Midmark Procenter Systems offer a comprehensive solution for dental professionals, providing versatility and efficiency in various clinical settings. These systems integrate seamlessly with Midmark chairs, featuring a range of capabilities to enhance patient care and streamline workflows. From air, water, and vacuum controls to assistant's unit integration, the Procenter Systems optimize dental procedures for both the clinician and the patient.

Midmark Procenter Systems offer a comprehensive solution for dental professionals, providing versatility and efficiency in various clinical settings. These systems integrate seamlessly with Midmark chairs, featuring a range of capabilities to enhance patient care and streamline workflows. From air, water, and vacuum controls to assistant's unit integration, the Procenter Systems optimize dental procedures for both the clinician and the patient.

1© Midmark Corporation 2003 SF-1664
DA134500i
Step 2: Remove safety bail cover.
Tip: Push cover IN & DOWN,
then pull cover OUT & UP.
Step 3: Remove small cover.
Step 4: Remove end cover.
003-1152-00 Rev. AA1 (9/17)
Parts Bag Contents
T-Fitting w/tubing......... 1
Screw ......................... 2
Lockwasher ................ 2
Cable Tie .................... 2
Cap Sleeve (red)......... 1
Toggle Cap (black) ...... 1
Water Q.D. Coupling ... 1
Special Tools
Level
LR Assistant’s Unit Installation
Step 1: Raise chair all the way up;
then unplug the chair.
Next
Style G
2© Midmark Corporation 2003 SF-1664
DA134600i
Step 5:
Secure extension plate to chair
with two screws & lockwashers.
Step 6:
Remove cover.
Adjust three leveling screws until arm is level.
Install cover.
Equipment Aler
t
Have an assistant secure the instrument arm at all pivot points to prevent damage to equipment.
Leveling Screws
Next
Back
3© Midmark Corporation 2003 SF-1664
Step 8: Pull out yellow tubing & cut.
Insert T-fitting w/yellow tubing
.
Step 9: Connect 1/8” tubing.
Note: 1/4” tubi
ng
is not used.
Step 10: Connect 5/8” grey tubing from Asst’s Unit
to 5/8” ribbed tubing from umbilical.
In J-Box (not shown):
Connect 5/8” ribbed
tubing from umbilical to central vacuum unit.
Note: 5/8” smooth tubing is for cuspidor drain.
Step 7: Remove plugged section from 1/8” blue tubing.
Next
Back
Color Page
4© Midmark Corporation 2003 SF-1664
Step 11: Connect wiring and tubing. Use two or more ties to
bundle wiring and vacuum tubing together
and attached to the chair frame.
Chair Umbilical
Assistant’s Unit
Umbilical
Controls
Power
Vacuum
Air
Water
Next
Back
Color Page
5© Midmark Corporation 2003 SF-1664
Midmark Corporation
60 Vista Drive
P.O. Box 286
Versailles, OH 45380-0286
937-526-3662
Fax 937-526-5542
www.midmark.com
DA135200i
Step 14: Slide extension plate cover over tabs.
Cable Ties
Step 12: Pull tubing/wiring toward extension plate to minimize slack.
Secure with two cable ties
.
Step 13: Install safety bail cover.
Tip: Position bottom “hooks” on lower mounting pins.
Align top “hooks” with upper mounting pins.
Push cover Down, In, then pull Upward
until upper “hooks” engage.
Tabs
“Hooks”
Back
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5

Midmark Procenter Systems Installation guide

Type
Installation guide

Midmark Procenter Systems offer a comprehensive solution for dental professionals, providing versatility and efficiency in various clinical settings. These systems integrate seamlessly with Midmark chairs, featuring a range of capabilities to enhance patient care and streamline workflows. From air, water, and vacuum controls to assistant's unit integration, the Procenter Systems optimize dental procedures for both the clinician and the patient.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI