8 Type Xerolyt Electrodes
© 04 / 16 Mettler-Toledo GmbH Type Xerolyt Electrodes
59 905 939 Printed in Switzerland
1 Introduction
Thesegeneralinstructionsmustbereadilyavailabletoope-
ratorsoftheelectrodeatalltimes.Beforeusingtheelectro-
de,theseinstructionsmustbereadandunderstoodbyall
operatingpersonnel.
Owingtotheirspecialcharacteristics,pH/redoxelectrodes
withaXerolytreferencesystemaresuitableforthefollowing
applications:
–Measurementsinhighlycontaminatingsolutions
Measurements in emulsions and suspensions
–Measurementsinefuents/wastewater
Measurements in solutions containing proteins
–Measurementsinsolutionswithahighsuldecontent
–Measurementsinhotalkalinesolutions(withTypeHApH
glass)
–Installationinmeasuringandcontrolloopswithwidepres-
sureuctuations.
Xerolyt electrodes are notsuitable forthe followingappli-
cations:
–Measurementswheretheelectrodeissteam-sterilized
–Measurementswithlarge, long-term, temperature varia-
tions
–MeasurementsinsolutionswithapHvalueofpH<2
–Measurementsinsolutionscontainingchlorine
2 Safety remarks
TheXerolyt electrodesareintendedsolelyfor thepurpose
describedin§1above.Theelectrodesconsistpartlyofglass.
Iftheyaretobecleanedorcalibratedusingacidoralkaline
solutions, protective clothing (goggles and gloves) must
beworn.Inaddition,localandin-house-safetyregulations
mustbeobserved.Werecommendthattheelectrodesonly
beputintooperationinconjunctionwithoriginalproducts/
accessoriesfromMETTLERTOLEDO.Theelectrodesmaybe
operatedandservicedonlybypersonnelfamiliarwiththe
equipmentinquestionandwhohavereadandunderstood
theseinstructions.
3 Product description
ThefollowingdataapplyonlytotheXerolytreferencesystem.
(With single-rod pH or redox electrodes assemblies, the
specicationoftheindicatorelectrodeshouldalsobetaken
intoconsideration.)
Themarkings oneach electrodeprovidethefollowingin-
formation:
XXX-DXK-S8/yyydesignation/lengthmm
xxx=405combinationelectrode
xxx=363referenceelectrode
pH2…14,0°…+110°C=pHmeasuringrange/tempe-
rature range
Ordernumber:xxxxxxxxx
en
en Instruction manual
General instructions
for Xerolyt-Electrodes
Type Xerolyt Electrodes 9
Type Xerolyt Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 59 905 939
en
4 Installation and preparation for use
a Carefullyinspecttheelectrodeondeliveryformechanical
damagetothemembraneortheshaft.Electrodesshowing
anydefectsshouldbesentbackintheoriginalpackaging,
accompaniedbytherelativeinvoicenumber.
b Removethewateringcapandrinseelectrodewithdistilled
water. After rinsing, a pH electrode unit should be dried
bycarefuldabbingwithapaperhandkerchieforKleenex
tissue.Donotrubtheglassmembrane,sincethiscanlead
toverysluggishresponse.
c Removeanyairbubblespresentinthemembranechamber
oftheglasselectrodebylightlyswingingroundinavertical
plane.
d pHelectrodeassemblieswithdryglassmembranesmust
beconditionedbyimmersionforseveralhoursin3mol/l
KClinthesuppliedwateringcap(useapprox.1ml).
5 Calibration and measurement
Xerolytelectrodesrequire2-pointcalibration.Tocalibrate,
rstremovethewateringcap.Theelectrodeisthendipped
insuccessionintotwodifferentbuffersolutionswithgiven
pHreferencevalues,andthetransmittercalibratedtothese
buffervalues(forexamplebufferpH4.01andpH7.00).
Pleaserefertotheinstructionmanualofthetransmitterfor
details.
6 Maintenance
The electrode with polymer electrolyte Xerolyt cannot be
relled.Theelectrodemustnotbestoreddry.Intheevent
thattheelectrodedriesout,itisnecessarytoreconditionit
beforeusebyimmersioninFriscolyt™(Orderno.51340
053)foratleast24hours.Theelectrodeshouldbestored
withttedwateringcapcontaining3mol/lKClorMettler-
Toledo Friscolyt. After each working cycle, the electrode
tip and the apertures (diaphragms) must be thoroughly
cleanedusingde-ionizedwater.Regenerationoftheglass
membraneofapHelectrodewithHF(hydrouoricacid)is
notrecommended,sinceHFwillattacktheXerolytreference
system.
7 Disposal
Ifnorelativeofcialregulationsareinforce,usedordefecti-
veelectrodeandthepackagingcanbedisposedasordinary
householdwaste.
8 Warranty
Intheeventofmanufacturingfaults,12monthswarranty
fromdateofproductionisgranted.
XerolytandFriscolytaretrademarksoftheMettler-ToledoGroup.
38 Type Xerolyt Electrodes
© 04 / 16 Mettler-Toledo GmbH Type Xerolyt Electrodes
59 905 939 Printed in Switzerland
Notes
Type Xerolyt Electrodes 39
Type Xerolyt Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 59 905 939
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-Toledo Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.59905939
www.mt.com/pro
Addresses|Impressum
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40

Mettler Toledo DXK pH electrodes (Xerolyt) Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI