Mettler Toledo pH electrode HA405-DPA-SC Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo pH electrode HA405-DPA-SC Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
HA405-DPA-SC pH Electrodes
53 500 086
Type HA405-DPA-SC
pre-pressurized gel-reference
pH Electrodes
Instruction Manual
cs Návod k obsluze 2
da Brugsanvisning 4
de Betriebsanleitung 6
en Instruction manual 8
es Manual de instrucciones 10
fr Instructions dutilisation 12
 Käyttöohje 14
hu Használatiutasítás 16
it Istruzioni per l‘uso 18
ja 20
ko 22
nl Gebruikershandleiding 24
pl Instrukcjaobsługi 26
pt Manual de instruções 28
ru Инструкция по эксплуатации 30
sv Bruksanvisning 32
th ำ�ำ� 34
zh
36
8 HA405-DPA-SC pH Electrodes
© 04 / 16 Mettler-Toledo GmbH HA405-DPA-SC pH Electrodes
53 500 086 Printed in Switzerland
1 Introduction
TheMETTLERTOLEDOTypeHA405-DPA-SCcombinationpH
electrodeisa low-maintenance,pre-pressurizedelectrode
withagel-typereference.IthasaspeciallyformulatedHA
(highalkalinity)pHmembranetominimizesodiumerrorin
extremelyalkalinemedia.Thereferenceelementincorporates
asilver-iontraptopreventsilversuldecontaminationofthe
referencejunction.Thegel-typereferenceisalsoavailablein
acombinationORPandasaseparatereferenceelectrode.
Fororderinformationortechnicalassistance,contactour
CustomerServicedepartmentat+1-800-352-8763.
2 Preparation
Afterunpacking,examinetheelectrodeforanymechanical
damage.
METTLERTOLEDO electrodes are factory tested and are
shippedready-to-use.Tomaintainfactorypressurization,
theType HA405-DPA-SCelectrodes haveasiliconeseal
coveringtheceramicdiaphragm;thismustberemovedwith
theenclosedbladepriortouse.Removethehydrationcap
andremoveanyelectrolytecrystalsbyrinsingtheelectrode
withdistilledwater.Patdry;donotrubasthismaybuildup
astaticchargeandlengthentheresponsetime.
Eliminateanyairbubblesaroundtheinternalelementinthe
pH-sensitivemembrane.
3 Operation
METTLERTOLEDOpHelectrodescanbeusedwithmostpH/
mVmeters.Followmanufacturer’sinstructionsforuse.
Atwo-pointcalibrationisrecommended,usingfreshbuffers
ofknownpHvalueswhichbracketthesamplepH;onebuffer
shouldbeclosetopH7tosetthezeropoint.Forfastroutine
analysis,astandardizationusingonebufferatavaluenear
thesamplepHcanbedone.
4 Maintenance
When properly cared for, METTLER TOLEDO electrodes
shouldprovideaccurateresultsoverthousandsofuses.
Storage:Electrodesshouldberinsedthoroughlyafteruse
andstoredin3MKClorpH4buffersolutionwithaddedKCl.
DONOTstoretheelectrodedryorindistilledortapwater;
thisdehydratesthepHmembraneandtheelectrodemust
berehydratedbeforeusebyimmersingforseveralhoursin
3MKCl.
Relling:TypeHA405-DPA-SCgel-referenceelectrodesare
notrellable.
Protein Contamination:CleantheelectrodewithpHElectro-
deCleanerforProteinsfollowinginstructionsonthebottle.
Reactivating pH membrane:UseReactivatingSolutionfollo-
wingdirectionsCAREFULLY.
Servicing a clogged or dry reference junction:Immersethe
electrodeinabeakerof3MKCl.Heatthesolutiontoappro-
ximately70°C(160°F)andallowelectrodetosoakinthis
solutionasitcoolstoroomtemperature.
en
en Instruction manual
Instructions for Type HA405-DPA-SC
pre-pressurized gel-reference pH Electrodes
HA405-DPA-SC pH Electrodes 9
HA405-DPA-SC pH Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 53 500 086
en
5 Troubleshooting
No response (symptom)
– Poor meter connection
 Checkconnectorsatthemeterandat
theelectrode
– Cracked glass
Replaceelectrode
Slow response (symptom)
–DehydratedpHmembrane
Rehydrate(seeMaintenance)
–Cloggedreferencejunction
SoakinwarmKCl(seeMaintenance)
– Protein contamination
UsepHElectrodeCleaner(seeMain-
tenance)
–Depletedgellayer
UseReactivatingSolution(seeMain-
tenance)
Noise (symptom)
– Poor meter connection
Checkconnector
–Cloggedreferencejunction
(Seeabove)
Low slope (symptom)
–DehydratedpHmembrane
Rehydrate(seeMaintenance)
– Contaminated membrane
(Seeabove)
Forfurtherassistance,callourCustomerServiceCenterat
+1-800-352-8763.
38 HA405-DPA-SC pH Electrodes
© 04 / 16 Mettler-Toledo GmbH HA405-DPA-SC pH Electrodes
53 500 086 Printed in Switzerland
Notes
HA405-DPA-SC pH Electrodes 39
HA405-DPA-SC pH Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 53 500 086
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-Toledo Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.53500086 www.mt.com/pro
Addresses | Impressum
/