  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo 3201i-UPW User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
3201-UPW / 3201 i-UPW
52 002 471 F
3201-UPW & 3201i-UPW
pH Combination Electrode
Instruction Manual
cs Návod k obsluze 2
da Brugsanvisning 6
de Betriebsanleitung 10
en Instruction manual 14
es Manual de instrucciones 18
 Käyttöohje 22
fr Instructions dutilisation 26
hu Használatiutasítás 30
it Istruzioni per l‘uso 34
jp
38
ko
42
nl Gebruikershandleiding 46
pl Instrukcjaobsługi 50
pt Manual de instruções 54
ru
Инструкция по эксплуатации 58
sv Bruksanvisning 62
th
 66
zh
70
14 pHCombinationElectrodeType3201-UPW/3201i-UPW
3201-UPW / 3201 i-UPW © 05 / 18 Mettler-Toledo GmbH
52 002 471 F Printed in Switzerland
1 Introduction ...................................................................15
2 Safetyinstructions ..........................................................15
3 Product description .........................................................15
4 Installation and commissioning .......................................16
5 Operation .......................................................................16
5.1 CalibrationofthesensorandpHtransmitter ......................16
5.2 Temperaturesensor .......................................................16
6 Maintenance ..................................................................16
7 Trouble-shooting ............................................................17
8 Environmentalprotection .................................................17
9 Warranty .......................................................................17
Contents
en
3201-UPW & 3201i
pH Combination Electrode
Instruction Manual
12 mm
0.47 in
a
1
2
3
4
6
7
8
5
1 ConnectortypeVPorISM
2 Referenceelectrolytegel
3 Argenthalreferenceelement
4 Silver-ion trap
5 Diaphragm
6 Lead-offelementofglass
electrode
7 pHsensitiveglass
8 Temperaturesensor
a Shaftlength
en Instruction manual
ISMisaregisteredtrademarkoftheMETTLERTOLEDOGroupin
Switzerland,USA,theEuropeanUnionandafurthervecountries.
pHureSensorisaregisteredtrademarkoftheMETTLERTOLEDO
GroupintheUSAandChina.
pHCombinationElectrodeType3201-UPW/3201i-UPW 15
© 05 / 18 Mettler-Toledo GmbH 3201-UPW / 3201 i-UPW
Printed in Switzerland 52 002 471 F
1 Introduction
Mettler-Toledo Thornton pH sensors type 3201-UPW
are low-maintenance electrodes withpressurized gel-
electrolyte,silver-iontrapandbuilt-intemperaturesensor.
2 Safety instructions
The sensor consists primarily of glass and should
behandledcarefully.Ifthesensoristobecleanedor
calibratedusingacidoralkalinesolutions,safetygoggles
andglovesshouldbeworn.Localsafetyregulationsare
applicable to the handling of the sensor. Please take
extra care during installation/tting of the sensor to
ensurethatthepolymerthread(Pg13.5)doesnotsuffer
anydamage.
Werecommendthatthesensormayonlybeinoperation
with the original accessories from Mettler-Toledo
Thornton.Operationandmaintenanceshouldbecarried
outonlybytrainedpersonnelandstafffamiliarwiththe
unitandwhohavereadandunderstoodthisinstruction
manual.
3 Product description
The markings on each sensor provide the following
information:
METTLER TOLEDO THORNTON Sensor manufacturer
3201-UPW / i / 120 / Pt 1000  Typedesignation/i=ISM
digital/shaftlengthin
mm / temperature sensor
combination pH Typeofsensor
pH111 pHmeasuringrange
0…80°C(176°F) Temperaturerangeforoperation
Order No. 52 00X XXX Article number for ordering
Additionally,thesensorissuppliedwithaserialnumber
ontheredconnectorpartforidenticationofeachsensor
produced.
en
16 pHCombinationElectrodeType3201-UPW/3201i-UPW
3201-UPW / 3201 i-UPW © 05 / 18 Mettler-Toledo GmbH
52 002 471 F Printed in Switzerland
4 Installation and commissioning
1. On unpacking, check the sensor for possible
mechanicaldamageofthepHsensitiveglass,the
ceramic reference junction/diaphragm, the shaft
andtheVarioPin(VP)orISMconnector.Reportany
damageimmediatelytoyourMettler-ToledoThornton
supplier.
2. Removethewateringcapandbrieyrinsethesensor
with deionized water. Important: Carefully and
thoroughlyremovetheprotectivedropletofsilicone
sealantonthereferencejunction/diaphragmwiththe
enclosedknife,asshownonthegreenpictureson
thepacking.
3. Againrinsethesensorwellwithdeionizedwater.After
rinsing,thesensorshouldonlybedabbeddrywith
atissue.Donot rub thepHsensitiveglass,since
thiscanleadtoelectrostaticchargingandsluggish
response times.
4. Carefullycheckthe spacebehindthepHsensitive
glassforthepresenceofanyairbubblesandremove
same by gently swinging the sensor in a vertical
plane.
5. Pleaserefertotheinstructionmanualfortherelevant
housinginordertoinstallthesensorcorrectly.
6. ConnectthesensortothepHpreamportransmitter,
usingacablewithVarioPin(VP)orISMconnector.
Please refer to the wiring and terminal diagram
accompanyingthecable.
5 Operation
5.1 Calibration of the sensor and pH transmitter
For 3201-UPW sensors, a 2-point calibration is
recommended. After removing the watering cap, the
sensorisdippedinsuccessionintotwodifferentbuffer
solutions with known pH reference values and the
transmitteriscalibratedtothesebuffervalues.
5.2 Temperature sensor
TheintegratedPt1000temperaturesensorisprimarily
intendedforautomatictemperaturecompensationofthe
pHsignalduringcalibrationandnormaloperation.For
accurateindicationorcontroloftheprocesstemperature,
a further sensor temperature calibration may be
performed.
6 Maintenance
1. Carefully rinse the sensor tip and reference
junction/diaphragm with deionized water after
removalfromtheprocess.Undernocircumstances
should measuring solution containing solids be
allowedtodryontheseparts!
2. When the sensor isnot in operation,storeit with
sensor tip and reference junction/diaphragm well
submergedinto3MKCl.
3. Ifthesensorisstoredmountedinitshousing,the
sameruleasin2)aboveapplies,butthestorage
electrolytehastobeslightlymodied,with2parts
en
pHCombinationElectrodeType3201-UPW/3201i-UPW 17
© 05 / 18 Mettler-Toledo GmbH 3201-UPW / 3201 i-UPW
Printed in Switzerland 52 002 471 F
ofbuffersolutionpH9to10partsofelectrolyte,this
inordertoavoidcorrosionofhousingparts.
4. Ifasensorisstoreddryforafewdaysbymistake,let
itsoakinthenormalstorageelectrolyteforseveral
hoursbeforeuse.
5. Occasionallychecktheconnectorforpossibletraces
ofmoisture.Ifnecessary,cleanwellwithdeionized
wateroralcohol,andafterwardsdrycarefully.
7 Trouble-shooting
Sluggish response time (e.g. after a long and dry
storageperiod)isanindicationthatreactivationofthepH
sensitiveglassmaybenecessary.DipthepHsensitive
glass(butnofurthersensorparts!)intoReconditioning
(HF)Solution(METTLERTOLEDOproduct,articleno.51
319001)for5min.Afterwards,rinsewellwithdeionized
waterandletsensortipanddiaphragmsoakinreference
electrolyteforminimum12hours.
8 Environmental protection
Waste electrical products should not be
disposedofwithhouseholdwaste.Please
recycle where facilities exist. Check with
yourLocalAuthorityorretailerforrecycling
advice.
9 Warranty
Intheeventofmanufacturingfaults,12monthswarranty
from date of production is granted.
en
74 pHCombinationElectrodeType3201-UPW/3201i-UPW
3201-UPW / 3201 i-UPW © 05 / 18 Mettler-Toledo GmbH
52 002 471 F Printed in Switzerland
zh
pHCombinationElectrodeType3201-UPW/3201i-UPW 75
© 05 / 18 Mettler-Toledo GmbH 3201-UPW / 3201 i-UPW
Printed in Switzerland 52 002 471 F
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mail mtbr@mt.com
France Mettler-Toledo Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mail mtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mail prozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444747
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mail mtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs