Midmark Elevance® Standard Delivery System Installation guide

Type
Installation guide
English - 1
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
DA3047i
Upper Pivot Cover
Elevance™ Cuspidor: Rear Mounted Installation
Language of origin: English
Upper Lift Arm Cover
Lower Lift Arm Cover
Switch
Bracket
Note
Cuspidor is wrapped and sealed in a box to protect the glass. Carefully inspect parts before installing.
003-2285-99 Rev. D (4/11/13)
Equipment Alert
Tap a lift arm or lower pivot cover three times to raise
the chair to the pre-programmed “home” position; a
convenient height for installation or patient access.
The chair beeps until it reaches the home position.
Lower Pivot Cover
Step 3: Remove one side then the other of the
lower pivot cover from the bracket
shoulder bolts and pins.
Step 4: Remove two screws from the
bottom of the switch bracket and
allow bracket to rotate down, out of
the way.
Step 1: Raise chair to comfortable working height.
WARNING
Always disconnect chair from the power source before removing
any covers. Failure to do so may result in personal injury.
Step 2: Remove two screws and the upper pivot cover.
Style G
midmark.com
English - 2
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
DA2418i
Upper Lift
Arm Cover
Mounting Pin
Lower Lift
Arm Cover
Note Only the bottom half of the chair is shown in illustrations for simplicity and focus on required tasks.
Step 5: Push in clip with screwdriver to release clip on each side.
Pull up to remove the upper lift arm cover.
Note: Upper lift arm cover has clips that fit over a dowel rod on front of the chair.
Step 6: Remove lower lift arm cover by working one side at a
time, moving cover to the side and back to free it from
mounting pins.
Style G
midmark.com
English - 3
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Set Screws
(4)
Set Screw
Adjustment
Hex Head
Screws (4)
Equipment Alert
Have an assistant support the cuspidor mount arm and
steady all pivot points to prevent equipment damage.
Step 7: Mount the cuspidor assembly on the chair pivot using four hex head screws and washers. Do not tighten hex head screws fully.
Step 8: Adjust four set screws to level
the assembly horizontally
Step 9: Level vertically by lifting either side
of the assembly (mounting plate)
and tightening the hex screws.
Style G
midmark.com
English - 4
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Constant Flow
Pilot Controlled Flow
Program Button
Touch (Pilot)
Controls
Limit Switch
Cuspidor Internal Components
Electric
Solenoid
Cuspidor Power
and Ground
Step 11: Install cup fill tube.
Alternate
Step 10: Install bowl and
strainer assembly.
Step 12: Install bowl rinse tube.
Note: May use alternate bowl rinse tube in facilities that have
a 1" (2.5 cm) air gap built into the drain system.
Style G
midmark.com
Color Page
English - 5
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 13: Remove bottom plate (two screws).
Bottom
Plate
1/8" Air - yellow
1/4" Water - blue
Ground
Cuspidor
Power
5/8” Drain
Bail Wire
Holder
Step 15: Keep all connections above the bail wire holder.
Use two or more ties to bundle wiring and vacuum
tubing together and strapped to the chair frame.
Install bottom plate (two screws).
Caution
Connect ground wire to ensure system ground continuity.
Failure to do so may result in electric shock.
Step 14: Connect tubing and cuspidor power connectors as
shown. Install hose clamp on drain tube. Connect
ground chair frame.
Style G
midmark.com
Color Page
English - 6
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
4
0
6
0
80
1
0
0
20
0
1
2
3
5
6
7
4
0
40
60
8
0
1
0
0
20
0
1
2
3
5
6
7
4
0
Step 16: Remove lid from J-box.
Step 18: Water valve connections
A) Connect 1/8” red pilot air tubing to T-Barb.
B) Replace 1/8" fitting on pilot valve with the 1/4" fitting from
the hardware kit. Connect 1/4" blue tubing.
C) Connect 1/8" blue tubing for city water to the T-Cross
fitting if city water option is being used.
Note: Replace 1/8” fitting on the T-Cross fitting with a plug from the hardware
kit, if CITY WATER option is not being used.
Step 17: Use hose clamp to install 5/8" tubing from umbilical
to facility drain.
Water Valve
Assembly
Step 19: Connect red, purple, clear with stripe and yellow
tubing to air valve assembly as shown.
Pilot Valve
T-Cross Fitting
T-Barb
To Water
Valve
Assembly
Air Valve
Assembly
From
Umbilical
From Air
Valve
Assembly
To
Power
Supply
Style G
midmark.com
Color Page
English - 7
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
DA3059i
Step 22: Install upper pivot cover (two screws in chair pivot).
Step 20: Return switch bracket to the upright position and
install two screws/lock washers in bottom.
Step 21: Install lower pivot cover (slip slotted holes over shoulder
bolts and hooks over pins on switch bracket).
Step 23: Install lift arm covers (reverse Steps 5 and 6) and
connect chair to power supply. Verify cuspidor
operation.
Upper Pivot
Cover
Lower Pivot Cover
Switch Bracket
Style G
midmark.com
English - 8
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Cup Fill
Button
Bowl Rinse
Button
Program
Button
Bowl rinse...
Press bowl rinse button to activate the bowl rinse function.
Note: The default bowl rinse function allows water to run for 15 seconds
then stops.
Program a new time for the cup fill function...
A) Press and release program button.
Wait to hear a beep to confirm it is in program mode.
B) Press and hold cup fill button for the desired amount
of cup fill time.
C) Listen for two beeps to confirm program.
Cup fill...
Press cup fill button to activate the cup fill
function. Press and hold button after a cup
fill to “top off” cup.
Note: The default cup fill function allows water to run
for 5 seconds then stops.
Program a new time for the bowl rinse function...
A) Press and release program button.
Wait to hear a beep to confirm it is in program mode.
B) Press and hold the bowl rinse button for the desired
amount of bowl rinse time.
C) Listen for two beeps to confirm program.
Cuspidor Lockout...
Tap program button three times to activate
cuspidor lockout. Listen for a beep to confirm
that the lockout is active.
Tap program button three times to inactivate
cuspidor lockout. Listen for two beeps to
confirm the lockout is inactive.
Cuspidor Operation
Style G
midmark.com
English - 9
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Test Verify the following:
1 Mount arm and cuspidor are level.
2 Mount arm rotates and moves easily.
3
Mount arm stays where placed without drifting. Adjust drag on pivot
points if necessary.
4
Chair lift arm covers are in place and limit switches activate. Chair
returns to home position when a cover is tapped three times.
5
Pivot covers are in place and limit switches activate. Chair
returns to home position when lower pivot cover is tapped three times.
6 No wires or tubing are pinched by covers or mountings.
7 Wipe components with a soft dry cloth to remove any dirt or smudges.
8 Pressing cup ll or bowl rinse button activates the selected function.
9
Cup Fill or Bowl Rinse functions can be programmed for new times.
(See page 8.)
10 There are no leaks in the ll or drain connections or tubing.
11
If cuspidor makes contact with an object while chair is being lowered, the
chair stops.
DA2424i
Mount Arm
pivot points
Allen wrench
To Adjust drag on pivot points...
Tighten or loosen the set screw near the desired pivot point to
modify the pivot’s drag.
Installation Checklist
Style G
midmark.com
English - 10
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
This page is intentionally left blank
Style G
midmark.com
Deutsch - 10
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Style G
midmark.com
Français - 10
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Style G
midmark.com
Español - 10
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Style G
midmark.com
Italiano - 10
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Style G
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50

Midmark Elevance® Standard Delivery System Installation guide