Mettler Toledo pH electrode InPro4550VP Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo pH electrode InPro4550VP Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
14 InPro 4550 VP pH electrode
InPro 4550 VP pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 400 Printed in Switzerland
1 Introduction ...................................................................15
2 Safetyinstructions ..........................................................15
3 Productdescription ......................................................... 15
4 Installationandcommissioning .......................................16
5 Cabling ..........................................................................16
6 Calibrationandmeasurement ..........................................16
7 Maintenance ..................................................................16
8 Diagnostics....................................................................17
9 Trouble-shooting(sluggishresponsetime,
insufcientsensitivity) ....................................................17
10 Disposal ........................................................................17
11 Guarantee ......................................................................17
Contents
en
InPro
4550 VP
pH Electrode
Instruction Manual
InProandXerolytareregisteredtrademarks
ofMettler-ToledoGmbHinSwitzerland,theUSA,
theEuropeanUnionandafurthervecountries.
en Instruction Manual
InPro 4550 VP pH electrode 15
© 04 / 16 Mettler-Toledo GmbH InPro 4550 VP pH electrode
Printed in Switzerland 52 002 400
1 Introduction
TheMETTLERTOLEDOelectrodeversionsInPro
®
4550 VP
are low-maintenance, temperature-compensated pH
combination electrodes containing polymer electrolyte
(Xerolyt
®
Plus). An integrated temperature sensor (RTD)
providesautomatic compensation of the electrode slope
toensureaccuratereadings.
The InPro 4550VP electrodes are suitable for pH
measurementsin:
–emulsionsandsuspensions
–sulde-bearingmedia
–solutions containing heavy metals (lead, mercury,
silver,etc.)
–processes with rapid pressure uctuations (within
thegivenrange)
–heavilycontaminatedmedia.
The InPro 4550VP electrodes are unsuitable for
sterilization/autoclaving procedures, for processes with
rapid temperature uctuations or food and beverage
applications.
Pleasereadthroughtheseoperatinginstructionscarefully
before commissioning, in order to ensure trouble-free
use.Thesegeneralinstructionsmustbereadilyavailable
tooperatorsoftheelectrodeatalltimes.
2 Safetyinstructions
TheInPro4550VPelectrodesareintendedsolelyforthe
purposedescribedin§1above.Theelectrodesconsistpartly
ofglassandshouldbehandledcarefully.Iftheelectrodeis
tobecleanedorcalibratedusingacidoralkalinesolutions,
safetygogglesandglovesshouldbeworn.Localsafety
regulationsareapplicabletothehandlingoftheelectrode.
We recommend that you only operate the electrode in
combination with original parts from METTLER TOLEDO.
Operationandmaintenanceshouldbecarriedoutonlyby
trainedpersonnelandstaffwhohavereadandunderstood
theoperatinginstructions.
3 Product description
The inscription on eachelectrode contains the following
information:
METTLER TOLEDO Manufactureroftheelectrode
InPro
®
4550 VP / xxx / SG ypedesignation/xxx=typeof
temperaturesensor(Pt100/
Pt1000)/SG=ancillaryelectrode
combinationpH Typeofelectrode
pH 0 … 14 pHmeasurementrange
0…130°C Temperaturerangeforoperation
Order No. 52 002 40X Ordernumber
SEV14ATEX0168X
Identicationandnumber
ofcerticate
IECExSEV14.0025X
Identicationandnumber
ofcerticate
IS/I,II,III/1/ABCDEFG/T6 FMapproved
Seeinstructionmanual! Instruction(Observetheoperating
instructions)
Inadditioneachelectrodeisallocatedaserialnumberon
theplug-inheadtoenableidentication.
en
16 InPro 4550 VP pH electrode
InPro 4550 VP pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 400 Printed in Switzerland
4 Installation and commissioning
1. Onunpacking,andagainpriortoinstallation,checkthe
electrodevisuallyforsignsofmechanicaldamagetothe
glassmembrane,theshaftandtheconnector.Report
any faults or defects immediately to your authorized
METTLERTOLEDOdealer.
2. Removethewateringcapandrinsetheelectrodewith
distilledordeionizedwater.Useonlysofttissuetodab
theelectrodedry.Donotrubtheglassmembranesince
this can lead to sluggish response behavior of the
electrode.
3. Electrodes with a dry glass membrane must be
conditioned forseveralhoursina3mol KClsolution
before being used.
4. Connectalwaysthesolutionground(bluecable)tothe
transmitter.
5. UsePTFEtapetosealtheNPTthread.
6. If possible use the protection sleeve (order no.
52401808).
5 Cabling
TheInPro4550VPisequippedwithaVarioPinconnector.
Toconnect thecable totheelectrode,attachandgently
rotateuntiltheconnectorisengaged(groove).Screwthe
cablecap nut onto the electrode until the connection is
secure. The plug assembly should be tted hand-tight
only; no tools should be used. For details of the exact
assignmentoftheindividualcableleadstotheVPpins,
please refer to the instruction leaet “Preparation of the
cableforconnectiontothetransmitter”suppliedtogether
withthecable.
6 Calibration and measurement
InPro 4550VP electrodes require two-point calibration.
Before calibration, the watering cap must be removed
from the pH electrode. The electrode is then dipped
successively into two different buffer solutions, each of
knownpHvalue,andthe transmittercalibratedtothese
buffervalues(forexample,bufferspH4.01andpH7.00).
Pleasenotetheinstructionsgivenintheoperatingmanual
ofthetransmitter.Calibrationshouldberepeatedasand
whenneeded.
7 Maintenance
–ElectrodeswithXerolyt
Pluspolymerelectrolytecannot
berelled.
–Electrodes shouldnotbe storeddryfor any periodof
time.
–Duringstorage,thewateringcap should beinplace,
lledwithFriscolyt™or3molKClsolution.
–Neverstoreanelectrodeindistilledordeionizedwater,
asthisshortensthelifetimeoftheelectrode.
–Shouldtheelectrodedryout,itmustbeconditionedfor
24hoursinFriscolyt(orderingno.51340053for250
ml,or51340054for63250ml)beforereuse.After
conditioning,recalibrationisrequired.
–Aftereachworkcycle,thetipoftheelectrodeshouldbe
cleanedusingdeionizedwater.
en
InPro 4550 VP pH electrode 17
© 04 / 16 Mettler-Toledo GmbH InPro 4550 VP pH electrode
Printed in Switzerland 52 002 400
Attention! Do notunderanycircumstanceuseMETTLER
TOLEDOreactivatingsolution(orderingno.51319053/in
USA439895125)asthiscontainshydrouoricacid(HF)
whichwouldattacktheXerolytreferencesystem.
8 Diagnostics
Theelectrodesareequippedwithagroundingringwhich
allows a diagnostic of the diaphragm condition. If this
diaphragm (open junction) should become clogged or
if the reference resistance exceeds a certain value (see
transmitter operation manual), it is then necessary to
replacetheelectrode.InPro4550VPelectrodesshouldasa
generalrulenotbecleanedwithacidiccleaningsolutions,
e.g.diaphragmcleaningsolution,reactivationsolution,etc.
However,byextremelycarefulapplication,itispossibleto
reactivatetheglassmembrane,seepoint9below.
9 Trouble-shooting(sluggishresponsetime,
insufcientsensitivity)
Attention! Theglassmembranemustbeverycautiously
dabbedwith HF solutionusinganebrush andtaking
care that solution does not come into contact with the
titanium pin (metallic solution ground). Also check the
electrodeconnector(VP)forpossibletracesofmoisture.
Ifnecessary,cleanwellusingdeionizedwateroralcohol,
andafterwardsdrycarefully.
10 Disposal
DisposeofInPro4550VPelectrodesinaccordancewith
local or state regulations.
11 Guarantee
Onmanufacturingdefects,12monthsafterdelivery.
en
74 InPro 4550 VP pH electrode
InPro 4550 VP pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 400 Printed in Switzerland
Notes
InPro 4550 VP pH electrode 75
© 04 / 16 Mettler-Toledo GmbH InPro 4550 VP pH electrode
Printed in Switzerland 52 002 400
Notes
Brazil Mettler-Toledo Ind. e Com. Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-ToledoAnalyseIndustrielleS.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-ToledoGmbH
Prozeßanalytik,Ockerweg3,DE-35396Gießen
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo(Schweiz)GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.52002400
www.mt.com/pro
Addresses|Impressum
/