Thermo Fisher Scientific CTS Rotea Counterflow Centrifugation System User guide

Type
User guide
For Research Use or Manufacturing of Cell, Gene, or Tissue- Based
Products. CAUTION: Not intended for direct administration into
humans or animals.
CTS Rotea Counterflow Centrifugation
System
USER GUIDE
Catalog NumberA44769
Publication Number MAN0018908
Revision C.0
Life Technologies Holdings Pte Ltd | Block 33 | Marsiling Industrial Estate Road 3 | #07-06, Singapore 739256
Revision history:Pub.No. MAN0018908
Revision Date Description
C.0 18 August 2022
Added Caution to Kit configuration; Warning to User maintenance shedule; Caution to Warnings
and precautions; updated Caution in Instrument safety, General; added Caution to Cleaning and
decontamination.
B.0 2 December 2020 MInor change to CTS Rotea Counterflow Centrifugation System specific disclaimer statement.
A.0 2 October 2020 New document. User guide for new workflow.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE
LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR
ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
The CTS Rotea Counterflow Centrifugation System is for use in further manufacturing applications only and not for fill and finish
applications. Drug manufacturers are responsible for ensuring the purity, potency, safety and identity of finished drug products and for
obtaining the appropriate regulatory approvals.
Important Licensing Information: This product may be covered by one or more Limited Use Label Licenses. By use of this product,
you accept the terms and conditions of all applicable Limited Use Label Licenses.
TRADEMARKS: All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.
©2022 Thermo Fisher Scientific Inc. All rights reserved.
Contents
CHAPTER1Productinformation .................................................. 7
Productdescription ............................................................. 7
Features ....................................................................... 7
Contents ....................................................................... 7
System overview ................................................................ 8
Instrumentdescription ....................................................... 8
Instrumentdimensions ..................................................... 10
Workingspace ............................................................ 10
Instrumentweight ......................................................... 10
Instrument properties ...................................................... 11
Push button controls ....................................................... 13
CTS Rotea Single-Use Kitschematic ...................................... 14
Wettedcomponents ....................................................... 15
Allowed chemicals for wetted surfaces ....................................... 15
CHAPTER2Installation andsetup ............................................... 16
Packaging contents ............................................................ 16
Instrumentinstallation .......................................................... 16
Power instrument "On" ......................................................... 18
Power instrument"O" ......................................................... 19
Installation .................................................................... 20
First timesetup ................................................................ 24
Download protocols ............................................................ 27
Create protocols ........................................................... 27
Load a protocol ........................................................... 27
Rotea GUI dashboard ..................................................... 31
UserManagement ......................................................... 32
Uninstall ...................................................................... 37
Logging ....................................................................... 38
CTS Rotea Counterflow Centrifugation System User Guide 3
CHAPTER3Basic instrument operation ........................................ 41
Open thedoor ................................................................. 41
Kit preparation, loading and removal ............................................. 43
Kitconfiguration ........................................................... 43
Pre-assembly of processingkit .............................................. 45
Connecting vessels, bags and reagents ...................................... 46
Bubble Trap ............................................................... 47
Kitinstallation ............................................................. 48
Initiate a protocol .......................................................... 54
Run a protocol ................................................................. 61
Rotea GUI navigation and instrument operation .............................. 61
Errormessages ............................................................ 67
Kit removal .................................................................... 69
Aseptic disconnection of the kitcomponents ...................................... 74
Systemsettings ................................................................ 75
Instrumentfirmware ........................................................ 75
OPC-UA .................................................................. 80
Settings .................................................................. 81
Licenses .................................................................. 82
Calibration .................................................................... 83
CHAPTER4OPC-UA interface ................................................... 87
Introduction ................................................................... 87
Definitions .................................................................... 87
Scope of system ............................................................... 88
Instrument hardwareconfiguration ........................................... 88
Single-UseKit ............................................................. 88
Protocolfile ............................................................... 89
Protocolmetadata ......................................................... 89
MES interface ................................................................. 89
Background ............................................................... 89
Electroniccommunications ................................................. 90
Communications protocol .................................................. 90
MES – Commands and reports overview .......................................... 91
Enable OPC-UA ........................................................... 92
Log in toinstrument ........................................................ 92
Load protocol ............................................................. 92
Loadkit .................................................................. 93
Transfermetadata ......................................................... 93
Initialize protocol on toinstrument ........................................... 93
Start/Stop/Pause/Resume protocol .......................................... 93
Instrument runstatus ....................................................... 94
Contents
4CTS Rotea Counterflow Centrifugation System User Guide
MES batch historydata ..................................................... 94
Alarm recovery tools ....................................................... 95
OPC-UA interface .............................................................. 96
Commands OPC-UA to Roteaapplication ................................... 96
Reports Rotea application to OPC-UA ...................................... 97
Use case sequence diagrams .................................................. 104
MES run a protocol ....................................................... 104
Protocol with parametricdata .............................................. 105
MES drives Rotea in response to external systems ........................... 106
MES drives steps within protocol ........................................... 106
MES manage error recovery events with admin user .......................... 107
Use list file to load protocol ............................................... 108
CHAPTER5Maintenance ....................................................... 110
User maintenanceschedule .................................................... 110
Instrumentcleaning ........................................................... 110
Inspectcasing ................................................................ 111
Open the door withoutpower .................................................. 111
Power andfuses .............................................................. 112
Powersupply ............................................................ 112
Fuses ................................................................... 112
CHAPTER6Troubleshooting ................................................... 113
CHAPTER7Compliance ........................................................ 115
CHAPTER8Safetyinstructions ................................................ 116
Labels ....................................................................... 116
Emergency procedures ........................................................ 116
Instrument emergency stopevents ......................................... 116
Emergency shutdown .................................................... 117
Restart after shutdown ................................................... 117
Protocol recovery process ................................................. 118
Warnings and precautions ..................................................... 119
Warnings ................................................................ 119
Cautions ................................................................ 120
Contents
CTS Rotea Counterflow Centrifugation System User Guide 5
APPENDIXASafety .............................................................. 121
Symbols on thisinstrument .................................................... 121
Standard safetysymbols .................................................. 122
Control and connectionsymbols ........................................... 122
Conformitysymbols ...................................................... 122
Safety information for instruments not manufactured by Thermo FisherScientific ..... 124
Instrumentsafety ............................................................. 124
General ................................................................. 124
Physical injury ............................................................ 125
Electricalsafety .......................................................... 125
Cleaning anddecontamination ............................................. 126
Safety and electromagnetic compatibility (EMC) standards ......................... 126
Safety standards ......................................................... 126
EMC standards ........................................................... 127
Environmental design standards ............................................ 127
Chemicalsafety .............................................................. 128
Biological hazardsafety ....................................................... 129
APPENDIXBDocumentation and support .................................... 130
Customer and technical support ................................................ 130
Limited product warranty ...................................................... 130
Contents
6CTS Rotea Counterflow Centrifugation System User Guide
Product information
IMPORTANT! Before using this product, read and understand the information in the “Safety” appendix
in this document.
Product description
The CTS Rotea Counterflow Centrifugation System is an automated, closed-flow benchtop
instrument that can perform a wide variety of cell processing steps across dierent cell types.
The CTS Rotea Single-Use Kit is specifically designed for a range of cell processing applications
such as separation, isolation, buer exchange, wash and concentrate. Each kit has 8 input/output
sterile weldable tubes, counterflow centrifuge and Carrier Frame for unparalleled setup flexibility.
The CTS Rotea Software is user-programmable and allows users to save, create, and modify
protocols depending on the application. The system's open design and integrated camera allows users
to visualize cells in real-time to more easily optimize protocols.
Features
CTS Rotea Counterflow Centrifugation System easily fits into existing workflows, from research to
commercial manufacturing.
Integrated camera provides real-time sample visualization
Patented technology that can process up to 20 L of starting volume to as little as 5mL of output
volume
Gentle fluidized bed results in unmatched cell recovery and viability
Seamlessly transition from process development to GMP manufacturing on the same system.
Contents
Table1CTS Rotea Counterflow Centrifugation System (Cat.No.A44769)
Contents Amount
CTS Rotea Centrifuge 1 ea
CTS Rotea Single-Use Kit Available separately
Protocol Development Application 1 ea
Dell 2-in-1 laptop for operation/programming 1 ea
1
CTS Rotea Counterflow Centrifugation System User Guide 7
System overview
Instrument description
2
1
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
Figure1Instrument – Front View
1Instrument Door
2Pinch Valves (J & K)
3Kit Location Button
42D Barcode Reader
5CFC Chamber Detector
6CFC Chamber Carrier
7Camera
8Moisture Sensor
9Pushbutton Controls
10 Optical Density Sensor & Pressure Sensor (P1)
11 Peristaltic Pump
12 Pump Clamp Arm
13 Pinch Valve (H)
14 Kit Location Button & Pressure Sensor (P2)
15 Door Latch
16 Pinch Valves (A, B, C, D, E, F, & G)
17 Bubble Sensors
18 Hanger Posts
Chapter1Product information
System overview
1
8CTS Rotea Counterflow Centrifugation System User Guide
8
7
6
5111
1432
Figure2Instrument – Rear View
1Power Switch
2Mains Connector
3CAN bus (2 o)
4Ethernet (RJ-45)
5USB-C
6Bag Hooks
7Hanger Post
8Hanger Post Adjustment
Chapter1Product information
System overview 1
CTS Rotea Counterflow Centrifugation System User Guide 9
Instrument dimensions
Working space
The Rotea instrument is designed for bench top use and requires a working space of approximately
Height: 30” (76 cm), Width: 25” (63.5cm), Depth: 20” (50.8cm).
Note: The 2in1 laptop is connected via a USB-C cable and hence can be located in a range of
positions. The width dimension of 25” assumes that the laptop is located on the bench beside the
instrument.
Instrument weight
The Rotea instrument weighs 44lbs (20kg) with hanger posts installed.
Chapter1Product information
System overview
1
10 CTS Rotea Counterflow Centrifugation System User Guide
Instrument properties
Category Property Value
Electrical
Supply voltage 100–240V AC ±10%, 50/60Hz
Phases Single
Maximum Rated Input Current 5A
Fuses 2 x 5A
Sound level
Maximum sound level 70 dBA (measured 1 m from instrument)
Typical sound level 65 dBA (measured 1 m from instrument)
Environmental ranges
Ambient temperature 15 to 30
Transport temperature 0 to 45
Storage temperature 15 to 30
Maximum relative humidity 80% (non-condensing)
Altitude (max.) 2000m
Indoor use only
Not intended for use in a wet
location
Intended for use in Pollution
degree 2 environment
Operating limits
Centrifuge speed 0 to 3000×g (7000 rpm)
Flow rate Standard: 5 to 110mL/min (LS19)
Hi-Flow: 30 to 160mL/min (LS16 tube)
System pressure 100 kPa
Liquid temperature 4 to 38
Fluid density 1.0 to 1.1g/mL
Minimum working volume Approx. 50mL
Note: Smaller volumes of input material can be
diluted to enable processing
Chapter1Product information
System overview 1
CTS Rotea Counterflow Centrifugation System User Guide 11
(continued)
Category Property Value
Minimum concentrate recovery
volume
5 mL (subject to protocol optimization)
Centrifuge chamber volume 10 mL chamber containing up to 5×109cells (cell
type and operating parameters dependent).
Note: Centrifuge chamber can be filled multiple
times to process larger batches.
Maximum bag weight per pole 4.4lb (2kg)
Output concentration Up to 300×106 cells/mL (cell type and operating
parameters dependent)
CAN bus connection
Purpose For connection and control of external devices
Connector TE Connectivity T4111402051-000 M12 Plug
Connections Pin 1 - VCC CAN - +5V out
Pin 2 - CAN-H - Dominant High
Pin 3 - CAN-L - Dominant Low
Pin 4 - GND - Ground
Pin 5 - GND - Shield, optional
Chapter1Product information
System overview
1
12 CTS Rotea Counterflow Centrifugation System User Guide
Push button controls
Once a protocol has been loaded and confirmed, the Rotea instrument can be operated almost
entirely from the push buttons on the front of the instrument.
Figure3Instrument push button controls
Light STOP START/PAUSE ADVANCE DOOR UNLOCK
Push button
operation
Solid Push button to stop
the instrument and
the current protocol
Push button to start
the Protocol, pause
the current step in
the protocol or re-
start a paused step
Push button to
advance to the next
step in the protocol
Push button to
unlock the door
Flashing Instrument has
stopped due to
an alarm condition.
User intervention is
required.
Instrument is in a
paused condition.
Pressing allows
the instrument to
continue the current
step of the protocol.
Button can be
pushed, advancing
to the next step of
the protocol
N/A
Chapter1Product information
System overview 1
CTS Rotea Counterflow Centrifugation System User Guide 13
CTS Rotea Single-Use Kit schematic
The CTS Rotea Single-Use Kit enables users to configure kits to suit dierent protocols by adding
input and output vessels. Unused fluid lines are simply left sealed o.
6
3
5
1
2
4
78
9
10
11
12
Figure4 CTS Rotea Single-Use Kit
1Valve Hole & Tube ID
2Carrier Frame
3Bubble Trap
4Location Hole
5CFC Chamber
6Rotary Coupling
72D Barcode
8Pump Tubing
9Tube Retainer
10 Valve H Hole
11 Location Hole
12 Kit Tubing (Input and Output)
Chapter1Product information
System overview
1
14 CTS Rotea Counterflow Centrifugation System User Guide
Wetted components
The materials used in the manufacturing of the CTS Rotea Single-Use Kit have been chosen for their
biological and chemical compatibility.
Component Description Material
Pump tubing
Pump tube Bioprene
Kit tubing
0.160” OD, 0.116” ID
(4 mm OD, 3mmID)
DEHP-FREE PVC
CFC ChamberC
Base Tritan MX711
Cone Tritan MX711
Rotary Coupling
O-Rings EPDM, 70 DURO, FDA compliant
Cannular 316SS
Cap Tritan MX711
Body Tritan MX711
Bush & Collar Ketron LSG Peek Classix, USP
Class VI
Bubble Trap
Base Tritan MX711
Cap Tritan MX711
Fittings
Triple Y Connector PVC
Barbed T-Connector Polyproylene
Allowed chemicals for wetted surfaces
Chemical Concentration Maximum exposure time
WFI 100% 5 minutes
Ethanol in water 50% 5 minutes
Culture media 100% 5 minutes
Chapter1Product information
System overview 1
CTS Rotea Counterflow Centrifugation System User Guide 15
Installation and setup
Packaging contents
Rotea instrument
Hanger posts (2 o)
Mains plug IEC cables (USA/EUROPE/AUSTRALIA)
Pre-fitted mains fuses (2 x 5A)
USB-C (male) to USB-C (female)
Dell2in1 laptop tablet (this will be in a separate carton)
Instrument installation
Remove outer packaging
Prepare bench space to receive Rotea instrument unit (see “Working space” on page10) ensuring
that there is easy access to the “On/O” switch and mains IEC connector at the rear of the
instrument
CAUTION! The Rotea instrument should be installed on a stable bench. If a mobile set up is used,
it is the user’s responsibility to ensure all safety hazards and functional risks are addressed including
adequately supporting the weight of the instrument and having sucient wheel span so that both the
trolley and instrument are stable.
Ensure that the bench is stable and level to within ±0.4" (10mm)
Position the instrument on the bench with a minimum of 4" (10m) between the rubber feet and the
edges of the bench
CAUTION! The Rotea instrument weighs 40 lbs (18 kg) excluding hanger bag posts. Use
appropriate care and seek assistance if necessary when lifting or moving the instrument. Lift the
instrument, holding between the rubber feet on the bottom of the instrument.
CAUTION! Do not position the instrument is such a way that it would make it dicult to disconnect
the power cord from the power inlet.
CAUTION! If using an alternate mains cable, ensure it is rated for at least 10A.
2
16 CTS Rotea Counterflow Centrifugation System User Guide
Note: Suitable means of protective earthing are provided for this device
12
Figure5Back view of instrument
1Mains IEC connector
2USB-C port
1. Connect mains IEC connector.
2. Plug mains into power GPO
3. Connect the laptop to USB-C port on the instrument using the USB-C cable.
4. Insert 2 o bag hanger posts, adjust height to suit Kit and turn nut clockwise to secure in place.
5. Adjust the instrument position on the bench where it is required.
Chapter2Installation and setup
Instrument installation 2
CTS Rotea Counterflow Centrifugation System User Guide 17
Power instrument "On"
1. Switch on using the mains IEC switch at the rear of the instrument.
1
2. The instrument button lights will sequence as the instrument powers up. If the door is closed, the
instrument will automatically check if a Single-Use Kit is installed.
Chapter2Installation and setup
Power instrument "On"
2
18 CTS Rotea Counterflow Centrifugation System User Guide
Power instrument "O"
Switch o the instrument using the mains IEC switch or mains power.
1
CAUTION! There is no automatic recovery procedure. If a loss of power event has occurred see
“Restart after shut down” on page117.
Chapter2Installation and setup
Power instrument "O" 2
CTS Rotea Counterflow Centrifugation System User Guide 19
Installation
The Dell2in1 laptop will be pre-installed with Windows OS. The latest CTS Rotea Graphical User
Interface (GUI) software will need to be installed onto the Dell2in1 laptop provided. New releases of
the Rotea GUI software will be available for download from thermofisher.com.
1. Download the Rotea-X.X.X.exe installer.
Chapter2Installation and setup
Installation
2
20 CTS Rotea Counterflow Centrifugation System User Guide
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132

Thermo Fisher Scientific CTS Rotea Counterflow Centrifugation System User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI