  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo InPro 42XX (i) Series pH Electrode User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
14 InPro 42XX (i) Series pH electrode
InPro 42XX (i) pH electrode © 04 / 16 Mettler-Toledo GmbH
52 005 365 Printed in Switzerland
1 Introduction ...................................................................15
2 Safetyinstructions ..........................................................15
3 Product description .........................................................15
4 Installation and commissioning .......................................16
5 Operation:Calibrationoftheelectrode
andpHtransmitter .........................................................16
6 Maintenance ..................................................................16
7 Removalofsourcesofmalfunction ..................................17
8 Disposal ........................................................................17
9 Guarantee ......................................................................17
10 Storage conditions ..........................................................17
Contents
en
InPro
42XX (i) Series
pH Electrode
Instruction Manual
InPro,ISMandXerolytareregisteredtrademarks
ofMettler-ToledoGmbHinSwitzerland,theUSA,
theEuropeanUnionandafurthervecountries.
en Instruction Manual
InPro 42XX (i) Series pH electrode 15
© 04 / 16 Mettler-Toledo GmbH InPro 42XX (i) pH electrode
Printed in Switzerland 52 005 365
1 Introduction
TheMETTLERTOLEDOInPro
®
42XX(i)pH-singlerod
measuringcellisalowmaintenance,high-pressure
resistantpHelectrodewithasolidpolymerelectrolyte
(Xerolyt
®
Extra) together with a silver ions barrier.
Thediaphragmisofanopenaperturetype,sothat
direct contact exists between the medium being
measuredandtheelectrolyte.
Please read through these operating instructions
carefully before commissioning, in order to ensure
trouble-freeuse.
2 Safety instructions
When cleaning or calibrating the electrode using
solutionscontainingacidsoralkalisbothprotective
spectaclesandprotectiveglovesshouldbeworn.
Donottouchtheelectrodeattheplug-incontacts,since
electrostaticdischargescandamagetheelectronics.
Werecommendthatyouonlyoperatetheelectrodein
combinationwithoriginalpartsfromMETTLERTOLEDO.
Operationandmaintenanceshouldbecarriedoutonly
by trained personnel and staff who have read and
understoodtheoperatinginstructions.
3 Product description
The inscription on each electrode contains the
followinginformation:
METTLER TOLEDO Manufactureroftheelectrode
InPro
®
42XX i / SG / yyy / zzz Typedesignation;XX=shaft
material,membraneglass;
i=ISMDigital/SG=ancillary
electrode/yyy=rodlengthin
mm/zzz=typeoftemperature
sensor
combination pH Typeofelectrode
pH 0 14 pH measurement range
pH 1 14 (InPro 4281 i)
0…100°C(130°C) Temperaturerangefor
operation (sterilization)
Order No. 52 00X XXX Order number
SEV14ATEX0168X Identicationandnumber
ofcerticate
IECExSEV14.0025X Identicationandnumber
ofcerticate
IS/I,II,III/1/ABCDEFG/T6 FMtested
Seeinstructionmanual! Instruction(Observethe
operating instructions)
In addition each electrode is allocated a serial
numberontheredpartoftheplug-inheadtoenable
identication.
en
16 InPro 42XX (i) Series pH electrode
InPro 42XX (i) pH electrode © 04 / 16 Mettler-Toledo GmbH
52 005 365 Printed in Switzerland
4 Installation and commissioning
1. While unpacking check the electrode for
mechanicaldamage.Pleasereportanydamage
immediatelytoyourMETTLERTOLEDOsupplier.
2.Removetheprotectivecapandrinsetheelectrode
for a short time with deionized water. If the
electrodehasalreadybeenusedthesensitivepart
canbecleanedwithasofttoothbrushandamild
soapsolution,andthenrinsedforashorttimewith
deionizedwater.Afterrinsingtheelectrodeshould
onlybedabbed.AnyrubbingofthepHsensitive
glasscanincreasetheresponsetimeasaresult
ofelectrostaticchargeaccumulation.
3.Check the spacebehind thepHsensitive glass
forairbubblesandremoveanyairbubblesthat
maybepresentbyshakingtheelectrodeslightly
in a vertical direction.
4.Placetheelectrodeintothehousingasdescribed
intheinstructionsforthehousing.
5.ConnecttheelectrodeandthepHtransmitterwith
thecorrespondingcable(VP6,VP8,AK9).Observe
the connections scheme that accompanies
thecableorthetransmitter.
6. Intelligent ISM
®
sensorssuchasInPro426Xienable
“PlugandMeasure” andexpandeddiagnostics.
Forinstallation,commissioningandoperationof
theISMsystempleaserefertoinstructionsforthe
transmitter, module and cable.
5 Operation: Calibration of the electrode
and pH transmitter
Before a calibration immerse the electrode for
10minutesinapH7.00orpH4.01buffersolution
whilst connected it to the transmitter. A 2-point
calibration is recommended, e.g. pH 7.00 and pH
4.01buffersolutions.Forfurtherdetailspleaseconsult
theoperatinginstructionsforthepHtransmitter.
6 Maintenance
1. After each working cycle the electrode tip and
diaphragm have to be carefully rinsed with
deionized water. Any remaning of the solution
beingmeasuredonthesepartsmustbeavoided
at all costs!
2. Iftheelectrodeisnotinuseithastobestoredwith
theelectrodetipandthediaphragmimmersedin
atankofelectrolyte(9823/3MKClorFriscolyt™
9848).).
en
InPro 42XX (i) Series pH electrode 17
© 04 / 16 Mettler-Toledo GmbH InPro 42XX (i) pH electrode
Printed in Switzerland 52 005 365
3. Iftheelectrodeisstoredinitshousingtheadvices
described under Item 2 apply, although the
electrolytehastobeslightlymodiedforstorage
purposes (2 parts of buffer solution pH 9.2 to
10 parts of electrolyte) in order to avoid any
corrosionofthehousingparts.
4. If an electrode is inadvertently stored in a dry
condition for several days it must be rinsed
forseveralhoursinastandardstorageelectrolyte.
5. Theplug-inheadshouldoccasionallybechecked
forpossibletracesofmoisture.Ifnecessaryclean
itthoroughlywithdeionizedwateroralcoholand
thendryitoffcarefully.
7 Removal of sources of malfunction
Withasofttoothbrushandwateranyresiduescan
easilyberemovedfromtheelectrode.Mildwashing
agents can also be used.
8 Disposal
Itistheusersresponsibilitytodisposeofthesensor
inaprofessional manner. The sensor can contain
electroniccomponentsthatrequiredisposalsuchthat
nopersonsortheenvironmentareendangered.
9 Guarantee
Onmanufacturingdefects12monthsafterdelivery.
10 Storage conditions
Werecommendthattheelectrodesarestoredatroom
temperature.
en
74 InPro 42XX (i) Series pH electrode
InPro 42XX (i) pH electrode © 04 / 16 Mettler-Toledo GmbH
52 005 365 Printed in Switzerland
Notes
InPro 42XX (i) Series pH electrode 75
© 04 / 16 Mettler-Toledo GmbH InPro 42XX (i) pH electrode
Printed in Switzerland 52 005 365
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-Toledo Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.52005365
www.mt.com/pro
Addresses|Impressum
/