Mettler Toledo InPro 5000 (i) Sensors Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo InPro 5000 (i) Sensors Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
14 InPro 5000 (i) CO
2
Sensor Series
© 06/2019Mettler-ToledoGmbH InPro5000(i)SensorSeries
30 313 393 B PrintedinSwitzerland
1 Introduction ................................................................... 15
2 Safetyinstructions .......................................................... 15
3 Productdescription ........................................................ 15
4 Sensorinstallation .......................................................... 16
4.1 Connection .................................................................... 16
5 Operation ..................................................................... 16
5.1 CalibrationofthepHsensor ............................................ 16
5.2 Mountingthemembranebody ......................................... 17
5.3 CalibrationoftheCO
2
sensor .......................................... 17
6 Maintenance .................................................................. 17
7 Disposal ....................................................................... 17
8 Warranty ....................................................................... 17
9 Storagecondition ........................................................... 17
10 ECDeclarationofConformity ........................................... 17
Contents
InPro 5000 (i)
CO
2
Sensors
Quick Setup Guide
InProandISMareregisteredtrademarksof
Mettler-ToledoGmbHinSwitzerland,theUSA,
theEuropeanUnionandafurthervecountries.
en
en QuickSetupGuide
InPro 5000 (i) CO
2
Sensor Series 15
InPro 5000 (i) Sensor Series © 06/2019Mettler-ToledoGmbH
PrintedinSwitzerland 30 313 393 B
1 Introduction
Thank you for buying this InPro 5000(i) sensor from
METTLER TOLEDO. InPro 5000(i) sensors are intended
solelyfor in-line measurementof dissolved CO
2
(carbon
dioxide).
2 Safety instructions
h
Notice:TheInPro5000(i)sensorshouldbeinstalled,oper-
ated,maintainedonlybypersonnelfamiliarwiththesensor
andwhoarequaliedforsuchwork.
h
Notice:Adefectivesensormustneitherbeinstallednorput
intoservice.
a
Caution: Only the maintenance work described in these
operatinginstructionsmaybeperformedonthesensors.
h
Notice:Whenchangingfaultycomponents,useonlyorig-
inal spare parts obtainable from your METTLER TOLEDO
supplier.
h
Notice:Nomodicationstothesensorortheaccessories
areallowed.Themanufacturer/supplieracceptsnorespon-
sibilityfordamagecausedbyunauthorizedmodications.
Theriskisborneentirelybytheuser.
3 Product description
Eachsensorissuppliedtogetherwith:
1bottleCO
2
-electrolyte25ml
1qualitycontrolcerticate
–1inspectioncerticate3.1(inaccordancewithEN
10 2 0 4. 3 / 3.1)
1quicksetupguide
1ECDeclarationofConformity
Membranebodyandcapsleevearenotmounted.Theinte-
riorbodyisprotectedagainstdryingoutwithawettingcap,
lledwithCO
2
-electrolyte.
Pleasecheckcompleteness.
en
16 InPro 5000 (i) CO
2
Sensor Series
© 06/2019Mettler-ToledoGmbH InPro5000(i)SensorSeries
30 313 393 B PrintedinSwitzerland
1 VPconnector
2 ISMconnector
3 Hexagonalscrew
4 ThreadedsleevePg13.5
5 WasherPTFE
6 O-Ring(10.77×2.62mm,
SiliconeFDA/USPVI)
7 Retainernut
8 Membranebody
9 Interiorbody
10 Capsleeve
11 Wettingcap
4 Sensor installation
4.1 Connection
Theinteriorbody(pHelectrode)isreadyforcalibration.
–Onunpacking,checkthepHelectrodeformechanicalda-
mage.ReportanydamageimmediatelytoyourMETTLER
TOLEDOsupplier.
–Removethewateringcap,capsleeve,membranebody
andbrieyrinsethesensorwithde-ionizedwater.
–Afterrinsing,thesensorshouldonlybedabbeddrywitha
tissue.DonotrubthepH-sensitiveglassasthiscanlead
toelectrostaticchargingandsluggishresponsetimes.
–CheckinsidethepH-sensitiveglassforthepresenceof
any air bubbles and remove same by gently swinging
theelectrodeinaverticalplane(shakingdownaswitha
thermometer).
–CalibratethepHsensorasdescribedinSection5.
–Placetheelectrodeintothehousingasdescribedinthe
instructionsforthehousing.
h
Notice:Mountingupside-downorhorizontallyisnotpos-
sibleduetotheinternalpHelectrode.Anangleof15°or
greatertohorizontalisneeded.
–Connect the electrode and the pH transmitter with the
correspondingcable(AK9fortheInPro5000(i),VPfor
the InPro 5000). Observe the connection scheme that
accompaniesthecableorthetransmitter.
ISM
®
sensorssuchasInPro5000ienablePlugandMea-
surestartupandexpandeddiagnostics.Forinstallation,
commissioningandoperationoftheISMsystemplease
refertoinstructionsforthetransmitter,moduleandcable.
5 Operation
5.1 Calibration of the pH sensor
(NotneededforInPro5000i)
a:Min.immersiondepth>25mm/1"
For the analog CO
2
sensor InPro 5000,we recommend
a2-pointcalibrationinpHbuffers7.00and9.21.Before
calibration,rstremovewettingcap.StartwithbufferpH
7.00fordeterminingzeropointandthenbufferpH9.21for
determiningtheslope.Pleaserefertoinstructionmanualof
thetransmitterforfurtherdetails.
h
Notice:Onthetransmitter,selectPt1000asthetempera-
turesensorfortheanalogInPro5000.TheInPro 5000i
doesnotrequireatemperaturesetting.
en
InPro 5000 (i) CO
2
Sensor Series 17
InPro 5000 (i) Sensor Series © 06/2019Mettler-ToledoGmbH
PrintedinSwitzerland 30 313 393 B
h
Notice:Pleasedonotleavethesensorinthecalibration
bufferforlongerthan1hour.Toolonganexposuremay
resultinhighsensordriftintheprocess.
5.2 Mounting the membrane body
–AftercalibrationofthepHelectroderinsethepHelectrode
withde-ionizedwaterandcarefullydabitwithapaper
tissue.
FillthenewmembranebodyhalfwaywithCO
2
electroly te.
–Makesurethatallairbubblesareremovedfromthemem-
brane body. Air bubbles can be removed by carefully
tappingonthemembranebody.
–Slip the membrane body over the pH electrode while
holding the sensor in a vertical position. The excess
electrolytethatwillbedisplacedshouldberemovedwith
apapertissue.
h
Notice: No electrolyte, sample media or contamination
shouldbepresentbetweenthemembranebodyandthe
capsleeve.Pleasecheckcarefully.
–Carefully slip the cap sleeve over the tted membrane
body,holdingthesensorinaverticalposition,andscrew
ittight.Thecapsleevemustbecleanandtight.
5.3 Calibration of the CO
2
sensor
TheInPro5000isinsitusterilizableorinanautoclave.After
sterilizationaprocesscalibrationstepisneededtoachieve
accuratemeasurements.
The process calibration can be performed by using the
transmitter’s calibration menu. For further details please
refertothemainmanual.
6 Maintenance
Werecommendusinganewmembranebodyaftereach
batch.AftereachcalibrationofthepHelectrodenewCO
2
electrolyteshouldbelledinthemembranebody.
IfthepHelectrodeshowsasluggishresponsetimeand/or
insufcientsensitivity,thepH-sensitiveglasscanbereac-
tivatedusinganHFsolution(p/n51319053).Forfurther
detailspleaserefertothemainmanual.
Typically,after10–20sterilizationcyclesthepHelectrode
shouldbereplaced.Pleaserefertotheinstructionmanual
ofthereplacementinteriorbodyInPro5000(i).
7 Disposal
Itistheuser’sresponsibilitytodisposeofthespentspare
partsandifnecessarythesensorinaprofessionalmanner.
Sensorscontainelectroniccomponentsthatrequirespecial
disposalwithoutanyhazardforhumans,andwithnorisk
ofharmtotheenvironment.Pleaseobservetheapplicable
localor national regulationsconcerning the disposal of
"wasteelectricalandelectronicequipment".
8 Warranty
Warrantedfreeofmanufacturingdefectsforatermof12
monthsafterdelivery.
9 Storage condition
Forstoragewerecommendllingthemembranebodywith
freshCO
2
electrolyteandreconnectingittothesensor.The
wetting cap, lled with CO
2
-electrolyte, will prevent dry-
ing-outofthemembrane.
10 EC Declaration of Conformity
Thecompletedeclarationisavailableinthedeclarationof
conformitycerticate.
en
74 InPro 5000 (i) CO
2
Sensor Series
© 06/2019Mettler-ToledoGmbH InPro5000(i)SensorSeries
30 313 393 B PrintedinSwitzerland
Notes
InPro 5000 (i) CO
2
Sensor Series 75
InPro 5000 (i) Sensor Series © 06/2019Mettler-ToledoGmbH
PrintedinSwitzerland 30 313 393 B
Notes
Mettler-ToledoGmbH,ProcessAnalytics
ImHackacker15,CH-8902Urdorf,Switzerland
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges.
©06/2019METTLERTOLEDO.Allrightsreserved.
PrintedinSwitzerland.30313393B
www.mt.com/pro