  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo InPro 3100(i) User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
14 InPro 3100 (i) pH electrode
InPro 3100 (i) pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 293 Printed in Switzerland
1 Introduction ...................................................................15
2 Safetyinstructions ..........................................................15
3 Product description ......................................................... 15
4 Installationandcommissioning .......................................16
5 Operation .......................................................................16
5.1 CalibrationoftheelectrodeandpHtransmitter ..................16
5.2 Sterilization of electrode ..................................................16
5.3 TemperaturesensorofInPro3100(analogversion) ...........16
6 Maintenance ..................................................................17
7 Trouble-shooting(sluggishresponsetime,insufcientsensiti-
vityetc.) ........................................................................17
8 Disposal
........................................................................17
9 Guarantee ......................................................................17
Contents
en
InPro
3100 (i)
pH Electrode
Instruction Manual
InProandISMareregisteredtrademarks
ofMettler-ToledoGmbHinSwitzerland,theUSA,
theEuropeanUnionandafurthervecountries.
12 mm
0.47"
Connectortype(here:VP)
Bridgeelectrolyte
Silver-iontrapanddiaphgragm
Referenceelectrolyte
Temperaturesensor
Argenthalreferenceelement
Diaphragm
pH sensitive glass
en Instruction Manual
InPro 3100 (i) pH electrode 15
© 04 / 16 Mettler-Toledo GmbH InPro 3100 (i) pH electrode
Printed in Switzerland 52 002 293
1 Introduction
The METTLER TOLEDO InPro 3100(i) pH electrodes are
low-maintenance, autoclavable and pressure-resistant
measuringelectrodeswithagel-typereferenceelectrolyte.
The reference element incorporates a silver-ion trap to
preventsilversuldeprecipitationonthediaphragm.Incase
ofInPro3100,atemperaturesensorisintegrated,mainlyfor
automatictemperaturecompensationpossibilitiesofthepH
signal during calibration and operation.
The InPro 3100(i) electrodes are mainly designed for
the measurement of pH value in biotechnical processes
at moderate pressures (<6bar). It can be used under
industrial fermentation, pilot and laboratory conditions,
and is particularly intended for those processes where
intermediatesterilizationorautoclavingisindispensable.
Pleasereadthroughtheseoperatinginstructionscarefully
beforecommissioning,inordertoensuretrouble-freeuse.
2 Safetyinstructions
The InPro 3100(i) electrodes are intended solely for the
purposedescribedin§1above.Theelectrodeconsistspartly
ofglassandshouldbehandledcarefully.Iftheelectrodeis
tobecleanedorcalibratedusingacidoralkalinesolutions,
safetygoggles and gloves shouldbeworn.Localsafety
regulationsareapplicabletothehandlingoftheelectrode.
Pleasefollowtheguidelinesenclosedforaninstallationin
theEx-zone.
We recommend that you only operate the electrode in
combination with original parts from METTLER TOLEDO.
Operationandmaintenanceshouldbecarriedoutonlyby
trainedpersonnelandstaffwhohavereadandunderstood
theoperatinginstructions.
3 Product description
The inscriptionon each electrode contains thefollowing
information:
METTLER TOLEDO Manufactureroftheelectrode
InPro
®
310XiUD/SG/yyy/zzzTypedesignation;X=membrane
glass;i=ISMDigital;UD=upside
down/SG=ancillaryelectrode/
yyy=rodlengthinmm/zzz=type
oftemperaturesensor
combinationpH Typeofelectrode
pH0…14 pHmeasurementrange
0…80°C/140°C(130°C) Temperaturerangeforoperation/
sterilization(forInPro3100UD)
OrderNo.5200XXXX Ordernumber
SEV14ATEX0168X Identicationandnumber
ofcerticate
IECExSEV14.0025X Identicationandnumber
ofcerticate
IS/I,II,III/1/ABCDEFG/T6 FMapproved
Seeinstructionmanual! Instruction(Observethe
operating instructions)
Inadditioneachelectrodeisallocatedaserialnumberon
theplug-inheadtoenableidenticationandtraceability.
en
16 InPro 3100 (i) pH electrode
InPro 3100 (i) pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 293 Printed in Switzerland
4 Installationandcommissioning
1. On unpacking, check the electrode for mechanical
damage of the pH sensitive glass and shaft. Report
any damage immediately to your METTLER TOLEDO
supplier.
2. Removethewateringcapandbrieyrinsetheelectrode
withdeionizedwater.Afterrinsing,theelectrodeshould
onlybedabbeddrywithatissue.DonotrubthepH
sensitive glass, since this can lead to electrostatic
chargingandsluggishresponsetimes.
3. CheckthespacebehindthepHsensitiveglassforthe
presenceofanyairbubblesandremovesamebygently
swingingtheelectrodeinaverticalplane.
4. Please refer to the instruction manual of the relative
housinginordertoinstalltheelectrode.
5. Connect the electrode to the pH transmitter, using a
cable with the appropriate connector (VP connector
forInPro3100).Pleaserefertothewiringandterminal
diagramaccompanyingthecable.
6. Intelligent ISM
®
electrodessuchasInPro310Xienable
“Plug and Measure” and extended diagnostics. For
installation, commissioning and operation of the
ISM system please observe the instructions for the
transmitter,moduleandcable.
5 Operation
5.1CalibrationoftheelectrodeandpHtransmitter
FortheInPro3100(i)electrodea2-pointpHcalibrationis
recommended.Beforecalibration,rstremovethewatering
cap.Tospeeduptheresponsetimeoftheelectrodeafter
longtransportandstorageperiods,andbeforecalibration,
diptheelectrodeintobufferpH7,bufferpH4andbufferpH
7again,eachtimeforaboutoneminute.Theelectrodeis
thendippedsuccessivelyintotwodifferentbuffersolutions
withgivenpHreferencevaluesandthetransmittercalibrated
tothesebuffervalues.NormallyyoustartwithbufferpH7
fordeterminingthe0-pointandthenbufferpH4or9for
determiningtheslope.Inthecaseofmicroprocessor-based
transmitters with buffer recognition,the order of the two
chosenbuffer solutionsis irrelevant.Whenever possible,
atemperaturesensor(forInPro3100theintegratedone)
should be connected to the transmitter and automatic
temperaturemeasurement”selectedthere.Pleasereferto
theinstructionmanualofthetransmitterforfurtherdetails.
5.2Sterilizationofelectrode
TheInPro3100(i)electrodecanbesterilizedeitherthrough
autoclaving (remove red protection cap) or through in-
situ sterilization with steam or superheated fermentation
medium. Sterilization temperatures above 140°Care not
allowed,andexcessivesterilizationtime(forinstancet>30
minat140°C)could notably shorten the life time of the
electrode.Repeatedsterilizationand/orautoclavingofthe
electrodecanleadtoaslightcolorationoftheelectrolyte.
This color change does not effect the functioning of the
electrode.
5.3TemperaturesensorofInPro3100(analogversion)
The integrated Pt100/1000 temperature sensor in the
InPro3100isprimarilyintendedforautomatictemperature
compensationofthepHsignalandnotforanyaccurate
andsafetemperatureindicationorcontroloftheprocess
temperature. If you anyhow want to try this, check that
yourtransmitter has the required temperature indication
accuracyandafree outletforthesignal.Furthermore,a
calibrationofthetemperaturesignalmightbenecessary.
en
InPro 3100 (i) pH electrode 17
© 04 / 16 Mettler-Toledo GmbH InPro 3100 (i) pH electrode
Printed in Switzerland 52 002 293
6 Maintenance
1. Carefully rinse the electrode tip and diaphragm with
deionizedwateraftereachproductioncycle.Underno
circumstancesmustmeasuringsolutionbeallowedto
dryontheseparts!
2. When the electrode is not in operation, store it with
electrode tip and diaphragm well submerged into
9823
/3MKClor9816/Viscolyt™electrolyte.
3. Iftheelectrodeisstoredmountedinitshousing,the
same rule as in 2) above applies, but the storage
electrolytehastobeslightlymodied,with2partsof
buffersolutionpH9.2to10partsofelectrolyte,thisin
ordertoavoidcorrosionofhousingparts.
4. Ifanelectrodeisstoreddryforafewdaysbymistake,
letitsoakinthenormalstorageelectrolyteforseveral
hoursbeforeuse.
5. Occasionallychecktheconnectorforpossibletracesof
moisture.Ifnecessary,cleanwellwithdeionizedwater
oralcohol,andafterwardsdrycarefully.
7 Trouble-shooting(sluggishresponsetime,
insufcientsensitivityetc.)
TheglassmembranecanbereactivatedusingaHFsolution
(orderno.51319053).Theelectrodetip(onlypHsensitive
glasspart)isimmersedintotheHFsolutionforaperiodof
about5minutes,isthenimmediatelyandthoroughlyrinsed
withdeionizedwaterandthereafterstoredinbuffersolution
pH4(seepoint6above)uptothediaphragmfor12hours.
Po s siblep rote inpre c ipita t ionandblocka geoft h e diaphra g m
canbetreatedwithourspecialelectrodecleaner(orderno.
51340068).However,useofthediaphragmcleaner(order
no.51340070)isnormallynotnecessaryforthiselectrode
type,asthebuilt-insilver-iontrappreventstheformationof
silversuldecontamination.
Also,checktheconnectorforpossibletracesofmoisture.If
necessary,cleanwellusingdeionizedwateroralcohol,and
afterwardsdrycarefully.
8 Disposal
If no relative ofcial regulations are in force, used or
defectiveInPro3100(i)electrodesandthepackagingcan
bedisposedofasordinaryhouseholdwaste.
9 Guarantee
Onmanufacturingdefects,12monthsafterdelivery.
en
74 InPro 3100 (i) pH electrode
InPro 3100 (i) pH electrode © 04 / 16 Mettler-Toledo GmbH
52 002 293 Printed in Switzerland
Notes
Brazil Mettler-ToledoInd.eCom.Ltda.
AvenidaTamboré,418–Tamboré,
BR-06460-000Barueri/SP,Brazil
Phone+551141667400
e-mailmettler@mettler.com.br;service@mettler.com.br
France Mettler-ToledoAnalyseIndustrielleS.A.S.
30,Bld.deDouaumont,FR-75017Paris,France
Phone+33147370600
e-mailmtpro-f@mt.com
Germany Mettler-ToledoGmbH
Prozeßanalytik,Ockerweg3,DE-35396Gien
Phone+49641507-444
e-mailprozess@mt.com
Switzerland Mettler-Toledo(Schweiz)GmbH
ImLangacher,CH-8606Greifensee
Phone+41449444760
e-mailProSupport.ch@mt.com
United States METTLER TOLEDO
ProcessAnalytics
900MiddlesexTurnpike,Bld.8,Billerica,MA01821,USA
Phone+17813018800
Freephone+18003528763(onlyUSA)
e-mailmtprous@mt.com
FormoreaddressesofMETTLERTOLEDOMarketOrganizationspleasegoto:
www.mt.com/pro-MOs
Mettler-ToledoGmbH,ProcessAnalytics,ImHackacker15,CH-8902Urdorf
Phone+41447296211,Fax+41447296636
Subjecttotechnicalchanges
04/2016©Mettler-ToledoGmbH
PrintedinSwitzerland.52002293
www.mt.com/pro
Addresses|Impressum
/