1
© 2016 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
TP201 20-42-FO-00013 Rev A1 C2169
Applies to Models:
153800 (-001 thru -003)
Special Tools:
none
Language of origin: English
Note
Use this kit to replace the Electronic Proportional Water Valve, for all Elevance
®
Standard and Continental-Style Delivery Units.
Electronic Proportional Water Valve Installation
[Kit # 002-2029-00]
003-2913-00 Rev. A (mm/yy)
Tray
Pad
Magnetic Lid
Metal Cover
Top Cover
Assembly
Screws (8)
Water
Shut Off
Air
Shut Off
Unplug Power
Supply
Water
Bottle
J-Box
WARNING
Always disconnect power from the chair / unit before removing any covers.
Failure to do so may result in personal injury.
Accessing the Proportional Water Valve on an Elevance
®
Standard Unit...
A) Turn off air and water using the Shut Offs in the J-Box.
B) Unplug Power Supply from electrical mains in the J-Box.
C) Turn Water Bottle CCW to release the air pressure in the water tubing.
D) Remove, Tray, Pad, and Magnetic Lid.
E) Remove two screws and lift off the Metal Cover.
F) Remove handpieces from holders (allow them to hang) and remove eight
screws so you can rotate the Top Cover Assembly upside down on top
of the instrument platform (leaving handpieces, tubing and wiring connected).
G) Disconnect the handpiece sensor harnesses, touch sensor harnesses, USB and
LIN connections from the ICM PC Board and remove the Top Cover Assembly.
ICM
PC Board
Elevance
®
Standard Unit