  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo pH electrodes InPro4501VP and InPro4501 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
InPro 4501 pH Combination Electrode 11
InPro 4501 pH Combination Electrode © 02 / 16 Mettler-Toledo GmbH
52 002 390 Printed in Switzerland
InPro 4501
pH Combination Electrodes
General instructions
1. Introduction
Thesegeneralinstructionsmustbereadilyavailabletoopera-
torsoftheelectrodeatalltimes.Beforeusingtheelectrode,
theseinstructionsmustbereadandunderstoodbyalloperat-
ingpersonnel.
The
METTLER
TOLEDO
electrode
versions
InPro
®
4501VPand
InPro4501xedcablearelow-maintenance,temperature-com-
pensatedpHcombinationelectrodescontainingpolymerelec-
trolyte (Xerolyt
®
Plus), and are hereafter referred to as the
InPro4501 family. An integrated temperaturesensor (RTD)
provides automatic compensation of the electrode slope to
ensureaccuratereadings.
T
heInPro4501familyissuitableforpHmeasurementsin:
e
mulsions and suspensions
– sulde-bearingmedia
– solutions
c
ontaining
h
eavy
m
etals
(
lead,
m
ercury,
s
ilver,
e
tc.)
– processes
w
ith
r
apid
p
ressure
uctuations
(
within
t
he
g
iven
r
ange)
– measurements
in
w
astewater
a
nd
in
d
econtamination
p
ro-
cessesforindustrialefuents
– heavilycontaminatedmedia
ATTENTION:TheInPro4501familyelectrodesareunsuitable
forsterilization/autoclavingprocedures,forpro-
cesseswithrapidtemperatureuctuationsorfood
andbeverageapplications.
2. Safety remarks
TheInPro4501familyofelectrodesisintendedsolelyforthe
purposedescribedin§1above.Theelectrodesconsistpartly
ofglass.Iftheyaretobecleanedorcalibratedusingacidor
alkaline solutions, protective clothing (goggles and gloves)
mustbeworn.Inaddition,localandin-house-safetyregula-
tionsmustbeobserved.
W
e
r
ecommend
t
hat
t
he
e
lectrodes
o
nly
b
e
p
ut
in
to
o
peration
in
c
onjunction
w
ith
o
riginal
p
roducts
/
a
ccessories
f
rom
M
ETTLER
T
OLEDO.
T
he
e
lectrodes
m
ay
b
e
o
perated
a
nd
s
erviced
o
nly
b
y
p
ersonnel
f
amiliar
w
ith
t
he
e
quipment
in
q
uestion
a
nd
w
ho
h
ave
r
ead
a
nd
u
nderstood
t
hese
in
struc-
tions.
3. Product description
Themarkingsoneachelectrodeprovidethefollowinginforma-
tion:
InPro 4501 VP/xxx/yyy Designation/temperaturesensor/
solutionground(SG)
InPro 4501/xxx / yyy/zzz/www Fixedcabledesignation/
temperature
sensor/solution
ground(SG)/cablelength/
cable end
C
ombination electrode typeofsensor
Pt100/Pt1000 Typeofbuilt-intemperaturesen-
sor
p
H 114/0…100 °C pHmeasuringrange/tempera-
ture range
O
rder no. 41 453 310x Ordering number VP
Order no. 10 453 30xx Orderingnumberxedcable
en
en Instruction Manual
12 InPro 4501 pH Combination Electrode
© 02 / 16 Mettler-Toledo GmbH InPro 4501 pH Combination Electrode
Printed in Switzerland 52 002 390
4. Installation and preparation for use
1.Onunpacking,andagainpriortoinstallation,checktheelec-
trodevisuallyforsignsofmechanicaldamagetotheglass
membrane,theshaftandtheconnector.Reportanyfaultsor
defectsimmediatelytoyour authorizedMETTLERTOLEDO
dealer.
2.Removethewateringcapandrinsetheelectrodewithdis-
tilled orde-ionized water.Useonlysofttissuetodab the
electrodedry.Donotrubtheglassmembranesincethiscan
leadtosluggishresponsebehavioroftheelectrode.
3.Electrodeswithadryglassmembranemustbeconditioned
forseveralhoursina3molKClsolutionbeforebeingused.
4.Connectalwaysthesolutionground(bluecable)tothetrans-
mitter.
5.UsePTFEtapetosealtheNPTthread.
6.Ifpossibleusetheprotectionsleeve(orderno.52401808).
5. Cabling
TheInPro4501VPisequippedwithaVarioPinconnector.To
connect thecableto theelectrode,attach and gently rotate
untiltheconnectorisengaged(groove).Screwthecablecap
nutontotheelectrodeuntiltheconnectionissecure.Theplug
assemblyshouldbettedhand-tightonly;notoolsshouldbe
used.Fordetailsoftheexactassignmentoftheindividualcable
leads to the VP pins, please refer to the instruction leaet
“Preparationofthecableforconnectiontothetransmitter”sup-
pliedtogetherwiththecable.
InPro4501
xed
c
able.
T
here
a
re
t
wo
o
ptions
f
or
t
ermination
f
ortheInPro4501withxedcable:
• BNC
c
onnector
f
or
p
H
c
ombination
e
lectrode
w
ith
s
tripped
a
ndtinnedleadsforRTDandsolutionground.
• StrippedandtinnedleadsforpHcombinationelectrode,for
RTDandsolutionground.
Refertoyourtransmitterinstructionmanualforcorrectwiring.
ATTENTION:Thesolutiongroundleadisdesignedtogotoa
galvanicallyisolatedterminalonthetransmitter
andshouldnotbeusedasanoutergroundshield.
Ifthetransmitterdoesnothavesensordiagnos-
tics,thisleadshouldbeleftunterminated.
Color Function
black/transparent Glass
red Coaxshield/reference
grey RTDsensing(temperaturesensor,
3-wire
connection)
blue Solutionground
white RTDreturn(temperaturesensor,
2-wire
connection)
green RTDin(temperaturesensor,
2-wire
connection)
green/yellow Cableshield
6. Calibration and measurement
InPro4501VPelectrodesrequiretwo-pointcalibration.Before
calibration,thewateringcapmustberemovedfromthepH
electrode.Theelectrodeisthendippedsuccessivelyintotwo
differentbuffersolutions,eachof knownpHvalue,and the
transmittercalibratedtothesebuffervalues(forexample,buf-
ferspH4.01andpH7.00).Pleasenotetheinstructionsgiven
intheoperatingmanualofthetransmitter.Calibrationshould
berepeatedasandwhenneeded.
en
InPro 4501 pH Combination Electrode 13
InPro 4501 pH Combination Electrode © 02 / 16 Mettler-Toledo GmbH
52 002 390 Printed in Switzerland
7. Maintenance
– ElectrodeswithXerolytPluspolymerelectrolytecannotbe
relled.
– Electrodesshouldnotbestoreddryforanyperiodoftime
– Duringstorage,thewateringcapshouldbeinplace,lled
withFriscolyt
®
or3molKClsolution.
– Neverstoreanelectrodeindistilledordeionizedwater,as
thisshortensthelifetimeoftheelectrode.
– Should the electrode dry out, it must be conditioned for
24hoursinFriscolyt(orderingno.512340053for250mL,
or51340054for6x250mL)beforereuse.Aftercondition-
ing,recalibrationisrequired.
– After each workcycle, the tip of the electrode should be
cleanedusingde-ionizedwater.
ATTENTION:
D
o
n
ot
u
nder
a
ny
c
ircumstance
u
se
M
ETTLER
T
OLEDO
r
eactivating
s
olution
(
ordering
n
o.
5
1319053
/
in
U
SA
4
39895125)
a
s
t
his
c
on-
tains
h
ydrouoric
a
cid
(
HF)
w
hich
w
ould
a
ttack
t
heXerolytreferencesystem.
8. Diagnostics
Theelectrodesareequippedwithagroundingringwhichal-
lowsadiagnosticofthediaphragmcondition.Ifthisdiaphragm
(openjunction)shouldbecomecloggedorifthereferencere-
sistance exceeds a certain value (see transmitteroperation
manual),itisthennecessarytoreplacetheelectrode.Elec-
trodesoftheInPro4501familyshouldasageneralrulenotbe
cleanedwithacidiccleaningsolutions,e.g.diaphragmclean-
ingsolution,reactivationsolution,etc.However,byextremely
carefulapplication,itispossibletoreactivatetheglassmem-
brane,see«point9»below.
9. Troubleshooting(sluggishresponsetime,insuf-
cientsensitivity,etc.)
ATTENTION: The glass membrane must be very cautiously
dabbedwithHFsolutionusinganebrushandtakingcarethat
solutiondoesnotcomeintocontactwiththetitaniumpin(me-
tallic solution ground). Also check the electrode connector
(onlyVPversion)forpossibletracesofmoisture.Ifnecessary,
cleanwellusingdeionizedwateroralcohol,andafterwards
drycarefully.
10. Disposal
DisposeofInPro4501familyelectrodesinaccordancewith
localorstateregulations.
11. Guarantee
METTLERTOLEDOgrants12monthswarrantycoveringmanu-
facturingdefects,validfromthedateofproductionoftheelec-
trode.
InPro,
X
erolyt
a
nd
F
riscolyt
a
re
t
rademarks
o
f
t
he
M
ettler
T
oledo
G
roup.
Allothertrademarksarethepropertyoftheirrespectiveholders.
en
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;
service@mettler.com.br
France Mettler-Toledo
Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo(Schweiz)GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-Toledo GmbH
ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
02/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.52002390
www.mt.com/pro