Steris Harmony Lc Surgical Lighting System Operating instructions

Type
Operating instructions
OPERATOR MANUAL
Harmony® LC500R Surgical Lighting System
(08/10/11) P129388-276
i
Introduction Operator Manual P129388-276
A WORD FROM STERIS CORPORATION
©2011, STERIS Corporation. All rights reserved. Printed in U.S.A.
This manual contains important information on proper use and care
of this surgical lighting system. All operators and department heads
are urged to carefully review and become familiar with the warnings,
cautions and instructions contained herein. Your new surgical
lighting fixture features an advanced, state-of-the-art design, with
cool, shadow-reduced light, and ease of maneuverability. It
produces light of a quality necessary for the most demanding and
complex of surgical procedures.
A thorough preventive maintenance program is essential for safe
and proper operation of your surgical light. You are encouraged to
contact STERIS concerning our annual maintenance agreement.
Under the terms of this agreement, preventive maintenance,
adjustments and replacement of worn parts are done on a
scheduled basis to verify lighting fixture performance according to
its specifications and to help avoid untimely or costly downtime.
STERIS maintains a nationwide staff of well-equipped, factory-
trained technicians to provide this service, as well as expert repair
services. Contact STERIS for details. STERIS is also able to provide
Customer training; contact STERIS Customer Service for more
information.
Indications for Use The Harmony® LC500R Surgical Lighting System is a variable
pattern, variable intensity surgical lighting fixture designed to
provide visible illumination of the surgical field or the patient for the
operating room staff.
Advisory The following is an important message from STERIS about the
advantages and limitations associated with the use of high
intensity surgical lighting systems.
Because of the variety of surgical procedures performed and the
wide range of individual preferences of surgical staffs, it is desirable
that a surgical lighting system be capable of selective control across
a wide range of illumination intensities. The Illuminating Engineering
Society (IES) stresses that in addition to providing control of
intensity, surgical lighting systems should provide shadow control,
correct color rendition, and a suitable depth of field to provide sharp,
consistent lighting into deep body cavities. As illumination levels
increase, however, radiant heat also increases. Therefore, the IES
cautions that for most operations, radiant heat should be kept to a
minimum. The user of surgical lights should utilize the lowest
possible illumination level suitable for the procedure, especially in
certain neurological or intestinal procedures on delicate, thin, dry or
abnormal tissue. Furthermore, for the protection of surgically
exposed tissues and for the comfort and efficiency of the surgeon
and assistants, radiant energy can be effectively controlled by
limiting the time of exposure at higher illumination levels. Extra care
must be taken when the light fields from multiple lightheads are
ii
P129388-276 Operator Manual Introduction
Manufactured by:
STERIS Corporation
2720 Gunter Park East
Montgomery, AL 36109 •
USA
Class 1 Equipment
Type B Equipment
Ordinary Equipment (enclosed equipment without protection
of ingress of water)
Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or oxygen or nitrous oxide.
Suitable for continuous operation.
EC Authorized Representative
STERIS Ltd.
Chancery House
190 Waterside Road
Hamilton Industrial Park
Leicester LE5 1QZ
United Kingdom
overlapped on the surgical site, since this condition creates a risk of
too much heat.
An international standard for the safety of surgical lights established
by the IEC sets minimum and maximum levels of illumination and
maximum levels of radiant heat that can be emitted from a single
surgical luminaire. The LC500R Surgical Lighting has been designed
to comply with this international standard and to provide a wide
range of illumination levels while minimizing the potentially damaging
infrared heat in the surgical field.
The illumination level of the Harmony LC500R Surgical Lights can be
adjusted through several intensity settings via conveniently located
controls on the lighthead handle. The illumination level also
decreases as the pattern size increases. Maximum illuminance can
exceed 12,000 fc for the smallest pattern size of the Harmony
LC500R lighthead and can be adjusted by intensity control or pattern
size control throughout the entire range specified by the IEC.
STERIS Corporation,
Montgomery, Alabama is an
ISO 13485 certified facility.
The base language of this document is
ENGLISH. Any translations must be made
from the base language document.
STERIS Surgical Technologies SAS,
Orleans, France is an
ISO 13485 certified facility.
iii
Table of Contents Operator Manual P129388-276
TABLE OF CONTENTS
Section
Number Title Page
1 Safety Precautions....................................................................................................1-1
2 Installation Verification.............................................................................................2-1
2.1 Pre-operation Checklist.............................................................................................................2-1
2.1.1 Check Suspension Movement..................................................................................................2-1
2.1.2 Check System Operation..........................................................................................................2-1
2.2 Harmony LC500R Optical Performance ...................................................................................2-2
3 Operating Instructions..............................................................................................3-1
3.1 Surgeon's Intensity Control Buttons..........................................................................................3-1
3.2 Harmony LC500R Lightheads Lamp Failure Indications ..........................................................3-2
3.3 Lighthead Positioning................................................................................................................3-3
3.4 Harmony LC500R Lighthead Pattern Adjustment.....................................................................3-4
3.5 Lighthandle and Lighthandle Cover..........................................................................................3-5
3.5.1 Installing and Using Lighthandle and Lighthandle Cover..........................................................3-5
3.5.2 Metal Sterilizable Handle..........................................................................................................3-5
4 Cleaning the Equipment...........................................................................................4-1
4.1 Cleaning Equipment .................................................................................................................4-1
4.2 General Cleaning/Disinfecting Procedure.................................................................................4-2
4.3 Areas To Be Cleaned Before Each Use ...................................................................................4-3
5 Operator Troubleshooting........................................................................................5-1
6 Maintenance...............................................................................................................6-1
6.1 Preventive Maintenance Record...............................................................................................6-1
6.2 Inspect lamp holder Module......................................................................................................6-2
6.3 Inspect Suspension...................................................................................................................6-3
6.4 Reattach Lighthead Access Hood.............................................................................................6-3
6.5 Optional Service Disconnect Switch.........................................................................................6-4
6.6 Lamp Replacement...................................................................................................................6-5
7 Replacement Parts....................................................................................................7-1
8 Waste Disposal Guidelines ......................................................................................8-1
9 AppendixEMC Compliance Technical Data..........................................................9-1
iv
P129388-276 Operator Manual Table of Contents
Healthcare
Capital Equipment
v
Table of Contents Operator Manual P129388-276
LIST OF FIGURES
Figure Title Page
Figure 3-1. Harmony LC500R Lighthead...............................................................................................................3-1
Figure 3-2. Lighthead Lamp-Failure Indicator........................................................................................................3-2
Figure 3-3. Lighthead Positioning ..........................................................................................................................3-3
Figure 3-4. Light Pattern Adjustment .....................................................................................................................3-4
Figure 3-5. Standard Lighthandle ..........................................................................................................................3-6
Figure 6-1. Reattach Lighthead Access Hood .......................................................................................................6-3
Figure 6-2. Optional Service Disconnect Switch....................................................................................................6-4
Figure 6-3. Lamp Replacement Procedure............................................................................................................6-6
vi
P129388-276 Operator Manual Table of Contents
Healthcare
Capital Equipment
vii
Table of Contents Operator Manual P129388-276
LIST OF TABLES
Table Title Page
Table 1-1. Definition of Symbols.......................................................................................................................1-4
Table 2-1. Harmony LC500R Lighthead Optical Performance.........................................................................2-2
Table 5-1. Operator Troubleshooting ...............................................................................................................5-1
Table 6-1. Preventive Maintenance Schedule – Harmony® LC500R Surgical Lighting System ......................6-1
Table 7-1. Harmony LC500R Lighting System Replacement Parts .................................................................7-1
Table 9-1. Guidance and Manufacturer’s Declaration – Electromagnetic Emissions for All Equipment
and Systems ...................................................................................................................................9-1
Table 9-2. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity for All Equipment
and Systems ...................................................................................................................................9-2
Table 9-3. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity for Equipment and
Systems that are Not Life-Supporting.............................................................................................9-3
Table 9-4. Recommended Separation Distances Between Portable and Mobile RF
Communications Equipment and the Harmony LC500R System ...................................................9-4
viii
P129388-276 Operator Manual Table of Contents
Healthcare
Capital Equipment
1-1
Safety Precautions Operator Manual P129388-276
1
The following Safety Precautions must be observed when operating or servicing this Harmony® LC500R Surgical
Lighting System. WARNING indicates the potential for personal injury and CAUTION indicates the potential for
damage to equipment. For emphasis, certain Safety Precautions are repeated throughout the manual. It is
important to review ALL Safety Precautions before operating or servicing the unit.
Strictly following these Safety Precautions enhances your ability to safely and effectively utilize the unit and helps
the Customer avoid improper maintenance methods which may damage the unit or render it unsafe. It is
important to understand that these Safety Precautions are not exhaustive; Customers are encouraged to develop
their own safety policies and procedures to enhance and complement these Safety Precautions.
WARNING PERSONAL INJURY HAZARD AND/OR EQUIPMENT DAMAGE HAZARD:
WARNING POSSIBLE PATIENT INJURY HAZARD:
WARNING STERILITY ASSURANCE HAZARD:
Safe and reliable operation of this equipment requires regularly scheduled preventive maintenance, in
addition to the regular performance of routine maintenance. Contact STERIS Service Engineering to
schedule preventive maintenance.
Repairs and adjustments to this equipment should be made only by fully qualified service personnel.
Maintenance performed by inexperienced, unqualified personnel or installation of unauthorized parts
could cause personal injury, invalidate the warranty or result in costly equipment damage. Contact
STERIS Service regarding service options.
Do not attempt to clean lighthead unless power is turned off and the lighthead has cooled sufficiently.
Do not attempt to adjust suspension system. Refer servicing to qualified service personnel.
Do not attempt to replace the lamp unless lighthead is turned off and has cooled sufficiently.
Avoid looking directly at high-intensity light, whether at the lamp or directly at the lighthead. Eye injury
may result.
The ON/OFF switch under the canopy only removes power from the load side of the system. Any system
wiring located between this switch and the utility junction box is still energized.
Failure to ensure that the lighthandle cover is securely attached prior to the start of a procedure may
result in cover falling from lighthead during a procedure.
Do not use the surgeon’s control buttons unless an unused disposable sterile cover is in place. If the
sterilizable lighthandle (metal) is used without a disposable cover, the surgeon’s control buttons are not
protected by a sterile covering.
!
!
!
!
!
!
!
!
!
SAFETY PRECAUTIONS 1
1-2
P129388-276 Operator Manual Safety Precautions
WARNING – BIOHAZARD:
WARNING SHOCK AND BURN HAZARD:
WARNING EXPLOSION HAZARD:
WARNING ELECTRIC SHOCK HAZARD:
WARNING PINCHING HAZARD:
WARNING – DISPOSAL HAZARD:
CAUTION POSSIBLE EQUIPMENT DAMAGE:
Sterile disposables are intended for single use only.
Universal precautions must be observed when disposing of any single use disposable item.
Disconnect all utilities to lighting fixture before servicing. Do not install the lighting fixture unless all
utilities have been properly locked out. Always follow OSHA Lockout-Tagout and electrical safety-related
work practice standards.
Do not use this lighting system in the presence of flammable anesthetics.
Do not remove covers or perform service other than as described in this operator manual. Refer servicing
to qualified service personnel (Maintenance Manual P764333-556).
Pinch points are created during extreme articulation of the suspension system. Do not place hands on or
near the suspension knuckle during lighthead articulations.
This product contains materials which may require disposal through appropriately licensed and permitted
hazardous waste management firms. The materials listed in Section 8 are contained within the Harmony
LC500R Surgical Lighting System. When disposing of the lighting fixture or its parts, ensure proper
disposal of hazardous and other regulated waste in compliance with national, state and local regulations.
Appropriate components of this lighting system have been tested and found in compliance with IEC
60601-1-2: 2001, Medical Electrical Equipment – Part 1: General Requirements for Safety;
Electromagnetic Compatibility (EMC). There is, however, a potential for electromagnetic or other
interference between this equipment and other devices. Should you experience interference, relocate this
device or minimize the use of the affected equipment while this device is in operation.
Do not bump lightheads into walls or other equipment.
!
!
!
!
!
!
!
!
!
1-3
Safety Precautions Operator Manual P129388-276
CAUTION POSSIBLE EQUIPMENT DAMAGE (CONT’D):
Use of any disinfectant solution on the lens surface other than Germicidal Surface Wipes Disinfecting/
Deodorizing/Cleaning Wipes may cause discoloration or deformation of the lens surface. Other cleaning
solutions have NOT been tested for compatibility or effectiveness. Always follow manufacturer
instructions for concentrations and use of cleaning products.
Use only recommended cleaning/disinfecting and/or anti-static agents on this light. Some degree of
staining, pitting and/or discoloration could occur if a phenolic-, iodophor- or glutaraldehyde-based
disinfectant is used on the surfaces of this light. Also, use of alcohol or aerosol spray cleaner/
disinfectants containing a substantial amount of alcohol in the formula can damage the polycarbonate
lens.
Do not scratch optical coating on accessible portions of optic assembly when cleaning; always wear
rubber gloves and use only a clean, white, lint-free cloth when wiping external surfaces.
Prevent leakage of fluids into interior of lighthead.
Do not touch the glass portion of the lamp with bare fingers. Skin oils can cause deterioration of material
leading to possible failure of the lamp.
Use only STERIS replacement lamps in this lighting fixture. Use of lamps other than those provided by
STERIS may damage the equipment.
Portable and mobile RF communications equipment used in close proximity to the canopy power supply
enclosures may temporarily affect the operation of the Harmony LC500R System equipment.
Use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement parts
for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the Harmony
LC500R System.
Accessories or replacement parts not listed in the Operator or Maintenance Manuals should not be used
adjacent to or stacked with the Harmony LC500R System. If adjacent or stacked use is necessary, the
equipment or system should be observed to verify normal operation in the configuration in which it will be
used.
Accessories or replacement parts not listed in the Operator or Maintenance Manuals should not be used
as they may affect EMC or result in equipment damage.
Electrical Medical Equipment may affect EMC within the surgical suite and therefore must be installed
and put into service according to the EMC information provided in this manual.
Avoid discoloration of system. Do not clean with Betadine®a solutions or allow such solutions to contact
system.
a. Betadine® is a registered trademark of Purdue Pharma, L.P.
!
!
!
!
!
!
!
!
!
!
!
!
1-4
P129388-276 Operator Manual Safety Precautions
The following symbols appear on the Harmony® LC500R Surgical Lighting System.
Table 1-1. Definition of Symbols
Symbol Definition
ON-OFF
Lighthead (designation and intensity)
Lamp Out
Protective Earth (Ground)
Attention, consult manual for further instructions
Hot, Potential Burn Hazard
SER. NO. Serial Number of Unit
V~ Voltage Rating of Unit, Alternating Current
A Amperage Rating of Unit
Hz Frequency Rating of Unit
+ Increase Intensity (Surgeon's Control Buttons)
- Decrease Intensity (Surgeon's Control Buttons)
Potential Impact Hazard
Potential Pinch-Point Hazard
1-5
Safety Precautions Operator Manual P129388-276
Brightness
Battery backup
System fault
Table 1-1. Definition of Symbols (Continued)
Symbol Definition
1-6
P129388-276 Operator Manual Safety Precautions
Healthcare
Capital Equipment
2-1
Installation Verification Operator Manual P129388-276
Equipment Drawings showing all of the space and utility
requirements were sent to the purchaser after the order for this
surgical light was received. The clearance space shown on the
drawing is necessary for proper installation, operation and
maintenance of this fixture.
Installation and Uncrating Instructions were furnished with the
Lighting Fixture.
If any of these documents are missing or misplaced, contact
STERIS, giving the serial and model numbers of the equipment.
Replacement copies will be sent to you promptly.
2.1 Pre-operation
Checklist
Before operating the equipment, complete the pre-operation
checklist. It is essential to the safe operation and continuing
maintenance of this equipment to verify that the installation is
complete and correct.
2.1.1 Check Suspension
Movement
Check all suspension joints for compromised integrity, such as loose
fasteners or components.
Verify that suspension system moves through all articulations
smoothly without binding. Lightheads should move smoothly and
easily. When positioned, the lighthead support arms should not
drift. If binding or drifting is present in suspension movements,
call your STERIS service representative to make adjustments.
2.1.2 Check System Operation Check Lamp Failure LED: If the LED on any of the lightheads is
blinking, one or both of the fixture’s lamps may have to be
replaced. (Refer to SECTION 6.6, LAMP REPLACEMENT, PAGE 6-5.)
Verify lamp-failure indicator on control center display is visible.
Check Intensity Controls at Each Surgeon's Control: Verify
that the intensity level can be increased and decreased at each
lighthead using the surgeon's control buttons. Turn OFF each
lighthead by holding - symbol until light goes out.
WARNING – ELECTRIC
SHOCK HAZARD: Do not
remove covers or
perform service other
than as described in this
operator manual. Refer
servicing to qualified
service personnel
(Maintenance Manual
P764333-556).
!
WARNING PINCHING HAZARD:
Pinch points are created during
extreme articulation of the
suspension system. Do not place
hands on or near the suspension
knuckle during lighthead
articulations.
!
Lamp-Failure Indicator
INSTALLATION VERIFICATION 2
2-2
P129388-276 Operator Manual Installation Verification
2.2 Harmony LC500R
Optical Performance
Important: Values are typical for the small pattern size at highest
intensity setting (unless otherwise noted) at 39-3/8" (1 m) from the
lighthead. Definitions and measurements are in accordance with IEC
60601-2-41
.
Table 2-1. Harmony LC500R Lighthead Optical Performance
Harmony LC500R
Lighthead
Central Illuminance
(range for intensity settings 1-7)
59,000–160,000 lux
(5,480–14,860 fc)
Pattern Size
D10 150–280 mm
(6–11")
D50 80–150 mm
(3–6")
Depth of Illumination 1090 mm (43")
Peak Total Irradiance < 700 W/m2
Color Temperature 4,400 K
Color Rendering Index (CRI) 94
Shadow Control
Single Mask: 31%
Double Mask: 43%
Cavity: 100%
Single mask w/ cavity: 32%
Double mask w/ cavity: 44%
Average Lamp Life:
at mid-range intensity
1,200 hrs
Operating Temperature
Non-Camera System 20°C to 24°C
(68°F to 75°F)
Storage Temperature -40°C to 70°C
(-40°F to 158°F)
Humidity 30 to 70%, non-condensing
3-1
Operating Instructions Operator Manual P129388-276
3.1 Surgeon's
Intensity Control
Buttons
Refer to Figure 3-1
1. Grasp the handle of the appropriate lighthead. A ring of
membrane buttons (bumps) are located on the control bezel
adjacent to the lighthead lens.
2. Press any of the surgeon's intensity control buttons molded with
+ symbols to increase the lighthead intensity. Press any of the
surgeon's control buttons molded with - symbols to decrease
lighthead intensity.
NOTE: For longer lamp life, use lowest intensity level suitable for
surgical procedure.
3. To turn lighthead OFF, press and hold any of the - membrane
buttons until the light goes out.
NOTE: Press the decrease intensity button (-) on the lighthead
surgeon's control for an additional two seconds to turn OFF all
lightheads in the system.
Figure 3-1. Harmony LC500R Lighthead
Lighthead Handle
OPERATING INSTRUCTIONS 3
3-2
P129388-276 Operator Manual Operating Instructions
3.2 Harmony LC500R
Lightheads Lamp
Failure Indications
Lamp status LED is located on each Harmony LC500R lighthead
(see Figure 3-2). When flashing, the lighthead LED indicates one or
both lighthead lamps should be replaced.
Check the lamp status LED each time the surgical light is used.
A secondary lamp inside Harmony LC500R lightheads
automatically turns on when the primary lamp fails.
If lamp status LED is blinking, replace the failed lamp(s). (See
SECTION 6.6, LAMP REPLACEMENT, PAGE 6-5.) After lamp is
replaced, primary lamp lights, and LED indicators on lighthead
stop blinking.
Figure 3-2. Lighthead Lamp-Failure Indicator
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44

Steris Harmony Lc Surgical Lighting System Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI