Midmark Elevance® Standard Delivery System Installation guide

Type
Installation guide
English - 1
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Console
Side Cover
Elevance™ Assistant’s Unit: Console Mounted
Installation
Language of origin: English
Step 3: On a LEFT side mounted console,
position mounting arm straight out
from the rear of the console.
Console
Cover
Step 4: Rotate ADAPTER clockwise before
loosely installing mounting screws.
Do not tighten mounting screws.
Stop Pin
Right side
pin hole
Left side
pin hole
Magnet
Bottom
Screw
003-2287-99 Rev. D (1/29/13)
Mounting
Screw
Leveling
Screw
Step 2: Install three leveling screws.
WARNING
Always disconnect chair from the power
source before removing any covers.
Failure to do so may result in personal injury.
Step 1: Remove magnetic side cover.
Loosen bottom screw and remove
inside console cover.
Tip: Console covers fit up under a lip at the top,
pull out from bottom, then down to remove.
Note:
The stop pin is factory set for a LEFT side mounted console (patient’s left).
Move the stop pin ONLY to install arm on a RIGHT side mounted console.
Adapter
Style G
midmark.com
English - 2
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Assistant’s
Umbilical
Leveling
Screw
Step 5: Route the assistant’s umbilical and
vacuum line into the console.
Note: Take care not to kink the umbilical or vacuum line.
Step 6: Adjust three leveling screws to level
the mount arm assembly.
Tighten three mounting screws.
Mounting
Screw
Vacuum
Line
Style G
midmark.com
English - 3
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Step 8: Connect like colored tubing in the console. Connect vacuum lines.
Note: No cover is required over tubing connections.
Keypad
Communication
Coupler
Optional
Heated
Syringe
Power
Console
Cover
Curing Light (option)
(red and black wires)
Ground
Wire
Quick Connects
Air
Water
Air
Water
Flow Control Knobs
Solids
Collector
HVE
Saliva
Ejector
Vacuum
Optional
Curing Light
Syringe
Tubing connections
Water (blue) /
Air (yellow)
Keypad
Communication
Line
Vacuum
Assistant’s Unit Tubing
Caution
Connect ground wire to ensure system ground continuity. Failure to do so may result in electric shock.
Step 7: Use velcro to mount keypad communication coupler to console wall and connect lines.
Connect any optional electrical components being used. Screw ground wire connections into
console. Place console cover over all electrical connections.
Style G
midmark.com
Color Page
English - 4
© 2012 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Installation Checklist
Test Verify the following:
1 Mounting arm and assistant’s unit are level.
2
Mount arm and assistant’s unit rotate and move
easily.
3
Assistant’s instruments work - no wires or tubing are
pinched by covers or mountings.
4
Wipe console and assistant’s unit with a soft dry cloth
to remove dirt or smudges.
5 (Optional) Curing light works.
Step 9: Install covers...
Install console cover over all electrical connections.
Install console side cover (align magnets).
Note: Be careful not to pinch or kink tubing or wires when
installing covers.
Step 10: Connect chair to power supply and
test assistant’s unit operation.
Style G
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20

Midmark Elevance® Standard Delivery System Installation guide

Type
Installation guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI