003-1120-00 Rev. PA2 4/2/2019
Console Mtd. Procenter, Asepsis 21 or
International (Whip) Delivery Unit Installation
Step 1: Remove console side cover.
Step 2: Install console bearing.
Secure with three screws.
Note: Position console bearing so that
set screws are accessible thru holes.
Step 3: Install bearing washer.
Insert mounting post into console bearing.
Step 5:
To adjust rotational “drag”:
Tighten / loosen console
bearing set screws.
Parts Bag Contents
Stop Pins .................... 2
Bearing Washer .......... 1
Screws ....................... 3
Special Tools
Level
Step 4: Rotate arm to verify proper operation.
For LH mtd. console only:
Move rotation stop screw to opposite hole.
Rotate Stop Screw
Note: Procenter delivery unit shown
Style G
2
© Midmark Corporation 2003 SF-1643
2
4
1
3
KA967402i
Equipment Alert
Stop pins must be installed correctly to ensure proper operation.
Step 6: Position delivery unit arms as shown.
(Delivery unit above foot end of chair)
Right Mounted Console
Left Mounted Console
Step 7: Install two stop pins.
Note: Rotation stop screw must be
between two stop pins.
Tip: Move pins under trim ring,
then down into holes.
3
© Midmark Corporation 2003 SF-1643
KA954701i
Step 8: Connect tubing as shown.
(match colors)
Do not remove plug!
4
© Midmark Corporation 2003 SF-1643
Accessories
Step 10: Connect accessory
wire harness to
J4 or J5 on PC Board.
Connect ground wire to
console.
Chair Position Touch Pad
Step 9: Connect touch pad
wire harness to
J2 or J3 on PC Board.
Connect ground from
umbilical.
5
© Midmark Corporation 2003 SF-1643
2
4
1
3
KA967501i
Step 13: Slide cover forward.
Replace end cap.
Step 12: Remove wire tie from lock pawls.
Step 11: Remove flex arm end cap.
Slide cover back.
Step 14: Remove bottom cover.
Step 16: Replace bottom cover.
Step 15: Loosen three button head screws.
Adjust three leveling screws to level unit.
Tighten three button head screws.
Note: Steps 11 - 13 apply to all delivery units. Steps 14 - 16 are Procenter delivery unit only.
6
© Midmark Corporation 2003 SF-1643
Midmark Corporation
60 Vista Drive
P.O. Box 286
Versailles, OH 45380-0286
937-526-3662
Fax 937-526-5542
www.midmark.com
For Units Having Deluxe Rheostat with Chip Air Foot
Controls Only:
Make Chip Air Connection in Delivery Head.
A) Cut green 1/8" Coolant Air Supply tubing.
B) Locate clear 1/8” tubing coming into the unit head.
C) Use barbed tee and sleeves to tee the 1/8" clear tubing
into the 1/8" green coolant air tubing.
Deluxe Rheostat with
Chip Air Foot Control
Chip Air
1/8” Green
1/8” Clear