Mettler Toledo InPro 8100/S, InPro 8000 Series, InPro 8050, InPro 8200/S(H) Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo InPro 8100/S Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
InPro 8000 Series
Turbidity Sensors
Quick Setup Guide
cs Průvodcerychlýmnastavením 2
da Lynvejledning 5
de Kurzanleitung 8
en QuickSetupGuide 11
es Guíadeconguraciónrápida 14
 Pika-asetusopas 17
fr Guidedeparamétragerapide 20
hu Gyorsbeállításiútmutató 23
it Guidaallacongurazionerapida 26
ja
29
ko 32
nl Beknoptehandleiding 35
pl Podręcznikszybkiejkonguracji 38
pt GuiadeConguraçãoRápida 41
ru
Руководство по быстрой настройке 44
sv Snabbguide 47
th
   50
zh 53
InPro8000Series
30313267B
Contents
InPro8000Series 11
©07/2019Mettler-ToledoGmbH InPro8000Series
PrintedinSwitzerland 30313267B
Quick Setup Guide
InPro 8000 Series
1 Introduction
ThankyouforbuyingthisInPro
®
8000Seriesturbiditysen-
sorfromMETTLERTOLEDO.InPro8000Seriesareoptical
sensors which measure the concentration of suspended
particlesinaliquidmediumonthebasisoftheprincipleof
backscatteredlight.ThesensorisoperatedwiththeM800
transmitterfromMETTLERTOLEDO.
2 Safety instructions
h
Notice:Before every start-up, the sensor must be
checkedfor:
– Damagetotheconnections,fastenings,etc.
– Leakage
– Perfectfunctioning
– Authorizationforuseinconjunctionwithotherplant
equipmentandresources.
h
Notice: The manufacturer/supplier accepts no re-
sponsibilityforanydamagecausedbyunauthorized
attachments orfor the incorporationof spareparts
whicharenotofMETTLERTOLEDOprovenance.The
riskisborneentirelybytheoperator.
a
Caution:Beforethesensorisputintooperation,the
operatormusthavealreadyclariedthatuseofthe
sensorinconjunctionwiththeotherassociatedequip-
mentandresourcesisfullyauthorized.
a
Caution:Adefectivesensormayneitherbeinstalled
norputintooperation.Faultycontainmentorinstalla-
tionoutofconformancewithregulationsandinstruc-
tionscanleadtotheescapeofmediumortopressure
surges (explosion),potentially harmfulboth to per-
sonsandtotheenvironment.
a
Caution:BeforeinstallationinEx-zones,pleaseread
andobservetheguidelines(frominstructionmanuals)
andadheretosiderequirements.
a
Caution:Theberopticcableshaveacoreofhard
cladsilica(HCS)bers.Whenlayingberopticca-
bles, the bend radiusmustalways exceed at least
10cm(4").Fiberopticcablesmayneverbeshortened
independently.FittingoftheSMAplugrequiresspecial
tooling.Thetransparentplasticcaponthesensoris
toberemovedbeforeuse.Anyformationofdeposits
ofsolidparticlesonthesurfaceistobeavoided.
a
Warning:Beforedismantlingthesensororcommenc-
inganymaintenanceworkonit,ensurethattheequip-
mentinwhichthesensorisinstalledisinasafecon-
dition(depressurize,noexplosionrisk,empty,rinse,
vent,etc.).Themetalbodyoftheturbiditysensorsas
wellasthesafetyweld-insocketsMETTLERTOLEDO
housingInFit
®
type76*-***andInTrac
®
7**-***are,
ifnecessary,tobeincludedintotherecurringpressure
testoftheunit.
3 Product description
InPro8100/8200seriesareATEXcertiedtobeinstalledin
Ex-classiedarea.Pleaseobservetheprecautionsandspe-
cialconditionsforsafeusebeforeinstallingthesensorinan
Ex-classiedarea.
en
en QuickSetupGuide
12 InPro8000Series
InPro8000Series ©07/2019Mettler-ToledoGmbH
30313267B PrintedinSwitzerland
Ex-classication:
II1/2GExopis/opprIIA95
°CGa/Gb
II1/2GExopis/opprIIB38
°CGa/Gb
Examinationcerticatenumber:SEV12ATEX0169X
Notiedbody:ElectrosuisseSEV,(No.:1258)
Technical Data
Technical
data
InPro
8200/S(H) InPro
8100/S InPro
8050
Meas.
principle
dualber singleber singleber
Shaft
material
DIN1.4435or
1.4404(316L)
orHastelloy
C276
(availableonly
withEpoxy
sealedwindow)
1.4435or
1.4404(316L)
PSU
(Polysulfone)
Shaft
diameter
12mm 12mm 12mm
Shaft
length
120,205,
297or407mm
120,205,
297or407mm
120mm
Window
material
Sapphire withoutwindow Sapphirespigot
O-ring
sealing
Viton
®
-FDA Viton-FDA Viton-FDA
Process
adaption
Pg13.5 Pg13.5 Pg13.5
Plug
2SMA 2SMA 2SMA
Meas.
range
5FTU…4000FTU
0…30g/l
10FTU…4000FTU
0…250g/l
10FTU…4000FTU
0…250g/l
Pressure
range
0…6bar
(0…87psi)with
Kalrez
®
-FDA
sealedwindow
0...16bar
(0...232psi)
withEpoxybon
-
dedwindow
0…6bar
(0…87psi)
0…2bar
(0...29psi)
Tempera-
turerange
–10…130°C
(14…266°F)
withKalrez-FDA
sealedwindow
–30…130°C
(–22…266°F)
withEpoxybon
-
dedwindow
–30…130°C
(–22…266°F)
0…60°C
(32…140°F)
Steam-
steriliz-
able
130°C(only
withKalrez-FDA
sealedwindow)
130°C non-sterilizable
Subjecttotechnicalchanges.
4 Sensor installation
Theturbiditysensorsaretobeinstalledatadistanceofat
least10cm(4")fromthewallsofthevessel,agitatororfrom
thepipewallsinordertopreventdisturbancesthroughre-
ections.Ifitisnotfeasibleinpipesofsmalldiameterthen
thesensorcanbeinstalledsloped,atanangleof1545°
or,alternatively,inakneebend.Thepipemustalwaysbe
completelyfullofliquid.
5 Operation
Iftheturbiditymeasuringsystemiscalibratedoff-line,itis
recommendedtousetheMETTLERTOLEDOcalibrationtool,
CaliCap
®
,(orderingno.52800210)forttingonthesensor
shaft.IfaCaliCapisnotused,aminimumdistanceof10cm
(4")tothevesselwallshastobeobservedaswell.During
calibration,thesampleshouldbestirredinordertoprevent
thesedimentationofsolidparticles.Incaseofanyfailureor
en
InPro8000Series 13
©07/2019Mettler-ToledoGmbH InPro8000Series
PrintedinSwitzerland 30313267B
malfunction,theequipmentinwhichthesensorisinstalled,
needstobeinasafecondition(depressurized,empty,rinse,
vent,etc.).
6 Maintenance: Cleaning the sensor
Dirtyorcontaminatedsensors candeliverincorrectmea-
surementvalues.Iffoulingispresumed,thesensoristobe
removedandcleaned.Suitablecleaningsolutionsaremild
detergentsorstronglydilutedacids(<0.5%byweight)such
ashydrochloricacid.
7 Disposal
Disposalistobecarriedoutbytheuserinlinewithvalid
localregulations.Theusermusteitherdeliverthesensorto
an authorized private or public collectionundertaking, or
disposeofithimselfinaccordancewiththeregulationsfor
suchitems/materials.
8 Warranty
Onmanufacturingdefects,12monthsafterdelivery.
9 Storage condition
Inspecttheinstrumentimmediatelyafterreceivingforany
transport damage.Thesensorshouldbestored in a dry,
cleanandprotectedareauntilthetimeofinstallation.Ifthe
sensorhasbeenremovedfromtheprocess,itshouldbe
thoroughlycleanedanddried.Itshouldbestoredinadry,
cleanandprotectedareauntiltimeofinstallation.
10 EC Declaration of Conformity
Thecompletedeclarationisavailableinthedeclarationof
conformitycerticate.
InPro, InFit, InTrac and CaliCap are registered trademarks of
Mettler-ToledoGmbHinSwitzerland,theUSA,theEuropeanUnion
andafurthervecountries.
VitonandKalrezareregisteredtrademarksofDuPontPerformance
ElastomersLLC.
Allothertrademarksarethepropertyoftheirrespectiveholders.
en
Mettler-ToledoGmbH,ProcessAnalytics
ImHackacker15,CH-8902Urdorf,Switzerland
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges.
©07/2019METTLERTOLEDO.Allrightsreserved.
PrintedinSwitzerland.30313267B
www.mt.com/pro
ForaddressesofMETTLERTOLEDO
MarketOrganizationspleasegoto:
www.mt.com/pro-MOs