Midmark Elevance® Dental Chair User guide

Type
User guide
English-1
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
TP200 Rév. A
Service Procedure Reference Guide
003-1917-00 Rev. G (4/11/13)
Contents:
PC Board Cover Removal ..................... 1
Chair Calibration .................................... 1
Remote Control Synchronization ........... 1
Hydraulic Valve Adjustment ................... 2
1
2
3
4
P
WARNING
Always disconnect power before
removing covers. Failure to comply
may result in electric shock.
PC Board Cover Removal:
A) Using a screwdriver, push in on release
clip on each side of lift arm cover.
B) Lift cover up to disengage front clips.
C) Remove two screws & PC board cover.
Release Clips
#4 Dipswitch
(shown in ON position)
WARNING
Chair will move automatically when “TEST” button is
pressed. Stand back and allow chair to complete all three
Up / Down cycles of the calibration routine. Failure to comply may
result in serious personal injury.
Chair Calibration:
A) Disconnect power to the chair.
B) Move #4 dipswitch on S2 switch block to ON.
C) Connect power, then press “TEST” button on S1.
D) Allow chair to complete entire calibration cycle (see Note).
E) Disconnect power, then move #4 dipswitch to OFF.
F) Connect power - Check for proper operation - Install covers.
Note: The chair emits a short “beep” every two seconds during calibration.
Three long “beeps” indicate successful calibration is complete.
One long “beep” indicates calibration has failed. (Repeat procedure)
WARNING
Use caution when operating chair with covers removed.
Failure to comply may result in electric shock.
Remote Control Synchronization:
A) Verify that chair is properly calibrated.
B) Connect power to the chair, then move #4 dipswitch on
S2 switch block to ON. (two “beeps”)
C) Press & hold buttons “2” and “3” on the hand control within
three seconds from setting the dipswitch 4 to the ON position.
(Three “beeps” from the chair, then three “beeps” from the remote control)
D) Move #4 dipswitch to OFF.
E) Check for proper operation - Install covers.
Press & Hold
Style A
English-2
© 2011 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 | midmark.com
TP200 Rév. A
Throttle Valve Factory Settings
(Number of Turns OPEN from the FULLY CLOSED position)
Throttle Valve Chair Only Chair w/ L/R or Console
Base
TV1 Base Up 3 Full turns 5
1/2 Full turns
TV2 Base Down 3
1/4 Full turns 2 3/4 Full turns
TV5 Base Damper 3 Full turns 4
1/2 Full turns
Back
TV3 Back Up 1
1/2 Full turns 1 1/2 Full turns
TV4 Back Down 5
1/2 Full turns 5 1/2 Full turns
TV6 Back Damper 5 Full turns 5 Full turns
ATTENTION: This procedure adjusts the speed of the Base and Back cylinders.
Hydraulic Valve Adjustment:
A) Use the illustration to locate the throttle valve for the desired function.
B) Use a screwdriver to adjust the throttle valve as necessary.
C) Check for proper operation.
D) Repeat Step B as necessary to achieve desired speed.
Note: Midmark does not recommend adjustment of TV5 or TV6.
Back Cylinder
Base Cylinder
Reservoir
Pump
Base Down
Throttle Valve
Base Up
Throttle Valve
Back Down
Throttle Valve
Back Up
Throttle Valve
Speed Adjustment
Increase Decrease
Style A
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10

Midmark Elevance® Dental Chair User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI