Thermo Fisher Scientific CTS Cellmation Software User guide

Type
User guide
For Research Use or Manufacturing of Cell, Gene, or Tissue- Based
Products.
CTS Cellmation Software for DeltaV
System
USER GUIDE
for use with: CryoMed Controlled Rate Freezer with OPC UA
Catalog NumbersA59196, A59198
Publication Number MAN0029867
Revision A.0
Life Technologies Holdings Pte Ltd | Block 33 | Marsiling Industrial Estate Road 3 | #07-06, Singapore 739256
Revision history:MAN0029867 A.0 (English)
Revision Date Description
A.0 5 October 2023 Release 2.0 of CTS Cellmation Software for DeltaV System.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE
LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR
ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
NOTICE TO PURCHASER: DISCLAIMER OF LICENSE: Purchase of this software product alone does not imply any license under any
process, instrument or other apparatus, system, composition, reagent or kit rights under patent claims owned or otherwise controlled by
Thermo Fisher Scientific, either expressly, or by estoppel.
TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. DeltaV is a
trademark of Emerson Process Management.
©2023 Thermo Fisher Scientific Inc. All rights reserved.
Contents
CHAPTER1Productinformation .................................................. 5
Productdescription ............................................................. 5
Operationaldescription .......................................................... 6
Licenses ................................................................... 6
Definitions ................................................................. 6
Toolbar buttons ............................................................. 8
Hardware overview .............................................................. 9
Physicalconnection ............................................................ 11
CryoMed instrument networksetup ............................................. 12
EIOC networksetup ............................................................ 14
EIOC OPC UAsetup ........................................................... 15
Network connectivitycheck ..................................................... 16
OPC UA connectivitycheck ..................................................... 17
Continuous Historiansetup ...................................................... 18
User interface ................................................................. 18
Splash screen ............................................................. 19
Overviewdisplay .......................................................... 20
Main CryoMed instrumentdisplay .......................................... 21
CryoMed instrument interface parametersdisplay ............................ 23
Contextual displays (Faceplates) ............................................. 26
CHAPTER2General system operation .......................................... 34
Userlogin ..................................................................... 34
Manual operation .............................................................. 35
Step 1: Acquire the equipmentmodule ....................................... 35
Step 2: Setup the equipmentmodule ......................................... 37
Step 3: Setup the equipment module to Run Protocol .......................... 39
Step 4: Monitor Protocol state automatically selected .......................... 40
Step 5: Protocolcompletion ................................................. 40
Recipe control operation ........................................................ 41
Phases ................................................................... 44
Phase recipe parameters ................................................... 48
Recipeconstruction ........................................................ 50
Recipe execution .......................................................... 54
CTS Cellmation Software for DeltaV System User Guide 3
CHAPTER3Alarmmanagement ................................................. 58
Alarmbanner .................................................................. 59
Faceplate alarmwindow ........................................................ 59
Alarmslists .................................................................... 60
Acknowledging and silencingalarms ............................................. 61
Suppressingalarms ............................................................ 62
Shelvingalarms ........................................................... 63
Alarm listshelve ........................................................... 63
Detail faceplateshelve ..................................................... 64
Out of Service (OOS)alarms ................................................ 64
Alarm listOOS ............................................................ 65
Detail faceplateOOS ....................................................... 66
Suppressed alarmlist ...................................................... 66
CHAPTER4Maintenance ......................................................... 68
Backup and recovery ........................................................... 68
Computercleanup ............................................................. 68
Poweroutage ................................................................. 69
CHAPTER5Data storage ......................................................... 70
APPENDIXAAbbreviations/acronyms .......................................... 73
APPENDIXBFaceplatedetails ................................................... 74
GT_MOD_FP .................................................................. 75
GT_EM_FP .................................................................... 76
GT_USM_DT .................................................................. 78
GT_EM_DT .................................................................... 79
GT_EM_FM_DT ............................................................ 80
GT_EM_Params_DT ........................................................ 81
GT_EM_Diag_DT .......................................................... 82
APPENDIXCCompatibilitymatrix ............................................... 83
APPENDIXDSafety ............................................................... 84
APPENDIXEDocumentation and support ...................................... 85
Customer and technical support ................................................. 85
Limited product warranty ........................................................ 85
Contents
4CTS Cellmation Software for DeltaV System User Guide
Product information
IMPORTANT! Before using this product, read and understand the information in the “Safety” appendix
in this document.
Product description
The Gibco CTS Cellmation Software (based on the Emerson DeltaV platform) is an open, fully-
configurable software solution designed to provide a flexible system for the control of Thermo Fisher
Scientific cell therapy manufacturing instruments. CTS Cellmation Software is powered by the
DeltaV system from Emerson Process Management, and is fully compatible with DeltaV system
version 14.LTS. The software has been developed according to GAMP5 methods and conforms to
regulatory requirements for use in current good manufacturing practice (cGMP)-compliant processes. It
is optimized to allow easy scaling to accommodate R&D, process development, and manufacturing.
The software must be installed on all of the DeltaV workstations controlling/viewing cell therapy
manufacturing operations. The exceptions are the application stations that may be used for purposes
other than control (e.g., continuous history storage).
The CTS Cellmation Software requires a DeltaV MQ, MX, SQ, SX or PK controller and an OPC
UA Client interface. The OPC UA Client can reside on an Ethernet Input Output Card (EIOC) –
recommended or on an Application Station.
Neither Emerson Process Management nor Thermo Fisher Scientific recommend or support installation
of any third-party software. Downloading files from the internet is not recommended. See Emerson KBA
(knowledge based article) AP-0800-0025.
The software system is intended for use by personnel with a reasonable familiarity with bioprocesses
and systems. The user is encouraged to become familiar with the manuals to obtain maximum
performance from the system.
The contents of this guide may be copied into customer documents such as standard operating
procedures (SOPs) as needed, but only for use with Thermo Fisher Scientific equipment.
Every eort has been made to ensure the accuracy of the information in this document. However,
Thermo Fisher Scientific reserves the right to change its products without notice.
The software is designed to provide the user with the ability to use a wide range of cell therapy
manufacturing instruments. It will also allow the user to:
Use dierent cell therapy manufacturing instruments in the same network with no change in the
operator interface.
Create user specific batch recipes to control workflows across multiple instruments.
Store historical data for retrieval later on.
Allow for 3rd party batch reporting packages to retrieve data to produce batch reports containing
alarm/events, continuous history and batch recipe information.
1
CTS Cellmation Software for DeltaV System User Guide 5
Operational description
CTS Cellmation Software provides a control solution for cell therapy process workflows that is
independent of process-control hardware. It can be configured by end users without having to learn
process control programming and resides on the DeltaV system by Emerson Process Management.
Licenses
CTS Cellmation Software is licensed per DeltaV system.
Definitions
DeltaV Books Online
A set of help files included with DeltaV systems. These files comprise the ocial documentation of the
standard DeltaV system and include parameter and feature definitions.
Process value
Also referred to as process variable or measured variable. In process control, a measurable quality or
quantity that, if changed, can cause other changes in the process or make such changes possible.
Common process values are pump speed and door position.
OPC UA
Industry standard interface (Open Platform Communications, formerly known as OLE for Process
Control Unified Architecture). This protocol is platform independent protocol allowing for greater
integration between devices. This is the communication protocol between the DeltaV system and
a cell therapy manufacturing instrument.
DeltaV Live
DeltaV Live is an application that provides the GUI to allow operation of the instrument and is the
replacement GUI to DeltaV Operate. From here a user can control the instrument by the use of
faceplates to monitor and enter data.
Contextual displays
Contexual display is a DeltaV Live terminology for a faceplate. The contextual display is a pop-up
picture that contains the graphics and controls necessary to perform normal or expected control.
The appearance of a faceplate varies depending upon the type of module, or device with which it is
associated.
Detail faceplate
A detail faceplate is another kind of contextual display. It shows more of the module details (parameter
limits, tuning constants, alarm information and diagnostic messages) than the faceplate.
Chapter1Product information
Operational description
1
6CTS Cellmation Software for DeltaV System User Guide
Batch list faceplate
A batch list faceplate is another kind of pop-up. It shows the current list of recipes that are loaded on
the Batch Executive. From here recipes can be added and started or aborted.
Chapter1Product information
Operational description 1
CTS Cellmation Software for DeltaV System User Guide 7
Toolbar buttons
Table1Toolbar button icons and functionality descriptions
Icon Function
Opens the suppressed alarm list. Only visible when there are suppressed alarms.
Exits operator interface.
Opens the tools list. The following tools are available, Print, Reset Layout, Layout Scale
and Themes.
Refreshes the current display.
Search button. Use this to search for a tag within the system. When entering the search
text, the list of matching tags is displayed. From this list contextual displays, primary
displays and process history displays can be selected for the listed tags.
Navigate forward to the previous Main Screen. Holding the button will display the display
history.
Navigate forward to the previous Main Screen. Holding the button will display the display
history.
Opens the Main Overview graphic, with links to each tower and vessel graphic. An easy
way to navigate quickly between vessels.
Opens the Alarm Summary, showing all active and unacknowledged alarms.
Opens the Alarm Filter page, excluding alarms from inactive vessels and showing only
alarms from active vessels.
DeltaV Utilities, provides access to Emerson Process Management functions.
Indicates if any errors are present on the current display. If an error exists, the icon turns
red.
Opens the Process History view.
Opens the Batch Operator Interface (BOI).
Opens the Batch History view. Only available if the Batch Historian is installed.
Chapter1Product information
Operational description
1
8CTS Cellmation Software for DeltaV System User Guide
Hardware overview
The CTS Cellmation Software is imported into an existing DeltaV system and comprises graphics
and controller software modules. The method of communication between the DeltaV system and an
instrument is OPC UA, with the instrument acting as the server and the DeltaV system acting as the
client. The OPC UA client functionality is only supported on an Application Station or an EIOC controller
(recommended). The CTS Cellmation Software has been developed to reside on a typical system
architecture as shown with the OPC UA client residing on the EIOC (recommended) or Application
Station (further validation may be required).
Chapter1Product information
Hardware overview 1
CTS Cellmation Software for DeltaV System User Guide 9
Figure1DeltaV system architecture
A: Oce/Plant network
B: DeltaV system Area Control Network (ACN)
C: OPC Network
D: Other OPC Instruments
1ProPlus workstation with Alarm/Events Chronicle, Batch Executive, Continuous Historian
2Base Stations with remote sessions, Operation UI
3Application Station OPC UA Client –Optional if EIOC is not utilized
4EIOC OPC UA Client (recommended)
5PK/MX/SX Controller
6CryoMed instrument
The intention is that the OPC UA network is a separate network from either the DeltaV system Area
Control Network or the oce/plant network for security reasons.
Chapter1Product information
Hardware overview
1
10 CTS Cellmation Software for DeltaV System User Guide
Physical connection
The connection between the DeltaV system and the CryoMed instrument is via an ethernet network,
and as such the architecture is dependent on the network infrastructure installed on the system.
To connect the CryoMed instrument to the network, a CAT6 cable can be inserted into the RJ45 port
situated on the rear of the instrument (OPC UA (optional port) with the other end of the cable connected
to a switch or directly to the EIOC if only one instrument is connected.
Chapter1Product information
Physical connection 1
CTS Cellmation Software for DeltaV System User Guide 11
CryoMed instrument network setup
The network adapter needs to be set up in the CryoMed instrument with an IPv4 address for
connection to the DeltaV system (see the example). This address forms part of the DeltaV system
OPC UA setup. From the CryoMed Connectivity screen, press the OPC UA button.
The following screen is displayed.
To enable the OPC UA Server move the Server Enabled toggle to the active position and press the
Save button. Once the server is enabled the Endpoint URL will display the address of the OPA UA
Server. This endpoint will subsequently be entered into the DeltaV OPC UA Client setup. Wireless
connections to this endpoint are not supported.
Enter the server name and port number as required.
Chapter1Product information
CryoMed instrument network setup
1
12 CTS Cellmation Software for DeltaV System User Guide
Lock UI: This option can be enabled to prevent the UI Panel from being accessed during an active run
if the run was started remotely via an OPC UA Client. If this option is enabled, a local user will only be
able to view the Active Run time series data and file metadata associated with the active run.
IPv4 Method: The following screen is displayed when you press the IPv4 Method button.
This screen allows you to choose between automatic and manual IPv4 address assignments. If
Automatic (DHCP) is selected, the device will be assigned an IPv4 address dynamically based on
the local network. If Manual is selected, you can enter a static IPv4 address, netmask and gateway to
assign to the device.
Chapter1Product information
CryoMed instrument network setup 1
CTS Cellmation Software for DeltaV System User Guide 13
EIOC network setup
The EIOC network adapter needs to be set up via DeltaV Explorer with an IPv4 address. Select
Properties of the EIOC port (P01) and select the Advanced tab. From this tab enter the required IPv4
address. Currently Redundant connection is not used so leave this blank.
In the Interface drop down box at the bottom, select OPC UA as the protocol to be used.
Chapter1Product information
EIOC network setup
1
14 CTS Cellmation Software for DeltaV System User Guide
EIOC OPC UA setup
Once the EIOC has been setup to use the OPC UA protocol, if an instrument requires the OPC Client
certificate then this can be generated by selecting the Certificate tab. Use the Generate button to
create a certificate.
Note: The CryoMed instrument currently does not support OPC UA authentication and operates in
Anonymous mode only.
The CTS Cellmation Software for DeltaV System installation files for the CryoMed instrument
contain a configuration file that has default settings for the instrument OPC UA setup. The default setup
creates a Physical Device Tag (PDT) named “PDT_CRFXXXX” with all the associated Logical Device
Tags underneath it. The PDT properties contain the Endpoint URL of the instrument OPC Server and
if this requires modifying, then select properties of the PDT and update the URL by entering the IPv4
address.
Note: The default port number is 4843, but this should match the port number setup in the instrument.
Chapter1Product information
EIOC OPC UA setup 1
CTS Cellmation Software for DeltaV System User Guide 15
Network connectivity check
Once the DeltaV system configuration has been downloaded, a quick test can be performed using a
PC connected to the OPC UA network. On the PC, right click the Windows icon in the left hand corner
of task bar and select Command Prompt from the list when it appears.
In the command window, type “ping” in the black box and hit the space bar. Type the IP address of the
EIOC (e.g., 192.168.0.5 in the following figure). If there is connectivity, then a reply will be seen. If not,
then the cabling and network settings will need to be checked.
Chapter1Product information
Network connectivity check
1
16 CTS Cellmation Software for DeltaV System User Guide
OPC UA connectivity check
Once the DeltaV software configuration has been downloaded, a quick test can be performed from the
Diagnostics DeltaV application. This can be opened by selecting the icon.
From the diagnostics application, browse to the required EIOC and port P01. Browse further to the
device, right click and select Test Endpoint Connection.
If the OPC UA communications are operating, the following screen will appear detailing the connection
diagnostics.
Chapter1Product information
OPC UA connectivity check 1
CTS Cellmation Software for DeltaV System User Guide 17
Continuous Historian setup
The trend chart on the main display is configured to extract the historical data from a continuous
historian. By default, this is set to ‘PROPLUS’.
This may not be the correct application station name to refer to. To set the correct application name
the following standard “SGT_CHist_Server” within Live needs to be updated with the actual Application
Station name of the historian.
User interface
The user interface consists of a number of screens that enable the user to interact with an instrument
and run instrument protocols. DeltaV Live is the only supported interface as DeltaV Operate is being
phased out. These screens are:
Splash/opening screen
Overview display
Main instrument display
Instrument parameter display
Chapter1Product information
Continuous Historian setup
1
18 CTS Cellmation Software for DeltaV System User Guide
Splash screen
The splash screen appears when the user interface is first started and from here the user can enter the
CTS Cellmation Software overview by clicking anywhere on this screen.
Note: If not using Cellmation splash screen, user will have to create an icon in Graphic Studio to open
CTS Cellmation software overview in DeltaV Live.
Chapter1Product information
User interface 1
CTS Cellmation Software for DeltaV System User Guide 19
Overview display
The Overview screen allows for an individual instrument to be selected to open more detailed
information for that instrument. On the initial installation, the overview screen has a single CryoMed
instrument selection point. As instruments are added to a system, this overview screen will be modified
to add the new instruments.
Chapter1Product information
User interface
1
20 CTS Cellmation Software for DeltaV System User Guide
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86

Thermo Fisher Scientific CTS Cellmation Software User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI