Mettler Toledo pH electrode InPro3100/InPro3100UD User manual

Type
User manual

This manual is also suitable for

14 InPro 3100 (i) pH electrode
InPro 3100 (i) pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 293 Printed in Switzerland
1 Introduction ...................................................................15
2 Safetyinstructions ..........................................................15
3 Product description ......................................................... 15
4 Installationandcommissioning .......................................16
5 Operation .......................................................................16
5.1 CalibrationoftheelectrodeandpHtransmitter ..................16
5.2 Sterilization of electrode ..................................................16
5.3 TemperaturesensorofInPro3100(analogversion) ...........16
6 Maintenance ..................................................................17
7 Trouble-shooting(sluggishresponsetime,insufcientsensiti-
vityetc.) ........................................................................17
8 Disposal
........................................................................17
9 Guarantee ......................................................................17
Contents
en
InPro
3100 (i)
pH Electrode
Instruction Manual
InProandISMareregisteredtrademarks
ofMettler-ToledoGmbHinSwitzerland,theUSA,
theEuropeanUnionandafurthervecountries.
12 mm
0.47"
Connectortype(here:VP)
Bridgeelectrolyte
Silver-iontrapanddiaphgragm
Referenceelectrolyte
Temperaturesensor
Argenthalreferenceelement
Diaphragm
pH sensitive glass
en Instruction Manual
InPro 3100 (i) pH electrode 15
© 04 / 16 Mettler-Toledo GmbH InPro 3100 (i) pH electrode
Printed in Switzerland 52 002 293
1 Introduction
The METTLER TOLEDO InPro 3100(i) pH electrodes are
low-maintenance, autoclavable and pressure-resistant
measuringelectrodeswithagel-typereferenceelectrolyte.
The reference element incorporates a silver-ion trap to
preventsilversuldeprecipitationonthediaphragm.Incase
ofInPro3100,atemperaturesensorisintegrated,mainlyfor
automatictemperaturecompensationpossibilitiesofthepH
signal during calibration and operation.
The InPro 3100(i) electrodes are mainly designed for
the measurement of pH value in biotechnical processes
at moderate pressures (<6bar). It can be used under
industrial fermentation, pilot and laboratory conditions,
and is particularly intended for those processes where
intermediatesterilizationorautoclavingisindispensable.
Pleasereadthroughtheseoperatinginstructionscarefully
beforecommissioning,inordertoensuretrouble-freeuse.
2 Safetyinstructions
The InPro 3100(i) electrodes are intended solely for the
purposedescribedin§1above.Theelectrodeconsistspartly
ofglassandshouldbehandledcarefully.Iftheelectrodeis
tobecleanedorcalibratedusingacidoralkalinesolutions,
safetygoggles and gloves shouldbeworn.Localsafety
regulationsareapplicabletothehandlingoftheelectrode.
Pleasefollowtheguidelinesenclosedforaninstallationin
theEx-zone.
We recommend that you only operate the electrode in
combination with original parts from METTLER TOLEDO.
Operationandmaintenanceshouldbecarriedoutonlyby
trainedpersonnelandstaffwhohavereadandunderstood
theoperatinginstructions.
3 Product description
The inscriptionon each electrode contains thefollowing
information:
METTLER TOLEDO Manufactureroftheelectrode
InPro
®
310XiUD/SG/yyy/zzzTypedesignation;X=membrane
glass;i=ISMDigital;UD=upside
down/SG=ancillaryelectrode/
yyy=rodlengthinmm/zzz=type
oftemperaturesensor
combinationpH Typeofelectrode
pH0…14 pHmeasurementrange
0…80°C/140°C(130°C) Temperaturerangeforoperation/
sterilization(forInPro3100UD)
OrderNo.5200XXXX Ordernumber
SEV14ATEX0168X Identicationandnumber
ofcerticate
IECExSEV14.0025X Identicationandnumber
ofcerticate
IS/I,II,III/1/ABCDEFG/T6 FMapproved
Seeinstructionmanual! Instruction(Observethe
operating instructions)
Inadditioneachelectrodeisallocatedaserialnumberon
theplug-inheadtoenableidenticationandtraceability.
en
16 InPro 3100 (i) pH electrode
InPro 3100 (i) pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 293 Printed in Switzerland
4 Installationandcommissioning
1. On unpacking, check the electrode for mechanical
damage of the pH sensitive glass and shaft. Report
any damage immediately to your METTLER TOLEDO
supplier.
2. Removethewateringcapandbrieyrinsetheelectrode
withdeionizedwater.Afterrinsing,theelectrodeshould
onlybedabbeddrywithatissue.DonotrubthepH
sensitive glass, since this can lead to electrostatic
chargingandsluggishresponsetimes.
3. CheckthespacebehindthepHsensitiveglassforthe
presenceofanyairbubblesandremovesamebygently
swingingtheelectrodeinaverticalplane.
4. Please refer to the instruction manual of the relative
housinginordertoinstalltheelectrode.
5. Connect the electrode to the pH transmitter, using a
cable with the appropriate connector (VP connector
forInPro3100).Pleaserefertothewiringandterminal
diagramaccompanyingthecable.
6. Intelligent ISM
®
electrodessuchasInPro310Xienable
“Plug and Measure” and extended diagnostics. For
installation, commissioning and operation of the
ISM system please observe the instructions for the
transmitter,moduleandcable.
5 Operation
5.1CalibrationoftheelectrodeandpHtransmitter
FortheInPro3100(i)electrodea2-pointpHcalibrationis
recommended.Beforecalibration,rstremovethewatering
cap.Tospeeduptheresponsetimeoftheelectrodeafter
longtransportandstorageperiods,andbeforecalibration,
diptheelectrodeintobufferpH7,bufferpH4andbufferpH
7again,eachtimeforaboutoneminute.Theelectrodeis
thendippedsuccessivelyintotwodifferentbuffersolutions
withgivenpHreferencevaluesandthetransmittercalibrated
tothesebuffervalues.NormallyyoustartwithbufferpH7
fordeterminingthe0-pointandthenbufferpH4or9for
determiningtheslope.Inthecaseofmicroprocessor-based
transmitters with buffer recognition,the order of the two
chosenbuffer solutionsis irrelevant.Whenever possible,
atemperaturesensor(forInPro3100theintegratedone)
should be connected to the transmitter and automatic
temperaturemeasurement”selectedthere.Pleasereferto
theinstructionmanualofthetransmitterforfurtherdetails.
5.2Sterilizationofelectrode
TheInPro3100(i)electrodecanbesterilizedeitherthrough
autoclaving (remove red protection cap) or through in-
situ sterilization with steam or superheated fermentation
medium. Sterilization temperatures above 140°Care not
allowed,andexcessivesterilizationtime(forinstancet>30
minat140°C)could notably shorten the life time of the
electrode.Repeatedsterilizationand/orautoclavingofthe
electrodecanleadtoaslightcolorationoftheelectrolyte.
This color change does not effect the functioning of the
electrode.
5.3TemperaturesensorofInPro3100(analogversion)
The integrated Pt100/1000 temperature sensor in the
InPro3100isprimarilyintendedforautomatictemperature
compensationofthepHsignalandnotforanyaccurate
andsafetemperatureindicationorcontroloftheprocess
temperature. If you anyhow want to try this, check that
yourtransmitter has the required temperature indication
accuracyandafree outletforthesignal.Furthermore,a
calibrationofthetemperaturesignalmightbenecessary.
en
InPro 3100 (i) pH electrode 17
© 04 / 16 Mettler-Toledo GmbH InPro 3100 (i) pH electrode
Printed in Switzerland 52 002 293
6 Maintenance
1. Carefully rinse the electrode tip and diaphragm with
deionizedwateraftereachproductioncycle.Underno
circumstancesmustmeasuringsolutionbeallowedto
dryontheseparts!
2. When the electrode is not in operation, store it with
electrode tip and diaphragm well submerged into
9823
/3MKClor9816/Viscolyt™electrolyte.
3. Iftheelectrodeisstoredmountedinitshousing,the
same rule as in 2) above applies, but the storage
electrolytehastobeslightlymodied,with2partsof
buffersolutionpH9.2to10partsofelectrolyte,thisin
ordertoavoidcorrosionofhousingparts.
4. Ifanelectrodeisstoreddryforafewdaysbymistake,
letitsoakinthenormalstorageelectrolyteforseveral
hoursbeforeuse.
5. Occasionallychecktheconnectorforpossibletracesof
moisture.Ifnecessary,cleanwellwithdeionizedwater
oralcohol,andafterwardsdrycarefully.
7 Trouble-shooting(sluggishresponsetime,
insufcientsensitivityetc.)
TheglassmembranecanbereactivatedusingaHFsolution
(orderno.51319053).Theelectrodetip(onlypHsensitive
glasspart)isimmersedintotheHFsolutionforaperiodof
about5minutes,isthenimmediatelyandthoroughlyrinsed
withdeionizedwaterandthereafterstoredinbuffersolution
pH4(seepoint6above)uptothediaphragmfor12hours.
Po s siblep rote inpre c ipita t ionandblocka geoft h e diaphra g m
canbetreatedwithourspecialelectrodecleaner(orderno.
51340068).However,useofthediaphragmcleaner(order
no.51340070)isnormallynotnecessaryforthiselectrode
type,asthebuilt-insilver-iontrappreventstheformationof
silversuldecontamination.
Also,checktheconnectorforpossibletracesofmoisture.If
necessary,cleanwellusingdeionizedwateroralcohol,and
afterwardsdrycarefully.
8 Disposal
If no relative ofcial regulations are in force, used or
defectiveInPro3100(i)electrodesandthepackagingcan
bedisposedofasordinaryhouseholdwaste.
9 Guarantee
Onmanufacturingdefects,12monthsafterdelivery.
en
74 InPro 3100 (i) pH electrode
InPro 3100 (i) pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 293 Printed in Switzerland
Notes
Brazil Mettler-ToledoInd.eCom.Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-ToledoAnalyseIndustrielleS.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-ToledoGmbH
Prozeßanalytik,Ockerweg3,DE-35396Gien
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo(Schweiz)GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.52002293
www.mt.com/pro
Addresses|Impressum
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76

Mettler Toledo pH electrode InPro3100/InPro3100UD User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI