44
9. DefinetheprocessvariablesPV,SV,TVandQV
Menupath:DeviceSetup>Measurements>UnitSetup
10.Calibratesensor.Menupath:DeviceSetup>SensorCalibration
11.Performfurthersettings.SeeOperationManualoftheM400transmitter.
12.Storeconfigurationtodevice.
HART handheld terminal
h
Note:TheDD“008E8E700101.hhd”isonthesuppliedCD-ROM.Youcanalsodownloadthe
DDviatheinternet:“www.mt.com/M400“.
Forstep1refertothedocumentationoftheHARThandheldterminal.
1. CheckiftheDDoftheM400transmitterhasalreadybeeninstalledontheHARThandheld
terminal.InstalltheDDifnecessary.
2. Thecommunicationisbuiltupautomatically.
3. Loadconfigurationfromdevice.Menupath:DeviceSetup>DetailedSetup
4. SetTagor/andLong Tag.
Menupath:DeviceSetup>HARTSetup
5. SetDateandTime.Setthetimeinthe24hourformat.Thetimeformatcannotbe
changed.Menupath:DeviceSetup>System
6. Setrangeoftheanalogoutputsignal.
Menupath:Detailedsetup>AnalogOutput
– URV(UpperRangeValue)andLRV(LowerRangeValue):Thevaluescanbechanged
tothecurrentmeasurementrange.Thevalueshavetobewithinthemeasurementlimits
ofthesensor.
– USL(UpperSensorLimit)andLSL(LowerSensorLimit):Thelimitsaredefinedbythe
sensorandcannotbechanged.
7. DefinetheprocessvariablesPV,SV,TVandQV
Menupath:DeviceSetup>Measurements>UnitSetup
8. Calibratesensor.Menupath:DeviceSetup>SensorCalibration
9. Performfurthersettings.SeeOperationManualoftheM400transmitter.
en