  • Hello! I am an AI chatbot trained to assist you with the Midmark 244 Barrier-Free™ Bariatric Power Treatment Table Installation guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
1© Midmark Corporation 2004
003-1457-00 Rev. (C) 5/24/16
Power Cord
Hand Control Cord
Power Switch
Step 2: Turn power switch to ON (I)
Step 1: Connect hand control cord.
Connect power cord.
Equipment Alert
Table must be leveled to ensure proper operation.
Step 3: Adjust four leveling feet until table is level.
For models:
244 (all)
Barrier-Free™ Bariatric Power
Treatment Table Installation
Before connecting the Midmark device power cord into an electrical
power source outlet, it is recommended to contact a local licensed
electrician or a local Authority Having Jurisdiction over the internal
building power distribution to verify that the device when powered
up will comply with all local electrical codes. Use the electrical
power rating marked on the device when determining the
appropriate electrical branch circuit and outlet requirements.
For 115V rated device models with a current rating at or above
12 Amps, a dedicated branch circuit may be required if the branch
circuit protection is provided by a 15 Amp breaker.
/