  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo pH electrodes type 465 Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
8 Type 465 pH Electrodes
© 04 / 16 Mettler-Toledo GmbH Type 465 pH Electrodes
59 905 937 Printed in Switzerland
Foralltype465combinationpHelectrodestobemounted
intothefollowinghousings:types763,764,776and796.
Werecommendthatthesensormayonlybeinoperation
withtheoriginalaccessoriesfromMETTLERTOLEDO.The
sensor may be operated and serviced only by personnel
familiarwiththeunit,andwhohavereadandunderstood
thisinstructionmanual.
1 Installation
a Whenunpacking,checkelectrodeforpossiblemechanical
damage to the pH sensitive glass, the diaphragm or the
body.Electrodessubjecttocomplaintsmustbereturnedto
usortoouragentintheoriginalboxindicatingtheinvoice
number.
b Takeoffwateringcapandwashtheelectrodetipwithdistilled
water.Afterrinsingwithdistilledwater,dabtheelectrodewith
atissue.Donotrubas anincreaseof theresponsetime
mightresult.
c Important: Reference electrolyte ller tube:The T-shaped
rubber stopper as well as the elastic securing band ha-
ve to be removed completely from the electrode,before
insertingtheelectrodeintothehousing.RemoveanyKCl
solutionresidinginthellingport.Re-assemblethehousing
asdescribedintheinstructionforthehousing.
d EliminateanyairbubblestrappedbehindthepHsensitive
glassbysmoothlyswingingtheunitintheverticalplane.
Thelead-offelementoftheglasselectrodemustdipintothe
innerbuffersolutionwithoutairbubbles.
e CombinationpHelectrodesstoreddryforseveraldayshave
tobesoakedintheappropriatestoragesolutionforseveral
hoursbeforeuse.Itisrecommendedtostorecombination
andreferenceelectrodesinthereferenceelectrolytesolution.
f Afteralonganddrystorageperiod,thepHsensitiveglasscan
bereactivatedwithhydrouoricacid(orderno.51319053):
Theelectrodemembraneisdippedfor1min.inthissolution
followed by rinsing with distilled water and soaking for
12hoursinthestoragesolution.
g Checkingelectrolytelevel:Combinationelectrodesofthetype
465containvariouselectrolytes.Forrellingalwaysusethe
typeofelectrolyteindicatedonthebodyoftheelectrode.The
imprint«rell9816»meanstorellwithViscolyt(orderno.
51340235);«rell9848»meanstorellwithFriscolyt-B
(orderno.51340053).
Theliquidlevelofaproperlylledelectrodeisabout1–2cm
belowthellingport.
2 Calibrating of electrode and pH transmitter
Thetransmitteriscalibratedusingtheelectrodeeitherready
formountingoralreadymountedintothehousing.Calibra-
tionismadeintwobuffersolutionsofaknownpHvalueby
dippingtheelectrode intothebufferandadjustingthepH
transmitter.ForfurtherstudiesinpHmeasurementthefollo-
wingbookletisavailablefreeofcharge:«pHTheoryGuide»
(orderno.30078149).Forotherlanguageversionsplease
contactyourMettler-Toledorepresentative.
en
en Instruction manual
General instructions
For all type 465 combination pH electrodes
Type 465 pH Electrodes 9
Type 465 pH Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 59 905 937
en
3 Maintenance
a Aftereachmeasuringcycletheelectrodetipandtheliquid
junctionmustbethoroughlyrinsedwithdistilledwater.Under
nocircumstancemusttheelectrodebeallowedtodrywith
measuringsolutioncoatingit.This is very important.
b On storage ofanassembledprobe, tip and liquid juncti-
onmustbeleftfullyimmersedinanalkalineKClsolution
(10partsof3MKClorViscolytplus2partsofbuffersolution
pH9.2).WithouttheadditionofbuffersolutionpH9.2the
housingmaycorrode.
c Theliquidlevelofthereferenceelectrolytemustbechecked
regularly.Ifithassunkbelowthebulbousshaftsectionof
theelectrode,thereferenceelectrolytemustbereplenished.
Attention is drawn to the instructions given in 1.g. above.
d Rapid rise in resistance of reference electrode: A rapid rise
inresistanceofthereferenceelectrode(upto1MΩ)after2
to 3 measuring cycles may occasionally be noticed in media
whicharehighlyenrichedwithproteins.Mosttransmitters
areverysensitivetothisriseinresistance,whichmanifests
itselfthroughadrift.Ifthishappens,theelectrodeshouldbe
treatedwithourelectrodecleaner(orderno.51340068).
Apreventiveaction is tousecombinationelectrodes with
Friscolyt-Breferenceelectrolyte(sufx/9848inproductde-
signation).
e Black diaphragm (liquid junction):If the solution to be
measuredcontainssuldecompounds,theliquidjunction
maybecomeblack.Simultaneously,itsresistancewillrise
signicantly.Suchanelectrodedoesnotworkproperlyany
longerandtreatingwithourdiaphragmcleaner(orderno.
51340070)becomesnecessary.
Apreventivemeasureistheuseofcombinationelectrodest-
tedwithasilver-iontrap(code–SC–inproductdesignation).
4 Cables
Makecertainthattheshieldofthecoaxialcablecarryingthe
reference electrode signal of the combined electrode is
never connected to ground.
CombinationelectrodewithscrewcapS7suitabletocable
socketAS9.
CombinationelectrodewithscrewcapK9suitabletocable
socketAK9.
38 Type 465 pH Electrodes
© 04 / 16 Mettler-Toledo GmbH Type 465 pH Electrodes
59 905 937 Printed in Switzerland
Notes
Type 465 pH Electrodes 39
Type 465 pH Electrodes © 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland 59 905 937
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-Toledo Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
Process Analytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.59905937 www.mt.com/pro
Addresses | Impressum
/