Mettler Toledo electrode LoT 406-M6-DXK-S7 /25, LoT406-M6-DXK-S7 / 25, LoT406-M6-DXK-S7/25 User manual

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo electrode LoT 406-M6-DXK-S7 /25 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
LoT 406 Sensor Series
59 906 021
Puncture-Type
Combination pH Electrodes
LoT406-M6-DXK-S7 / 25
Instruction Manual
cs Návod k obsluze 2
da Brugsanvisning 6
de Betriebsanleitung 10
en Instruction Manual 14
es Manual de instrucciones 18
 Käyttöohje 22
fr Instructions dutilisation 26
hu Használatiutasítás 30
it Istruzioniperl‘uso 34
ja 38
ko 42
nl Gebruikershandleiding 46
pl Instrukcjaobsługi 50
pt Manual de instruções 54
ru Инструкция по эксплуатации 58
sv Bruksanvisning 62
th
 66
zh 70
14 LoT 406 Sensor Series
LoT 406 Sensor Series © 04 / 16 Mettler-Toledo GmbH
59 906 021 Printed in Switzerland
1 Introduction ................................................................... 15
2 Importantnotes .............................................................. 15
3 Connectingtheelectrode ................................................. 15
4 Operation ...................................................................... 15
4.1 Gettingready ................................................................. 15
4.2 Calibratingtheelectrode .................................................. 15
4.3 Measuring ..................................................................... 16
5 Maintenance .................................................................. 16
5.1 Cleaning ....................................................................... 16
5.2 Storingtheelectrode ....................................................... 16
6 Trouble-shooting ............................................................ 16
7 Electrodespecication .................................................... 17
8 Factorsthatshortenelectrodelife ..................................... 17
9 Literature ....................................................................... 17
Contents
Operating Instruction for puncture-
type combination pH electrodes
LoT406-M6-DXK-S7 / 25
Argenthal,XerolytandFriscolytaretrademarksoftheMettler-ToledoGroup.
en
Description of the electrode
A End cap
B Plug-inconnectorS7
(totcableAS7)
C ElectrodeshaftofPBTorglass
D Referencesystem
(Argenthal™
/polymerelectrolyte
Xerolyt
®
)
E Aperture(noceramicdiaphragm)
F pH-sensingglass(needle-shaped)
G Watering cap
a Penetrationdepth=25mm
b Diameter=6mm
A
B
C
E
F
G
D
a
b
en Instruction Manual
LoT 406 Sensor Series 15
© 04 / 16 Mettler-Toledo GmbH LoT 406 Sensor Series
Printed in Switzerland 59 906 021
1 Introduction
METTLERTOLEDOpuncture-typepHtheelectrodeswithXerolyt
polymer electrolyte are used for direct measurement at pH
infoods, for example fresh or curedmeat, cheeseand sh.
Beforeusingtheelectrode,readinstructionscarefully.
2 Important notes
Onreceipt,checkthatthepHelectrodeisundamagedandin
goodcondition. For further information, ordering and service
(advice),pleaserefertoyoursupplier.
Safety
Theelectrodestipismadeofglass.Iftheelectrodebreaks,
thereisariskofinjuryandofglasssplintersinthemeasured
material.
Wearglovesandgoggleswhenhandlingsolutions,e.g.when
attachingthewateringcaplledwithelectrolyte,andwhen
calibratingwithpHbuffersolutions.
3 Connecting the electrode
UseanAS7cabletoconnectthepuncturepHelectrodedirectly
toapHmeterorpHtransmitter (Fig. 4).
4 Operation
4.1 Getting ready
1 2
3
4
4.2 Calibrating the electrode
5 6
7
8
Please read the instructionsfor thepH meteror transmitter.
Werecommendcalibratingatthetemperatureofthemeasured
material.ApH/temperaturetableisprintedonthebufferlabel.
Automatic temperature compensationis possible by using a
separatetemperaturesensor.Ifhighlyaccuratemeasurementis
required,inserttheelectrodethematerial3–4minutesbefore
starting to take readings.
en
16 LoT 406 Sensor Series
LoT 406 Sensor Series © 04 / 16 Mettler-Toledo GmbH
59 906 021 Printed in Switzerland
4.3 Measuring
9 10
11
12
FollowtheinstructionsforthepH/mVmeterortransmitter.
5 Maintenance
5.1 Cleaning
Werecommendregularcleaning
Slight contamination
13
14
Fat residues
15
Ethanol
16
17
5.2 Storing the electrode
FriscolytorKCl3mol/l
h
Important:
–DonotkeeppHelectrodesindemineralizedordistilledwater,
asthiswillshortentheirlife.
–Do not store pHelectrodesdry.Iftheelectrodehasbeen
storeddry,regenerateitovernightinthewateringcapwith
KCl3mol/lorFriscolyt.
6 Trouble-shooting
Problem Cause Remedy
Air bubbles
inside
electrode tip
Bubblesformed
whilestored
horizontally
–Withtippointingdown,shake
theelectrodewithshorticking
movements(likeaclinical
thermometer)
Storeelectrodeinuprightposition
en
LoT 406 Sensor Series 17
© 04 / 16 Mettler-Toledo GmbH LoT 406 Sensor Series
Printed in Switzerland 59 906 021
Problem Cause Remedy
Air bubbles
on aperture
diaphragm;
Slope not
adjustableat
pH4
Temperature
uctuations
Placeelectrodeintapwaterat40°C
until bubbles disappear
Temperature
setting on
pHmeter
MakesurepHmeterissettothe
buffertemperature
Buffer solution
isdirtyorold
Freshbuffersolution
Electrodeoily
ordirty
Ifheavilyfouled,cleanelectrode
withethanol/ether1:1orhand-hot
(40°C)soapsolution
ThepH
metershows
unstable
readings
Electrodeoily
ordirty
Ifheavilyfouled,cleanelectrode
withethanol/ether1:1orhand-hot
(40°C)soapsolution
Problemswith
cable
Checkthatcableisproperly
connected,orchangethecable
Unusal
measured
values
pHmeter,
electrode,cable
–Checkcontactsatelectrode/cable
andcable/pHmeter
–Checkcableoruseadifferentone
–Changeelectrodeandrecalibrate
yourpHmeter
7 Electrode specication
Designation LoT406-M6-DXK-S7/25
OrderNo. 104063123
pHmeasurementrange 2…11
Temperaturerange 0…80°C
Referencesystem Argenthal
Referenceelectrolyte Xerolyt
Diaphragmtype Aperture
Plug-inconnectortype S7
Penetrationdepth 25mm
Diameter 6mm
Shaft PBT(FDAapproved)
PBT=Polybutyleneterephthalate
8 Factors that shorten electrode life
–Drystorage
–pH<2
–Prolongedimmersioninorganicsolvents
–Storageindistilledordemineralizedwater
–Temperatureuctuations
9 Literature
PracticeandtheoryofthepHmeasurement
206061531-E
ApplicationpHMeasurementinmeat
52 000 455
Application
pHMeasurementincheesemaking
52 000 459
ApplicationpHMeasurementinrawmeatprocessing
52 000 467
en
70 LoT 406 Sensor Series
LoT 406 Sensor Series © 04 / 16 Mettler-Toledo GmbH
59 906 021 Printed in Switzerland
1 ............................................................................ 71
2
..................................................................... 71
3 ..................................................................... 71
4 ............................................................................ 71
4.1 ............................................................................ 71
4.2 ..................................................................... 71
4.3 ............................................................................ 72
5 ............................................................................ 72
5.1 ............................................................................ 72
5.2 ..................................................................... 72
6 ..................................................................... 72
7 ..................................................................... 73
8
............................................ 73
9 ............................................................................ 73
zh
LoT406-M6-DXK-S7 / 25
pH
ArgenthalXerolytFriscolyt-
A
B S7
AS7
C PBT
D
Argenthal™
/
Xerolyt
®
E
F pH
G
a
=25mm
b =6mm
A
B
C
E
F
G
D
a
b
zh
LoT 406 Sensor Series 75
© 04 / 16 Mettler-Toledo GmbH LoT 406 Sensor Series
Printed in Switzerland 59 906 021
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-Toledo Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.59906021
www.mt.com/pro