English - 1
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Console
Side Cover
Elevance™ Assistant’s Unit: Console Mounted
Installation
Language of origin: English
Step 3: On a LEFT side mounted console,
position mounting arm straight out
from the rear of the console.
Console
Cover
Step 4: Rotate ADAPTER clockwise before
loosely installing mounting screws.
Do not tighten mounting screws.
Stop Pin
Right side
pin hole
Left side
pin hole
Magnet
Bottom
Screw
003-2287-99 Rev. D (1/29/13)
Mounting
Screw
Leveling
Screw
Step 2: Install three leveling screws.
WARNING
Always disconnect chair from the power
source before removing any covers.
Failure to do so may result in personal injury.
Step 1: Remove magnetic side cover.
Loosen bottom screw and remove
inside console cover.
Tip: Console covers fit up under a lip at the top,
pull out from bottom, then down to remove.
Note:
The stop pin is factory set for a LEFT side mounted console (patient’s left).
Move the stop pin ONLY to install arm on a RIGHT side mounted console.
Adapter