Midmark Integra™ Operatory Cabinetry - Treatment Consoles Installation guide

Type
Installation guide
1
003-1890-00 Rev. D (1/12)
Freestanding Treatment Console Installation
Applies to Models:
FTC4472
FTC4452
Special Tools:
Level
3/8” masonry drill bit
9/32” wood drill bit
Hammer drill
Note
This Midmark casework unit must be connected to a dedicated circuit rated 20 A, 120 VAC.
Failure to comply could result in an overload of the electrical circuit and/or components. All
electrical and plumbing connections must be made by a licensed electrician and plumbing
contractor following applicable local, city, and national codes.
Midmark casework unit contains four electrical duplex outlets, which are rated at a combined 15
amps. This rating must be taken into account when plugging device(s) into the available outlets.
2113i
Note
Inspect all components for shipping damages. A concealed damage
report must be filed with the carrier (by the person recieving the goods)
within 15 days of delivery.
© 2009 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
NextNext
BackBack
Style A
midmark.com
2
21"
(53.3 cm)
42-1/2"
(108 cm)
2.5" (6.4 cm)
(max.)
2" (5.1 cm)
(min.)
5/8" (1.6 cm)
O.D.
2.5" (6.4 cm)
(max.)
2" (5.1 cm)
(min.)
1/2" (1.3 cm) NPT
16-1/2"
(41.9 cm)
10-7/8"
(27.6 cm)
10-7/8"
(27.6 cm)
12-3/8"
(31.4 cm)
3-1/16"
(7.8 cm)
2-3/8"
(6.0 cm)
3-1/16"
(7.8 cm)
Wall
22-5/8"
(57.5 cm)
3/4"
(1.9 cm)
44-1/4"
(112.4 cm)
7/8"
(2.2 cm)
2-7/8"
(7.3 cm)
Front
Floor Layout (as applicable)
A. The floor layout.
Note: Full size Floor Template and Wall Template are available upon request,
or refer to rough-in dimensions on next page.
B. Have licensed electrical and plumbing contractors position and
install electrical outlet receptacles and related plumbing using
the pre-installation template or layout.
Note: Be sure floor construction is sufficient to adequately support / anchor the
casework unit. If necessary, contact a licensed contractor to reinforce the floor.
Electrical
Supply
Top View of Base
120 VAC / 20 Amp / 60 Cycle
3-wire, Dedicated Circuit
Electrical
Terminal Box
Air Supply
Vacuum
Supply
Right Side Plumbing
Rough-In Perimeter
Water
Supply
Secure Terminal Box to cabinet
after electrical supply is connected
Left Side Plumbing
Rough-In Perimeter
NextNext
BackBack
3
31 -1/8"
(79.1 cm)
44-3/8"
(112.7 cm)
3-1/4"
(8.3 cm)
1-1/16"
(2.7 cm)
15"
(38.1 cm)
10"
(25.4 cm)
10"
(25.4 cm)
10"
(25.4 cm)
3"
(7.6 cm)
3"
(7.6 cm)
3-1/4"
(8.3 cm)
4-1/4"
(10.8 cm)
8-3/8"
(21.3 cm)
42-1/2"
(108 cm)
44-1/4"
(112.4 cm)
7/8"
(2.2 cm)
Wall Layout (as applicable)
Right Side Plumbing
Rough-In Perimeter
Left Side Plumbing
Rough-In Perimeter
Center
Rough-In Perimeter
Right Side Plumbing
Rough-In Perimeter
Left Side Plumbing
Rough-In Perimeter
Center
Rough-In Perimeter
Floor
NextNext
BackBack
4
DA2116i
5
4
3
1
2
Setting Up
A. If necessary, remove the drawers by pulling
the drawer outward until it meets resistance.
B. On both sides of drawer, pull outward until locking
slots (1) on drawer are free of tabs (2) on drawer
glides (3).
C. Lift drawer upward (4) while pulling outward to free
rear of drawer from hooks (5) on glide.
D. On base sections that have doors, open door and
remove bottom shelf.
Note
If applicable, refer to 2” Drawer Removal
on next page as removal is different.
Note
When reinstalling drawers assure back of drawer is
positioned beneath hooks (4) on glides and tabs (2)
on glide snap into locking slots (1) on drawer.
NextNext
BackBack
5
Removing the 2” (5 cm) Drawer:
To remove drawer and front slide...
A. While lifting or pushing down release levers
(on each side), slide rail and drawer from slide.
To remove drawer tray and front...
B. Remove four screws, then slide drawer
tray and front from channel.
To remove slides...
C. Remove four screws, then remove slides.
DA134900i
NextNext
BackBack
6
Side Storage Unit
A. Open the doors and lift out the bottom panel.
Note
You must remove the bottom panel, to install the
anchor mounting clips.
DA2128i
NextNext
BackBack
7
Locating Anchor Mounting Clips
A. Place the anchor mounting clips within 4” (10 cm)
of each corner. Be sure clips do not interfere with
the leveling bolts. [suggested locations of the four
mounting clips].
B. Using the anchor mounting clips as templates,
place location marks for the hold-down bolts on
the floor.
21"
(53.3 cm)
42.5"
(108 cm)
4 in. (10 cm)
maximum
4 in. (10 cm)
maximum
4 in. (10 cm)
maximum
4 in. (10 cm)
maximum
Anchor Mounting Clip
Note
The hold-down bolts for the anchors should be
placed as close as possible to the sub-base lip.
NextNext
BackBack
8
Installing Anchor Mounting Clips & Leveling
A. Using the appropriate drill bit (masonry: 3/8”)
(wood: 9/32”), drill holes at each location per
specification.
B. Loosely install the anchor mounting clips.
C. Using a 1/4” socket, adjust leveling screws in
each corner until cabinet is level.
D. Insert shims beneath area of anchor and
tighten each bolt.
E. Trim any excess shim material that may
protrude beyond base.
Note
To prevent instability, do not adjust leveling
screws more than 1/2” (1.3 cm).
Anchor Mounting Clip
3/8” diameter
9/32” diameter
Sub-base
2” (5.1 cm)
Sub-base
Concrete
Non-Concrete
Shim
1/2” (1.3 cm)
maximum
Lower Raise
NextNext
BackBack
9
2207i
Securing Electrical Box
A. Secure the electrical box to the inside
of the cabinet base with (#10 x 3/8”) screws.
Screws
NextNext
BackBack
10
C
D
DA2124i
A
B
E
E
A
F
G
Installing Overhead Section (if required)
A. Remove side panels and receptacle enclosure covers.
B. Position overhead section on top of midsection.
Be sure to align mounting holes.
C. Connect the plugs from the overhead section
and midsection.
D. Secure overhead section to midsection with
six 10-24 x 5/8” screws.
E. Install the receptacle enclosure and side panels.
F. Remove task light cover to access holes in upper
unit by taking out thumb screws.
G. Install particle board screws.
Equipment Alert
The particle board screws
must be installed for stability.
NextNext
BackBack
11
Installing Clock / Timer
A. Remove clock / timer from box located in cabinet and
push clock / timer into opening.
Installation of Power Supply for
Clock / Timer (if required)
B. Plug power supply to electrical outlet.
C. Install grommet into upper cabinet.
D. Route power supply wire through cabinet.
E. If necessary, install grommet into task light cover.
F. Plug power supply cord into the back of the
clock / timer.
A
F
DA2215i
D
C
B
E
NextNext
BackBack
12
DA2216i
B
D
A
C
Removal of Accessory Panel
A. Remove side panels and then pull out on
accessory panel.
Relocating Pull-out Shelf and
Swing-out Shelf or Tub
B. Remove six screws from shelf slides and
relocate pull-out shelf in desired position.
C. Loosen screws so the shelf or tub is loose.
D. Move shelf forward to remove. Remove screw,
then re-install at desired location (do not tighten).
Position shelf, then tighten screw.
Side Panel
Side Panel
Shelf Slide
NextNext
BackBack
13
Installing and Adjusting Doors / Panels / Shelves
A. Check doors for proper adjustment. [Gap between
doors should be approximately 1/16” (0.16 cm)].
B. Install shelf brackets at desired height of shelf.
Note
Refer to Hinge Adjustments for
proper adjustment procedures.
DA2129i
A
B
NextNext
BackBack
14
Hinge Adjustments
A. Side Adjustment
Turn screw (A) until door is in desired position.
B. Height Adjustment
Loosen screw (B) approximately 1/4 turn,
reposition door, then tighten screw.
C. Depth Adjustment
Loosen screw (C), reposition door,
then tighten screw.
A
A
A
A
B
B
C
C
A
B
B
A
A
C
Base Doors
Upper Doors
Bi-Fold Doors
Base Doors
Figure 12
NextNext
BackBack
15
WHITE
BLACK
GREEN/ YELLOW
WHITE
WHITE
BLACK
BLACK
WHITE
WHITE
GREEN
BLACK
BLACK
WHITE
GREEN
GRN
WHITE
BLACK
GREEN/ YELLOW
BLACK
RESET
BLACK
WHITE
GREEN
WHITE
BLACK
G/Y
G/Y
BLACK
WHITE
GREEN/YELLOW
BLACK
GRN/YEL
BLACK
GRN/YEL
BLACK
WHITE
GREEN
BLACK
WHITE
GREEN/YELLOW
BLACK
WHITE
GREEN
GRN/YLW
GREEN/YELLOW
G/Y
TEST
GREEN/YELLOW
BLACK
WHITE
GREEN
GREEN/YELLOW
GREEN/YELLOW
BLACK
BALLAST
1
2
1
2
3
3
3
2
1
3
2
1
3
2
1
3
2
1
N
L
BLACK
BLACK
BLACK
BLACK
BLACK
BLACK
WHITE
WHITE WHITE
WHITE
WHITE
BLACK
DA2195i
G
G
G
G
G
G
3
2
1
3
1
2
G
Optional Task Light
Mid Section of 12:00 Unit
User
Electrical
Connections
Freestanding Treatment Console
(FTC-4472)
Note
Only Licensed Electricians following local
and applicable electrical codes should
install and connect the electrical supply power.
Midmark cannot be held responsible for
installations that do not conform to local codes.
Equipment Alert
Dedicated branch circuit rated at
120VAC, 50/60 Hz, 20 amp.
NextNext
BackBack
16
WHITE
BLACK
GREEN/ YELLOW
WHITE
WHITE
BLACK
BLACK
WHITE
WHITE
GREEN
BLACK
BLACK
WHITE
GREEN
WHITE
BLACK
GREEN/ YELLOW
BLACK
N
L
G
DA2196i
G
GREEN
Freestanding Treatment Console
(FTC-4452)
Note
Only Licensed Electricians following local
and applicable electrical codes should
install and connect the electrical supply power.
Midmark cannot be held responsible for
installations that do not conform to local codes.
Equipment Alert
Dedicated branch circuit rated at
120VAC, 50/60 Hz, 20 amp.
NextNext
BackBack
Midmark Corporation
For contact information, go to:
www.midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16

Midmark Integra™ Operatory Cabinetry - Treatment Consoles Installation guide

Type
Installation guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI