InPro 86X0 i 17
© 03 / 18 Mettler-Toledo GmbH InPro 86X0 i
30 421 373 A
1 Introduction ................................................................... 18
2 Safetyinstructions .......................................................... 18
3 Product description ........................................................ 18
4 Sensor Installation .......................................................... 19
5 Operation ..................................................................... 20
6 Maintenance:Cleaningthesensorhead ............................ 20
7 Disposal ....................................................................... 20
8 Warranty ....................................................................... 21
9 Storage conditions ......................................................... 21
10 ECDeclarationofConformity ........................................... 21
Contents
InPro 86X0 i Sensors
InstructionManual
InProisaregisteredtrademarkoftheMETTLERTOLEDOGroupin
Switzerland,USAandafurtherelevencountries.
ISMisaregisteredtrademarkoftheMETTLERTOLEDOGroupin
Switzerland,USA,theEuropeanUnionandafurthervecountries.
Appendix
TechnicalDrawings .................................................................. 92
en Instruction manual
en
18 InPro 86X0 i
InPro 86X0 i © 03 / 18 Mettler-Toledo GmbH
30 421 373 A
1 Introduction
Thank you for buying this InPro
®
86X0 i sensor from
METTLERTOLEDO.TheInPro86X0isensorisusedforthe
determination of undissolved particles or substances in
solutions.Theyaredesignedformeasuringlowtomedium
levelsofturbidityandcoloration.
2 Safetyinstructions
h
Notice:Beforeeverystart-up,thesensormustbechecked
for:
–Damagetotheconnections,fastenings,etc.
– Leakage
– Perfect functioning
–Authorizationforuseinconjunctionwithotherplantequip-
ment and resources.
h
Notice:Themanufacturer/supplieracceptsnoresponsibil-
ityforanydamagecausedbyunauthorizedattachments
or for the incorporation of spare parts which are not of
METTLERTOLEDOprovenance.Theriskisborneentirelyby
theoperator.
a
Caution:Beforethesensorisputintooperation,theoperator
musthavealreadyclariedthatuseofthesensorinconjunc-
tionwiththeotherassociatedequipmentandresourcesis
fullyauthorized.
a
Caution:A defectivesensormayneitherbeinstallednor
putintooperation.Faultycontainmentorinstallationoutof
conformancewithregulationsandinstructionscanleadto
theescapeofmediumortopressuresurges(explosion),
potentiallyharmfulbothtopersonsandtotheenvironment.
a
Caution:Thesensorisnotequippedwithheatprotection.
Duringsteam-sterilizationproceduresinpipes,thesurface
ofthesensorcanreachhightemperaturesandcauseburns.
a
Caution:Somecomponentsinsidethesensorareenergized
withvoltageswhichcancauselethalshocksincaseofcon-
tact.Theinstrumentneedstobeswitchedtozeropotential
beforeanyoperationonthewiringterminal.
a
Warning: Beforedismantlingthe sensororcommencing
anymaintenanceworkonit,ensurethattheequipmentin
whichthesensorisinstalledisinasafecondition(pressur-
ize,noexplosionrisk,empty,rinse,vent,etc.).
3 Product description
Thelabelonthehousingcontainsthefollowinginformation:
InPro 86X0 i
x: 1=single-angleversion(25°)
3=dual-anglepluscolourversion(25°and90°plus
color)
Serialnumber
Theserialnumbercanbefoundonthelabelonthesensor.
Sensorcertications
The label locatedon the sensor provides the information
aboutthesensorcertications.
en
InPro 86X0 i 19
© 03 / 18 Mettler-Toledo GmbH InPro 86X0 i
30 421 373 A
Examplelabel
P/N:
S/N:
30421361
1752001
METTLER TOLEDO
www.mt.com/pro
Made in China by
METTLER-TOLEDO
0036
1.4404/316L
PS:16 bar
TS:120 °C
梅特勒
-
托利多仪器
(
上海
)
有限公司
地址:上海市桂平路
589
邮编:
200233
Mettler-Toledo GmbH
Im Hackacker 15
8902 Urdorf,Switzerland
MODELTurbidity Sensor InPro8630i
型号:浊度传感器 InPro8630i
CableconnectiontoM800Process1-Channeltransmitter
TheRS-485datacable(5-pin)isavailablefromMETTLERTOLEDO
invarious lengths.Connectthe datacabletotheM800 Process
1-Channeltransmitterasdescribedbelow.
M800Process1-Channel TB3
Terminal Function 5-pincablewirecolor
1 1-Wire –
2 GND5V Yellow
3 RS485B Blue
4 RS485A White
5 GND5V Grey
6 5V –
7 24V Brown
8 GND24V Black
9 – 16 Not used
Inordertoensureoptimummeasurementresultsthefollo-
wingpointsmustbeconsidered:
–Runtheconnectioncablethroughthecablegland.
–Connectthewirestothesensor.
–Hand-tightenthecablegland.
4 Sensor Installation
ThesensorisequippedwithaTuchenhagen-Variventprocess
adapter,whichisusedtoattachthesensortotheprocess
pipeindimensionsfromDN40toDN150.Priortothein-
stallation,thesealingsurfaces(sensorandpipeadapter)
needtobecleanedandcheckedfordamages.Further,it
needstobeveriedthatnoobstaclesarepresentwithinthe
installationcavityofthesensor.
Theinstallationofthesensorshouldbeinavertical stand-
pipe.Ifthesensorwillbeinstalledinahorizontalpipe,it
mustbeinahorizontalpositionandtheprocesspipehas
tobelledcompletelywithliquidduringthemeasurement.
Installation on a vertical line (preferred)
en
20 InPro 86X0 i
InPro 86X0 i © 03 / 18 Mettler-Toledo GmbH
30 421 373 A
Installationonahorizontalline
Thereisamarkonthesensorheadwhichshowstheow
directionofthemedia.Thesensorneedstobeinstalledin
thisdirection.
Markfortheinstallationintoapipe
Thesensormustbettedtothepipeatleast2meters(6.6ft)
awayfromstraylightsources(suchassightglasses).
5 Operation
Incaseofanyfailureormalfunction,theequipmentinwhich
thesensorisinstalled,needstobeinasafecondition(de-
pressurized,empty,rinse,vent,etc.).
6 Maintenance:Cleaningthesensorhead
Themeasurementprincipleprovidesalargedegreeofcom-
pensationforcontaminationofthesensorhead.However,
thecontaminationcanbecomesosevere,dependingonthe
operatingconditionsandmedia,thatthecompensationis
nolongeradequateandthesensorheadmustbecleaned.
1.Placethesystemorequipmentinasafestate(depressu-
rized,empty,rinsed,ventedorventilated,etc.)
2.Disconnectelectricalpower
3.Dismountthesensorfromprocess
4.Cleanthesensorheadusing a soft,lint-freecloth and
a mild cleaning agent
5.Installthesensor
7 Disposal
Itistheuser’sresponsibilitytodisposeofthe
sensorinaprofessionalmanner.Thissensor
contains electronic components that require
special disposal without any hazard for hu-
mans,and with noriskof harm tothe envi-
ronment.Pleaseobservetheapplicablelocalornational
regulationsconcerningthedisposalof“wasteelectricaland
electronic equipment.
en
InPro 86X0 i 21
© 03 / 18 Mettler-Toledo GmbH InPro 86X0 i
30 421 373 A
8 Warranty
Onmanufacturingdefects,12monthsafterdelivery.
9 Storageconditions
Inspecttheinstrument immediatelyafter receiving forany
transportdamage.The sensor shouldbestoredina dry,
cleanandprotectedareauntilthetimeofinstallation.Ifthe
sensorhasbeen removed fromtheprocess,itshould be
thoroughlycleanedanddried.Itshouldbestoredinadry,
clean and protected area until time of installation.
10 ECDeclarationofConformity
Thecompletedeclarationisavailableinthedeclarationof
conformitycerticate.
en
92 InPro 86X0 i
InPro 86X0 i © 03 / 18 Mettler-Toledo GmbH
30 421 373 A
TechnicalDrawings
Appendix
Ø 60 mm (2.36")
163.5 mm
(6.44")
Ø 84 mm (3.31")
41.3 mm (1.63")
29.3 mm
(1.15")
29 mm
(1.14")
95.5 mm
(3.76")
Connection
Varivent
®
N 50/40
87.7 mm
(3.45")
Ø 92 mm
(3.62")
Flow direction (media)
59 mm
(2.32")
OPL 34
InPro 86X0 i 93
© 03 / 18 Mettler-Toledo GmbH InPro 86X0 i
30 421 373 A
Notes
Brazil Mettler-ToledoInd.eCom.Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mail mtbr@mt.com
France Mettler-ToledoAnalyseIndustrielleS.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mail mtpro-f@mt.com
Germany Mettler-ToledoGmbH
Prozeßanalytik,Ockerweg3,DE-35396Gien
Phone+49641507-444
e-mail prozess@mt.com
Switzerland Mettler-Toledo(Schweiz)GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444747
e-mailProSupport.ch@mt.com
UnitedStates METTLERTOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mail mtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
03/2018©METTLERTOLEDO.Allrightsreserved
30 421 373 A
www.mt.com/pro
0036
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96

Mettler Toledo InPro 86X0i Turbidity Sensor The InPro 86X0 i sensor Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI