad to turn the lamp ON.
[symbol] Press this membrane touch pad to turn the lamp OFF.
Press this membrane to DECREASE the pattern intensity level.
Press this membrane to INCREASE the pattern intensity level.
Wall Control
» Illumination Intensity
» Lamp
WARNING PERSONAL
INJURY HAZARD: Eye in-
jury may result from expo-
sure to this lamp. Avoid
looking directly at the high-
intensity light, whether at
lamp, directly at the light-
head, or fiber optic cable.
CAUTION: This device has
been tested and found in
compliance with IEC
60601-1-2 (1st Edition
1993-04), Medical Electri-
cal Equipment Part 1:
General Requirements for
Safety; Electromagnetic
Compatibility (EMC). There
is, however, a potential for
electromagnetic or other
interference between this
equipment and other de-
vices. Should you experi-
ence interference, relocate
this device or minimize the
use of the affected equip-
ment while this device is
in operation.
WARNING EXPLOSION
HAZARD: Do not use this
lighting fixture in the pres-
ence of flammable anes-
thetics.
LAMP
INTENSITY
LAMP LIFE
SYSTEM
POWER
REPLACE LAMP
Fiber Optic Light
THERMAL
SHUTDOWN
DeepSite
TM
Lamp
ON/OFF
Control
Illumination
Intensity
Controls
System Power
Indicator
Thermal
Shutdown
Indicator
Illumination
Intensity
Indicator
Lamp-life
Indicator
Figure 3-1. DeepSite Wall Control
Lamp ON
Indicator
3-2
P-129380-133 Operator Manual Operating Instructions
Check the lamp-life indicator before the beginning of any or all procedures.
If the AMBER (non-flashing) LED is illuminated, the lamp should be changed
at the earliest opportunity.
If the AMBER (flashing) LED is illuminated, the lamp must be changed before
using the lighting system in a procedure.
Refer to Lamp Replacement procedure in Section 6 of this manual when it
becomes necessary to change the system lamp.
Lighthead positioning gooseneck (see Figure 3-3) can be adjusted into a
stable curved position. Lighthead may also: (1) rotate 340° around vertical
suspension tube; (2) rotate 360° around centra mount; (3) move up or down by
pivoting at suspension fork 15° up, and 90° down until vertical (i.e., full
movement up and down through 105°).
Check Lamp-Life
Indicator
Lighthead
Positioning
WARNING PINCHING
HAZARD: Pinch points are
created during extreme ar-
ticulation of the suspen-
sion system. Do not place
hands on or near the sus-
pension knuckle during
lighthead articulations.
CAUTION: Do not bump
lighthead into walls or
other equipment.
Figure 3-2. Lamp-life Indicator
LAMP LIFE
REPLACE LAMP
Amber
LED
(Change Lamp
Soon)
Flashing
Amber
LED
(Change Lamp
NOW)
Figure 3-3. Lighthead Positioning Gooseneck
Lighthead
Positioning
Gooseneck
3-3
Operating Instructions Operator Manual P-129380-133
The light pattern size is adjusted by grasping the lighthead with one hand and
gently rotating in a clockwise (larger) or counterclockwise (smaller) direction.
Refer to Figure 3-4.
NOTE: Pattern diameter can also be adjusted by changing the distance
between the lighthead and the patient. Do not position the lighthead closer than
15" (381 mm) from the patient.
» Light Pattern
Adjustment
WARNING PERSONAL
INJURY HAZARD: Eye in-
jury may result from expo-
sure to this lamp. Avoid
looking directly at the high-
intensity light, whether at
lamp, directly at the light-
head or fiber optic cable.
STERIS does not recom-
mend using the light closer
than 15" (381 mm) from
the patient.
Figure 3-4. Pattern Size Adjustment at Lighthead
Rotate this Section of
Lighthead to Adjust
Light Pattern Size
Minimum Diameter
(Smaller pattern can also be
achieved by positioning lighthead
closer to patient.)
Maximum Diameter
(Larger pattern can also be
achieved by positioning lighthead
further from patient.)
3-4
P-129380-133 Operator Manual Operating Instructions
Disposable Sterile
Sleeve
Each time the lighting system is used for a surgical procedure, a sterile sleeve
should be used to cover the lighthead and adjustable suspension arm. Refer
to Figure 3-5.
1. Remove protective packaging from the sleeve.
2. Slip the sleeve over the lens holder end of the lighthead.
3. Securely fit the sleeve's end ring to the lighthead's lens cover. Ensure the
end ring is snug (parallel) to the lens cover.
4. Slowly unfold sleeve until the suspension arm is covered to the knuckle.
NOTE: Allow enough slack in the sleeve for lighthead positioning.
5. Once the sleeve is fully unfolded, and a suitable area of the lighthead and
suspension is covered; tear the adhesive tape off the sleeve and remove the
white backing. Wrap the tape tightly around the sleeve, anywhere along the
suspension arm assembly.
NOTE: The tape must be securely wrapped around the sleeve, only. Tape
MUST NOT be applied to the surface of the suspension arm.
IMPORTANT: The sterile sleeve is intended as a single-use item. Sleeve must
be removed from lighting fixture at the end of each procedure and disposed
of following facility waste protocols.
WARNING BIOHAZARD:
Sterile sleeve is intended
for single use only.
WARNING BIOHAZARD:
Universal precautions
must be observed when
disposing of the sterile
sleeve.
Sleeve
Adhesive
Tape
Lighthead
Lens Cover
Adjustable
Suspension
Arm
Figure 3-5. Fit Disposable
Sterile Sleeve to Lighthead
Knuckle
4-1
Cleaning the Equipment Operator Manual P-129380-133
4
CLEANING THE EQUIPMENT
Pail
Sponge
Cloth Wipes
Rubber Gloves
Coverage® Spray HB*
Germicidal Cloths*
Any other protective equipment recommended by cleaning product manu-
facturer
NOTE: Always follow manufacturer instructions for concentrations and use of
cleaning products.
*Available from STERIS Corporation, please contact your sales representative.
WARNINGPERSONAL
INJURY HAZARD: Read
and follow warnings and
instructions found on de-
tergent labels and Mate-
rial Safety Data Sheets.
CAUTION: Use only a soft,
clean cloth dampened in a
solution of mild detergent
and water to clean the ex-
terior of this light. Always
follow manufacturer in-
structions for concentra-
tions and use of cleaning
products.
CAUTION: Prevent fluid
leakage into the lighting
system; do not use excess
amounts of fluid during
cleaning/disinfecting pro-
cedure.
General Cleaning/
Disinfecting
Procedure
1. Wear rubber gloves.
2. Use sponge and a mild detergent and water solution.
3. Prepare cleaning solution in accordance with directions on the container
labels (follow manufacturer instructions for concentrations when using any
cleaning products).
4. Using a soft cloth and the cleaning solution, thoroughly wipe the areas to be
cleaned (see Figure 4-1). Make sure to wring out excess solution before
wiping.
5. Rinse all surfaces with a soft cloth wipe and clear water.
6. Wipe all surfaces dry with a clean, dry cloth.
Cleaning Equipment
4-2
P-129380-133 Operator Manual Cleaning the Equipment
Areas to be Cleaned
Before Each Use
Figure 4-1. Areas to Be Cleaned
CAUTION: Use only a soft,
clean cloth dampened in a
solution of mild detergent
and water to clean the ex-
terior of this light. Always
follow manufacturer in-
structions for concentra-
tions and use of cleaning
products.
CAUTION: Prevent fluid
leakage into the lighting
system, do not use excess
amounts of fluid during
cleaning/disinfecting pro-
cedure.
Suspension
Knuckle
Vertical
Suspension
Tube Lighthead
Upper
Suspension
Arm
Refer to Figure 4-1.
The following areas of the lighthead and suspension system must be cleaned
and disinfected before each use of the lighthead (see General Cleaning/
Disinfecting Procedure for instructions):
Suspension Arm Wipe the entire suspension arm, including the suspen-
sion knuckle and adjustable arm section.
Lighthead Wipe the entire outer surface of the lighthead.
Inboard Hub of
Centra Mount
Adjustable
Positioning
Gooseneck
Adjustable (Lower)
Suspension Arm
Outboard Hub of
Centra Mount
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32

Steris Deepsite Fiber Optic Light Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI