English-4
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Equipment Alert
If the chair has a console, you will have to
cut the vacuum tubing routed to the console
and connect it to the vacuum tubing going to
the rear mounted assistant’s unit.
Bottom
Plate
1/8" Air - yellow
1/8" Water - blue
Optional Curing
Light (Red & Black)
3/4" Vacuum
Tubing
Ground
Wires
Bail Wire
Holder
Vacuum Power
Relay Harness
LIN Coupler
LIN Wire
From Chair
Note If the chair has a console, disconnect LIN wire labeled
“To Console” from LIN coupler and connect the LIN
wire from Assistant’s unit.
Step 10: Connect tubing and wiring harnesses.
A) Remove bottom plate (two screws).
B) Connect like colored tubing and mating harness
connectors.
C) Adhere LIN coupler inside chair frame cavity with velcro
strips.
D) Connect LIN lines to coupler box.
Caution
Connect ground wire to ensure system ground continuity.
Failure to do so may result in electric shock.
E) Connect ground wires to chair frame.
F) Connect the vacuum power to the relay harness.
G) Install bottom plate (two screws).
Note: Keep all connections above the bail wire holder. Use two or more
ties to bundle wiring and tubing together and strapped to the
chair frame.