  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo Pre- Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
59 905 940
Pre-Operation Treatment
of Electrodes
cs Návod k obsluze 2
da Brugsanvisning 3
de Betriebsanleitung 4
en Instruction Manual 5
es Manual de instrucciones 6
fr Instructions dutilisation 7
 Käyttöohje 8
hu Használatiutasítás 9
it Istruzioni per l‘uso 10
ja 11
ko 12
nl Gebruikershandleiding 13
pl Instrukcjaobsługi 14
pt Manual de instruções 15
ru Инструкция по эксплуатации 16
sv Bruksanvisning 17
th  18
zh 19
Pre-Operation of Electrodes 5
Pre-Operation of Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 59 905 940
en
a Whenunpacking,checkelectrodeforpossible
mechanicaldamagetothemembrane(1),the
diaphragmorthebody.Electrodessubjectto
complaints must be returned to us or to our
agentintheoriginalboxindicatingtheinvoice
number.
b Takeoffwateringcap(7)andushtheelectro-
de tip with distilled water. After rinsing with
distilledwater,wipetheelectrodewithatissue.
Donotrubasanincreaseoftheresponsetime
mightresult.
c Removerubbercapwhichcoverstheelectroly-
tellingaperture (4)orcutoffnipple(T)to
ensure pressure equalization with the atmo-
sphere.
d Eliminateanyairbubblesinsidethemembrane
space(6)ofthemeasuringelectrodebysmoo-
thlyswingingtheunitintheverticalplane.
e Torenewadriedoutglassmembrane,soak
it for several hours in the appropriate stora-
gesolution.Itisrecommendedtostoreglass
electrodesindist.water,combinedandrefe-
rence electrodes in the reference electrolyte
solution(asunderh).
f Afteralonganddrystorageperiod,theglass
membranecanbereactivatedwithhydrouoric
acid(orderno.51340073):Theelectrode
membraneisdippedfor1min.inthissoluti-
onfollowedbyrinsingwithdistilledwaterand
soakingfor12hoursinthestoragesolution.
g Electrodes with ground sleeve diaphragm:
Slightlyliftthegroundsleevesothatsomere-
ferenceelectrolytedropsout.
h Checklevelofelectrolyte(5)whichshouldbe
approx.10mmbelowrellingaperture(4).
Rell,ifnecessarywith:
–3MKCl+AgCl(orderno.51340045)for
electrodeswithAg/AgClwire
–3MKCl(orderno.51340049)orViscolyt™
(orderno.51340235)forhighthemperature
electrodeswithArgenthalcartridge
i Electrodeswithgelliedelectrolytewithoutrel-
lingportaretreatedasper§§a,b,d,e,fabove.
Forstoragewerecommendasolution3mol/l
KCl.
k Before calibrating or taking measurements,
attentionshouldbepaidtotheinstructionsof
therelativepHinstrumentsmanufacturer.Ad-
ditionally,pleaserefertotheMETTLERTOLEDO
brochure“PracticeandTheoryofpHMeasure-
ment”.
Pre-operation treatment of electrodes
ViscolytisatrademarkoftheMettler-ToledoGroup.
en Instruction Manual
4
(T)
2
5
3
6
7
1
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-Toledo Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.59905940 www.mt.com/pro