Midmark 224 Barrier-Free® Exam Chair (-001 thru -003, -011) Installation guide

Type
Installation guide

This manual is also suitable for

Midmark 224 Barrier-Free® Exam Chair (-001 thru -003, -011) combines versatility and functionality for various medical examinations. Its 340-pound weight capacity and power articulating movement allow smooth positioning of patients. It features an articulating backrest with gas-spring assistance for Trendelenburg and reverse Trendelenburg procedures, and adjustable stirrups for patient comfort. The chair's design facilitates easy patient transfer and accommodates a wide range of patient sizes.

Midmark 224 Barrier-Free® Exam Chair (-001 thru -003, -011) combines versatility and functionality for various medical examinations. Its 340-pound weight capacity and power articulating movement allow smooth positioning of patients. It features an articulating backrest with gas-spring assistance for Trendelenburg and reverse Trendelenburg procedures, and adjustable stirrups for patient comfort. The chair's design facilitates easy patient transfer and accommodates a wide range of patient sizes.

1
© 2018 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Bottom
Actuator Pin
Actuator Cap Assembly Installation (002-10244-00)
003-10064-00 AA1 4/12/18
Caution
Always disconnect power to chair before removing
covers or performing any service procedures.
Style A
Step 1: Remove Back Cylinder.
A) Remove upholstery.
B) Remove Top Cover by removing two screws lifting up and off.
C) Remove Bottom Actuator Pin by pushing Spring Tab inward
and pulling pin out.
D) Rotate chair back onto the seat of the chair.
E) Loosen nuts and unthread Back Cylinder from Actuator Cap
Assembly.
Back
Actuator
Spring
Tab
Actuator
Cap
Top
Cover
midmark.com
2
© 2018 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
MA10843i
Set Screws
Step 2: Remove Handle, Rod and Bushing.
A) Remove set screw from Actuator Shaft.
B) Pull Actuator Handle, Release Rod, and Release Rod
Bushing out of chair back. Repeat for opposite side.
Actuator
Handle
Release Rod
Bushing
Release Rod
MA10843i
Step 3: Remove Actuator Cap Assembly.
A) Rotate Actuator Cap Assembly 90° as shown, slide off Actuator Cap Top.
B) Remove the E-Clip from one side of Actuator Shaft.
C) Press down on Actuator Plunger to compress Spring.
D) Pull Actuator Shaft out of Pivot Bearing, chair back and Actuator Cap Body.
E) Remove Pivot Bearings, Actuator Cap Body. Actuator Plunger and Spring.
Actuator Cap
Assembly
Actuator
Cap Top
E-Clip
Actuator
Plunger
Spring
Actuator
Cap Body
Pivot
Bearing
Actuator
Shaft
Actuator
Shaft
midmark.com
3
© 2018 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
Pivot
Bearing
MA10897i
Step 4: Reassemble Actuator cap (Reverse Order).
A) Install Spring and Actuator Plunger into the Actuator Cap Body and position into chair back. Align shaft cut outs.
B) Press down on Actuator Plunger compressing spring, Insert Actuator Shaft through chair back and Actuator Cap Body.
C) Insert Pivot Bearings and E-Clips locking Actuator shaft in place.
D) Install Actuator Cap Top and Rotate Actuator Cap Assembly to line up with cylinder.
E) Install Actuator Handle, Release Rod, and Release Rod Bushing into chair back and Actuator Shaft, Secure with Set
Screw. Repeat on opposite side.
Note
The old Actuator Plunger can be turned upside down and
used to push on the new one compressing the spring so
Actuator Shaft can be inserted into the Actuator Body.
Actuator
Cap Top
E-Clip
Old Actuator
Plunger
Spring
Actuator
Cap Body
Pivot
Bearing
E-Clip
Actuator
Plunger
Actuator
Shaft
midmark.com
4
© 2018 Midmark Corp. | 60 Vista Drive Versailles, OH 45380 USA | 1-800-643-6275 | 1-937-526-3662 |
MA10898i
MA10454i
Plunger
If Back section ‘locked up’,
or difficult to move...
A) Remove clevis pin from bottom of cylinder.
B) Loosen jam nuts.
C) Screw cylinder in slightly.
D) Tighten jam nuts.
E) Install clevis pin and test operation.
If Back section drifts up or down...
A) Remove clevis pin from bottom of cylinder.
B) Loosen jam nuts.
C) Unscrew cylinder slightly.
D) Tighten jam nuts.
E) Install clevis pin and test operation.
Step 5: Install Cylinder, Cover and Upholstry.(Reverse Order).
A) Position Back Actuator Jam Nut 5/8” to 11/16”” from top of the Plunger.
B) Rotate Actuator Cap and thread in Back Actuator.
C) Rotate chair back up, position and secure bottom of Back Actuator
with Bottom Actuator Pin.
D) Test back section adjustment. If it drifts up or down, or is difficult to
move proceed to step 6, If adjustment is good then proceed to Step E.
E) Install Top Cover and secure with two screws.
F) Install upholstery.
Step 6: Back Adjustment (If Needed).
A) If Back section drifts up or down, or is difficult to move perform
appropriete steps listed below.
5/8” - 11/16”
Jam
Nuts
Jam
Nuts
midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

Midmark 224 Barrier-Free® Exam Chair (-001 thru -003, -011) Installation guide

Type
Installation guide
This manual is also suitable for

Midmark 224 Barrier-Free® Exam Chair (-001 thru -003, -011) combines versatility and functionality for various medical examinations. Its 340-pound weight capacity and power articulating movement allow smooth positioning of patients. It features an articulating backrest with gas-spring assistance for Trendelenburg and reverse Trendelenburg procedures, and adjustable stirrups for patient comfort. The chair's design facilitates easy patient transfer and accommodates a wide range of patient sizes.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI