Midmark VetPro Solo Installation guide

Type
Installation guide

This manual is also suitable for

English-1
© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
TP200 Rev. A
004-1080-00 Rev. A (9/13)
Pre-installation Guide for VetPro
®
5000 Dental System
Note
Please allow time for basic equipment training after the installation (approx 10 minuntes) The complete training is available
in-service. To schedule an in-service please contact Vet Customer Service at Midmark, 1-800-Midmark (1-800-643-6275).
Name of Clinic___________________________________Contact Name__________________________________
Address_____________________________________________________________________________________
Phone_________________________________Email:_________________________________________________
1. What method will the delivery system be mounted? (Wall, cabinet, or table)_________________________
2. Will a junction box or solo unit be utilized? __________________
Note: Outlet must be within 5 ft of the Solo unit; If a junction box is used please refer to the junction box template.
3. Is power available? ______________________
Note: Review power requirements within this pre-install guide.
4. If “NO” to Q.3 what date will the power be available? ______________
5. Will this be a dedicated line?____________Midmark strongly recommends a dedicated line.
Note: Electrician will need a template to determine exact location for the electrical wiring to be installed when using a junction
box.
6. Is there anything that might interfere with the dental system in its home position? _________________________________
Refer to minimum space requirements in this guide.
7. Is the stud material wood or metal? ________________
Note: Site must have two studs available to complete the install for wall mounted units.
Style A
midmark.com
English-2
© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
TP200 Rev. A
Suggested Mounting Location From Procedure Table
Delivery Unit can be installed on left or right side of table or cabinet.
180° Arm Length
Standard - 28”
Short - 13 1/2”
250°
Arm Length
Standard - 45.3”
Short - 37 5/16”
Mounting Requirements
Mounting Structure Note: The mounting structure must be capable of supporting the
weight and torque listed below without deection.
Unit / Arm Assembly Maximum
Weight 50lbs (22.6 kgs)
Torque 170.8 ft. lbs. (231.5 N•M)
Note: To prevent unit from drifting, the mounting plate must deect no more than 1/16” (1.6 mm)
over a 12” (30.5) distance.
Cabinet or Wall
Cabinet or Wall
Note
WHIP head folded against wall measures 30.6” from center of main axis to outermost point.
Specications
Water Module
Reservoir One - 1 Liter
City Water Toggled with Reservoir
Pressurized System 30 - 50 PSI
Aseptic Non-Retracting HPC Coolant
Handpiece Flush System Manual
Recommended Pressure Settings (Junction Box)
“City” Water Regulator Valve 30 PSI
Air Regulator Valve 80 PSI
Water Bottle Regulator Valve 30 PSI
Electrical
Voltage 115 VAC
Circuit Breaker Protection 15 amps
Frequency 60 Hz
Space Requirements
Style A
midmark.com
English-3
© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
TP200 Rev. A
Range of Motion
Style A
midmark.com
English-4
© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
TP200 Rev. A
Junction Box
Equipment Alert
When installing 115 VAC receptacle in the junction box,
the ground must be positioned as shown on diagram/template
(in this manual) to accommodate power cord.
Electrical Requirements
• 115 VAC models only - Grounding reliability can only be achieved if this equipment is connect
to an equivalent receptacle marked Hospital Grade (NEMA 5-15R, HOSPITAL GRADE).
• 1/2” (13mm) conduit box with quad receptacle.
• Wire box per local codes and regulations.
Water Supply Requirements
• 1/2” (13 mm) pipe (supplied by contractor) protruding 2” (51 mm) above finished surface of floor.
• Manual shut off valve ( supplied by Midmark) to be installed by contractor.
• Water pressure must be 30-50 (max.) PSI at the shut-off valve. Flush water plumbing clean before
making final connections to dental equipment.
Air Supply Requirements
• 1/2” (13 mm) pipe (supplied by contractor) protruding 2” (51 mm) from
finished surface of floor.
• Manual shut off valve ( supplied by Midmark) to be installed by contractor.
Air pressure must be 80 - 100 (max.) PSI at the shut-off valve.
• Flush air plumbing clean before making final connections to dental equipment.
15 1/2 “
(39.4 cm)
7.3“
(18.5 cm)
15 1/2 “
(39.4 cm)
ATTENTION: Junction Box Must be Positioned
Within Reach of the Umbilical.
Style A
midmark.com
English-5
© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
TP200 Rev. A
4.75"
(12.1 cm)
1.5"
(3.8 cm)
1.75"
(4.4 cm)
Water
Air
Electrical
15.48"
(39.3 cm)
15.48"
(39.3cm)
Universal Power Supply
7.74"
(19.7cm)
Umbilical
7.74"
(19.7 cm)
.25"
(0.6 cm)
2.00"
(5.1 cm)
Junction Box Template - Not to Scale
Note: Full Size Templates is included with Junction Box
Style A
midmark.com
English-6
© 2013 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
TP200 Rev. A
Solo
Fan
Dimensions
Width 12.75 inches (32.4 cm)
Length 20.89 inches (53 cm)
Height 11.40 inches (29 cm)
Weight 54 lbs ( 24.5 kg)
Shipping Weight 61 lbs ( 27.7 kg)
Electrical
Voltage 115 VAC
Frequency 60 Hz
Circuit Breaker Protection 15 amps
Hospital Grade Wiring NEMA type: 14 and 18 AWG
Accessory Outlet (Side Mounted,
User Specs)
GFCI 5 amps, 125 VAC
Receptacles (Internal) NEMA Outlet, Snap AC Connector,
15 amp, 125 VAC
Plug 15 amps, 125 VAC
Power Cord Length 5 ft. NEMA 5-15R
Compressor
Motor Oil Free 1/3 HP
Working Pressure 80 - 100 PSI
Flow Rate 2.6 scfm @ 45 PSI
Sound Level (full load) < 55 dB
Air Tank 1 gal 3.79 Liter
Water Connection
Water Inlet Length 10 ft. (3 m)
Water Inlet Thread Size 3/8” FNPT
Saddle Valve Adapter Length 2 ft. (.61 m)
Saddle Clamp Range
3/8” - 1
5/16
” OD Copper Pipe
Accessory Outlet
On / Off Switch
Drain Valve
City Water Inlet
Connection
Umbilical & Foot
Control Cord
Entry
Specications
Style A
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6

Midmark VetPro Solo Installation guide

Type
Installation guide
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI