  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo Type 405-DPAS-SC-VP pre-pressurized gel-reference pH Electrode User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
8 405-DPAS-SC-VP pH Electrodes
© 04 / 16 Mettler-Toledo GmbH 405-DPAS-SC-VP pH Electrodes
53 500 197 Printed in Switzerland
1 Introduction
TheMETTLERTOLEDOType405-DPAS-SC-VPcombination
pH electrode is a low-maintenance, pre-pressurized,
autoclavable electrode with a gel-type reference and
aVarioPin(VP)connector.Thereferenceelementincorporates
asilver-iontraptopreventsilversuldecontaminationof
thereferencejunction.
2 Preparation
Afterunpacking,examinetheelectrodeforanymechanical
damage.
METTLERTOLEDO electrodes are factory tested and are
shippedready-to-use.(Tomaintainfactorypressurization,
theType405-DPAS-SC-VPelectrodeshaveasiliconeseal
coveringtheceramicdiaphragm;thismustberemovedwith
theenclosedbladepriortouse.)Removethehydrationcap
andremoveanyelectrolytecrystalsbyrinsingtheelectrode
withdistilledwater.Patdry;donotrubasthismaybuildup
astaticchargeandlengthentheresponsetime.
Eliminateanyairbubblesaroundtheinternalelementinthe
pH-sensitive membrane.
3 Operation
METTLERTOLEDO pH electrodes can be used with most
pH/mVmeters.Followmanufacturer’sinstructionsforuse.
Atwo-pointcalibrationisrecommended,usingfreshbuffers
ofknownpHvalueswhichbracketthesamplepH;onebuffer
shouldbeclosetopH7tosetthezeropoint.Forfastroutine
analysis,astandardizationusingonebufferatavaluenear
thesamplepHcanbedone.
4 Sterilization
Type 405-DPAS-SC-VP electrodes are steam-sterilizable
in-situandauto-clavable.Followoperatinginstructionsfor
theautoclave.Itisnotnecessarytouseaprotectivecapon
theconnectorwhenautoclaving.
Repeatedsterilizationmaycausetheelectrolytetodiscolor
slightlyduetotheelevatedtemperatures.Thisdoesnotaffect
theelectrodeperformance.
5 Maintenance
When properly cared for, METTLER TOLEDO electrodes
shouldprovideaccurateresultsoverthousandsofuses.
Storage:Electrodesshouldberinsedthoroughlyafteruse
andstoredin3MKClorpH4buffersolutionwithaddedKCl.
DONOTstoretheelectrodedryorindistilledortapwater.
Iftheelectrodehasbeenstoreddryorinwater,itmustbe
reconditionedbysoakingin3MKClforseveralhours.
Relling: Type 405-DPAS-SC-VP gel-reference electrodes
arenotrellable.
Protein Contamination:CleantheelectrodewithpHElectrode
CleanerforProteinsfollowinginstructionsonthebottle.
Reactivating pH membrane: Use Reactivating Solution
followingdirectionsCAREFULLY.
en
en Instruction manual
Instructions for Type 405-DPAS-SC-VP
pre-pressurized gel-reference pH Electrodes
405-DPAS-SC-VP pH Electrodes 9
405-DPAS-SC-VP pH Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 53 500 197
en
Servicing a clogged or dry reference junction: Immerse
theelectrodeinabeakerof3MKCl.Heatthe solutionto
approximately70°C(160°F)andallowelectrodetosoak
inthissolutionasitcoolstoroomtemperature.
6 Troubleshooting
No response (symptom)
– Poor meter connection
 Checkconnectorsatthemeter
andattheelectrode
– Cracked glass
Replaceelectrode
Slow response (symptom)
–DehydratedpHmembrane
Rehydrate(seeMaintenance)
–Cloggedreferencejunction
SoakinwarmKCl(seeMaintenance)
– Protein contamination
UsepHElectrodeCleaner(see
Maintenance)
–Depletedgellayer
UseReactivatingSolution(see
Maintenance)
Noise (symptom)
– Poor meter connection
Checkconnector
–Cloggedreferencejunction
SoakinwarmKCI(seeMaintenance)
Low slope (symptom)
–DehydratedpHmembrane
Rehydrate(seeMaintenance)
– Contaminated membrane
UsepHelectrodecleaner(see
Maintenance)
Forfurtherassistance,callourCustomerServiceCenterat
+1-800-352-8763.
38 405-DPAS-SC-VP pH Electrodes
© 04 / 16 Mettler-Toledo GmbH 405-DPAS-SC-VP pH Electrodes
53 500 197 Printed in Switzerland
Notes
405-DPAS-SC-VP pH Electrodes 39
405-DPAS-SC-VP pH Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 53 500 197
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-Toledo Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mail mtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mail prozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mail mtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.53500197 www.mt.com/pro
Addresses | Impressum
/