Thermo Fisher Scientific CTS Cellmation Incubator Software User guide

Type
User guide
For Research Use or Manufacturing of Cell, Gene, or Tissue- Based
Products.
CTS Cellmation Incubator Software for
DeltaV System Installation
USER GUIDE
for use with: Thermo Scientific HeraCell VIOS CR CO2
Incubator
Catalog Number A59197
Publication Number MAN0029996
Revision A.0
Life Technologies Holdings Pte Ltd | Block 33 | Marsiling Industrial Estate Road 3 | #07-06, Singapore 739256
Revision history: (English)
Revision Date Description
A.0 25 September
2023 Initial release.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE
LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR
ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
NOTICE TO PURCHASER: DISCLAIMER OF LICENSE: Purchase of this software product alone does not imply any license under any
process, instrument or other apparatus, system, composition, reagent or kit rights under patent claims owned or otherwise controlled by
Thermo Fisher Scientific, either expressly, or by estoppel.
TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.
©2023 Thermo Fisher Scientific Inc. All rights reserved.
CTSTM CellmationTM Incubator Software for DeltaVTM System Installation User Guide 3
TABLE OF CONTENTS
1 PURPOSE AND SCOPE ............................................................................................................................ 4
2 REFERENCES ............................................................................................................................................ 4
3 DEFINITIONS AND ACRONYMS ............................................................................................................... 4
4 SAFETY REQUIREMENTS ........................................................................................................................ 4
5 SOFTWARE INSTALLATION PROCEDURE ............................................................................................ 5
5.1 SIGNATURE LOG ........................................................................................................................................ 5
5.2 PREREQUISITES ......................................................................................................................................... 5
5.3 SOFTWARE INSTALLATION .......................................................................................................................... 7
6 SOFTWARE INSTALLATION VERIFICATION CHECKS........................................................................ 30
7 SAFETY .................................................................................................................................................... 31
8 DOCUMENTATION AND SUPPORT ....................................................................................................... 31
4 CellmationTM Incubator Software for DeltaVTM System Installation User Guide
1 PURPOSE AND SCOPE
The purpose of this document is to provide instructions for installing the GibcoCTSCellmation
Incubator OPC UA Interface DeltaVsoftware and DeltaV Live functionality.
This document will also detail the tests required to verify that the installed configuration has been
imported correctly.
2 REFERENCES
Document Name
CellmationCommon Module Software Installation Instructions
CTSCellmationXenonSoftware Release Notes
CTSCellmationCommon Software Release Notes
3 DEFINITIONS AND ACRONYMS
Term & (Acronym) Definition
GT Gene Therapy
SN Sentinel Monitor
HM Hold Monitor
EM Equipment Module
USM Unit Support Module
LND Landing Module
PH Phase
OAR Operator Action Request
MON Monitor
PCC Process Cell Class
INC Incubator
OPC UA Open Platform Communications United Architecture
I/O Input / Output
N/A Not Applicable
4 SAFETY REQUIREMENTS
N/A
CTSTM CellmationTM Incubator Software for DeltaVTM System Installation User Guide 5
5 SOFTWARE INSTALLATION PROCEDURE
These installation instructions are designed to be undertaken by a Thermo Fisher Service Engineer or
an engineer fully trained in DeltaV.
5.1 Signature Log
The signature log in Table 5-1 needs to be completed by all personnel involved in installing this
software before any work is commence.
Table 5-1 – Signature Log
Name (print)
Title
Initials
Signature
Date
5.2 Prerequisites
The prerequisites in Table 5-2 should be completed and signed-off before the software installation can
commence.
Table 5-1 – Prerequisites
Item
Number
Prerequisites Initials / Date
1. User installing the software must have full Windows Administrator Privileges
and DeltaV Administrator Privileges.
2. User must have access to ProfessionalPLUS Station.
3. If applicable, confirm that it is safe to proceed with the installation by following
the change management process.
4.
Ensure that there are no pending downloads required. Before installing
software, conduct a risk assessment to evaluate the impact of any required
downloads.
5. Ensure that the gateway is updated to the latest version.
Software Backup - ProfessionalPLUS
6 CellmationTM Incubator Software for DeltaVTM System Installation User Guide
Item
Number Prerequisites Initials / Date
6.
Export the DeltaV Database.
In DeltaV Explorer, right-click on the topmost level (i.e., the database
name).
Note the location of the exported database below (this will differ
between clients and systems)
Path: ____________________________________________________
____________________________________________________
7.
Export the DeltaV Live Database.
In DeltaV Graphics Studio, select the topmost level (i.e., the
database name) from the Graphics Explorer tab.
From the “File” menu item, select “Export” and check the following:
Note the location of the exported database below
Path: ____________________________________________________
____________________________________________________
Check for other CTS Cellmation Software installations if applicable.
8.
If there are no Thermo Fisher Scientific instruments on-site, refer to the
installation guide of the CTS Cellmation Common Software Installer and
confirm that the CTS Cellmation Common Software Module installation is
complete in the system.
CTSTM CellmationTM Incubator Software for DeltaVTM System Installation User Guide 7
Item
Number Prerequisites Initials / Date
9.
If there are previous CTS Cellmation Software installations on the system,
then a decision must be made as to which files need updating.
To check for installs, open Windows Control Panel and select
“Programs ->“Programs and Features
Check for any “gibco_CTS_Cellmation_...............” installs and note them
below
1. ____________________________________________________
2. ____________________________________________________
3. ____________________________________________________
4. ____________________________________________________
5. ____________________________________________________
10.
If a previous CTS Cellmation Common Software installation is present, then
refer to the latest release of the Common Installer and release note for any
updated common items to be installed.
Common Software items can be installed by following the instructions in the
Installation Guide of the Common Installer. Only the common items should be
overwritten.
11.
If any other previous CTS Cellmation Software installations are present, then
refer to the latest release of the instrument specific installer and release note
for any updated items to be installed.
Instrument specific software items can be installed by following the
instructions in the Installation Guide of the instrument specific installers; only
those items should be overwritten.
CTS Cellmation Incubator software items can be installed by following this
guide; only those items should be overwritten.
IMPORTANT: If you already have DeltaV configured for communication with other instruments, only
select “NO” when prompted about an update unless installing an updated version of the
module to ensure Cellmations installation does not negatively impact existing DeltaV
connections! This is the safest way to install Cellmation without risking existing configurations.
5.3 Software Installation
The steps shown in Table 5-3 should be performed on the DeltaV ProfessionalPLUS (ProPlus) station
using DeltaV Explorer.
This revision of the installation instructions is applicable to release 2.0.0.
8 CellmationTM Incubator Software for DeltaVTM System Installation User Guide
IMPORTANT: The order of component installation listed below must be followed because there
are interdependencies to the imported files! If your installation is unsuccessful, you can retry your
import, but the step sequence should be maintained. Additionally, DO NOT delete any files on the
DeltaV computer that you did not import from the Cellmation Installation Folder.
Table 5-2 – Software Installation Instructions
Step Instructions Initials / Date
1.
Copy the Cellmation Incubator Installer Package into the folder
D:\DeltaV\DVData\Import-Exporton the DeltaV computer. The Installer package
will include the installer .exe file and other required files/folders. DO NOT delete or
rename any files/folders in the installer package.
Execute the “gibco_CTS_Cellmation_Incubator_2.0.0.exe” and perform the
following to complete the installation:
Right-click on the “gibco_CTS_Cellmation_Incubator_2.0.0.exe” file and
select “Run as administrator”.
The welcome screen will appear. Click “Next >” once displayed on the
popup screen.
Accept the terms of the License Agreement to proceed with the installation.
Click “Next >.”
The pop-up displays the destination location for the software installation.
Click “Next >”.
Enter Username, Company Name, and serial number (Reg Code received
with the installer). Click “Next >”.
After entering a valid Reg Code, the “Install” button will be available. Click to
proceed with the installation. Click “Finish” once complete.
If an invalid Reg Code is entered, it will show the error message “Invalid
registration code.” Click “OK” and re-enter the correct Reg Code to proceed.
Note: If the same version of the installer is already present on the system, then a
prompt will appear to uninstall the previous version. Select “OK” to continue
uninstallation and “Cancel” to exit uninstallation. Choose the appropriate option.
Otherwise, uninstall the previous installer before proceeding with the new
installation. Select “Control Panel > Programs and Features” to uninstall the
previous installer.
2.
After successful installation, a folder will be created entitled
gibco_CTS_Cellmation_Incubator_2.0.0” with two subfolders, “Displays” and
“FHX,” in the Import-Export folder.
Also, this will automatically copy all the chart files present in the
“Displays/Charts_Incubator” folder to “D:\DeltaV\DVData\Charts\lib”.
Note: Upon uninstallation of the installer, all the created folders during installation
will be automatically deleted (The entire Installer folder
gibco_CTS_Cellmation_Incubator_2.0.0” with subfolders and chart files copied
under “D:\DeltaV\DVData\Charts\lib will get deleted).
CTSTM CellmationTM Incubator Software for DeltaVTM System Installation User Guide 9
Step Instructions Initials / Date
3.
Steps 4 through 12 provide instructions to import various packages of code directly
into the DeltaV library. The standard import flow is as follows:
Open DeltaV Explorer (from Start menu under DeltaV Engineering)
Navigate to File Import Standard DeltaV Format
FHX files to import are located in D:\DeltaV\DVData\Import-Export\
gibco_CTS_Cellmation_Incubator_2.0.0\FHX
Make sure to uncheck ‘Include the license information’ tick box before
confirming to import a component. Click Open.
This will typically take around one minute. When done, “Import complete”
will appear.
Note: When importing .fhx files into the database, there may be some file
conflicts that open a pop-up prompt.
If importing for the first time, select “No to All”.
If re-importing a file for any reason without deleting its previously
imported copy, make sure to read the pop-up.
4.
From DeltaV Explorer, import the Setup .fhx file from the “FHX” subfolder:
Module Name
gibco_Named_Sets_Incubator
10 CellmationTM Incubator Software for DeltaVTM System Installation User Guide
Step Instructions Initials / Date
Confirm the following are imported into “ThermoGT_Incubator” Named Sets folder in
the database:
Named Set Name
INC_Active_Inactive
INC_Commands
INC_Cylinder_Status
INC_Door_Status
INC_HEPA_Config
INC_Link_Status
INC_ON_OFF
INC_SteriRun_Status
INC_Water_Level
Located at DeltaV_System System Configuration Setup Named
Sets ThermoGT_Incubator
5.
From DeltaV Explorer, import the composite templates .fhx file from “FHX”
subfolder:
Module Name
gibco_Composites_Incubator
CTSTM CellmationTM Incubator Software for DeltaVTM System Installation User Guide 11
Step Instructions Initials / Date
Confirm the following modules have been imported into theThermoGT_Incubator
folder:
IMPORTANT: The internal revision, in the INTERNAL_REV parameter, for each
module should match the revision documented below:
Module Name Rev Installed Rev
GT_INC_ABORT 0.2
GT_ INC_FAIL_MON 0.2
GT_ INC_HOLD 0.2
GT_ INC_RESTART 0.2
GT_ INC_STOP 0.2
GT_ INC_ALM_MSG 0.3
Located at DeltaV_System Library Composite Templates
ThermoGT_Incubator
To locate the internal revision values, select each module and find the
INTERNAL_REV parameter from the listed parameters.
6.
From DeltaV Explorer, import the Control Modules Classes .fhx file from the “FHX”
subfolder:
Module Name
gibco_Control_Module_Classes_Incubator
Confirm the following modules have been imported into the “ThermoGT” folder:
Note the internal revision in the INTERNAL_REV parameter for each module which
should match the revision documented below:
Module Name Rev Installed Rev
GT_INCUBTR_LAND 0.4
Located at DeltaV_System Library Advanced Definitions Control
Module Classes ThermoGT
To locate the internal revision values, select each module and find the
INTERNAL_REV parameter from the listed parameters.
12 CellmationTM Incubator Software for DeltaVTM System Installation User Guide
Step Instructions Initials / Date
7.
From DeltaV Explorer, import the Equipment Modules Classes .fhx file from the
“FHX” subfolder:
Module Name
gibco_Equipment_Module_Class_Incubator
Confirm the following modules have been imported into the “ThermoGT” folder:
Note the internal revision in the INTERNAL_REV parameter for each module should
match the revision documented below:
Module Name Rev Installed Rev
GT_INCUBTR_EM 0.3
Located at DeltaV_System Library Advanced Definitions
Equipment Module Classes ThermoGT
To locate the internal revision values, select each module and find the
INTERNAL_REV parameter from the listed parameters.
8.
From DeltaV Explorer, import the Phase Classes .fhx file from the “FHX” sub folder:
Module Name
gibco_Phase_Classes_Incubator
Confirm the following phases have been imported into the “ThermoGT” folder:
Note the internal revision in the INTERNAL_REV parameter for each module should
match the revision documented below:
Module Name Rev Installed Rev
GT_INC_INIT 0.2
GT_INC_PROMPT 0.2
Located at DeltaV_System Library Advanced Definitions Phase
Classes ThermoGT
To locate the internal revision values, select each module and find the
INTERNAL_REV parameter from the listed parameters.
9.
From DeltaV Explorer, import the Unit Classes .fhx file from the “FHX” subfolder:
Module Name
gibco_Unit_Class_Incubator
CTSTM CellmationTM Incubator Software for DeltaVTM System Installation User Guide 13
Step Instructions Initials / Date
Confirm the following modules have been imported into the “ThermoGT” folder:
Note the internal revision in the INTERNAL_REV parameter for each module should
match the revision documented below:
Module Name Rev Installed Rev
INCUBATR_UC 0.2
Located at DeltaV_System Library Advanced Definitions Unit
Classes ThermoGT
To locate the internal revision values, select each module and find the
INTERNAL_REV parameter from the listed parameters.
10.
From DeltaV Explorer, import the Operations .fhx file from the “FHX” subfolder:
File Name
ThermoGT_Operations_Incubator
ThermoGT_UnitProcedures_Incubator
Confirm the following recipes have been imported into the “ThermoGT” folder:
Note the version for each recipe should match the version documented below:
Recipe Name
OP_INC_ACQ_REL
OP_INC_INIT_SETUP
OP_INC_STERI_RUN_STRT
OP_INC_STERI_RUN_END
Located at DeltaV_System System Configuration Recipes
Operations ThermoGT
Recipe Name
UP_INC_RUN
Located at DeltaV_System System Configuration Recipes Unit
Procedures ThermoGT
14 CellmationTM Incubator Software for DeltaVTM System Installation User Guide
Step Instructions Initials / Date
Create Unit Instance
11.
Edit the file ‘gibco_INCXXXX-AREA.fhx’ in Notepad.
Note: XXXX is correct nomenclature at this point.
Replace “INCXXXX” with the incubator tag specified for the instrument
within DeltaV and save the file with a new name identifying it with the new
tag.
It is highly recommended to use Find and Replace to catch all instances to
updateINCXXXX” to “INCplus the last 4 digits of the instrument serial
number.
Repeat this step for additional incubator unit instances as required.
12.
Import the file with the new tag name into DeltaV.
To confirm the import, verify that a new area and included modules are
located at DeltaV Explorer DeltaV System System Configuration
Control Strategies
13.
Assign the following modules to the required controller. (Note INCXXXX below is
replaced with the new tag name)
INCXXXX
INCXXXX-INC-EM
INCXXXX-INC-USM
To assign modules:
Drill-down to the unit module (DeltaV_System System Configuration
Control Strategies INCXXXX-AREA [area] INCXXXX-PC
[process cell] INCXXXX [unit]
Right-click on each module listed above and select Properties.
Navigate to the Tools tab and select Assign Now.
Assign to the appropriate controller and answer Yes to confirm the
prompt.
14.
Assign the following module to the required EIOC/Application Station. (Note
INCXXXX below is replaced with the new tag name)
INCXXXX-INC-LND
To assign modules:
Drill-down to the unit module (DeltaV_System System Configuration
Control Strategies INCXXXX-AREA [area] INCXXXX-PC
[process cell] INCXXXX [unit]
Right-click on the module listed above and select Properties.
Navigate to the Tools tab and select Assign Now.
Assign to the appropriate EIOC controller and answer Yes to confirm
the prompt.
CTSTM CellmationTM Incubator Software for DeltaVTM System Installation User Guide 15
Step Instructions Initials / Date
15.
Assign the following Area to the Alarms and Events of the ProPlus/required
Workstation/Sever, Continuous Historian and Batch Executive (Note INCXXXX
below is replaced with the new tag name.)
INCXXXX-AREA
Assign area to the Alarms and Events:
Drill-down to the ProPlus/Workstation (Physical Network Control
Network ProPlus Station/Workstation/Server Alarms and
Events
Right-click and select Assign Area.
Select INCXXXX-AREA and select Yes when prompted.
Assign area to the Continuous Historian:
Drill-down to the Continuous Historian (Physical Network Control
Network ProPlus Station/Workstation/Server Continuous
Historian
Right-click and select Assign Area
Select INCXXXX-AREA and select Yes when prompted
Assign area to Batch Executive:
Drill-down to the Batch Executive (Physical Network Control
Network ProPlus Station/Workstation/Server Batch Executive
Assigned Areas
Right-click and select Assign Area.
Select INCXXXX-AREA and select Yes when prompted.
16.
Assign the following Recipes to the Batch Executive:
Assign Recipe to the Batch Executive:
Drill-down to the Batch Executive (Physical Network Control
Network ProPlus Station/Workstation Batch Executive
Assigned Recipes
Right-click and select Recipe.
Select OP_INC_ACQ_REL and select Yes when prompted.
Similarly assign other operations OP_INC_ACQ_REL,
OP_INC_INIT_SETUP, OP_INC_STERI_RUN_STRT,
OP_INC_STERI_RUN_END and Unit Procedure “UP_INC_RUN”
16 CellmationTM Incubator Software for DeltaVTM System Installation User Guide
Step Instructions Initials / Date
17.
Download as follows:
Download the Setup data of the ProPlus/Workstation/Server for the area
added under Alarms and Events.
Right-click on the ProPlus/Workstation/Server and select Download
Setup Data
Select Yes when prompted to confirm the download.
Download the Continuous Historian for the area added:
Right-click on the Continuous Historian and select Download
Continuous Historian
Select Yes when prompted to confirm the download.
Download the Batch Executive for the area added and recipe assigned:
Right-click on the Batch Executive and select Download Batch
Executive
Select Yes when prompted to confirm the download.
OPC UA IO installation
18.
The OPC UA PDT/LDT configuration is contained within the file
‘gibco_Incubator_PDT_INCXXXX.fhx in the “FHX” subfolder. This file and the area
file can be edited so that it imports the required tag information. Using the Find and
Replace search function is recommended to make all required changes within the
gibco_Incubator_PDT_INCXXXX.fhx file.
The OPC UA I/O has been assigned to an EIOC named “EIOC1” by default.
If the I/O is to be assigned to another EIOC/App Station, then replace all
instances of “EIOC1” in the .fhx file with the desired name of the EIOC or
Application Station.
Replace “INCXXXX” with the Incubator tag specified for the instrument
within DeltaV.
Save the file with a new name identifying it with the new tag.
Repeat this step for additional Incubator unit instances as required.
19.
Ensure the OPC UA Client has been set up for the OPC UA protocol.
If the OPC UA I/O is being assigned to an EIOC, ensure the ethernet port P01
protocol has been set to OPC UA.
Navigate to DeltaV_System System Configuration Physical Network
Control Network I/O Network EIOC Name P01
Right-click P01 and select Properties.
Under the Advanced tab, select OPC UA from the Interface Protocol
drop-down.
If the OPC UA I/O is being assigned to an Application Station:
Navigate to DeltaV_System System Configuration Physical Network
Control Network Control Network Application Station Name
OPC UA Client
Right-click OPC UA Client and select Properties.
Under the General tab, set the checkbox to “enable OPC UA client”.
CTSTM CellmationTM Incubator Software for DeltaVTM System Installation User Guide 17
Step Instructions Initials / Date
20. Import the ‘gibco_OPCUA_Incubator_PDT_INCXXXX.fhx’ file with the assigned tag
name into DeltaV.
21.
Set the IP address of the Incubator OPC UA server.
If the OPC UA I/O is being assingned to an EIOC
Navigate to DeltaV_System System Configuration Physical Network
Control Network I/O Network EIOC Name P01
If the OPC UA I/O is being assigned to an Application Station
Navigate to DeltaV_System System Configuration Physical Network
Control Network Control Network Application Station Name
OPC UA Client
From here right click on the Rotea PDT that has been imported and select the
Primary tab.
Change the default IP address to match the IP address set in the Incubator.
Note the port number 4840 is the Incubator default port and should not be
changed.
22.
Download the unit module and contained modules. No errors should be reported.
(Note INCXXXX below is replaced with the new tag name)
Navigate to the INCXXXX unit module (DeltaV Explorer DeltaV_System
System Configuration Control Strategies INCXXXX-AREA
INCXXXX-PC INCXXXX
Right-click the INCXXXX unit module and select Download Unit module
and contained modules
Select Yes when prompted to confirm the download.
23.
If ProPlus/Workstation/Server Setup data has already been downloaded in step 17,
skip this step.
Download the Setup data of the ProPlus/Controllers/Workstation/Server for all the
named sets.
Navigate to DeltaV Explorer DeltaV System System Configuration
Physical Network ProPlus/Controllers/Workstation/Server
Right-click the ProPlus/Controllers/ Workstation/Server and select
Download Setup Data
Select Yes when prompted to confirm the download.
18 CellmationTM Incubator Software for DeltaVTM System Installation User Guide
Step Instructions Initials / Date
24.
Download the OPC IO. There should not be errors except for with a secodary
connection and/or Application Certificate, which can be ignored.
EIOC assigned IO
Navigate to DeltaV Explorer DeltaV System System Configuration
Physical Network Control Network I/O Network EIOC Name P01
Right-click the port P01 and select Download Ethernet I/O Port
Application Station assigned IO
Navigate to DeltaV Explorer DeltaV System System Configuration
Physical Network Control Network Application Station Name
OPC UA Client
Right-click OPC UA Client and select Download OPC UA Client.
Establishing OPC UA Communication
25.
On the DeltaV Computer, using DeltaV Explorer
EIOC assigned IO
Assign the predetermined IP address to the ethernet port P01 and generate a
certificate.
Navigate to DeltaV_System System Configuration Physical Network
Control Network I/O Network EIOC Name P01
Right-click P01 and select Properties.
In the Advanced tab, input the predetermined IP address of the DeltaV
System under Primary Connection IP Address.
In the Certificate tab, select Generate to create a new certificate.
Click OK to save the changes and close the pop-up dialog box.
Application Station assigned IO
Open the Windows Control Panel and select Control Panel\Network and
Internet\Network Connections.
Select the Network Interface Card (NIC) to be used as the OPC UA network
connection and set the TCP/IPv4 IP address to the predetermined IP
address of the DeltaV System.
Navigate to DeltaV Explorer DeltaV System System Configuration
Physical Network Control Network Application Station Name
OPC UA Client
Right-click OPC UA Client and select Properties.
In the Certificate tab, select Generate to create a new certificate.
Click OK to save the changes and close the pop-up dialog box.
Note: If the ethernet port P01 or Application Station OPC UA Client is already
assigned to the predetermined IP address and a certificate is generated, skip this
step.
CTSTM CellmationTM Incubator Software for DeltaVTM System Installation User Guide 19
Step Instructions Initials / Date
26.
Perform a factory reset of the incubator gateway device
a. Connect the USB Micro (slot C) on the incubator Gateway to the USB Type
A port of a Windows PC.
b. Press the reset button (S1) lightly (you should be able to feel a ‘click’ when
pressing it). Then, press and hold for > 20 seconds. Observe that the LED
light will blink in red, until it turns green. Release S1 button.
20 CellmationTM Incubator Software for DeltaVTM System Installation User Guide
Configure the static IP address of the incubator Gateway and download
gateway OPC UA certificate
a. From the PC, open a web browser (Microsoft Edge, Google Chrome,
Mozilla Firefox, etc.) and enter the address https://192.168.7.1 and press
Enter.
Note: You may need to refresh a few times until the page below is loaded.
b. Select “Proceed to 192.168.7.1 (unsafe)”
c. In the web interface, the OPC UA Server IP address, the user name and the
password can be configured. This is important so that the OPC UA Server
can be integrated in the customer network and an OPC UA Client, e.g.,
DeltaV system, can read out the data.
d. Enter the predetermined information for the following fields:
o IP-configuration (Select: Static)
o IPv4 Address
o SubNet Mask
o OPC-Username
o OPC-Password
o Update Server Password
o Repeat Update Server Password
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32

Thermo Fisher Scientific CTS Cellmation Incubator Software User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI