Roche Cedex HiRes Reference guide

Type
Reference guide

Roche Cedex HiRes is a high-resolution analyzer that can measure the size and concentration of particles in a sample. It is commonly used in quality control and manufacturing processes to ensure product quality and consistency. With its multi-sampler, the device can analyze up to 20 samples automatically, saving time and increasing efficiency. Additionally, the Cedex HiRes software allows users to modify data entry, change samples, and view results during the analysis process.

Roche Cedex HiRes is a high-resolution analyzer that can measure the size and concentration of particles in a sample. It is commonly used in quality control and manufacturing processes to ensure product quality and consistency. With its multi-sampler, the device can analyze up to 20 samples automatically, saving time and increasing efficiency. Additionally, the Cedex HiRes software allows users to modify data entry, change samples, and view results during the analysis process.

1 / 2
For use in quality control/manufacturing processes only.
Working with the Multi Sampler — (April 2014)
Moving parts of the Multi Sampler may cause hand injuries. The Multi Sampler needle
may cause injuries during operation. Never use the Cedex HiRes Analyzer without the
protective cover in place for the Multi Sampler.
The Multi Sampler DefaultCup
The position of the Cedex Sample Cup used for the Prime, LM Shutdown, calibration and
cleaning routines is called the DefaultCup. An empty Cedex Sample Cup must be manually
placed in the position set for the DefaultCup before running LM Shutdown, focus adjustment,
Prime and all cleaning routines. This Cedex Sample Cup does not differ from the usual sample
cups. The standard (default) position for the DefaultCup is position 1 of the Multi Sample
Tray.
Using the Multi Sample Tray
To insert the tray, gently push the tray into the notch until the automatic tray rope starts
pulling it in. To eject the tray, gently push on the tray. The tray will be automatically ejected.
The tray can only be ejected when the system is idle. It cannot be ejected if the system is
moving or the needle is down in a cup.
Carrying out a Measurement with the Multi Sampler
1
Click on the MULTI RUN button on the Cedex Control Center to open the Multi Run
dialog box.
2
Fill 300 µL +/- 10% of the samples to be measured into individual Cedex Sample Cups
(maximum 20) and place them in the Multi Sample Tray according to the desired sequence.
Insert the tray into the Multi Sampler.
3
Click anywhere in the row of the Multi Run preparation area that
corresponds to the
location of a cup containing a sample. The row will be converted to edit mode, and the
sample and process parameters for the measurement can be defined.
4
Enter the sample data and parameter information as described in the section “Carry out a
single measurement ” in the Quick Guide “Performing a Measurement ”.
5
Repeat steps 3 — 4 for the remaining samples to be measured.
6
Activate the status of each sample row by double clicking in the Status field of the row to
be activated. The row appears pale blue, and further modifications of sample information is
impossible. Return to the Edit status by double clicking on the Status area of the sample
row to allow for further changes, if desired.
7
Alternatively, activate or deactivate all samples simultaneously by double clicking on the word
Status in the column heading at the top of the Multi Run preparation area. Confirm the
safety query with YES to accept the modification.
8
After some or all rows have been activated, click on the START button on the right-hand side
of the Multi Run dialog box to start the analysis of the activated measurements.
Cedex HiRes Software 2.3.2 Quick Guide
2 / 2
07267088001
3
042014
For use in quality control/manufacturing processes only.
CEDEX is a trademark of Roche.
All other product names and trademarks are the property of their respective owners.
Working with the Multi Sampler — (April 2014)
Modifying Data Entry, Changing Samples
Information about the operation and process parameters for samples that have not yet been
analyzed can be modified during the run as follows:
1
Double-click on the Status field of the measurement to be modified. The status field will
indicate that the sample is in edit mode.
2
Make desired changes to the measurement.
3
Double-click on the Status field to re-activate the measurement
(status: active or scheduled).
Adding Measurements to the List while Measurements are Running
Sample rows in which data has not yet been entered can be filled out at any time while the
Multi Run is running. In addition, sample rows containing finished measurements (status:
done) can be cleared and prepared for a new measurement by double-clicking in the Status
field of the row with the finished measurement. Sample data for a measurement cannot be
modified after that measurement has begun.
Viewing results via the Multi Run Result List
At the bottom of the Multi Run dialog box, the Analyzed Measurements field displays the
list of results for measurements previously carried out. Measurements are listed in order of
process time, from the newest measurements at the top of the list to the oldest measurements at
the bottom. All measurements carried out using the Multi Run dialog box will be displayed in
this field unless the Clear Resultlist button has been pressed, which will clear all measurement
results from the list.
In addition, measurements displayed in this field can be restricted to only the measurements
carried out by the user logged into the system (mine), measurements that have not yet been
printed out (unprinted), and/or measurements carried out on that particular day (today).
Other results, such as Average Diameter, can be added for display in the result list via the User
Preferences (see “Adjusting Result Display in the Measurement List and Multi Run dialog boxes
in the Cedex HiRes Operator’s Guide). Double-click on a measurement to view all results.
If the result list has been cleared, previously carried out measurements can still be viewed,
printed out and exported via the Measurement List dialog box (see Using the Measurement
List in the Cedex HiRes Operator’s Guide).
Published by:
Roche Diagnostics GmbH
Sandhofer Strasse 16
68305 Mannheim
Germany
Cedex HiRes Software 2.3.2 Quick Guide
  • Page 1 1
  • Page 2 2

Roche Cedex HiRes Reference guide

Type
Reference guide

Roche Cedex HiRes is a high-resolution analyzer that can measure the size and concentration of particles in a sample. It is commonly used in quality control and manufacturing processes to ensure product quality and consistency. With its multi-sampler, the device can analyze up to 20 samples automatically, saving time and increasing efficiency. Additionally, the Cedex HiRes software allows users to modify data entry, change samples, and view results during the analysis process.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI