Mettler Toledo InPro 5500 i Series Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo InPro 5500 i Series Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
14 InPro 5500 i CO
2
Sensors
© 06/2019Mettler-ToledoGmbH InPro5500iCO
2
Sensors
30 313 397 B PrintedinSwitzerland
1 Introduction ................................................................... 15
2 Safetyinstructions .......................................................... 15
3 Productdescription ........................................................ 15
4 Installation .................................................................... 16
4.1 Purgeairsupply ............................................................ 16
4.2 Connectingthecabletothetransmitter ............................. 16
5 Operation ..................................................................... 16
6 Maintenance:Cleaningthesensorhead ........................... 17
7 Disposal ....................................................................... 17
8 Warranty ....................................................................... 17
9 Storageconditions ......................................................... 17
10 ECDeclarationofConformity ........................................... 17
Contents
InPro 5500 i
CO
2
Sensors
Quick Setup Guide
InProandISMareregisteredtrademarksof
Mettler-ToledoGmbHinSwitzerland,theUSA,
theEuropeanUnionandafurthervecountries.
en
Appendix
TechnicalDrawings ....................................................................74
en QuickSetupGuide
InPro 5500 i CO
2
Sensors 15
InPro 5500 i CO
2
Sensors © 06/2019Mettler-ToledoGmbH
PrintedinSwitzerland 30 313 397 B
1 Introduction
ThankyouforbuyingthisInPro5500iSeriessensorfrom
METTLERTOLEDO.TheInPro5500iSeriesofsensorsare
used for the determination of dissolved carbon dioxide
(CO
2
)insolutions.Theyaredesignedformeasuringlowto
mediumlevelsofturbidityandcoloration.
2 Safety instructions
h
Notice:Beforeeverystart-up,thesensormustbechecked
for:
–Damagetotheconnections,fastenings,etc.
–Leakage
–Perfectfunctioning
–Authorization for use in conjunction with other plant
equipmentandresources.
h
Notice:Themanufacturer/supplieracceptsnoresponsibility
foranydamagecausedbyunauthorizedattachmentsorfor
theincorporationofsparepartswhicharenotofMETTLER
TOLEDO provenance. The risk is borne entirely by the
operator.
a
Caution:Beforethesensorisputintooperation,the
operatormusthavealreadyclariedthatuseofthesensor
in conjunction with the other associated equipment and
resourcesisfullyauthorized.
a
Caution:Adefectivesensormayneitherbeinstallednor
putintooperation.Faultycontainmentorinstallationoutof
conformancewithregulationsandinstructionscanleadto
theescapeofmediumortopressuresurges(explosion),
potentiallyharmfulbothtopersonsandtotheenvironment.
a
Caution: Thesensorisnotequippedwithheatprotection.
Duringsteam-sterilizationproceduresinpipes,thesurface
ofthesensorcanreachhightemperaturesandcauseburns.
a
Caution: Somecomponentsinsidethesensorareenergized
with voltages which can cause lethal shocks incase of
contact. The instrument needs to be switched to zero
potentialbeforeanyoperationonthewiringterminal.
a
Warning: Beforedismantlingthesensororcommencingany
maintenanceworkonit,ensurethattheequipmentinwhich
thesensorisinstalledisinasafecondition(pressurize,no
explosionrisk,empty,rinse,vent,etc.).
3 Product description
The InPro 5500iisavailable with three different process
connectionsforCO
2
measurementinbreweryandcarbonated
softdrinkapplications.
–VariventDN40
–TriClamp2"
–28mm;M42
TheInPro5500icanbeusedinapplicationswhereCIPand
SIPcyclesareapplied.Thesensorhasaprotectivefunction
wheretheCO
2
-measurementloopshutsoffduringelevated
temperatures (>55°C) and turns back on when it cools
down(<50°C).ThisfunctionensuresthattheTC-chipwill
notbedamagedduringCIP/SIP.
en
16 InPro 5500 i CO
2
Sensors
© 06/2019Mettler-ToledoGmbH InPro5500iCO
2
Sensors
30 313 397 B PrintedinSwitzerland
4 Installation
4.1 Purge air supply
Thesensorneedsapurgeairsupplyforcorrectfunctioning.
Thepurgegasalsokeepsthemeasuringgaschamberdry
andpreventshumiditybuildupthatcouldharmthesensor.
Requirements of the purge gas supply:
Anyairsupplythatfullllstherequirementsbelowissuitable
tobeusedwiththissensor:
–Deliveredpressure:min.1–2barg
–Particulatefree(40micrometer)
–Condensatefree(dewpointbelow–10°C)
Connecting the purge air to the sensor
Toconnectthepurgegassupplytothesensor
use a exible nylon, PVC or a stainles steel hose.
Thesensorissuppliedwitha6mmSwagelockconnection
forthispurpose.
4.2 Connecting the cable to the transmitter
CO
2
Transmitter VPorAK9cable
RS 485 cable Transmitter
M400 M800
Channel2 Channel4
Color Function TB4 TB2 TB4
brown 24DC+ 1 9 9
black 24DC– 2 10 10
gray shield 6 12 12
yellow shield 6 15 15
blue RS485– 7 13 13
white RS485+ 8 14 14
h
Notice: Cable assignement can be found in the
METTLERTOLEDOcableinstructionmanual.
h
Notice: For connecting the cable to the terminals of the
transmitter,pleasealsorefertotheinstructionsgiveninthe
METTLERTOLEDOtransmittermanual.
5 Operation
In case of any failure or malfunction, the equipment in
whichthesensorisinstalledneedstobeinasafecondition
(depressurized,empty,rinse,vent,etc.).
en
InPro 5500 i CO
2
Sensors 17
InPro 5500 i CO
2
Sensors © 06/2019Mettler-ToledoGmbH
PrintedinSwitzerland 30 313 397 B
6 Maintenance: Cleaning the sensor head
The measurement principle provides a large degree of
compensationforcontaminationofthesensorhead.However,
contamination can become so severe, depending on the
operatingconditionsandmedia,thatthecompensationis
nolongeradequateandthesensorheadmustbecleaned.
1.Place the system or equipment in a safe state
(depressurized,empty,rinsed,ventedorventilated,etc.)
2.Disconnectelectricalpower
3.Dismountthesensor
4.Cleanthesensorheadusingasoft,lint-free clothand
amildcleaningagent
5.Reinstallthesensor
7 Disposal
Itistheuser’sresponsibilitytodisposeofthesensorina
professionalmanner.Sensorscontainelectroniccomponents
thatrequirespecialdisposalwithoutanyhazardforhumans,
andwithnoriskofharmtotheenvironment.Pleaseobserve
theapplicablelocalornationalregulationsconcerningthe
disposalof"wasteelectricalandelectronicequipment".
8 Warranty
Onmanufacturingdefects,12monthsafterdelivery.
9 Storage conditions
Please inspect the instrument immediately after receiving
foranytransportdamage.Thesensorshouldbestoredin
adry,cleanandprotectedareauntilthetimeofinstallation.
Ifthesensorhasbeenremovedfromtheprocess,itshould
bethoroughlycleanedanddried.Itshouldbestoredinadry,
cleanandprotectedareauntiltimeofinstallation.
10 EC Declaration of Conformity
Thecompletedeclarationisavailableinthedeclarationof
conformitycerticate.
en
74 InPro 5500 i CO
2
Sensors
© 06/2019Mettler-ToledoGmbH InPro5500iCO
2
Sensors
30 313 397 B PrintedinSwitzerland
107 mm
4.21"
59 mm
2.32"
15.3 mm
0.60"
13.5 mm/
0.53"
6 mm/0.24"
33.5 mm
1.32"
28 mm
1.10"
1
2
3
4
5
6
7
8
Technical Drawing
Appendix
1 Konektor5pinů
2 Hlavicesenzoru
3 Výstupprofukovacíhoplynu
4 Vstupprofukovacíhoplynu
5 Procesnípřipojení
6 Dřík28mm
7 Difuzor
8 MembraCap
cs
1 5-benetstik
2 Sensorhoved
3 Purgegas-udløb
4 Purgegas-indløb
5 Procesforbindelse
6 28mmskaft
7 Diffusor
8 MembraCap
da
1 Gegenstecker5-pol.
2 Sensorkopf
3 AuslassSpülgas
4 EinlassSpülgas
5 Prozessanschluss
6 28-mm-Schaft
7 Diffusor
8 MembraCap
de
1 5-pinconnector
2 Sensorhead
3 Purgegasoutlet
4 Purgegasinlet
5 Processconnection
6 28mmshaft
7 Diffusor
8 MembraCap
en
1 Conectorde5pines
2 Cabezaldelsensor
3 Salidadegasdepurga
4 Entradadegasdepurga
5 Conexióndeproceso
6 Ejede28mm
7 Difusor
8 MembraCap
es
1 Connecteur5broches
2 Têtedesonde
3 Tuyaudepurgedugaz
4 Tuyaud’admissiondugaz
5 Raccordprocédé
6 Corpsdel’électrode28mm
7 Diffuseur
8 MembraCap
fr
1 5-napainenliitin
2 Anturinpää
3 Huuhtelukaasunlähtö
4 Huuhtelukaasuntulo
5 Prosessiliitäntä
6 28mmakseli
7 Diffuusori
8 MembraCap
Addresses|Impressum
Mettler-ToledoGmbH,ProcessAnalytics
ImHackacker15,CH-8902Urdorf,Switzerland
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges.
©06/2019METTLERTOLEDO.Allrightsreserved.
PrintedinSwitzerland.30313397B
www.mt.com/pro
ForaddressesofMETTLERTOLEDO
MarketOrganizationspleasegoto:
www.mt.com/pro-MOs
/