Steris Service Pc Application - Surgical Lighting Software Operating instructions

Type
Operating instructions
1
STERIS Service PC Application – Surgical Lighting Software Quick Start Guide 10066159
© 2015, STERIS Coporration. All Rights Reserved. Printed in U.S.A.
STERIS Service PC Application – Surgical Lighting Software Quick Start Guide
1. Hardware Required
Figure 1. Service Hardware
USB-RS232 Adapter
(P755717-735)
RS232Female-to-Male Straight-Through Cable
(P764333-680)
Calibration Screwdriver (
P
136
824-317)
Laptop
QUICK START GUIDE
STERIS SERVICE PC APPLICATION –
Surgical Lighting Software
(02/27/15) 10066159
2
10066159 Quick Start Guide STERIS Service PC Application – Surgical Lighting Software
2. Connect PC to HarmonyAIR Surgical Lighting System M Series
1. Ensure HarmonyAIR™ Surgical Lighting System M Series is POWER OFF.
2. Install USB-RS232 adapter (755717-735) to PC.
3. At Touch Panel Control Unit (TPCU), gently pry TPCU bezel away from wall with your fingers using the recessed
indentations at the bottom of the bezel.
4. Connect PC USB-RS232 adapter to TPCU serial port adapter harness using a straight through RS232 cable
(764333-680) as shown in Figure 2.
Figure 2. Connect PC To TPCU through A Straight Through Cable
USB-RS232
Adapter
Straight Through
Cable
3
STERIS Service PC Application – Surgical Lighting Software Quick Start Guide 10066159
3. Launch STERIS Service PC Application – Surgical Lighting
Software and Connect Serial Port
1. Ensure HarmonyAIR Surgical Lighting System M Series is POWER OFF.
2. Double-click Service PC Application – Surgical Lighting Software shortcut icon on the PC desktop as shown
in Figure 3.
The application can be also launched by going to Start > Programs > STERIS Service PC Application –
Surgical Lighting > STERIS Service PC Application – Surgical Lighting.
Figure 3. STERIS Service PC Application – Surgical Lighting Software Desktop Shortcut Icon
4
10066159 Quick Start Guide STERIS Service PC Application – Surgical Lighting Software
3. The software is running as shown in Figure 4. The status of the selected serial port is displayed in the Status bar.
Figure 4. STERIS Service PC Application – Surgical Lighting Software
5
STERIS Service PC Application – Surgical Lighting Software Quick Start Guide 10066159
4. From the Serial Port dropdown list, select the serial port to be used to communicate with the HarmonyAIR Surgical
Lighting System M Series and click Connect as shown in Figure 5. The Status bar updates the serial port status
after the port is connected as shown in Figure 6.
Figure 5. Serial Port Status
Figure 6. Software Is Ready to Communicate with HarmonyAIR Surgical Lighting System
M Series
Serial Port Status
ConnectSelect Serial Port
6
10066159 Quick Start Guide STERIS Service PC Application – Surgical Lighting Software
4. Establish Communication with HarmonyAIR Surgical Lighting
System M Series
1. Verify software is ready to communicate with HarmonyAIR Surgical Lighting System M Series as described in
Section 3.
2. Power up HarmonyAIR Surgical Lighting System M Series.
3. Verify software has established the communication to the surgical lighting system as shown in Figure 7.
Figure 7. Communication Established Between Software and Surgical Lighting System
Healthcare
Equipment
7
STERIS Service PC Application – Surgical Lighting Software Quick Start Guide 10066159
8
10066159 Quick Start Guide STERIS Service PC Application – Surgical Lighting Software
Healthcare
Equipment
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8

Steris Service Pc Application - Surgical Lighting Software Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI