Midmark European Assistant's Unit Installation guide

Type
Installation guide

Midmark European Assistant's Unit is a versatile and efficient piece of equipment designed to enhance your dental practice. Here are some of its key features and use cases:

  • Easy installation: The unit can be easily mounted on an Elevance™ console or LR model chair, providing you with a convenient and accessible solution.

  • Ergonomic design: The unit's ergonomic design allows for comfortable and efficient operation, reducing strain and fatigue during extended use.

  • Multiple instrument integration: The unit can accommodate a variety of instruments, including HVE instruments, providing you with a comprehensive solution for your dental procedures.

Midmark European Assistant's Unit is a versatile and efficient piece of equipment designed to enhance your dental practice. Here are some of its key features and use cases:

  • Easy installation: The unit can be easily mounted on an Elevance™ console or LR model chair, providing you with a convenient and accessible solution.

  • Ergonomic design: The unit's ergonomic design allows for comfortable and efficient operation, reducing strain and fatigue during extended use.

  • Multiple instrument integration: The unit can accommodate a variety of instruments, including HVE instruments, providing you with a comprehensive solution for your dental procedures.

English-1
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
003-2290-99 Rev. D (4/4/13)
European Assistant’s Unit: Rear Mounted Installation
Note
The rear mounted European Assistant’s Unit can be mounted on an Elevance™ console or LR model chair.
Switch
Bracket
Language of origin: English
Upper Pivot
Cover
Step 3: Remove one side then the other of
the lower pivot cover from the bracket
shoulder bolts and pins.
Step 2: Remove two screws and the upper pivot cover.
Step 4: Remove two screws from the bottom of
the switch bracket and allow bracket to
rotate down, out of the way.
Equipment Alert
Tap a lift arm or lower pivot cover three times to raise
the chair to the pre-programmed “home” position; a
convenient height for installation or patient access.
The chair beeps until it reaches the home position.
Upper Lift Arm Cover
Lower Lift Arm Cover
Lower Pivot Cover
Lower Pivot
Cover
Step 1: Position equipment.
A) Raise chair to a comfortable working height.
WARNING
Always disconnect power before removing covers.
Failure to do so may result in personal injury.
B) Disconnect power.
midmark.com
Style G
English-2
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
DA2418i
Mounting Pin
Step 5: Push in clip with screwdriver to release clip on each side.
Lift up to remove the upper lift arm cover.
Note: Upper lift arm cover has clips that fit over a dowel rod on front of the chair.
Upper Lift
Arm Cover
Step 6: Remove lower lift arm cover by working one side at a
time, moving cover to the side and back to free it from
mounting pins.
Lower Lift
Arm Cover
Note Only the bottom half of the chair is shown in illustrations for simplicity and focus on required tasks.
midmark.com
Style G
English-3
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Equipment Alert
Have an assistant support the mount arm and
steady all pivot points to prevent damage to the
equipment.
Set Screw
Adjustment
Step 7: Mount the rear assistant’s unit assembly on the chair pivot using four hex head screws and washers. Do not tighten hex head screws fully.
Step 8: Adjust four set screws to level
the assembly horizontally
Step 9: Level vertically by lifting either side
of the assembly (mounting plate)
and tightening the hex screws.
Set Screws
(4)
Hex Head
Screws (4)
midmark.com
Style G
English-4
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Equipment Alert
If the chair has a console, you will have to
cut the vacuum tubing routed to the console
and connect it to the vacuum tubing going to
the rear mounted assistant’s unit.
Bottom
Plate
1/8" Air - yellow
1/8" Water - blue
Optional Curing
Light (Red & Black)
3/4" Vacuum
Tubing
Ground
Wires
Bail Wire
Holder
Vacuum Power
Relay Harness
LIN Coupler
LIN Wire
From Chair
Note If the chair has a console, disconnect LIN wire labeled
“To Console” from LIN coupler and connect the LIN
wire from Assistant’s unit.
Step 10: Connect tubing and wiring harnesses.
A) Remove bottom plate (two screws).
B) Connect like colored tubing and mating harness
connectors.
C) Adhere LIN coupler inside chair frame cavity with velcro
strips.
D) Connect LIN lines to coupler box.
Caution
Connect ground wire to ensure system ground continuity.
Failure to do so may result in electric shock.
E) Connect ground wires to chair frame.
F) Connect the vacuum power to the relay harness.
G) Install bottom plate (two screws).
Note: Keep all connections above the bail wire holder. Use two or more
ties to bundle wiring and tubing together and strapped to the
chair frame.
midmark.com
Style G
Color Page
English-5
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
4
0
60
8
0
1
0
0
20
0
1
2
3
5
6
7
4
0
Step 11: Remove J-box lid. Connect relay harness to the vacuum power relay box (located in the installation hardware kit).
A) Have an electrician connect facility power to the vacuum relay box and place relay box inside the J-box.
B) Connect tubing to facility air and water supplies as shown.
Step 12: If no other components are to be installed, finish installation by installing covers.
A) Install J-box lid to cover all tubing and wiring connections.
B) Install console side covers. (Refer to Step 1.)
Note: Be careful not to pinch or kink tubing or wires when installing covers.
Vacuum
Relay
Equipment Alert
Reference product specifications in
the European Assistant’s Unit User
Guide for electrical requirements for
the vacuum.
Facility Power
for Vacuum
midmark.com
Style G
Color Page
English-6
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Switch
Bracket
Step 13: Secure wiring and tubing.
A) Rotate switch bracket up to pivot and install two screws.
B) Install HVE instruments.
Upper Pivot Cover
Step 14: Install covers.
A) Install lower pivot cover (slip slotted holes over shoulder
bolts and hooks over pins on switch bracket).
B) Install upper pivot cover (two screws in chair pivot).
C) Install lift arm covers (reverse Steps 5 and 6).
Lower Pivot Cover
HVE Instrument
midmark.com
Style G
English-7
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Installation Checklist
To Adjust drag on pivot points...
Tighten or loosen the set screw near the desired pivot point to
modify the pivot’s drag.
DA2424i
Mount Arm
pivot points
Allen wrench
Test Verify the following:
1 Mount arm and European assistant’s unit are level.
2
Mount arm and European assistant’s unit rotate and move
easily.
3
Mount arm stays where placed without drifting. Adjust drag
on pivot point is necessary.
4
Chair lift arm covers are in place and limit switches activate.
Chair raises when a cover is tapped three times.
5
Pivot covers are in place the limit switch activates to raise
chair when it is lowered into an obstruction.
7
Assistant’s instruments work properly. No wires or tubing
are pinched by covers or mountings.
8
Wipe components with a soft dry cloth to remove any dirt or
smudges.
Step 15: Connect chair to power supply and verify operation.
midmark.com
Style G
English-8
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
This page is intentionally left blank
midmark.com
Style G
Deutsch - 8
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | +1-937-526-3662 | midmark.com
Style G
Français - 8
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 États-Unis | 1-800-643-6275 | 1-937-526-3662 | midmark.com
Style G
Español - 8
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 EE. UU. | 1-800-643-6275 | 1-937-526-3662 | midmark.com
Style G
Italiano - 8
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
Style G
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40

Midmark European Assistant's Unit Installation guide

Type
Installation guide

Midmark European Assistant's Unit is a versatile and efficient piece of equipment designed to enhance your dental practice. Here are some of its key features and use cases:

  • Easy installation: The unit can be easily mounted on an Elevance™ console or LR model chair, providing you with a convenient and accessible solution.

  • Ergonomic design: The unit's ergonomic design allows for comfortable and efficient operation, reducing strain and fatigue during extended use.

  • Multiple instrument integration: The unit can accommodate a variety of instruments, including HVE instruments, providing you with a comprehensive solution for your dental procedures.

Ask a question and I''ll find the answer in the document