8 InPro3300Non-GlasspHSensor
© 02 / 16 Mettler-Toledo GmbH InPro 3300
Printed in Switzerland 52 002 258
InPro 3300
Non-Glass pH Combination Electrode
Instruction Manual
1. Introduction
METTLERTOLEDOpHsensortypeInPro
®
3300isalow-main-
tenance,autoclavableandpressure-resistantelectrodewitha
glass-free sensing head and thickened reference electrolyte.
Pleasereadtheseinstructionscarefullybeforeputtingtheelec-
trode into operation, in order to ensure perfect operation. All
wettedpartsaremadeofmaterialsincompliancewiththeFDA
directives(FoodandDrugAdministration).Werecommendthat
thesensormayonlybeinoperationwiththeoriginalaccessories
fromMETTLERTOLEDO.Thesensormaybeoperatedandser-
vicedonlybypersonnelfamiliarwiththeunit,andwhohaveread
andunderstoodthisinstructionmanual.
2. Safety remarks
If an electrode is to be cleaned or calibrated using acid or alka-
linesolutions,safetygogglesandglovesshouldbeworn.
3. Product designation
Themarkingsonthepolymershaftincludethefollowinginfor-
mation:
METTLER TOLEDO Electrode/sensormanufacturer
InPro3300/120/Ptxxx Designation/shaftlengthinmm/
temperature
sensor
CombinationpH Typeofelectrode
pH0...14 pHmeasurementrange
0...80°C(130°C) Temperaturerangeforoperation/
(sterilization)
O
rderNo.52xxxxxx Articlenumberforordering
Additionally, the electrode is supplied with a serial number
markedontheconnectorcaptoenableindividualidentication
ofeachelectrodeproduced.
4. Installation and preparation for use
1. Donottouch,holdormakeanycontactwiththeplugand/or
pinsoftheelectrode.Electrostaticdischargecandamagethe
chip.
52 910077
52 mm
2.04"
5.6mm
0.22"
12.0 mm
0.47"
15 mm
0.59"
a=120mm
4.70"
1)
2)
3)
4)
5)
6)
7)
1:VPconnector
2:Pg13.5thread
3:WasherPTFE
4:O-ring(SiliconeFDA)
5:PEEK(FDA)shaft
6:Ceramicdiaphragm
7:ISFET
en
en Instruction Manual
InPro3300Non-GlasspHSensor 9
InPro 3300 © 02 / 16 Mettler-Toledo GmbH
52 002 258 Printed in Switzerland
2. Onunpacking,checktheelectrodeformechanicaldamage.
ReportanysignsofdamageimmediatelytoyourMETTLER
TOLEDOsupplier.
3. Removethewateringcapandbrieyrinsetheelectrodewith
de-ionizedwater.Analreadyusedelectrodeshouldrstbe
cleaned by lightly scrubbing the sensing element with a
soft-bristledtoothbrushandamildsoapsolution,followed
byrinsingwithde-ionizedwater.
4. Connect the electrode to the pH transmitter, using cable
VP6-ST/ISFET.Pleaserefertothewiringandterminaldiagram
accompanyingthecable.
5. Priortocalibration,soaktheelectrodeinbufferpH7.00or
pH4.01foratleast10minutes,withtheelectrodeconnected
tothetransmitter.
6. Beforeputtingtheelectrodeintooperation,carryoutazero
pointcalibration.Exactinstructionsforzeropointcalibration
aregivenintheinstructionmanualoftheassociatedpHtrans-
mitter.
5. Operation
5.1 Calibration of the electrode and the pH transmitter
A2-pointpHcalibrationisrecommendede.g.usingbuffers
pH7.00andpH4.01.Pleaserefertotheinstructionmanu-
alofthepHtransmitterforfurtherdetails.
5.2 CIP-cleanability (Cleaning-In-Place)
Astandardprocedureformaintaininghygienicconditions
in food processing is Cleaning-In-Place (CIP). Periodic
ushingwithhotcausticsoda(NaOH)willhowevershorten
the lifetime of the electrode.Wetherefore recommend to
retracttheelectrodefromtheprocessandtocleanitwith
hotsteam(sterilizationprocedure).
6. Maintenance
1. Carefullyrinsethesensortipanddiaphragmwithde-ionized
wateraftereachproductioncycle.Undernocircumstances
mustmeasuringsolutionbeallowedtodryontheseparts!
2. Whenthesensorisnotinoperation,storeitwithsensortip
anddiaphragmwellsubmergedinto9823/3MKClor9816/
Viscolyt™electrolyte.
3. Ifthesensorisstoredmountedinitshousing,thesamerule
asin2)aboveapplies,butthestorageelectrolytehastobe
slightlymodied,with2partsofbuffersolutionpH9.2to10
partsofelectrolyte,thisinordertoavoidcorrosionofhousing
parts.
4. Ifasensorisstoreddryforafewdaysbymistake,letitsoak
inthenormalstorageelectrolyteforseveralhoursbeforeuse.
5. Occasionallychecktheconnectorforpossibletracesofmois-
ture.Ifnecessary,cleanwellwithde-ionizedwateroralcohol,
andafterwardsdrycarefully.
6. ThesiliconeO-ring(markedNo.4inthedrawing)istobe
replacedatleasteverysixmonthsor,ifnecessary,atashort-
erinterval.
7. Trouble-shooting
Forcleaning,lightlyscrubthesensingelementoftheelectrode
withasofttoothbrushandwatertoremoveanypossibleresi-
dues.Usemilddetergentifnecessary.Possibleproteinprecipi-
tationandblockageofthediaphragmcanbetreatedwithclean-
er51340070.
a
Attention:HF-containingsolutionswilldestroythepHsen-
sitiveelementoftheelectrode.
8. Disposal
Ifnoofcialregulationsareinforce,usedordefectiveelectrodes
and the packaging can be disposed as ordinary household
waste.
9. Warranty
Intheeventof manufacturingfaults,awarranty periodof12
monthsfromdateofproductionisgranted.
InProandViscolytaretrademarksoftheMettler-ToledoGroup.
en
38 InPro3300Non-GlasspHSensor
© 02 / 16 Mettler-Toledo GmbH InPro 3300
Printed in Switzerland 52 002 258
Notes
InPro3300Non-GlasspHSensor 39
InPro 3300 © 02 / 16 Mettler-Toledo GmbH
52 002 258 Printed in Switzerland
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;
service@mettler.com.br
France Mettler-Toledo
Analyse Industrielle S.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-Toledo GmbH
Prozeßanalytik,Ockerweg3,DE-35396Gien
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo (Schweiz) GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
w w w . m t . c o m / p r o - M O s
Mettler-ToledoGmbH
ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
02/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.52002258
w w w. m t . c o m / p r o
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40

Mettler Toledo InPro 3300 User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI