Mettler Toledo InPro 3300, pH electrode InPro3300-ISFET User manual

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo InPro 3300 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
InPro 3300
Non-Glass pH Electrode
Instruction Manual
Contents
InPro 3300
52 002 258
cs Návod k obsluze 2
da Brugsanvisning 4
de Betriebsanleitung 6
en Instruction Manual 8
es Manual de instrucciones 10
 Käyttöohje 12
fr Instructions d’utilisation 14
hu Használatiutasítás 16
it Istruzioni per l‘uso 18
ja
20
ko
22
nl Gebruikershandleiding 24
pl Instrukcjaobsługi 26
pt Manual de instruções 28
ru Инструкция по эксплуатации 30
sv Bruksanvisning 32
th
 34
zh
36
8 InPro3300Non-GlasspHSensor
© 02 / 16 Mettler-Toledo GmbH InPro 3300
Printed in Switzerland 52 002 258
InPro 3300
Non-Glass pH Combination Electrode
Instruction Manual
1. Introduction
METTLERTOLEDOpHsensortypeInPro
®
3300isalow-main-
tenance,autoclavableandpressure-resistantelectrodewitha
glass-free sensing head and thickened reference electrolyte.
Pleasereadtheseinstructionscarefullybeforeputtingtheelec-
trode into operation, in order to ensure perfect operation. All
wettedpartsaremadeofmaterialsincompliancewiththeFDA
directives(FoodandDrugAdministration).Werecommendthat
thesensormayonlybeinoperationwiththeoriginalaccessories
fromMETTLERTOLEDO.Thesensormaybeoperatedandser-
vicedonlybypersonnelfamiliarwiththeunit,andwhohaveread
andunderstoodthisinstructionmanual.
2. Safety remarks
If an electrode is to be cleaned or calibrated using acid or alka-
linesolutions,safetygogglesandglovesshouldbeworn.
3. Product designation
Themarkingsonthepolymershaftincludethefollowinginfor-
mation:
METTLER TOLEDO Electrode/sensormanufacturer
InPro3300/120/Ptxxx Designation/shaftlengthinmm/
temperature
sensor
CombinationpH Typeofelectrode
pH0...14 pHmeasurementrange
0...80°C(130°C) Temperaturerangeforoperation/
(sterilization)
O
rderNo.52xxxxxx Articlenumberforordering
Additionally, the electrode is supplied with a serial number
markedontheconnectorcaptoenableindividualidentication
ofeachelectrodeproduced.
4. Installation and preparation for use
1. Donottouch,holdormakeanycontactwiththeplugand/or
pinsoftheelectrode.Electrostaticdischargecandamagethe
chip.
52 910077
52 mm
2.04"
5.6mm
0.22"
12.0 mm
0.47"
15 mm
0.59"
a=120mm
4.70"
1)
2)
3)
4)
5)
6)
7)
1:VPconnector
2:Pg13.5thread
3:WasherPTFE
4:O-ring(SiliconeFDA)
5:PEEK(FDA)shaft
6:Ceramicdiaphragm
7:ISFET
en
en Instruction Manual
InPro3300Non-GlasspHSensor 9
InPro 3300 © 02 / 16 Mettler-Toledo GmbH
52 002 258 Printed in Switzerland
2. Onunpacking,checktheelectrodeformechanicaldamage.
ReportanysignsofdamageimmediatelytoyourMETTLER
TOLEDOsupplier.
3. Removethewateringcapandbrieyrinsetheelectrodewith
de-ionizedwater.Analreadyusedelectrodeshouldrstbe
cleaned by lightly scrubbing the sensing element with a
soft-bristledtoothbrushandamildsoapsolution,followed
byrinsingwithde-ionizedwater.
4. Connect the electrode to the pH transmitter, using cable
VP6-ST/ISFET.Pleaserefertothewiringandterminaldiagram
accompanyingthecable.
5. Priortocalibration,soaktheelectrodeinbufferpH7.00or
pH4.01foratleast10minutes,withtheelectrodeconnected
tothetransmitter.
6. Beforeputtingtheelectrodeintooperation,carryoutazero
pointcalibration.Exactinstructionsforzeropointcalibration
aregivenintheinstructionmanualoftheassociatedpHtrans-
mitter.
5. Operation
5.1 Calibration of the electrode and the pH transmitter
A2-pointpHcalibrationisrecommendede.g.usingbuffers
pH7.00andpH4.01.Pleaserefertotheinstructionmanu-
alofthepHtransmitterforfurtherdetails.
5.2 CIP-cleanability (Cleaning-In-Place)
Astandardprocedureformaintaininghygienicconditions
in food processing is Cleaning-In-Place (CIP). Periodic
ushingwithhotcausticsoda(NaOH)willhowevershorten
the lifetime of the electrode.Wetherefore recommend to
retracttheelectrodefromtheprocessandtocleanitwith
hotsteam(sterilizationprocedure).
6. Maintenance
1. Carefullyrinsethesensortipanddiaphragmwithde-ionized
wateraftereachproductioncycle.Undernocircumstances
mustmeasuringsolutionbeallowedtodryontheseparts!
2. Whenthesensorisnotinoperation,storeitwithsensortip
anddiaphragmwellsubmergedinto9823/3MKClor9816/
Viscolyt™electrolyte.
3. Ifthesensorisstoredmountedinitshousing,thesamerule
asin2)aboveapplies,butthestorageelectrolytehastobe
slightlymodied,with2partsofbuffersolutionpH9.2to10
partsofelectrolyte,thisinordertoavoidcorrosionofhousing
parts.
4. Ifasensorisstoreddryforafewdaysbymistake,letitsoak
inthenormalstorageelectrolyteforseveralhoursbeforeuse.
5. Occasionallychecktheconnectorforpossibletracesofmois-
ture.Ifnecessary,cleanwellwithde-ionizedwateroralcohol,
andafterwardsdrycarefully.
6. ThesiliconeO-ring(markedNo.4inthedrawing)istobe
replacedatleasteverysixmonthsor,ifnecessary,atashort-
erinterval.
7. Trouble-shooting
Forcleaning,lightlyscrubthesensingelementoftheelectrode
withasofttoothbrushandwatertoremoveanypossibleresi-
dues.Usemilddetergentifnecessary.Possibleproteinprecipi-
tationandblockageofthediaphragmcanbetreatedwithclean-
er51340070.
a
Attention:HF-containingsolutionswilldestroythepHsen-
sitiveelementoftheelectrode.
8. Disposal
Ifnoofcialregulationsareinforce,usedordefectiveelectrodes
and the packaging can be disposed as ordinary household
waste.
9. Warranty
Intheeventof manufacturingfaults,awarranty periodof12
monthsfromdateofproductionisgranted.
InProandViscolytaretrademarksoftheMettler-ToledoGroup.
en
38 InPro3300Non-GlasspHSensor
© 02 / 16 Mettler-Toledo GmbH InPro 3300
Printed in Switzerland 52 002 258
Notes
InPro3300Non-GlasspHSensor 39
InPro 3300 © 02 / 16 Mettler-Toledo GmbH
52 002 258 Printed in Switzerland
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;
service@mettler.com.br
France Mettler-Toledo
Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gien
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
w w w . m t . c o m / p r o - M O s