Mettler Toledo InPro 7100-VP Series Conductivity Electrode Installation guide

Type
Installation guide
InPro 7100-VP Series
Conductivity Electrode
Quick Setup Guide
cs Průvodcerychlýmnastavením 2
da Lynvejledning 5
de Quick-Setup-Leitfaden 8
en QuickSetupGuide 11
es Guíadeconguraciónrápida 14
 Pika-asetusopas 17
fr Guidedeparamétragerapide 20
hu Gyorsbeállításiútmutató 23
it Guidaallacongurazionerapida 26
ja
29
ko 32
nl Beknoptehandleiding 35
pl Podręcznikszybkiejkonguracji 38
pt GuiadeConguraçãoRápida 41
ru Руководство по быстрой настройке 44
sv Snabbinstallationsguide 47
th
 50
zh 53
InPro7100-VPSeries
30313390
©01/18Mettler-ToledoGmbH InPro 7100-VPSeries
PrintedinUSA 30313390
InPro7100-VPSeries 11
InPro 7100-VP Series (Conductivity Electrode)
Quick Setup Guide
1 Introduction
ThankyouforbuyingthisInPro
®
7100-VPSeries4-electrodeconductivitysensorfrom
ME
TTLER
T
OLEDO.
I
nPro
7
100-VP
S
eries
4
-electrode
c
onductivity
s
ensors
a
re
i
ntended
tobeusedtomeasureconductivityinmediumtohighconductivitywater.Thesensors
a
re
available
in
a
wide
selection
of
process
connections
to
meet
every
application
need.
2 Safety instructions
h
Notice:Beforeeverystart-up,thesensormustbecheckedfor:
Damagetotheconnections,fastenings,etc.
L
e a k a g e
Perfectfunctioning
Authorizationforuseinconjunctionwithotherplantequipmentandresources.
h
Notice: The manufacturer/supplier accepts no responsibility for any damage
c
aused
b
y
u
nauthorized
a
ttachments
o
r
f
or
t
he
i
ncorporation
o
f
s
pare
p
arts
w
hich
arenotofMETTLERTOLEDOprovenance.Theriskisborneentirelybytheoperator.
a
Caution:Beforethesensorisputintooperation,theoperatormusthavealready
c
lariedthatuseofthesensorinconjunctionwiththeotherassociatedequipment
andresourcesisfullyauthorized.
a
Caution:Adefectivesensormayneitherbeinstallednorputintooperation.Faulty
c
ontainment
ori
nstallation
o
ut
o
f
c
onformance
w
ith
r
egulations
a
nd
i
nstructions
canleadtotheescapeofmediumortopressuresurges(explosion),potentially
h
armfulbothtopersonsandtotheenvironment.
a
Caution:BeforeinstallationinEx-zones,pleasereadandobservetheguidelines
(
frominstructionmanuals)andadheretosiderequirements.
a
Warning:Beforedismantlingthesensororcommencinganymaintenanceworkon
i
t,
ensure
that
the
equipment
in
which
the
sensor
is
installed
is
in
a
safe
condition
(depressurize,noexplosionrisk,empty,rinse,vent,etc.).
3 Product description
The printed markings oneach InPro 7100-VP Series 4-electrode conductivity sensor
c
ontainthefollowinginformation:
MettlerToledo Sensormanufacturer
InPro710X-VP Typeofsensor
CellM:XXXXXX Cellconstant
TempM:XXXXXX Temperatureconstant
Serial
N
o:
XXXXXXXXX
Serial
n
umber
Ex-classication
III/2GExiaIIBT3/T4/T5/T6Ga/Gb
(plasticsensors)
Examinationcerticatenumber:SEV14ATEX0129X
Notiedbody: Notiedbody:ElectrosuisseSEV,(No.:1258)
en
en QuickSetupGuide
InPro 7100-VPSeries ©01/18Mettler-ToledoGmbH
30313390 PrintedinUSA
12 InPro7100-VPSeries
4 Sensor installation
Thesensoristobemountedinsuchawaythatthesamplemediumowsdirectlyacross
t
he
s
ensor
s
urface.
A
ny
o
ther
i
nstallation
p
osition
o
f
t
he
s
ensor
m
ay
c
reate
t
he
r
isk
o
f
formationofairlocks,orofcontaminationthroughdepositsofsolidmatter.
Flowofmediummustbeacrossthefaceofthesensor.Airbubblesandsolidmatter
d
eposits
are
to
be
avoided.
A
minimum
clearance
of
30
mm
(1.18”)
between
sensor
end
andpipewallmustbemaintained(seedrawing).
Verticalmounting(topentry)inapipeisrecommendedonlyifthepipeisfullandnoair
b
ubblesareabletodevelopintheow.
In
t
he
e
vent
o
f
s
ide
m
ounting
(
side
e
ntry)
o
f
t
he
s
ensor,
v
ertical
u
pward
ow
o
f
s
ample
processmediummustbeensured.
Connection of the sensor
AlloftheInPro7100-VPSeries4-electrodeconductivitysensorscanbeconnectedtothe
a
ssociatedtransmitterusingtheappropriatemulti-wirecablefromMETTLERTOLEDO.
Colorandfunctionofthecablestrands(suppliedseparately)
4-electrodesensor Cablecolor
Cndouter1 w h i t e / b l u e
Cndinner1 white
Cndinner2 blue
Cndouter2 black
RTD green
RTDsense red
RTDret/GND Bareshield
5 Operation
Foreachsensor,thecellconstantandtemperatureconstanthavebeenmeasured/es-
tablished
i
ndividually
a
nd
t
he
v
alues
d
ocumented
i
n
t
he
a
ccompanying
Q
uality
C
erticate.
Ingeneral,thecellconstantandtemperatureconstantwillnotchange.Itisnotnecessary
t
o
p
erform
s
ensor
c
ell
c
onstant
c
orrection
orc
alibration
w
ith
s
tandard
s
olutions.
H
owev-
er,dependingonplantoperatingprocedures,thecellconstantcanbeveried/corrected
a
fter
t
he
p
eriod
o
f
i
nitial
c
alibration
h
as
e
xpired
b
y
u
sing
c
onductivity
s
tandard
s
olutions.
T-piece
Flow
direction
Flow direction
Sensor
> R 30
Examplesensorinstallation.
en
©01/18Mettler-ToledoGmbH InPro 7100-VPSeries
PrintedinUSA 30313390
InPro7100-VPSeries 13
6 Maintenance: cleaning the sensor
Incaseofanyfailureormalfunction,theequipmentinwhichthesensorisinstalled,needs
t
obeinasafecondition(depressurized,empty,rinse,vent,etc.).
C
ontaminated
c
onductivity
c
ells
c
an
g
ive
f
alse
i
ndications.
I
f
c
ontamination
i
s
s
uspected,
dismounttheconductivitycellandcleantheelectrodesandtheinsulationbetweenthe
e
lectrodes
w
ith
a
s
oft
c
loth.
A
m
ild
d
etergent
o
r
a
h
ighly
d
iluted
s
olution
(
less
t
han
0
.5%
byweight)ofacid,suchasnitricacid,issuitabletouseasacleaningsolution(always
h
andleacidswithcaution!).
Only
u
se
c
leaning
a
gents
t
hat
a
re
c
ompatible
w
ith
t
he
m
aterials
t
o
b
e
c
leaned.
I
n
a
ddition,
donotusecoarsematerialssuchassteelwoolforcleaningastheywilldamagethe
s
urface.
R
insetheconductivitycellthoroughlywithdistilledwaterbeforeinstallation.
7 Disposal
Theuserisresponsiblefortheproperdisposaloftheconductivitysensor.The
s
ensor
c
ontains
el
ectronic
c
omponents
t
hat
m
ust
b
e
d
isposed
o
f
c
orrectly
i
n
ordertoavoidhazardstopeopleortotheenvironment.
8 Warranty
Warrantedfreeofmanufacturingdefectsforatermof12monthsafterdelivery.
9 Storage conditions
Inspecttheinstrumentimmediatelyafterreceivingforanytransportdamage.Thesensor
s
hould
b
e
s
tored
i
n
a
d
ry,
c
lean
a
nd
p
rotected
a
rea
u
ntil
t
he
t
ime
o
f
i
nstallation.
I
f
t
he
sensorhasbeenremovedfromtheprocess,itshouldbethoroughlycleanedanddried.
I
tshouldbestoredinadry,cleanandprotectedareauntiltimeofinstallation.
10 EC Declaration of Conformity
Thecompletedeclarationisavailableinthedeclarationofconformitycerticate.
InProisaregisteredtrademarkofMettler-ToledoGmbHinSwitzerland,theUSA,theEuropean
U
nionandafurthervecountries.
en
InPro 7100-VPSeries ©01/18Mettler-ToledoGmbH
30313390 PrintedinUSA
56 InPro7100-VPSeries
Lead
Mercury
cadmium
六价铬
Hexavalent
chromium
多溴联苯
Polybrominated
biphenyls
多溴二苯醚
Polybrominated
diphenyl ethers
(Pb) (Hg) (Cd) (Cr6+) (PBB) (PBDE)
插头
Elect. Connector
×
:表示该有害物质在该部件所有均质材料中的含量均在GB/T 26572规定的限量要求以下
Indicates that the content of the hazardous substance in all homogeneous materials of the part is below the limit specified in GB / T 26572
×:表示该有毒物质至少在该部件的某一均质材料中的含量超出GB/T 26572规定的限量要求。
×Indicates that the content of the toxic substance in at least one of the homogeneous materials of the part exceeds the limits specified in GB/T 26572.
本表依据SJ/T 11364的规定编制。 本产品符合以下标志规范:
Table composed in accordance with SJ/T 11364 (CN). This product is bearing the following symbol:
产品中有害物质的名称及含量
Toxic and hazardous substance name and containment in product
有毒有害物质或元素
Toxic and hazardous substances
部件名称
Part Name
zh
©01/18Mettler-ToledoGmbH InPro 7100-VPSeries
PrintedinUSA 30313390
InPro7100-VPSeries 57
Notes:
InPro 7100-VPSeries ©01/18Mettler-ToledoGmbH
30313390 PrintedinUSA
58 InPro7100-VPSeries
Notes:
©01/18Mettler-ToledoGmbH InPro 7100-VPSeries
PrintedinUSA 30313390
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmtbr@mt.com
France Mettler-Toledo Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mail[email protected]
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444747
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
01/2018.Rev.B.©Mettler-ToledoGmbH
PrintedinUSA.30313390
www.mt.com/pro
1258
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60

Mettler Toledo InPro 7100-VP Series Conductivity Electrode Installation guide

Type
Installation guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI