Midmark Elevance® Standard Delivery System: A versatile dental system designed to enhance your practice's efficiency and patient comfort. Equipped with a range of features including adjustable height, easy-to-use controls, and various mounting options for accessories, it adapts seamlessly to your specific needs. Its sleek design and durable construction ensure longevity, making it a reliable investment for your dental practice.

Midmark Elevance® Standard Delivery System: A versatile dental system designed to enhance your practice's efficiency and patient comfort. Equipped with a range of features including adjustable height, easy-to-use controls, and various mounting options for accessories, it adapts seamlessly to your specific needs. Its sleek design and durable construction ensure longevity, making it a reliable investment for your dental practice.

English - 1
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Seat
Shroud
Seat
Frame
Bolts &
Washers
Elevance™ Console Installation
Equipment Alert
Console mount arm is attached to chair. Remove shipping bolt (take it out from top)
before installing mounting bolts, which is located in a bag tied to the console arm.
Language of origin: English
Step 3: Remove shipping bolt from seat casting. Position the
console mounting arm to desired left or right side.
Install mounting bolts, with lock patches, and torque
to 45-55 ft/lbs (61-75 Nm).
Step 4: Install seat frame and shrouds.
Mounting
Bolts
Shipping
Bolt
003-2283-99 Rev. C (3/28/12)
Step 1: Raise chair to comfortable working height.
Equipment Alert
Tap a lift arm or lower pivot cover three times to raise
the chair to the pre-programmed “home” position; a
convenient height for installation or patient access.
The chair beeps until it reaches the home position.
Upper Lift
Arm Cover
Lower Lift Arm Cover
Lower Pivot Cover
WARNING
Always disconnect chair from the power source before removing
any covers. Failure to do so may result in personal injury.
Step 2: Remove three screws and washers, then the
seat shroud from right and left sides.
Remove two bolts (with washers) and remove seat frame.
Style G
midmark.com
English - 2
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 6: Fit the console hub into the console mounting arm opening.
Console should be mounted parallel to the chair with the
water bottle towards foot end.
Step 7: Level console by tightening, alternately, the eight leveling bolts
on console mounting arm.
Magnets
Leveling Bolts
Mounting Ports
Step 5: Remove console side cover.
Loosen bottom screw and remove inside
console cover.
Tip: Console covers fit up under a lip at the top. Pull out from
bottom, then down to remove.
Vacuum and
Drain Lines
Tubing
and
Wiring
Electrical
Connections
Tubing
Connections
Console
Cover
Console
Side
Cover
for Delivery
Unit
for Light or
Monitor
for Cuspidor
Note
Tubing, wiring harnesses and vacuum lines will be routed
through console arm into the console as shown.
Style G
midmark.com
English - 3
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Note
Tubing and connectors are provided for connecting a delivery
unit and an assistant’s unit. Leave connectors capped if you are
not installing these components.
An additional 1/8" piece of blue tubing is provided that can be
used to connect to a city water supply.
Step 8: Connect like colored tubing for all installed
components and cap off all unused tubing.
Standard Console Bottled Water Tubing
OPTIONAL Console Bottled Water or
City Water Toggle Switch Tubing
Style G
midmark.com
Color Page
English - 4
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 9: All components have unique electrical connectors
that only fit into properly mated connectors.
Note: Not all installations use all the connectors provided.
Step 11: Keep electrical connections inside the middle and water
bottle side compartments so the cover will fit over them.
Console
Cover
Caution
Connect ground wire to ensure system ground continuity.
Failure to do so may result in electric shock.
Step 10: Screw all grounds into the two holes in the console
frame as shown.
Style G
midmark.com
Color Page
English - 5
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Test Verify the following:
1 Console is level.
2 All unused access ports are covered.
3
Electrical connections have been properly grounded to the
console frame.
4
All electrical connections are covered by the metal cover
inside the console.
5 Water bottle is screwed in tightly.
6
Covers are all installed. No wires or tubing are pinched by
covers or mountings.
7
Wipe console with a soft dry cloth to remove any dirt or
smudges.
Installation Checklist
Step 12: Install remaining covers, connect to power
and air supplies and test.
Style G
midmark.com
English - 6
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
This page is intentionally left blank
Style G
midmark.com
Deutsch - 6
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Style G
midmark.com
Français - 6
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Style G
midmark.com
Español - 6
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Style G
midmark.com
Italiano - 6
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Style G
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30

Midmark Elevance® Standard Delivery System Installation guide

Type
Installation guide

Midmark Elevance® Standard Delivery System: A versatile dental system designed to enhance your practice's efficiency and patient comfort. Equipped with a range of features including adjustable height, easy-to-use controls, and various mounting options for accessories, it adapts seamlessly to your specific needs. Its sleek design and durable construction ensure longevity, making it a reliable investment for your dental practice.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI