Mettler Toledo pH electrode InPro4550VP Operating instructions

Type
Operating instructions
14 InPro 4550 VP pH electrode
InPro 4550 VP pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 400 Printed in Switzerland
1 Introduction ...................................................................15
2 Safetyinstructions ..........................................................15
3 Productdescription ......................................................... 15
4 Installationandcommissioning .......................................16
5 Cabling ..........................................................................16
6 Calibrationandmeasurement ..........................................16
7 Maintenance ..................................................................16
8 Diagnostics....................................................................17
9 Trouble-shooting(sluggishresponsetime,
insufcientsensitivity) ....................................................17
10 Disposal ........................................................................17
11 Guarantee ......................................................................17
Contents
en
InPro
4550 VP
pH Electrode
Instruction Manual
InProandXerolytareregisteredtrademarks
ofMettler-ToledoGmbHinSwitzerland,theUSA,
theEuropeanUnionandafurthervecountries.
en Instruction Manual
InPro 4550 VP pH electrode 15
© 04 / 16 Mettler-Toledo GmbH InPro 4550 VP pH electrode
Printed in Switzerland 52 002 400
1 Introduction
TheMETTLERTOLEDOelectrodeversionsInPro
®
4550 VP
are low-maintenance, temperature-compensated pH
combination electrodes containing polymer electrolyte
(Xerolyt
®
Plus). An integrated temperature sensor (RTD)
providesautomatic compensation of the electrode slope
toensureaccuratereadings.
The InPro 4550VP electrodes are suitable for pH
measurementsin:
–emulsionsandsuspensions
–sulde-bearingmedia
–solutions containing heavy metals (lead, mercury,
silver,etc.)
–processes with rapid pressure uctuations (within
thegivenrange)
–heavilycontaminatedmedia.
The InPro 4550VP electrodes are unsuitable for
sterilization/autoclaving procedures, for processes with
rapid temperature uctuations or food and beverage
applications.
Pleasereadthroughtheseoperatinginstructionscarefully
before commissioning, in order to ensure trouble-free
use.Thesegeneralinstructionsmustbereadilyavailable
tooperatorsoftheelectrodeatalltimes.
2 Safetyinstructions
TheInPro4550VPelectrodesareintendedsolelyforthe
purposedescribedin§1above.Theelectrodesconsistpartly
ofglassandshouldbehandledcarefully.Iftheelectrodeis
tobecleanedorcalibratedusingacidoralkalinesolutions,
safetygogglesandglovesshouldbeworn.Localsafety
regulationsareapplicabletothehandlingoftheelectrode.
We recommend that you only operate the electrode in
combination with original parts from METTLER TOLEDO.
Operationandmaintenanceshouldbecarriedoutonlyby
trainedpersonnelandstaffwhohavereadandunderstood
theoperatinginstructions.
3 Product description
The inscription on eachelectrode contains the following
information:
METTLER TOLEDO Manufactureroftheelectrode
InPro
®
4550 VP / xxx / SG ypedesignation/xxx=typeof
temperaturesensor(Pt100/
Pt1000)/SG=ancillaryelectrode
combinationpH Typeofelectrode
pH 0 … 14 pHmeasurementrange
0…130°C Temperaturerangeforoperation
Order No. 52 002 40X Ordernumber
SEV14ATEX0168X
Identicationandnumber
ofcerticate
IECExSEV14.0025X
Identicationandnumber
ofcerticate
IS/I,II,III/1/ABCDEFG/T6 FMapproved
Seeinstructionmanual! Instruction(Observetheoperating
instructions)
Inadditioneachelectrodeisallocatedaserialnumberon
theplug-inheadtoenableidentication.
en
16 InPro 4550 VP pH electrode
InPro 4550 VP pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 400 Printed in Switzerland
4 Installation and commissioning
1. Onunpacking,andagainpriortoinstallation,checkthe
electrodevisuallyforsignsofmechanicaldamagetothe
glassmembrane,theshaftandtheconnector.Report
any faults or defects immediately to your authorized
METTLERTOLEDOdealer.
2. Removethewateringcapandrinsetheelectrodewith
distilledordeionizedwater.Useonlysofttissuetodab
theelectrodedry.Donotrubtheglassmembranesince
this can lead to sluggish response behavior of the
electrode.
3. Electrodes with a dry glass membrane must be
conditioned forseveralhoursina3mol KClsolution
before being used.
4. Connectalwaysthesolutionground(bluecable)tothe
transmitter.
5. UsePTFEtapetosealtheNPTthread.
6. If possible use the protection sleeve (order no.
52401808).
5 Cabling
TheInPro4550VPisequippedwithaVarioPinconnector.
Toconnect thecable totheelectrode,attachandgently
rotateuntiltheconnectorisengaged(groove).Screwthe
cablecap nut onto the electrode until the connection is
secure. The plug assembly should be tted hand-tight
only; no tools should be used. For details of the exact
assignmentoftheindividualcableleadstotheVPpins,
please refer to the instruction leaet “Preparation of the
cableforconnectiontothetransmitter”suppliedtogether
withthecable.
6 Calibration and measurement
InPro 4550VP electrodes require two-point calibration.
Before calibration, the watering cap must be removed
from the pH electrode. The electrode is then dipped
successively into two different buffer solutions, each of
knownpHvalue,andthe transmittercalibratedtothese
buffervalues(forexample,bufferspH4.01andpH7.00).
Pleasenotetheinstructionsgivenintheoperatingmanual
ofthetransmitter.Calibrationshouldberepeatedasand
whenneeded.
7 Maintenance
–ElectrodeswithXerolyt
Pluspolymerelectrolytecannot
berelled.
–Electrodes shouldnotbe storeddryfor any periodof
time.
–Duringstorage,thewateringcap should beinplace,
lledwithFriscolyt™or3molKClsolution.
–Neverstoreanelectrodeindistilledordeionizedwater,
asthisshortensthelifetimeoftheelectrode.
–Shouldtheelectrodedryout,itmustbeconditionedfor
24hoursinFriscolyt(orderingno.51340053for250
ml,or51340054for63250ml)beforereuse.After
conditioning,recalibrationisrequired.
–Aftereachworkcycle,thetipoftheelectrodeshouldbe
cleanedusingdeionizedwater.
en
InPro 4550 VP pH electrode 17
© 04 / 16 Mettler-Toledo GmbH InPro 4550 VP pH electrode
Printed in Switzerland 52 002 400
Attention! Do notunderanycircumstanceuseMETTLER
TOLEDOreactivatingsolution(orderingno.51319053/in
USA439895125)asthiscontainshydrouoricacid(HF)
whichwouldattacktheXerolytreferencesystem.
8 Diagnostics
Theelectrodesareequippedwithagroundingringwhich
allows a diagnostic of the diaphragm condition. If this
diaphragm (open junction) should become clogged or
if the reference resistance exceeds a certain value (see
transmitter operation manual), it is then necessary to
replacetheelectrode.InPro4550VPelectrodesshouldasa
generalrulenotbecleanedwithacidiccleaningsolutions,
e.g.diaphragmcleaningsolution,reactivationsolution,etc.
However,byextremelycarefulapplication,itispossibleto
reactivatetheglassmembrane,seepoint9below.
9 Trouble-shooting(sluggishresponsetime,
insufcientsensitivity)
Attention! Theglassmembranemustbeverycautiously
dabbedwith HF solutionusinganebrush andtaking
care that solution does not come into contact with the
titanium pin (metallic solution ground). Also check the
electrodeconnector(VP)forpossibletracesofmoisture.
Ifnecessary,cleanwellusingdeionizedwateroralcohol,
andafterwardsdrycarefully.
10 Disposal
DisposeofInPro4550VPelectrodesinaccordancewith
local or state regulations.
11 Guarantee
Onmanufacturingdefects,12monthsafterdelivery.
en
74 InPro 4550 VP pH electrode
InPro 4550 VP pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 400 Printed in Switzerland
Notes
InPro 4550 VP pH electrode 75
© 04 / 16 Mettler-Toledo GmbH InPro 4550 VP pH electrode
Printed in Switzerland 52 002 400
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-ToledoAnalyseIndustrielleS.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-ToledoGmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo(Schweiz)GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.52002400
www.mt.com/pro
Addresses|Impressum
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76

Mettler Toledo pH electrode InPro4550VP Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI